K961606 · Metro Optics of Austin, Inc. · LPL · Nov 20, 1996 · Ophthalmic
Device Facts
Record ID
K961606
Device Name
METRO FOCAL ASPHERIC (POLYMACON)
Applicant
Metro Optics of Austin, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Nov 20, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The METRO FOCAL, 38 ASPHERIC (polymacon) soft (hydrophilic) contact lenses for daily wear are available in clear and with a blue visibility tint. The lenses are indicated for visual acuity in non-aphakic patients with non-diseased eyes. The lenses are available in powers of -20.00 up to +10.00 Diopters of and up to 2.50 Diopters of add. These lenses will mask 2.00 Diopters of astigmatism where it does not interfere with visual acuity.
Device Story
Lathe-cut soft (hydrophilic) contact lens; material polymacon (62% polymer, 38% water); aspherical geometry; daily wear. Placed on cornea to act as refracting medium, focusing light on retina. Used by patients for vision correction. Equivalent to existing aspheric polymacon lenses.
Clinical Evidence
No clinical data provided. Equivalence is based on the established safety profile of the polymacon polymer and the similarity of the device to existing aspheric contact lenses produced by eighteen other manufacturers.
Technological Characteristics
Material: polymacon (62% polymer, 38% water); Geometry: aspherical; Manufacturing: lathe-cut; Sterilization: autoclave; Power range: -20.00 to +10.00 Diopters; Add power: up to 2.50 Diopters.
Indications for Use
Indicated for visual acuity correction in non-aphakic patients with non-diseased eyes; corrects up to 2.50 Diopters of add and masks up to 2.00 Diopters of astigmatism.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Related Devices
K964902 — CONTACT LENSES METRO-G 3X (HIOXIFILCON B) · Metro Optics of Austin, Inc. · Mar 11, 1997
K960380 — SEESOFT (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS · Unilens Corp., USA · Mar 28, 1996
K960376 — FREESOFT (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS · Unilens Corp., USA · Mar 28, 1996
K023530 — HORIZON 38 (CLEAR) AND HORIZON 38 WESTINT (TINTED) · Westcon Contact Lens Co., Inc. · Nov 26, 2002
K233221 — Metro Soft, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) · Metro Optics of Austin, Inc. · Feb 6, 2024
Submission Summary (Full Text)
{0}
K 961606
Supplement to K961606 METRO FOCAL, 38 ASPHERIC (polymacon) Soft (Hydroophilic) Contact Lenses for Daily Wear AUG. 28. 1996
NOV 20 1996
SECTION 8 SUMMARY OF SAFETY AND EFFECTIVENESS
I. General Information:
A. Generic (USAN) Name: polymacon
B. Device Trade Name: METRO FOCAL 38 ASPHERIC (polymacon)
C. Sponsor's Name and Address: METRO OPTICS of AUSTIN, Inc.
15802 Vision Drive
Pflugerville, TX 78660
(512) 251-2382
D. 510(k) Number: K961606
II. Indications
The METRO FOCAL, 38 ASPHERIC (polymacon) soft (hydrophilic) contact lenses for daily wear are available in clear and with a blue visibility tint. The lenses are indicated for visual acuity in non-aphakic patients with non-diseased eyes. The lenses are available in powers of -20.00 up to +10.00 Diopters of and up to 2.50 Diopters of add. These lenses will mask 2.00 Diopters of astigmatism where it does not interfere with visual acuity.
III. Device Description
The METRO FOCAL, 38 ASPHERIC (polymacon) soft (hydrophilic) contact lenses are lathe cut into an aspherical contact lens geometry which becomes a hemispherical shell. The polymer, polymacon was the first soft (hydrophilic) material and is one of the most frequently used polymers of HEMA at this time. It consists of 62% polymacon and 38% water by weight when fully hydrated in normal saline solution buffered with sodium bicarbonate.
The METRO FOCAL, ASPHERIC (polymacon) soft (hydrophilic) contact lenses when hydrated and placed on the cornea acts as a refracting medium to focus light rays on the retina.
IV. Alternative Practices or Procedures
Alternative practices or procedures are in the use of available prescription products soft (hydrophilic) contact lenses, RGP or Rigid Gas Permeable contact lenses or spectacles.
V. Technical Summaries
A. Pre Clinical
K961606
{1}
Supplement to K961606 METRO FOCAL 38 ASPHERTC (polymacon) Soft (Hydroophilic) Contact Lenses for Daily Wear AUG. 28, 1996
1. Toxicology
This had been run on the polymer (polymacon) and included in another submission. Results were within normal limits.
2. Microbiology
Included in same submission as above Toxicology results. Results were within normal limits.
3. Compatibility
This had been run on this polymacon polymer in 1990. Results were within normal limits.
4. Stability
This had been run on the polymer polymacon in another submission in 1990. Results were within normal limits. Metro Optics has on-going stability studies validating the expiration date of four years at this time for the clear and a lesser time for the blue visibility tint
5. Preservative Uptake Tests
This had been run on polymer polymacon in another submission in 1990. Results were within normal limits.
B. Clinical
The Sponsor is referencing the fact that there are eighteen Contact Lens companies producing aspheric contact lenses for the correction of presbyopia. Thirteen of these lenses are made from the same polymer, polymacon, as is the lens in question. All these lenses are generated on similar equipment in a similar manner. The equipment manufacturers of lathes, autoclaves, laminar flow units, polishing equipment may be different between any two companies however the end contact lens products are for all intents equivalent.
The manufacturing materials, manufacturing methods, and quality control methods employed by most companies are equivalent. The results are all approximately the same.
This contact lens manufacturing facility has been producing soft (hydrophilic) contact lenses for over fourteen years and was last inspected by the FDA on October 8, 1993
68
{2}
Supplement to K961606 METRO FOCAL 38 ASPHERIC (polymacon) Soft (Hydrophilic) Contact Lenses for Daily Wear AUG. 28. 1996
C. Potential Safety and Effectiveness as well as Adverse Effects of the Device would be about the same as that of any other aspheric polymacon lens.
D. Conclusion drawn from Study
This lens should be the equivalent of any other lens of aspheric configuration.
E. A favorable recommendation from the FDA is anticipated for the manufacture and distribution of aspheric soft (hydrophilic) contact lenses made by Metro Optics of Austin, Inc. Any additional information will be supplied by the individual whose name and address is on the cover letter of this 510(k) Notification.
69
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.