Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lens; Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lens

K181349 · Visco Vision, Inc. · LPL · Sep 19, 2018 · Ophthalmic

Device Facts

Record IDK181349
Device NameVexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lens; Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lens
ApplicantVisco Vision, Inc.
Product CodeLPL · Ophthalmic
Decision DateSep 19, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

The Vexillum (olifilcon B) with Tangible Polymers Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The Vexillum (olifilcon B) with Tangible Polymers Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lens for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eves and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Vexillum (olifilcon B) with Tangible Polymers Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear single use soft contact lens for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is requried. The Daily Breeze (olifilcon B) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eves who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The Daily Breeze (olifilcon B) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Daily Breeze (olifilcon B) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is reguried.

Device Story

Silicone hydrogel soft contact lenses (olifilcon B) designed for daily wear; single-use; no cleaning/disinfection required. Vexillum lenses feature Tangible Polymers surface coating; Daily Breeze lenses lack this coating. Both utilize Reactive Blue19 tint for handling and benzotriazole UV-absorbing monomer. Lenses supplied sterile in blister packs with isotonic buffered saline. Prescribed by eye care practitioners for correction of myopia, hyperopia, astigmatism, and presbyopia. Cast-molded manufacturing; steam sterilized. Intended to provide vision correction while maintaining ocular health in non-diseased eyes.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and compliance with FDA-recognized consensus standards (ISO 18369, ISO 10993, ISO 17665, ASTM D792).

Technological Characteristics

Material: olifilcon B (53%) and water (47%). UV blocker: benzotriazole. Handling tint: Reactive Blue19. Manufacturing: cast molded. Sterilization: steam autoclave. Packaging: blister pack with isotonic buffered saline (Vexillum includes Tangible Polymers coating). Connectivity: N/A. Dimensions: 13.0-15.0 mm diameter, 8.0-9.2 mm base curve.

Indications for Use

Indicated for correction of refractive ametropia (myopia/hyperopia), astigmatism, and presbyopia in phakic or aphakic persons with non-diseased eyes. Powers range -20.00 to +20.00D; astigmatic correction -0.75 to -2.25D; add powers +0.75 to +2.75D. Daily wear, single-use disposable.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 19, 2018 Visco Vision Inc. Evan Huang Director of Global OA No. 1, Xingye St., Guishan Dist., Taoyuan City, 33341 TW Re: K181349 Trade/Device Name: Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses; Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 27, 2018 Received: August 27, 2018 Dear Evan Huang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # J. Angelo Green for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181349 ### Device Name Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses ### Indications for Use (Describe) The Vexillum (olifilcon B) with Tangible Polymers Spherical Silicone Hydrogel Soft Contact Lenses are daily wear single use soft contact lenses indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The Vexillum (olifilcon B) with Tangible Polymers Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lenses for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Vexillum (olifilcon B) with Tangible Polymers Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear single use soft contact lenses for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required. The Daily Breeze (olifilcon B) Spherical Silicone Hydrogel Soft Contact Lenses are daily wear single use soft contact lenses indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. The Daily Breeze (olifilcon B) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Daily Breeze (olifilcon B) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is horizontally aligned and centered. ## 510(k) Summarv--K181349 | Company Name: | Visco Vision Inc | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company Address: | No. 1, Xingye St., Guishan Dist.,<br>Taoyuan City, 33341, TAIWAN | | Telephone: | +886-3-359-6868 | | Fax: | +886-3-349-0202 | | Contact Person: | Evan Huang | | Summary Preparation Date: | 2018.5.15 | | Device Name: | | | Trade Name: | Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft<br>Contact Lenses<br>Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses | | Classification Name: | Soft (hydrophilic) contact lens. | | Regulation Number: | 886.5925 | | Product Code: | LPL, MVN | | Device Class: | Class 2 | | Panel: | Ophthalmic | # PREDICATE DEVICE: K 160344, Si-Hy (olifilcon B) Silicone Hydrogel Soft (Hydrophilic) Contact Lenses # Device Description for Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses The Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon B, with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 nm and less than 5% in the range of UVB range of 280-315 nm. Lenses are supplied sterile in sealed blister packs containing Tangible Polymers (coating on the lens surface during sterilization process) with isotonic buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is centered in the image. The font is sans-serif. autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date. # Intended Use: The Vexillum (olifilcon B) with Tangible Polymers Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The Vexillum (olifilcon B) with Tangible Polymers Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lens for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eves and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Vexillum (olifilcon B) with Tangible Polymers Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear single use soft contact lens for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is requried. All comparison table for applied devices are as following, and the substantial equivalence determination is based on the 510(k) Substantial Equivalence Decision-Making Process Flowchart which includes the comparison and discussion of indications for use, technology, and performance specifications. # Device Description for Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses The Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon B, with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is simple and clear, with a focus on the company name and its expertise in silicone hydrogels. by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 mm and less than 5% in the range of UVB range of 280-315 nm. Lenses are supplied sterile in sealed blister packs containing Sodium Hyaluronate and Sodium Alginate with isotonic buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date. ### Intended Use: The Daily Breeze (olifilcon B) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eves who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The Daily Breeze (olifilcon B) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The Daily Breeze (olifilcon B) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be dscarded upon removal. Therefore, no cleaning or disinfecting is reguried. All comparison table for applied devices are as following, and the substantial equivalence determination is based on the 510(k) Substantial Equivalence Decision-Making Process Flowchart which includes the comparison and discussion of indications for use, technology, and performance specifications. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is centered and the background is white. The image appears to be a logo or branding material. | Category | Vexillum<br>(olifilcon B) with<br>Tangible Polymers<br>Silicone Hydrogel<br>Soft Contact Lenses<br>New device | Daily Breeze<br>(olifilcon B) Silicone<br>Hydrogel Soft<br>Contact Lenses<br>New device | Si-Hy (olifilcon B)<br>Silicone Hydrogel<br>Soft Contact Lenses<br>K160344 | Result of<br>Comparasion | |--------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------| | Applicant | Visco Vision Inc | Visco Vision Inc | Visco Vision Inc | Same | | Classification | class II | class II | class II | Same | | Regulation<br>number | 886.5925 | 886.5925 | 886.5925 | Same | | Product code | LPL, MVN | LPL, MVN | LPL, MVN | Same | | Intended use | Myopia, Hyperopia,<br>astigmatism,<br>Presbyopia | Myopia, Hyperopia,<br>astigmatism,<br>Presbyopia | Myopia, Hyperopia,<br>astigmatism,<br>Presbyopia | Same | | Replacement<br>Schedule | Daily Disposable<br>(Single use) | Daily Disposable<br>(Single use) | Daily Disposable<br>(Single use) | same | | USAN Name | olifilcon B | olifilcon B | olifilcon B | same | | FDA Category<br>(Group) | Group 5C (Nonionic,<br>Water < 50 wt %) | Group 5C (Nonionic,<br>Water < 50 wt %) | Group 5C (Nonionic,<br>Water < 50 wt %) | same | | Manufacturing<br>Method | Cast Molded | Cast Molded | Cast Molded | same | | Lens Design | Spherical, toric or<br>multifocal | Spherical, toric or<br>multifocal | Spherical, toric or<br>multifocal | same | | Water Content | 47% | 47% | 47% | same | | Light<br>Transmittance | 94% | 94% | 94% | same | | Refractive Index | 1.410 (hydrated) | 1.410 (hydrated) | 1.410 (hydrated) | same | | Oxygen<br>Permeability (DK,<br>35°C) | 120<br>(Fatt method) | 120<br>(Fatt method) | 120<br>(Fatt method) | Same | | Diameter Range | 13.0 to 15.0 mm | 13.0 to 15.0 mm | 13.0 to 15.0 mm | same | | Power Range | - 20.00D~ +20.00D | - 20.00D~ +20.00D | - 20.00D~ +20.00D | same | | Center Thickness | 0.08mm @ -3.00D<br>(Varies with Power) | 0.08mm @ -3.00D<br>(Varies with Power) | 0.08mm @ -3.00D<br>(Varies with Power) | same | | Base Curve | 8.0 mm to 9.2 mm | 8.0 mm to 9.2 mm | 8.0 mm to 9.2 mm | same | | Blue handling tint | Reactive Blue19 | Reactive Blue19 | Reactive Blue19 | same | | Category | Vexillum<br>(olifilcon B) with<br>Tangible Polymers<br>Silicone Hydrogel<br>Soft Contact Lenses<br>New device | Daily Breeze<br>(olifilcon B) Silicone<br>Hydrogel Soft<br>Contact Lenses<br>New device | Si-Hy (olifilcon B)<br>Silicone Hydrogel<br>Soft Contact Lenses<br>K160344 | Result of<br>Comparasion | | Leachability | no leachable<br>monomers and<br>addictive residues | no leachable<br>monomers and<br>addictive residues | no leachable<br>monomers and<br>addictive residues | same | | Packaging | Blister Pack | Blister Pack | Blister Pack | same | | Packaging<br>Solution | sterile isotonic borate<br>buffered saline with<br>Tagible polymers | sterile isotonic borate<br>buffered saline | sterile isotonic borate<br>buffered saline | different | | Coating on the<br>lens surface | Yes, coating with<br>Tangible Polymer<br>during the<br>sterilization process | No | No | different | | Sterilization<br>method | Steam | Steam | Steam | same | | Shelf Life | 5 years | 5 years | 5 years | same | | Sterility of Device | SAL= 10-6 | SAL= 10-6 | SAL= 10-6 | same | | Tensile strength<br>(Mpa) | 0.80±0.2 | 0.80±0.2 | 0.80±0.2 | same | | Modulus (Mpa) | 0.50 ± 0.1 | 0.50 ± 0.1 | 0.50±0.1 | same | | Elongation at<br>break (%) | 200± 20% | 200± 20% | 200± 20% | same | | Toughness (J/m3) | 0.90±0.05 | 0.90±0.05 | 0.90±0.05 | same | | Compliance standard | | | | | | Biocompatibility | ISO10993-1 | ISO10993-1 | ISO10993-1 | same | | | ISO10993-5 | ISO10993-5 | ISO10993-5 | | | | ISO10993-10 | ISO10993-10 | ISO10993-10 | | | | ISO10993-11 | ISO10993-11 | ISO10993-11 | | | Sterilization and<br>Shelf life | ISO 17665-1 | ISO 17665-1 | ISO 17665-1 | Same | | | ISO11737-1 | ISO11737-1 | ISO11737-1 | | | | ISO 11987 | ISO 11987 | ISO 11987 | | | performance | ISO18369-3 | ISO18369-3 | ISO18369-3 | Same | | Category | Vexillum<br>(olifilcon B) with<br>Tangible Polymers<br>Silicone Hydrogel<br>Soft Contact Lenses<br>New device | Daily Breeze<br>(olifilcon B) Silicone<br>Hydrogel Soft<br>Contact Lenses<br>New device | Si-Hy (olifilcon B)<br>Silicone Hydrogel<br>Soft Contact Lenses<br>K160344 | Result of<br>Comparasion | | | ISO18369-4<br>ASTM D792-13 | ISO18369-4<br>ASTM D792-13 | ISO18369-4<br>ASTM D792-13 | | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image contains the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The SIIIcone Hydrogel Expert" in a smaller, gray font. The text is aligned to the left and the background is white. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is horizontally aligned and centered. # Substantial Equivalence Comparison The Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses, The Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses, and the The Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lenses, have same intended use, Replacement Schedule, classification name/product code, len materials, manufacturing and sterilization method, perfomance parameter ranges, lens design, mechanical properties, physical properties, mechanical properties, biocompatibility, shelf life, Elongation at break and other reference devices have same technical characteristic as new devices to support the substantial equivalence. The differences among those three devices are Packaging Solution and Coating on the lens surface. # Non-clinical tests The safety tests, such as biocompatibility have been performed and meet the requirement of following FDA Recognized Consensus Standards. Regarding the performance, the bench tests were performed in accordance with following FDA Recognized Consensus Standards. All tests passed the requirement to demonstration that new device is as substantial equivalent as the predicate device. - ISO18369-3 Ophthalmic optics - Contact lenses - Part 3: Measurement Methods - ISO18369-4 Ophthalmic optics - Contact lenses - Part 4: Physicochemical properties of contact lens materials - ISO18369-2 Ophthalmic optics - Contact lenses - Part 2: Tolerances - ISO11987 Ophthalmic optics -- Contact lenses and contact lens care products --Determination of physical compatibility of contact lens care products with contact lenses - ISO17665-2 terilization Of Health Care Products - Moist Heat - Part 2: Guidance on the Application of ISO 17665-1. (Sterility) - ISO17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "VISCO" in large, bold, blue letters. Below the word "VISCO" is the phrase "The Silicone Hydrogel Expert" in a smaller, gray font. The text is centered in the image. The image appears to be a logo. - ASTM D792-13 Standard Test Methods For Density And Specific Gravity (Relative Density) Of Plastics By Displacement. (Materials) # Clinical study This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply. # Conclusions: Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses and Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses have the same intended use and technological characteristics as the above predicate devices. Moreover, information contained in this submission supplied demonstrates that any differences in their packing solution and Coating on the lens surface do not raise any new questions of safety or effectiveness. Thus, Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses and Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lenses are substantially equivalent to the predicate devices with respect to its intended use, technological characteristics and performance characteristics.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...