BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens

K220143 · Shenzhen Dashicheng Optical Technology Co., Ltd. · LPL · Sep 8, 2022 · Ophthalmic

Device Facts

Record IDK220143
Device NameBARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens
ApplicantShenzhen Dashicheng Optical Technology Co., Ltd.
Product CodeLPL · Ophthalmic
Decision DateSep 8, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens are indicated for the correction of visual acuity in aphakic and not-aphakic persons with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients. Frequent/Planned Replacement Wear: Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. Disposable Wear: Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Device Story

Spherical, soft (hydrophilic) contact lens; fabricated from polymacon (62% polymer, 38% water); cosmetic tinting via sandwich structure using listed color additives. Acts as refractive medium to focus light on retina; conforms to corneal curvature. Prescribed by eyecare practitioners for daily wear; available for frequent/planned replacement (with cleaning/disinfection) or daily disposable use. Supplied sterile in foil blister packs with saline solution. Benefits include vision correction and cosmetic eye color alteration.

Clinical Evidence

No clinical data. Substantial equivalence established through non-clinical bench testing, including physical, optical, physicochemical, and mechanical performance tests (ISO 18369-2/3/4, ASTM D882-18) and biocompatibility testing (ISO 10993).

Technological Characteristics

Material: Polymacon (HEMA cross-linked with EGDMA). Water content: 38%. Oxygen permeability: 9.01x10^-11 (cm²/s)(ml O₂/(ml·mmHg)) at 35°C. Refractive index: 1.435. Modulus: 0.90 MPa. Diameter: 13.50-14.50 mm. Base curve: 8.00-9.00 mm. Sterilization: Moist heat. Non-ionic, low water content (Group I).

Indications for Use

Indicated for correction of visual acuity in aphakic and non-aphakic persons with myopia and refractive astigmatism of 0.50 diopters or less. Available tinted for cosmetic eye color enhancement.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 8, 2022 Shenzhen Dashicheng Optical Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Shenzhen, Guangdong 518000 China Re: K220143 Trade/Device Name: BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 3, 2022 Received: August 3, 2022 Dear Grace Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation titled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220143 #### Device Name BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens #### Indications for Use (Describe) BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens are indicated for the correction of visual acuity in aphakic and not-aphakic persons with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients. #### Frequent/Planned Replacement Wear: Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. Disposable Wear: Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use. Type of Use (Select one or both, as applicable) | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |------------------------------------------------| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary - K220143 # 1. Contact Details #### 1.1 Applicant information | Applicant Name | Shenzhen Dashicheng Optical Technology Co., Ltd. | |----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address | 101, 201 and 301, Unit 3, Building 1, Wanjiening Industrial<br>Park, No. 1450, Guanguang Road, Jutang Community,<br>Fucheng Sub-district, Longhua District, Shenzhen,<br>Guangdong, China | | Contact person | Jiayi Li | | Phone No. | +86-136 3296 6073 | | E-mail | lijiayi6073@dingtalk.com | | Date Prepared | 2022-01-11 | #### 1.2 Submission Correspondent Image /page/3/Picture/5 description: The image shows a logo with a green circle containing a stylized white tiger head. Above the circle are three pairs of green wings, symmetrically arranged. Below the circle, there are four Chinese characters, also in green, which appear to be the name of the organization or company associated with the logo. | | Shenzhen Joyantech Consulting Co., Ltd. | |-------------------------|-------------------------------------------------------| | | 1713A, 17th Floor, Block A, Zhongguan Times Square, | | | Nanshan District, Shenzhen, Guangdong Province, China | | Phone No. | +86-755-86069197 | | Contact person | Grace Liu; Field Fu; | | Contact person's e-mail | grace@cefda.com; field@cefda.com | | Website | http://www.cefda.com | #### 2. Device Information` | Trade name | BARBIE (polymacon) Daily Wear Soft (Hydrophilic)<br>Contact Lens | |---------------------|------------------------------------------------------------------| | Common name | Soft (Hydrophilic) Contact Lens | | Classification | II | | Classification name | Lenses, Soft Contact, Daily Wear | | Product code | LPL, MVN | | Regulation No. | 21 CFR 886.5925 | #### 3. Legally Marketed Predicate Device | Trade Name | VASSEN COLOR (polymacon) Daily Wear Soft (hydrophilic)<br>Contact Lenses (Clear & Tinted) | |---------------|-------------------------------------------------------------------------------------------| | 510(k) Number | K141699 | | Product Code | LPL, MVN | | Manufacturer | Vassen Co., Ltd. | ## 4. Device Description BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is a spherical lens. It is fabricated from polymacon which has been adopted by the United Stated Adopted {4}------------------------------------------------ Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The nonionic lens material, polymacon, is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) cross-linked with ethylene glycol dimethacrylate (EGDMA), initiated by 2, 2'-azobisisobutyronitrile (AIBN). The lens consists of 62% polymacon and 38% water by weight when immersed in standard saline solution. BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is available cosmetic tinted to enhance or alter the apparent color of the eye. It is tinted in an annular pattern, providing a clear optic zone, using a combination of one or more of the following 'listed' color additives: | Color additive | Listing No. | |----------------------|-----------------| | Reactive black 5 | 21 CFR 73.3121 | | Iron oxides | 21 CFR 73.3125 | | Solvent Yellow 18 | 21 CFR 73.3122 | | Titanium dioxide | 21 CFR 73.3126 | | Carbazole violet | 21 CFR 73.3107 | | Pigment Blue 36 | 21 CFR 73.3110a | | Phthalocyanine green | 21 CFR 73.3124 | The proposed lenses contain only the amount of color additives to accomplish the intended coloring effect. The proposed color lenses are manufactured by sandwiching the pattern layer containing color additives between two layers of lens materials (polymacon). As part of the manufacturing process, the color lenses are thoroughly washed to remove the unbound reactive color additives. Because of the 'sandwich' structure, the color additives will not be removed by lens handling and cleaning/disinfecting procedures. The cosmetic tinting pattern has a minimum clear pupil diameter of 6.0 mm. In the hydrated state, BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus, and it acts as a refracting media to focus light rays on the retina. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution compatible with the eye. BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is available in the spherical configuration with the following features and properties: | Total diameter: | 13.50 mm to 14.50 mm, step:0.10 mm | |--------------------|------------------------------------| | Base curve: | 8.00 mm to 9.00 mm, step: 0.10 mm | | Center thickness: | 0.050 mm to 0.120 mm | | Back vertex power: | -10.00D to 0.00D, step: 0.25D | The physical properties of the proposed lens are: {5}------------------------------------------------ | Refractive index: | 1.435 | |----------------------|------------------------------------------------------------------------| | Light Transmission: | ≥95% (at clear region corresponding to pupil, minimum 6.0 mm diameter) | | Surface character: | Hydrophilic | | Water content: | 38% | | Oxygen permeability: | 9.01×10-11 (cm²/s) [ml O₂/(ml·mmHg)] at 35°C (revised Fatt method) | BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is supplied sterile in the foil blister pack containing standard saline solution. ## 5. Intended Use/Indication for Use BARBIE (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients. Frequent/Planned Replacement Wear: Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. Disposable Wear: Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use. # 6. Substantial Equivalence Comparison | Comparison<br>item | Proposed Device (K220143) | Predicate Device (K141699) | Comment | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Manufacturer | Shenzhen Dashicheng Optical<br>Technology Co., Ltd. | Vassen Co., Ltd. | None | | Product Name | BARBIE (polymacon) Daily Wear<br>Soft (Hydrophilic) Contact Lens | VASSEN COLOR (polymacon)<br>Daily Wear Soft (hydrophilic)<br>Contact Lenses (Clear & Tinted) | None | | Product Code | LPL, MVN | LPL, MVN | Same | | Regulation<br>Number | 21 CFR 886.5925 | 21 CFR 886.5925 | Same | | Classification | Class II | Class II | Same | | Prescription Use | Yes | Yes | Same | | Intended Use /<br>Indications for<br>Use | BARBIE (polymacon) Daily Wear<br>Soft (Hydrophilic) Contact Lens<br>are indicated for the correction of<br>visual acuity in aphakic and<br>not-aphakic persons with<br>non-diseased eyes with myopia.<br>The lens may be worn by<br>persons who exhibit refractive<br>astigmatism of 0.50 diopters or<br>less where the astigmatism does<br>not interfere with visual acuity.<br>The lens is available tinted and<br>used to enhance or alter the<br>apparent color of the eye.<br>Daily wear replacement<br>schedules may vary from patient<br>to patient and should be decided<br>by eyecare practitioners in<br>consultation with their patients.<br>Frequent/Planned Replacement<br>Wear:<br>Eyecare practitioners may<br>prescribe any of the above lenses<br>for frequent/planned replacement<br>wear, with cleaning, disinfection<br>and scheduled replacement.<br>When prescribed for<br>frequent/planned replacement<br>wear, the lens may be disinfected<br>using a chemical disinfecting<br>system.<br>Disposable Wear:<br>Eyecare practitioners may<br>prescribe any of the above lenses<br>for Daily Disposable Wear. When<br>prescribed for Daily Disposable<br>Wear, the lenses are not to be<br>used with disinfecting systems as<br>they are to be discarded after a<br>single use. | The VASSEN COLOR<br>(polymacon) Spherical Soft<br>Contact Lenses for daily wear are<br>indicated for the correction of<br>visual acuity in aphakic and not<br>aphakic persons with<br>non-diseased eyes with myopia<br>or hyperopia. The lens may be<br>worn by persons who exhibit<br>refractive astigmatism of 0.50<br>diopters or less where the<br>astigmatism does not interfere<br>with visual acuity. The lens is<br>available clear or tinted and may<br>be used to enhance or alter the<br>apparent color of the eye.<br>Daily wear replacement<br>schedules may vary from patient<br>to patient and should be decided<br>by eyecare practitioners in<br>consultation with their patients.<br>Frequent/Planned Replacement<br>Wear:<br>Eyecare practitioners may<br>prescribe any of the above lenses<br>for frequent/planned replacement<br>wear, with cleaning disinfection<br>and scheduled replacement.<br>When prescribed for<br>frequent/planned replacement<br>wear, the lens may be disinfected<br>using a chemical disinfecting<br>system.<br>Disposable Wear:<br>Eyecare practitioners may<br>prescribe any of the above lenses<br>for Daily Disposable Wear. When<br>prescribed for Daily Disposable<br>Wear the lenses are not to be<br>used with disinfecting systems as | Similar | | | | they are to be discarded after a single use. | | | Functionality | The contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina. | The contact lenses act as a refractive medium that focus light rays from near and distant objects on the retina. | Same | | Lens Design | Spherical | Spherical | Same | | Manufacturing<br>Process | Fully-molded | Fully-molded | Same | | Lens Material | Polymacon | Polymacon | Same | | Material<br>Classification | Group I<br>Low water content (<50%),<br>Nonionic | Group I<br>Low water content (<50%),<br>Nonionic | Same | | Color Additives | Reactive Black 5<br>Iron Oxides<br>Solvent Yellow 18<br>Titanium Dioxide<br>Carbazole Violet<br>Pigment Blue 36<br>Phthalocyanine Green | C.I. Reactive black 5<br>Iron Oxide (Red)<br>D&C Yellow 10<br>Titanium Dioxide<br>D&C Green 6<br>C.I. Reactive Blue 19 | Different | | Storage Solution | Standard saline solution | Standard saline solution | Same | | Primary<br>Packaging | Foil blister pack | Foil blister pack | Same | | Total Diameter | 13.50 mm to 14.50 mm | 13.00 mm to 15.00 mm | Similar | | Base Curve | 8.00 mm to 9.00 mm | 8.0 mm to 9.0 mm | Same | | Center<br>Thickness<br>(varies with<br>power) | 0.050 mm to 0.120 mm | 0.06 mm to 0.17 mm | Similar | | Power Range | -10.00D to 0.00D | -10.00D to +3.00D | Similar | | Water Content | 38% ±2 % | 38% ±2 % | Same | | Light<br>Transmission | ≥95% | ≥95% | Same | | Refraction Index | 1.435 | 1.439…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...