← Product Code [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL) · K191121

# Apioc (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc P for Presbyopia (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc A for Astigmatism (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens (K191121)

_Lentechs, LLC · LPL · Sep 6, 2019 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K191121

## Device Facts

- **Applicant:** Lentechs, LLC
- **Product Code:** [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL.md)
- **Decision Date:** Sep 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Apioc, sphere/aspheric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity. The Apioc A, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters. The Apioc P, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 0.75 diopters and are presbyopic requiring add power of up to +4.00 diopters. The Apioc AP, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters and are presbyopic requiring add power of up to +4.00 diopters. Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

## Device Story

Soft silicone hydrogel contact lenses (efrofilcon A) designed for daily wear; intended for refractive error correction (myopia, hyperopia, astigmatism, presbyopia). Lenses feature lenticular zone for upper eyelid attachment; fabricated via lathe-cut method. Material is non-ionic, 74% water content, includes UV-blocking monomer. Prescribed by eyecare practitioners for frequent/planned replacement; requires patient-performed cleaning/disinfection using chemical or hydrogen peroxide systems. Lenses act as refractive medium to focus light on retina. Benefits include vision correction comparable to existing soft contact lenses.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including stability, lens design verification, and manufacturing parameter compliance.

## Technological Characteristics

Material: efrofilcon A (silicone hydrogel, non-ionic, 74% water). UV blocker: Benzophenone monomer. Production: Lathe cut. Oxygen permeability: 59.8 x 10^-11. Refractive index: 1.380. Standards: ANSI Z80.20 (2016). Sterilization: Sterile in buffered saline vials.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- IntelliWave 3, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) ([K100221](/device/K100221.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 6, 2019

Lentechs, LLC % Mr. Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Lane West Linn, OR 97068

Re: K191121

Trade/Device Name: Apioc (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc P for Presbyopia (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc A for Astigmatism (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens, Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: July 9, 2019 Received: July 10, 2019

Dear Mr. Andre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K191121

### Device Name

Apioc (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens: Apioc P for Presbyonia (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens; Apioc A for Astignatism (efrofilic) Silicone Hydrogel Contact Lens; Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens

### Indications for Use (Describe)

The Apioc, sphere/aspheric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity.

The Apioc A, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

The Apioc P, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myperopia) and presbyopia in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractism not exceeding 0.75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The Apioc AP, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters and are presbyopic requiring add power of up to +4.00 diopters.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

| Type of Use (Select one or both, as applicable)                                                                                    |                                                                                                                                   |
|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">      ❌     </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <span style="font-size: 16px;">      ❍     </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### The assigned 510(k) number is: K191121

#### I. SUBMITTER

- August 28th, 2019 Date Prepared: Lentechs, LLC Name:
- Address: 1275 Kinnear Road, Columbus, Ohio 43212 United States Contact Person: Mr. Robin Sears President and CEO Phone number: (614) 487-3700 Consultant: Bret Andre
- EyeReg Consulting, Inc. 6119 Canter Ln. West Linn, OR 97068 Phone number: (503) 372-5226

#### II. DEVICE

| Trade Name:          | Apioc (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens<br>Apioc P for Presbyopia (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel<br>Contact Lens<br>Apioc A for Astigmatism (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel<br>Contact Lens<br>Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lens |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Contact Lens, Daily Wear                                                                                                                                                                                                                                                                                                                         |
| Classification Name: | Soft (hydrophilic) Contact Lens (21 CFR 886.5925)                                                                                                                                                                                                                                                                                                |
| Regulatory Class:    | Class II                                                                                                                                                                                                                                                                                                                                         |
| Product Code:        | LPL                                                                                                                                                                                                                                                                                                                                              |

{4}------------------------------------------------

#### III. PREDICATE DEVICE

The Apioc, Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are substantially equivalent to the following predicate device:

- I "IntelliWave 3, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A)" Art Optical Contact Lens, Inc. 510(k) number; K100221 Primary Predicate
#### IV. DEVICE DESCRIPTION

The Apioc , Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are designed with a lenticular zone for attachment to the upper eyelid. The contact lenses are fabricated from (efrofilcon A), which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The non-ionic lens material, (efrofilcon A) is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The material consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).

In the Apioc, Apioc P, Apioc A, and Apioc AP Contact Lens with UV Blocker and center thickness of 0.18mm. a Benzophenone UV absorbing monomer is used to block UV radiation. The UV blocking for lenses averages > 98% in the UVB range of 280nm - 315nm and >70% in the UVA range of 316 -380nm.

The Apioc, Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses will be manufactured in the sphere, aspheric, toric, multifocal, and multifocal toric design configurations. The material properties and available parameters of the finished lenses are as follows:

| Parameter               | Range                                  | Tolerance*                                                                                             |
|-------------------------|----------------------------------------|--------------------------------------------------------------------------------------------------------|
| Chord Diameter          | 9.00 mm to 16.00 mm                    | ±0.25 mm                                                                                               |
| Center Thickness        | 0.01 mm to 0.50 mm                     | When ≤ 0.10 mm → ±0.010 mm + 10%<br>When > 0.10 mm → ±0.015 mm + 5%                                    |
| Base Curve              | 7.00 mm to 10.0 mm                     | ±0.25 mm                                                                                               |
| Back Vertex Power (F'v) | +20.00D to -20.00D<br>(in 0.25D steps) | When 0.00 <  F'v  ≤ 10.00 D → ±0.25 D<br>When 10.00 <  F'v  ≤ 20.00 D → ±0.50 D                        |
| Cylinder Power (F'c)    | -0.25D to -10.00D<br>(in 0.25D steps)  | When 0.00 <  F'c  ≤ 2.00 D → ±0.25 D<br>When 2.00 <  F'c  ≤ 4.00 D → ±0.37 D<br>F'c ≤ 4.00 D → ±0.50 D |
| Cylinder Axis           | 10° to 180°<br>(in 10° steps)          | When 0.00 <  F'c  ≤ 1.50 D → ± 8°<br>When  F'c  > 1.50 D → ± 5°                                        |

{5}------------------------------------------------

| Multifocal Add                                                | Up to +4.00 D (in 0.50D steps) | ±0.37 D                                                        |
|---------------------------------------------------------------|--------------------------------|----------------------------------------------------------------|
| Surface Appearance                                            | -                              | Lenses should be clear with no surface defect                  |
| Oxygen Permeability<br>(x 10-11(cm²/sec)(mlO2)/(ml<br>x hPa)) | 59.8                           | ±20%                                                           |
| Light Transmission - Tinted<br>(@ 380-780nm)                  | >97%                           | ±5%                                                            |
| Ultraviolet Radiation<br>Transmittance                        | <2% TUVB<br><30% TUVA          | TUVB (280 to 315 nm) < 0.05TV<br>TUVA (316 to 380 nm) < 0.50TV |
| Water Content                                                 | 74%                            | ±2%                                                            |
| Refractive Index                                              | 1.380 (hydrated)               | ±0.005                                                         |

* ANSI Z80.20, Ophthalmics – Contact Lenses – Standard Terminology, Tolerances, Measurements And Physicochemical Properties (2016)

The lens is supplied sterile in vials containing a buffered saline solution. Vial labeling is printed with appropriate lot numbering, expiration dating and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.

### V. INDICATIONS FOR USE

The Apioc, sphere/aspheric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of 0.75 diopters or less where the astigmatism does not interfere with visual acuity.

The Apioc A, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.

The Apioc P, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 0.75 diopters and are presbyopic requiring add power of up to +4.00 diopters.

The Apioc AP, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) and presbyopia in aphakic and nonaphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters and are presbyopic requiring add power of up to +4.00 diopters.

Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.

{6}------------------------------------------------

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The Apioc , Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are substantially equivalent to IntelliWave 3, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) (cleared under K100221) in terms of the following:

- FDA classification (21 CFR 886.5925)
- 트 FDA group 2 (>50% H2O, non-ionic polymer)
- USAN contact lens material (efrofilcon A)
- 트 Intended use - daily wear contact lenses
- 트 Indications for use
- 트 Actions
- Lathe cut production method

The Apioc , Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are designed with aspheric optics and a lenticular zone for attachment to the upper eyelid, which differs from the predicate device design (cleared under K100221). These differences in the dimensional parameters of the lens are non-significant and will not affect the safety and effectiveness of the new device.

The following matrix illustrates the production method, lens function and material characteristics of the Apioc , Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses, as well as the predicate device.

|                     | Lentechs, LLC<br>Apioc , Apioc P, Apioc A, and Apioc<br>AP (efrofilcon A) Soft (hydrophilic)<br>Silicone Hydrogel Contact Lenses<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Art Optical Contact Lens, Inc.<br>IntelliWave 3, Silicone Hydrogel Daily<br>Wear Soft Contact Lens (efrofilcon A)<br>(Predicate)<br>510(k) K100221                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Apioc, sphere/aspheric (efrofilcon A)<br>Silicone Hydrogel Soft Contact Lenses for<br>daily wear are indicated for the correction of<br>refractive ametropia (myopia and hyperopia) in<br>aphakic and non-aphakic persons with non-<br>diseased eyes. The lens may be worn by<br>persons who exhibit refractive astigmatism of<br>0.75 diopters or less where the astigmatism<br>does not interfere with visual acuity.<br>The Apioc A, toric (efrofilcon A) Silicone<br>Hydrogel Soft Contact Lenses for daily wear<br>are indicated for the correction of refractive<br>ametropia (myopia and hyperopia) in aphakic<br>and non-aphakic persons with non-diseased<br>eyes and/or possesses refractive astigmatism<br>not exceeding 12 diopters.<br>The Apioc P, multifocal (efrofilcon A) Silicone<br>Hydrogel Soft Contact Lenses for daily wear<br>are indicated for the correction of refractive<br>ametropia (myopia and hyperopia) and<br>presbyopia in aphakic and non-aphakic<br>persons with non-diseased eyes and/or<br>possesses refractive astigmatism not<br>exceeding 0.75 diopters and are presbyopic<br>requiring add power of up to +4.00 diopters. | The IntelliWave 3, sphere (efrofilcon A)<br>Silicone Hydrogel Soft Contact Lenses for<br>daily wear are indicated for the correction of<br>ametropia (myopia and hyperopia) in aphakic<br>and non-aphakic persons with non-diseased<br>eyes. The lens may be worn by persons who<br>exhibit refractive astigmatism of .75 diopters or<br>less where the astigmatism does not interfere<br>with visual acuity. The IntelliWave 3, toric<br>(efrofilcon A) Silicone Hydrogel Soft Contact<br>Lenses for daily wear are indicated for the<br>correction of ametropia (myopia and<br>hyperopia) in aphakic and non-aphakic persons<br>with non-diseased eyes and/or possesses<br>refractive astigmatism not exceeding 12<br>diopters. The IntelliWave 3, multifocal<br>(efrofileon A) Silicone Hydrogel Soft Contact<br>Lenses for daily wear are indicated for the<br>correction of ametropia (myopia and<br>hyperopia) in aphakic and non-aphakic persons<br>with non-diseased eyes and/or possesses<br>refractive astigmatism not exceeding 0.75<br>diopters and are presbyopic requiring add<br>power of up to +4.00 diopters. The<br>IntelliWave 3, multifocal toric (efrofilcon A) |

# Substantial Equivalence Matrix

{7}------------------------------------------------

|                                                                   | The Apioc AP, multifocal toric (efrofilcon A)<br>Silicone Hydrogel Soft Contact Lenses for<br>daily wear are indicated for the correction of<br>ametropia (myopia and hyperopia) and<br>presbyopia in aphakic and non-aphakic<br>persons with non-diseased eyes and/or<br>possesses refractive astigmatism not<br>exceeding 12 diopters and are presbyopic<br>requiring add power of up to +4.00 diopters.<br>Eyecare practitioners may prescribe the lenses<br>for frequent/planned replacement wear, with<br>cleaning, disinfection and scheduled<br>replacement. When prescribed for<br>frequent/planned replacement wear, the lens<br>may be cleaned and disinfected using a<br>chemical (not heat) lens care system or<br>hydrogen peroxide disinfection systems. | Silicone Hydrogel Soft Contact Lenses for<br>daily wear are indicated for the correction of<br>ametropia (myopia and hyperopia) in aphakic<br>and non-aphakic persons with non-diseased<br>eyes and/or possesses refractive astigmatism<br>not exceeding 4 diopters and are presbyopic<br>requiring add power of up to +4.00 diopters.<br>The IntelliWave3 irregular cornea (efrofilcon<br>A) Silicone Hydrogel Soft Contact Lenses for<br>daily wear are indicated for the correction of<br>ametropia (myopia and hyperopia) in aphakic<br>and non-aphakic persons and may be<br>prescribed in otherwise nan-diseased eyes that<br>require a Soft Contact Lens for the<br>management of irregular corneal conditions<br>such as keratoconus and post graft fitting.<br>Eyecare practitioners may prescribe the lenses<br>for frequent/planned replacement wear, with<br>cleaning, disinfection and scheduled<br>replacement. When prescribed for<br>frequent/planned replacement wear, the lens<br>may be cleaned and disinfected using a<br>chemical (not heat) lens care system or<br>hydrogen peroxide disinfection systems. |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functionality                                                     | The contact lenses act as a refractive medium<br>that focus light rays from near and distant<br>objects on the retina                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The contact lenses act as a refractive medium<br>that focus light rays from near and distant<br>objects on the retina                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| FDA Classification                                                | Soft (hydrophilic) Contact Lens (21 CFR<br>886.5925)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Soft (hydrophilic) Contact Lens (21 CFR<br>886.5925)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| FDA Group                                                         | FDA Group 2<br>(>50% H2O, non-ionic polymer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | FDA Group 2<br>(>50% H2O, non-ionic polymer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Production Method                                                 | Lathe Cut                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Lathe Cut                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| USAN name                                                         | Efrofilcon A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Efrofilcon A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Water Content (%)                                                 | 74±2%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 74±2%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Oxygen Permeability<br>x 10-11 (cm²/sec)(mlO₂)/(ml x hPa @ 35°C)) | 59.8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 59.8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Refractive Index<br>(hydrated)                                    | 1.380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 1.380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| UV Blocker                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{8}------------------------------------------------

#### VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

# Non-clinical Testing

Non-clinical testing for finished contact lenses manufactured from efrofilcon A blanks has been demonstrated previously. Additional non-clinical testing was conducted to support the claim that the Apioc, Apioc P, Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are substantially equivalent to the currently marketed predicate devices. A summary of the results from the non-clinical studies is presented below.

# Stability:

Testing was performed to demonstrate the stability of the parameters and features of the Apioc. Apioc P. Apioc A, and Apioc AP (efrofilcon A) finished contact lenses over the proposed shelf life. The data presented supports substantial equivalence of the Apioc . Apioc A, and Apioc AP (efrofilcon A) finished contact lenses to the predicate device.

# Lens Design/Manufacturing Verification:

Bench testing was performed to verify the ability of Lentechs. Inc. to manufacture the Apioc P. Apioc A, and Apioc AP (efrofilcon A) finished contact lenses to a variety of prescribed parameters within manufacturing tolerances.

# Clinical Testing

Clinical testing is not required. The clinical performance of soft (hydrophilic) contact lenses manufactured from efrofilcon A materials has been demonstrated previously.

{9}------------------------------------------------

# VIII. CONCLUSIONS

# Validity of Scientific Data

Laboratories under Good Laboratory Practice regulations conducted toxicology, microbiology, and shelf-life stability studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7.

### Substantial Equivalence

Information presented in this Premarket Notification establishes that the Apioc , Apioc A, and Apioc AP (efrofilcon A) Soft (hydrophilic) Silicone Hydrogel Contact Lenses are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.

### Risks and Benefits

The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) daily wear contact lenses. The benefits to the patient are the same as those for other soft (hydrophilic) daily wear contact lenses.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K191121](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K191121)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
