← Product Code [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL) · K160803

# Avaira Vitality (fanfilcon A) Soft (Hydrofilic) Contact Lens (K160803)

_CooperVision, Inc. · LPL · Jul 13, 2016 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K160803

## Device Facts

- **Applicant:** CooperVision, Inc.
- **Product Code:** [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL.md)
- **Decision Date:** Jul 13, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

AVAIRA VITALITY SPHERE and ASPHERE (fanfilcon A) Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. AVAIRA VITALITY (fanfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters. AVAIRA VITALITY (fanfilcon A) MULTIFOCAL Soft lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. AVAIRA VITALITY (fanfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optic correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes in powers of -20.00 to +20.00 diopters with add powers from +0.25 to +4.00 diopters and astigmatism corrections from -0.25 to -10.00 diopters. Eye Care Practitioners may prescribe the Avaira Vitality (fanfilcon A) Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear. When prescribed for frequent/planned replacement, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection only. When prescribed for single-use disposable wear, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be discarded after each removal.

## Device Story

Avaira Vitality (fanfilcon A) are silicone hydrogel soft contact lenses designed for daily wear. Lenses are molded, tinted with Reactive Blue #246 for visibility, and incorporate a Norbloc UV blocker. Supplied sterile in buffered saline. Prescribed by eye care practitioners for either frequent/planned replacement (with chemical disinfection) or single-use disposable wear. Lenses correct ametropia (myopia, hyperopia, astigmatism) and presbyopia. Patients benefit from vision correction consistent with standard soft contact lenses. Device is used by the patient after professional fitting.

## Clinical Evidence

Three-month, multi-center, randomized, masked, concurrent control study with 90 subjects (60 test, 30 control). Primary endpoints: slit lamp biomicroscopy, adverse event rates, and visual acuity. Results demonstrated test lens safety and efficacy equivalent to control lens.

## Technological Characteristics

Silicone hydrogel (fanfilcon A, 45% polymer/55% water). Molded manufacturing. Moist heat sterilization. Visibility tint: Reactive Blue #246. Includes Norbloc UV blocker. Oxygen permeability (Dk) 90. Refractive index 1.40. Specific gravity 1.026.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- CooperVision AVAIRA (enfilcon A) ([K113759](/device/K113759.md))
- CooperVision MyDay (stenfilcon A) ([K131378](/device/K131378.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, possibly black or dark blue. The image is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 13, 2016

CooperVision, Inc. Mr. Dan O'Mara Regulatory Affairs Manager 6150 Stoneridge Mall Rd.. Suite 370 Pleasanton, CA 94588

Re: K160803

Trade/Device Name: Avaira Vitality (fanfilcon A) Soft (hydrophilic) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: June 1, 2016 Received: June 2, 2016

Dear Mr. O'Mara:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160803

#### Device Name

Avaira Vitality (fanfilcon A) Soft (Hydrophilic) Contact Lens

#### Indications for Use (Describe)

#### Sphere/Asphere:

AVAIRA VITALITY SPHERE and ASPHERE (fanfilcon A) Soft Contact lenses are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.

#### Toric:

AVAIRA VITALITY (fanfilcon A) TORIC Soft Contact lenses are indicated for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and astigmatic corrections from -0.25 to -10.00 diopters.

Multifocal:

AVAIRA VITALITY (fanfilcon A) MULTIFOCAL Soft lenses are indicated for the correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 diopters and with add powers from +0.25 to +4.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Multifocal Toric:

AVAIRA VITALITY (fanfilcon A) MULTIFOCAL TORIC Soft Contact lenses are indicated for the optic correction of distance and near vision in presbyopic phakic or aphakic persons with non-diseased eyes in powers of -20.00 to +20.00 diopters with add powers from +0.25 to +4.00 diopters and astigmatism corrections from -0.25 to -10.00 diopters.

Eye Care Practitioners may prescribe the Avaira Vitality (fanfilcon A) Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear.

When prescribed for frequent/planned replacement, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection only.

When prescribed for single-use disposable wear, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be discarded after each removal.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| Date:                  | 6 July 2016                                                                                                   |                                                                                                                            |
|------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Submitter Information: | CooperVision, Inc.<br>6150 Stoneridge Mall Road<br>Suite 370<br>Pleasanton, CA 94588<br>USA<br>(800) 972-6724 |                                                                                                                            |
| Contact:               | Dan O'Mara<br>Regulatory Affairs Manager<br>domara@coopervision.com<br>(585) 756-9589                         |                                                                                                                            |
| Device Identification: | Common Name:<br>Trade Name:<br>Class. Name:                                                                   | Soft Contact Lens<br>AVAIRA VITALITY<br>(fanfilcon A) Soft Contact Lens<br>Soft (hydrophilic) Contact Lens –<br>Daily Wear |
|                        | Classification:                                                                                               | Class II [21 CFR 886.5925 (b) (1)]                                                                                         |
|                        | Product Code:                                                                                                 | LPL, MVN                                                                                                                   |
| Predicate Device(s):   | Primary (Indication, Wear Schedule and Material)                                                              |                                                                                                                            |
|                        | CooperVision AVAIRA (enfilcon A)<br>K113759<br>(Silicone Hydrogel, Daily wear - frequent replacement)         |                                                                                                                            |
|                        | Secondary (Indication, Wear schedule and Material)                                                            |                                                                                                                            |
|                        | CooperVision MyDay (stenfilcon A)<br>K131378<br>(Silicone Hydrogel, Daily wear - single use)                  |                                                                                                                            |

{5}------------------------------------------------

## Description of Device

- . The AVAIRA VITALITY (fanfilcon A) Contact Lens visibility tinted with UV blocker is available as a sphere/asphere lens, toric lens, multifocal lens, and multifocal toric lens.
- . The lenses are made of a silicone hydrogel material which is not surface treated and is characterized by high oxygen permeability (Dk).
- . The AVAIRA VITALITY (fanfilcon A)) Contact Lens is tinted blue using Reactive Blue #246 to make the lens more visible for handling.
- . A Norbloc UV blocker is used to reduce the amount of ultraviolet light transmitted into the eye.
- . The AVAIRA VITALITY (fanfilcon A) lens is supplied sterile, packaged in a buffered saline solution.
- . The composition of the lens is 45% fanfilcon A and 55% water by weight when hydrated and stored in buffered saline solution.

{6}------------------------------------------------

# Indications for Use

| Lens Design      | Indication                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sphere/Asphere   | AVAIRA SPHERE and ASPHERE (fanfilcon A) Soft<br>Contact lenses are indicated for the correction of ametropia (myopia<br>and hyperopia) in aphakic and non-aphakic persons with non-<br>diseased eyes in powers from -20.00 to +20.00 diopters. The lenses<br>may be worn by persons who exhibit astigmatism of 2.00 diopters<br>or less that does not interfere with visual acuity.                                                                                          |
| Toric            | AVAIRA VITALITY (fanfilcon A) TORIC Soft Contact lenses are<br>indicated for the correction of ametropia (myopia or hyperopia with<br>astigmatism) in aphakic and non- aphakic persons with non-diseased<br>eyes in powers from -20.00 to +20.00 diopters and astigmatic<br>corrections from -0.25 to -10.00 diopters.                                                                                                                                                       |
| Multifocal       | AVAIRA VITALITY (fanfilcon A) MULTIFOCAL Soft lenses<br>are indicated for the correction of refractive ametropia<br>(myopia and hyperopia) and emmetropia with presbyopia in<br>aphakic and non-aphakic persons with non-diseased eyes in<br>powers from -20.00 to +20.00 diopters and with add powers from<br>+0.25 to +4.00 diopters. The lenses may be worn by persons who<br>exhibit astigmatism of 2.00 diopters or less that does not interfere<br>with visual acuity. |
| Multifocal Toric | AVAIRA VITALITY (fanfilcon A) MULTIFOCAL TORIC Soft<br>Contact lenses are indicated for the optic correction of distance and<br>near vision in presbyopic phakic or aphakic persons with non-<br>diseased eyes in powers of -20.00 to +20.00 diopters with add<br>powers from +0.25 to +4.00 diopters and astigmatism corrections<br>from -0.25 to -10.00 diopters.                                                                                                          |

- Eye Care Practitioners may prescribe the Avaira Vitality (fanfilcon A) Soft Contact lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement or for single-use disposable wear.
When prescribed for frequent/planned replacement, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be cleaned, rinsed and disinfected each time the lens is removed. The contact lens is to be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When prescribed for frequent/ planned replacement wear, the lenses may be disinfected using a chemical disinfection only.

When prescribed for single-use disposable wear, the Avaira Vitality (fanfilcon A) Soft Contact lens is to be discarded after each removal.

{7}------------------------------------------------

# Technological Characteristics:

The technological characteristics of the AVAIRA VITALITY (fanfilcon A) Contact Lenses are compared to the characteristics of the predicate device, CooperVision AVAIRA (enfilcon A) Contact Lens, in the following tables.

| Material Comparison  |                        |                                 |  |
|----------------------|------------------------|---------------------------------|--|
|                      | Predicate Device       | Subject Device                  |  |
| Product Name         | CooperVision<br>AVAIRA | CooperVision<br>AVAIRA VITALITY |  |
| Material USAN Name   | enfilcon A             | fanfilcon A                     |  |
| 510(k) Number        | K113759                | TBD - Current Submission        |  |
| FDA Category (Group) | Silicone Hydrogel      | Silicone Hydrogel               |  |
| Manufacturing Method | Molded                 | Molded                          |  |
| Sterilization        | Moist Heat             | Moist Heat                      |  |
| Packaging            | Blister                | Blister                         |  |
| Visibility Tint      | Phthalocyanine Blue    | Reactive Blue #246              |  |

| Parameter Comparison         |                                                   |                                                                              |  |
|------------------------------|---------------------------------------------------|------------------------------------------------------------------------------|--|
|                              | Predicate Device                                  | Subject Device                                                               |  |
|                              | CooperVision<br>AVAIRA<br>(enfilcon A)<br>K113759 | CooperVision<br>AVAIRA VITALITY<br>(fanfilcon A)<br>TBD - Current Submission |  |
| Water Content, %             | 46%                                               | 55%                                                                          |  |
| Refractive Index @ 20℃       | 1.40                                              | 1.40                                                                         |  |
| Specific Gravity g/mL        | 1.06                                              | 1.026                                                                        |  |
| Oxygen Permeability<br>(Dk)* | 100                                               | 90                                                                           |  |
| Base Curve, mm               | 8.5                                               | 8.4                                                                          |  |
| Diameter, mm                 | 14.2                                              | 14.2                                                                         |  |

*Dk units: x 10-11 (cm²/sec) x (ml O2)/(ml x mm Hg)

{8}------------------------------------------------

# Technological Characteristics:

The technological characteristics of the AVAIRA Vitality (fanfilcon A) Contact Lenses are compared to the characteristics of the predicate device, CooperVision MyDay (stenfilcon A) Contact Lens, in the following tables.

| Material Comparison  |                       |                                 |  |
|----------------------|-----------------------|---------------------------------|--|
|                      | Predicate Device      | Subject Device                  |  |
| Product Name         | CooperVision<br>MyDay | CooperVision<br>AVAIRA VITALITY |  |
| Material USAN Name   | stenfilcon A          | fanfilcon A                     |  |
| 510(k) Number        | K131378               | TBD – Current Submission        |  |
| FDA Category (Group) | Silicone Hydrogel     | Silicone Hydrogel               |  |
| Manufacturing Method | Molded                | Molded                          |  |
| Sterilization        | Moist Heat            | Moist Heat                      |  |
| Packaging            | Blister               | Blister                         |  |
| Visibility Tint      | Reactive Blue #246    | Reactive Blue #246              |  |

| Parameter Comparison         |                                                    |                                                                              |
|------------------------------|----------------------------------------------------|------------------------------------------------------------------------------|
|                              | Predicate Device                                   | Subject Device                                                               |
|                              | CooperVision<br>MyDay<br>(stenfilcon A)<br>K131378 | CooperVision<br>AVAIRA VITALITY<br>(fanfilcon A)<br>TBD - Current Submission |
| Water Content, %             | 54%                                                | 55%                                                                          |
| Refractive Index @ 20oC      | 1.40                                               | 1.40                                                                         |
| Specific Gravity g/mL        | 1.033                                              | 1.026                                                                        |
| Oxygen Permeability<br>(Dk)* | 80                                                 | 90                                                                           |
| Base Curve, mm               | 8.4                                                | 8.4                                                                          |
| Diameter, mm                 | 14.2                                               | 14.2                                                                         |

*Dk units: x 10-11 (cm²/sec)x(ml O2)/(ml x mm Hg)

{9}------------------------------------------------

## Non-clinical Testing

A series of in-vitro and in-vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the contact lens. All tests were conducted in accordance with the GLP regulation (21 CFR Part 56) or according to valid scientific protocols.

The results of the non-clinical testing/evaluation demonstrate that:

- The lens material and/or extracts are non-toxic, non-irritating and non-sensitizing under the experimental conditions; and
- The lens physical and material properties are consistent with currently marketed lenses.

# Clinical Testing

A three-month clinical study was completed to evaluate the safety and efficacy of the AVAIRA VITALITY (fanfilcon A) Contact Lens for daily wear with a replacement schedule of one month (up to 30 days).

The study was a multi-center, randomized, masked, concurrent control study that evaluated ninety (90) male and female subjects who were dispensed lenses in a 2:1 ratio with 60 subjects dispensed into the Test lenses and 30 subjects dispensed into the Control lenses. The primary outcome measures were slit lamp (biomicroscopy) findings and adverse event rates along with lens visual acuity comparisons between the Test and the Control contact lenses. Secondary outcome measures included average lens wearing times and subjective lens comfort assessed by frequency or symptoms.

The Test contact lens was found to be substantially equivalent to the Control contact lens for safety and efficacy.

{10}------------------------------------------------

## Conclusion Drawn from Studies

Validity of Scientific Data:

A contract laboratory under Good Laboratory Practice Regulations conducted toxicology studies. Microbiology, chemistry, shelf-life stability, and leachability studies were conducted by Cooper Vision laboratories and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7.

Substantial Equivalence:

Information presented in this Premarket Notification establishes that the AVAIRA VITALITY (fanfilcon A) Contact Lens is as safe and effective as the predicate device when used in accordance with the labeled directions for use and the requested indication.

Risk and Benefits:

The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear basis. The benefits to the patient are the same as those for other soft (hydrophilic) contact lenses.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K160803](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K160803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
