MyDay (Stenfilcon A) ASPHERE, MyDay (Stenfilcon A) TORIC, MyDay (Stenfilcon A) MULTIFOCAL, MyDay (Stenfilcon A) MULTIFOCAL TORIC
K191763 · CooperVision, Inc. · LPL · Mar 23, 2020 · Ophthalmic
Device Facts
Record ID
K191763
Device Name
MyDay (Stenfilcon A) ASPHERE, MyDay (Stenfilcon A) TORIC, MyDay (Stenfilcon A) MULTIFOCAL, MyDay (Stenfilcon A) MULTIFOCAL TORIC
Applicant
CooperVision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Mar 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity. MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity. MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.
Device Story
Daily disposable soft contact lens; material stenfilcon A (copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide). Functions as refracting medium on cornea to focus light on retina. Available in Asphere, Toric, Multifocal, and Multifocal Toric designs. Prescribed by eye care professionals for daily wear. Includes 100 ppm Epsilon-Polylysine (e-PL) in blister packaging solution as preservative. Lens features blue visibility tint and UV-absorbing monomer. Benefits include vision correction for ametropia and presbyopia.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including physicochemical properties (water content, refractive index, light transmittance, mechanical properties per ISO/ANSI standards), biocompatibility (cytotoxicity, systemic toxicity, sensitization, irritation per ISO 10993), and preservative efficacy testing (ISO 14730).
Technological Characteristics
Material: stenfilcon A (silicone hydrogel). Manufacturing: Molded. Sterilization: Moist heat. Packaging: Polypropylene blisters with aluminum foil. Additives: 100 ppm Epsilon-Polylysine (preservative), benzotriazolyl methacrylate (UV blocker), Reactive Blue #246 (tint). Oxygen permeability: 80x10-11 (cm²/sec)x(ml O2)/(ml x mm Hg). Water content: 54%. Refractive index: 1.401.
Indications for Use
Indicated for correction of ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic and non-aphakic persons with non-diseased eyes. Powers range -20.00 to +20.00D; astigmatic correction up to -10.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens (K131378)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
March 23, 2020
CooperVision, Inc. Ms. Marie Dutton Senior Regulatory Affairs Specialist 5870 Stoneridge Drive, Suite 1 Pleasanton, CA 94588
Re: K191763
Trade/Device Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 24, 2020 Received: February 25, 2020
Dear Ms. Marie Dutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidanceregulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-devicesafety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-andconsumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191763
Device Name
MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens
### Indications for Use (Describe)
MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuty.
MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I. SUBMITTER:
CooperVision, Inc. 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588
## Contact Person:
Marie Dutton Senior Regulatory Affairs Specialist CooperVision, Inc. 5870 Stoneridge Drive. Suite 1 Pleasanton, CA 94588 Phone: Phone: (925) 251-6645 Fax: (925) 251-6643 MDutton@coopervision.com E-mail:
## Date Prepared:
June 28, 2019
#### II. DEVICE:
Trade Name: MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens Common Name: Soft (hydrophilic) Contact Lens Classification Name: Lens, Contact, (Disposable) [21 CFR 886.5925 (b) (1)] Regulatory Class: II LPL, MVN Product Code: Classification Panel: Ophthalmic
#### III. PREDICATE DEVICE:
The predicate device is CooperVision's MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens, submitted as SUS (stenfilcon A) Soft (hydrophilic) Contact Lenses for Single Use Daily Wear under 510(k) K131378 and received FDA clearance on August 30, 2013.
#### IV. DEVICE DESCRIPTION:
The device description for the subject device and the predicate device is identical.
MyDay (stenfilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens is available as an Asphere, Toric, Multifocal, and Multifocal Toric lens designs.
The MyDay material, stenfilcon A, is primarily a random copolymer of polydimethylsiloxane methacrylate and vinylmethyl acetamide. The UV blocker used is a benzotriazolyl methacrylate. The lenses have a blue tint which is added to make the lens more visible for handling. The lenses also contain a UV absorbing monomer which is used to block UV radiation.
{4}------------------------------------------------
When placed on the cornea in its hydrated state, the MyDay Soft (Hydrophilic) Contact Lens acts as a refracting medium to focus light rays on the retina.
MyDay (stenfilcon A) contact lens parameters are:
| • Chord Diameter: | 13.0 mm to 15.5mm |
|---------------------|-----------------------------------------|
| • Base Curve: | 8.4 ± 0.5 mm and 8.7 ± 0.5 mm |
| • Center Thickness: | 0.08 mm to 0.218 mm (varies with power) |
| • Powers: | -20.00D to +20.00D |
| • Cylinder Powers: | -0.25D to -10.00D |
| • Axis: | 0° to 180° in 10° increments |
| • Add Power Range: | +0.50 to +4.00 |
The physical/optical properties of the lens are:
| • | Specific Gravity: | 1.033 |
|---|----------------------|-------------------------------------------|
| • | Refractive Index: | 1.401 |
| • | Light Transmittance: | 96% |
| • | Surface Character: | Hydrophilic |
| • | Water Content: | 54% |
| • | Oxygen Permeability: | 80x10-11 [(cm²/sec)x(ml O2)/(ml x mm Hg)] |
#### V. INDICATIONS FOR USE:
There is no difference between the subject device and the predicate device with respect to indications and intended use.
MyDay (stenfilcon A) ASPHERE Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.
MyDay (stenfilcon A) MULTIFOCAL Soft Contact lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of -2.00 diopters or less that does not interfere with visual acuity.
MyDay (stenfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optical correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes who may have -10.00 diopters of astigmatism or less.
{5}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The technological principle for both the subject and the predicate device is identical. The subject and predicate device are based on the same technological elements. The only technological difference that exists between the subject and predicate device is the addition of 100 ppm Epsilon-Polylysine (also known as epolylysine or e-PL) to the blister packing solution formulation.
The e-PL is being added to the packaging solution as a preservative. Changes to the labeling are being made to reflect the addition of e-PL to the packaging solution.
Technology/Material Comparison Predicate Device Subject Device Product Name MyDay (stenfilcon A) Soft MyDay (stenfilcon A) Soft (Hydrophilic) Daily (Hydrophilic) Daily Disposable Contact Lens Disposable Contact Lens (with e-PL in the packaging solution) Material USAN Name stenfilcon A Same 510(k) Number K131378 TBD - Current Submission FDA Category (Group) Silicone Hydrogel Same Manufacturing Method Molded Same Sterilization Moist Heat Same Packaging Materials Injection molded Same polypropylene blisters covered by aluminum foil laminate; blister strips are packed into printed cartons Packaging Solution Phosphate Buffered Saline Phosphate Buffered Saline Solution with Tween Solution with Tween and 100 ppm e-PL Visibility Tint Reactive Blue #246 (RB246) Same UV Blocker Norbloc Same
The technological characteristics of the subject device and the predicate device are compared in the table below.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA:
Results from non-clinical studies were provided in support of the substantial equivalence determination.
### Performance testing - bench:
In accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, issued on May 12, 1994 and amended June 28, 1994, the following battery of physicochemical testing was performed. All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols. Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated:
- Water content per ISO18369-2:2012 and ISO 18369-4:2006 ●
- Refractive index per ISO 18369-2:2012 and ISO 16369-4:2006 ●
- Light transmittance per ISO 18369-2:2012 and ISO 18369-3:2006
- Non-polymeric residuals in lens and packaging solution per ISO 18369o 4:2006
- Total extractable per ANSI Z80.20-2010 and ISO 18369-4:2006 ●
- o Contact angle per ANSI Z80.20-2010
- Mechanical properties per ANSI Z80.20-2010 and ASTM D1708-02a .
Additionally, preservative efficacy testing (PET) per ISO 14730:2014, analytical testing (uptake and release) per ISO 11986:2010, and in-vitro microbiological testing (lens handling) were performed to support the function of e-PL as that of a preservative.
### Biocompatibility testing:
In accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, issued on May 12, 1994 and amended June 28, 1994 and with the GLP regulation (21 CFR Part 58), the following battery of biocompatibility testing was performed:
- Cytotoxicity per ISO 10993-5:2009 ●
- Systemic Toxicity per ISO 10993-11:2006 ●
- Sensitization per ISO 10993-10:2010 ●
- Irritation per ISO 10993-10:2010 ●
- o Irritation (22 Day) per ISO 9394:2012
{7}------------------------------------------------
## Clinical testing:
Data provided in this 510(k) is sufficient to adequately characterize the subject device in terms of its physical/mechanical/optical and toxicological performance characteristics when compared to the predicate device. The results are equivalent, determining that additional clinical performance data was not required to complete the substantial equivalence determination. Additionally, the technical characteristics and manufacturing and sterilization processes of the subject lens are equivalent to MyDay (stenfilcon A) contact lens currently marketed by CooperVision; therefore, it was confirmed that no clinical data is required.
#### VIII. CONCLUSIONS:
Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physical/mechanical/optical properties of the lens. Based on the performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). The performance testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.