Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens

K213216 · Unicon Optical Co., Ltd. · LPL · Jan 18, 2022 · Ophthalmic

Device Facts

Record IDK213216
Device NameQualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens
ApplicantUnicon Optical Co., Ltd.
Product CodeLPL · Ophthalmic
Decision DateJan 18, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and 1.00 diopter (D) or less of astigmatism. Eye Care Professionals may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, the lens is to be discarded after each removal. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical disinfection system only.

Device Story

Qualis is a soft, daily-wear contact lens fabricated from linofilcon A (silicone-containing monomer/hydrophilic monomer copolymer). It acts as a refractive medium to focus light on the retina. The lens is a hemispherical shell with a molded spherical base curve and front surface. It contains a blue handling tint and a benzotriazole UV-blocking monomer. It is prescribed by eye care professionals for either single-use disposable wear or frequent/planned replacement wear (requiring chemical disinfection). The device benefits patients by providing optical correction for ametropia. It is used by the patient in a daily-wear context.

Clinical Evidence

Open-label, multi-center, randomized concurrent-control study with 91-day follow-up. 75 subjects enrolled (50 test, 25 control). Primary endpoints included biomicroscopy findings, symptoms, and vision safety measures. 73 subjects completed the study. No serious adverse events reported. Results established substantial equivalence in clinical performance between Qualis and Acuvue Vita.

Technological Characteristics

Material: linofilcon A (62% polymer, 38% water). Design: Spherical, cast-molded. UV protection: Benzotriazole monomer (blocks >95% UVB, >70% UVA). Oxygen permeability: 100 (x 10^-11). Refractive index: 1.415 (hydrated). Packaging: Injection-molded polypropylene blister. Biocompatibility testing per ISO 10993 series.

Indications for Use

Indicated for correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and ≤ 1.00 D astigmatism.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 18, 2022 Unicon Optical Co., Ltd. % Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Lane West Linn, OR 97068 Re: K213216 Trade/Device Name: Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: December 6, 2021 Received: December 9, 2021 Dear Bret Andre: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) not yet assigned ### Device Name Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens ### Indications for Use (Describe) The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and 1.00 diopter (D) or less of astigmatism. Eye Care Professionals may prescribe the for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, the lens is to be discarded after each removal. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical disinfection system only. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Premarket Notification # 510 (k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ## The assigned 510(k) number is: ### SUBMITTER I. | Date Prepared: | September 24th, 2021 | |-----------------|--------------------------------------------------------------------------------------------------------------------------| | Name: | UNICON OPTICAL CO., LTD. | | Address: | No.16, Gongye E. 9th Rd.<br>Hsinchu Science Park, Baoshan Township<br>Hsinchu County Hsinchu – District,<br>TAIWAN 30075 | | Contact Person: | Nick Liao<br>Director | | Phone number: | +886-4-23551233 | | Consultant: | Bret Andre<br>EyeReg Consulting, Inc.<br>6119 Canter Ln.<br>West Linn, OR 97068 | (503) 372-5226 ### II. DEVICE Phone number: | Trade Name: | Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens | |-------------------------|------------------------------------------------------------------| | Common<br>Name: | Contact Lens, Daily Wear | | Classification<br>Name: | Soft (hydrophilic) Contact Lens (21 CFR 886.5925) | | Regulatory<br>Class: | Class II | | Product Codes: | LPL; MVN | ## Purpose of 510(k) Submission: ~ New Device ~ {4}------------------------------------------------ 510(k) Premarket Notification #### III. PREDICATE DEVICE The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is substantially equivalent to the following predicate device(s): - "ACUVUE® VITA™ (senofilcon C) Brand (Soft) Contact Lens" I By Johnson & Johnson Vision Care, Inc. 510(k) number; K160212 Primary Predicate ### IV. DEVICE DESCRIPTION The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is a hemispherical shell with a molded spherical base curve and a molded front surface. The hydrophilic characteristics allow aqueous solutions to enter the lens. The lenses are fabricated from linofilcon A, which is a random co-polymer of silicone containing monomers and hydrophilic monomers. The lens consists of 62.0% linofilcon A and 38.0% water by weight when immersed in saline solution. The linofilcon A name has been adopted by the United States Adopted Names Council (USAN). The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens contains C.I. Reactive Blue No. 19 (21 CFR Part 73.3127) for visibility and handling. The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens incorporates a benzotriazole UV blocking monomer to help protect against transmission of harmful UV radiation. The lens blocks >95% in the UVB range (280mm -315nm), and >70% in the UVA range (316nm - 380nm). The Oualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is manufactured in a spherical design configuration. The material properties and available parameters of the finished lenses are as follows: | Parameter | Range | Tolerance | |-------------------------------------------------------------|---------------------------|---------------------------------------------------------------------------------| | Chord Diameter | 13.00 mm to 15.00 mm | ±0.20 mm | | Center Thickness | 0.080 mm @ -3.00 D | When ≤ 0.10 mm → ±0.010 mm + 10%<br>When > 0.10 mm → ±0.015 mm + 5% | | Base Curve | 8.00 mm to 9.50 mm | ±0.20 mm | | Back Vertex Power (F'v) | +20.00D to -20.00D | When 0.00 < F'v ≤ 10.00 D → ±0.25 D<br>When 10.00 < F'v ≤ 20.00 D → ±0.50 D | | Surface Appearance | - | Lenses should be clear with no surface defect | | Oxygen Permeability<br>(x 10-11(cm²/sec)(mlO2)/(ml x mmHg)) | 100 | ±20% | | Light Transmittance - Tinted<br>(@ 380-780nm) | 95% | ±5% | | Ultraviolet Radiation Transmittance | < 5 % TUVB<br>< 30 % TUVA | TUVB (280 to 315 nm) < 0.05TV<br>TUVA (316 to 380 nm) < 0.50TV | | Water Content | 38% | ±2% | | Refractive Index | 1.415 (hydrated) | ±0.005 | {5}------------------------------------------------ 510(k) Premarket Notification ### V. INDICATIONS FOR USE The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and 1.00 diopter (D) or less of astigmatism. Eye Care Professionals may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, the lens is to be discarded after each removal. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical disinfection system only. ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE VI. DEVICE The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is substantially equivalent to the primary predicate device identified (K160212) regarding the following features: - ' FDA category - Group V - I FDA classification - Soft (hydrophilic) Contact Lens (21 CFR 886.5925) - I Intended use - daily wear contact lenses - 트 Actions - I Indications for use - 트 Design configurations available (spherical) - I Cast molded production method - I Injection molded polypropylene blister packaging {6}------------------------------------------------ 510(k) Premarket Notification The following matrix illustrates the production method, lens function and material characteristics of the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens, as well as the predicate device. | | Unicon<br>Qualis<br>(Subject Device) | Johnson & Johnson<br>Vision Care<br>Acuvue Vita<br>(K160212) | |---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Daily wear,<br>Soft (hydrophilic) contact lens | Daily wear,<br>Soft (hydrophilic) contact lens | | Actions | In its hydrated state, the contact lens, when<br>placed on the cornea, acts as a refracting<br>medium to focus light rays on the retina | In its hydrated state, the contact lens, when<br>placed on the cornea, acts as a refracting medium<br>to focus light rays on the retina | | FDA Classification | Soft (hydrophilic) Contact Lens (21 CFR<br>886.5925) | Soft (hydrophilic) Contact Lens (21 CFR<br>886.5925) | | FDA Group | FDA Group V | FDA Group V | | Production Method | Fully molded | Fully molded | | USAN name | linofilcon A | senofilcon C | | Water Content (%) | 38±2% | 41±2% | | Oxygen Permeability<br>x 10-11 (cm²/sec)(mlO2)/(ml x mmHg @<br>35°C)) (revised Fatt method) | 100 (edge-corrected) | 103 (edge-corrected) | | Refractive Index<br>(hydrated) | 1.415 | 1.420 | | UV Blocker | Yes - benzotriazole | Yes - benzotriazole | | Manufacturing | Cast-Molded | Cast-Molded | | Color | Blue Handling Tint<br>C.I. Reactive Blue No. 19 | Blue Handling Tint<br>Reactive Blue Dye #4 | {7}------------------------------------------------ 510(k) Premarket Notification The following matrix compares the indications for use of the Qualis (Linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens with the predicate device. | | Indications for Use | |-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Unicon<br>Qualis<br>(Subject<br>Device) | The Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is indicated for the<br>correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-<br>diseased eyes and 1.00 diopter (D) or less of astigmatism.<br><br>Eye Care Professionals may prescribe the lenses either for single-use disposable wear or<br>frequent/planned replacement wear with cleaning, disinfection and scheduled replacement.<br>When prescribed for daily disposable wear, the lens is to be discarded after each removal.<br>When prescribed for frequent/planned replacement wear, the lens may be cleaned and<br>disinfected using a chemical disinfection system only. | | Johnson &<br>Johnson<br>Vision Care<br>Acuvue Vita<br>(K160212) | ACUVUE® (senofilcon C) Soft (hydrophilic) Contact Lens (spherical) is indicated for daily<br>wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or<br>aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.<br><br>ACUVUE® (senofilcon C) Soft (hydrophilic) Contact Lens – TORIC is indicated for daily<br>wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or<br>aphakic persons with non-diseased eyes who may have 10.00D or less of astigmatism.<br><br>ACUVUE® (senofilcon C) Soft (hydrophilic) Contact Lens – MULTIFOCAL is indicated for<br>daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or<br>presbyopia in phakic or aphakic persons with non-diseased eyes who may need up to 4.00D of<br>ADD power and may have 0.75D or less of astigmatism.<br><br>ACUVUE® (senofilcon C) Soft (hydrophilic) Contact Lens – MULTIFOCAL-TORIC is<br>indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia<br>and/or astigmatism) and presbyopia in phakic or aphakic persons with non-diseased eyes who<br>may need up to 4.00D of ADD power and may have 10.00D or less of astigmatism.<br><br>These lenses contain a UV Blocker to help protect against transmission of harmful UV<br>radiation to the cornea and into the eye.<br><br>Eye Care Professionals may prescribe the lenses either for daily disposable wear or<br>frequent/planned replacement wear with cleaning, disinfection and scheduled replacement.<br>When prescribed for daily disposable wear, lenses should be discarded upon removal. When<br>prescribed for frequent/planned replacement wear, the lenses may be disinfected using a<br>chemical disinfection system only. The contact lens is to be discarded after the recommended<br>wearing period as prescribed by the Eye Care Professional. | {8}------------------------------------------------ 510(k) Premarket Notification ### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. ## Non-Clinical Performance Testing A series of non-clinical performance tests were performed to demonstrate the safety and effectiveness of the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens. The results support the claim that the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is substantially equivalent to the currently marketed predicate device. A summary of the results from the non-clinical studies is presented below. ## Toxicology: All biocompatibility/toxicology tests were conducted in accordance with the GLP regulation. - · In-Vitro Cytotoxicity: Cytotoxicity testing was performed in accordance with ISO 10993-5 with results indicating that the finished lenses and packaging materials are non-toxic. - · Systemic Toxicity: The finished lenses and packaging materials meet the requirements of the systemic injection test in accordance with ISO 10993-11 and are considered nontoxic. - · Acute Ocular Irritation: Acute ocular irritation testing was performed in accordance with ISO 10993-10, and the extracts from finished lenses and packaging materials produced no ocular irritation. - · Skin Sensitization Study (Maximization Test): The skin sensitization study was conducted on the finished lenses in accordance with ISO 10993-10, and the contact lens extracts did not produce skin sensitization. - · 22-Day Ocular Irritation: The 22-day ocular irritation test was conducted in accordance with ISO 9394 on the finished lenses, and the contact lens extracts produced no ocular irritation. ## Shelf Life: Testing was performed to evaluate the stability, sterility, and package integrity of the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens over the duration of the labeled expiration date. The data presented supports establishment of the proposed shelf life. ## Solution Compatibility The physical compatibility of Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lenses with commonly available cleaning and disinfection solutions (peroxide and MPDS) was confirmed following the methodology described in ISO 11981:2017, Ophthalmic optics -Contact lenses and contact lens care products - Determination of physical compatibility of contact lens care products with contact lenses. {9}------------------------------------------------ 510(k) Premarket Notification ## Preservative Uptake and Release Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lenses were analyzed for uptake and release of preservatives found in lens care products. Testing was conducted according to ISO 11986:2017, Ophthalmic optics - Contact lenses and contact lens care products - Determination of preservative uptake and release. Uptake and release profiles of linofilcon A contact lenses for polyaminopropyl biguanide (PAPB) and Polyquaternium-1 (PQ-1) demonstrate sub-detection limit amounts of release at each time point evaluated. ## Performance Testing - Bench: The following bench tests were completed: refractive index, water content, Dk, % transmission (visible & UV), tensile strength, modulus, % elongation to break, specific gravity and quantification of polymerization residuals. Results of physicochemical and mechanical property testing demonstrate consistent material properties between the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens and the predicate device. ## Clinical Testing An open-label. multi-center. randomized concurrent-control study with 91-day treatment follow-up was completed. Seventy-five (75) subjects were enrolled in the study-of which 50 subjects wore the test lenses (Qualis (linofilcon A) Silicone Hydrogel Soft Contact Lens for Daily Wear) and 25 subjects wore the control lenses (Acuvue Vita (senofilcon C) Monthly Contact Lens). Seventy-three (73) of the 75 enrolled subjects completed the study. No serious adverse events were reported. The study results establish substantially equivalent clinical performance between the test (Qualis) and control (Acuve Vita) lenses with respect to biomicroscopy findings, symptoms and vision safety measures; therefore, this clinical investigation supports the claim of substantial equivalence between the two lens types with regard to clinical safety and effectiveness. {10}------------------------------------------------ 510(k) Premarket Notification ## VIII. CONCLUSIONS ### Validity of Scientific Data Laboratories under Good Laboratory Practice regulations conducted toxicology, microbiology, and shelf-life stability studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7. ### Substantial Equivalence Information presented in this Premarket Notification establishes that the Qualis (linofilcon A) Soft (Hydrophilic) Daily Wear Contact Lens is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication. ### Risks and Benefits The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) daily wear contact lenses. The benefits to the patient are the same as those for other soft (hydrophilic) daily wear contact lenses.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...