POWER PLUS (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS
Device Facts
| Record ID | K960926 |
|---|---|
| Device Name | POWER PLUS (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS |
| Applicant | Unilens Corp., USA |
| Product Code | LPL · Ophthalmic |
| Decision Date | Apr 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 886.5925 |
| Device Class | Class 2 |
Indications for Use
The Power Plus (polymacon) Soft (Hydrophilic) Aspheric Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes that may exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Device Story
Power Plus contact lens; polymacon (38% water) hydrophilic polymer; front surface aspheric design with multiple aspheric zones and spherical base; center provides most plus power, transitioning to more minus toward periphery; designed for ~1 mm movement on blink to facilitate tear exchange, oxygenation, and removal of corneal metabolic wastes; daily wear modality; prescribed by eye care professionals for refractive error correction; benefits include established material safety and aspheric design for visual correction.
Clinical Evidence
No clinical data provided. Safety and effectiveness are based on the established history of the polymacon material (used since 1972) and the similarity of the aspheric design to multiple legally marketed predicate devices.
Technological Characteristics
Material: Polymacon (2-HEMA), 38% water content when hydrated. Design: Front surface asphere with multiple aspheric zones and spherical base. Power profile: Plus power center, minus power periphery. Movement: Designed for ~1 mm movement on blink for tear exchange.
Indications for Use
Indicated for daily wear correction of myopia and hyperopia in aphakic and non-aphakic patients with non-diseased eyes and astigmatism ≤ 2.00 diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
- Unilens 38 (polymacon) Aspheric Contact Lens (K941836)
- Ideal Optics PS (polymacon) Aspheric Contact Lens (P830012-S003)
- Fulfocus (polymacon) Aspheric Contact Lens (P830006-S005)
- Allvue (polymacon) Aspheric Contact Lens (P840006-S007)
- VX Soft Lens (polymacon) Aspheric Contact Lens (P780013-S004)
Reference Devices
- Unilens (hefilcon A) Soft (Hydrophilic) Aspheric Contact Lens (P850002-S005)
Related Devices
- K960376 — FREESOFT (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS · Unilens Corp., USA · Mar 28, 1996
- K960380 — SEESOFT (POLYMACON) SOFT (HYDROPHILIC) ASPHERIC CONTACT LENS · Unilens Corp., USA · Mar 28, 1996
- K110099 — PROCLEAR TORIC XR SOFT (HYDROPHILIC) CONTACT LENS; PROCLEAR MULTIFOCAL XR SOFT (HYDROPHILIC) CONTACT LENS; PROCLEAR MULT · CooperVision, Inc. · Jun 29, 2011
- K041608 — C-VUE 55 TORIC (METHAFILCON A) MULTIFOCAL CONTACT LENS · Unilens Corp., USA · Jul 7, 2004
- K982904 — ESSTECH MULTI ASPHERIC (MULTIFOCAL) (HIOXIFILCON B) SOFT (HYDROPHILIC) CONTACT LENS · Les Laboratoires Blanchard · Sep 1, 1998