The One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The lens is intended to be worn once and then discarded at the end of each wearing period on a daily basis. The patient should be instructed to start the next wearing period with a new lens.
Device Story
One Day Flat Pack is a daily disposable soft contact lens made of hioxifilcon A (non-ionic copolymer of 2-HEMA and GMA). The device functions as a refractive medium to focus light on the retina. The primary innovation is the 'Flat Pack' packaging system, which is approximately 1mm thick and designed to present the lens anterior side up to facilitate easier handling and correct orientation for patient insertion. The lens is intended for single-use by patients with refractive ametropia. It is supplied in a buffered saline solution. The device is used by the patient for daily vision correction; the healthcare provider prescribes the lens based on refractive needs. The packaging design aims to improve user compliance and ease of use compared to traditional blister packs.
Clinical Evidence
No clinical data was required for this submission. Safety and efficacy of the hioxifilcon A lens material were previously established in K080632. Substantial equivalence for the new packaging system was demonstrated through non-clinical bench testing, including packaging integrity, stress tests, and biocompatibility/toxicology studies.
Technological Characteristics
Material: hioxifilcon A (42% polymer, 58% water). Design: Spun-cast, spherical, single vision. Dimensions: 14.2mm diameter, 8.6mm base curve. Optical: >95% light transmission, 1.4011 refractive index. Packaging: Polypropylene and laminated foil 'Flat Pack' system. Sterilization: Sterile. Connectivity: None.
Indications for Use
Indicated for daily wear, single-use optical correction of myopia and hyperopia in phakic or aphakic persons with non-diseased eyes and 1.00D or less of astigmatism.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
K180819 — One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens · Menicon Co, Ltd. · May 16, 2018
K130931 — EASY DAY (METHAFILCON A) 1-DAY SOFT (HYDROPHILIC) CONTACT LENS · Jiangsu Horien Contact Lens Co., Ltd. · Aug 30, 2013
K243909 — Precision1; Precision1 for Astigmatism; Dailies Total1; Dailies Total1 for Astigmatism; Dailies Total1 Multifocal; Dailies Total1 Multifocal Toric · Alcon Laboratories, Inc. · Jan 13, 2025
K963487 — DAILIES (NELFILCON A) ONE-DAY SOFT CONTACT LENS · Ciba Vision Corporation · Nov 27, 1996
K082880 — EZVUE (OCUFILCON D) 1-DAY SOFT (HYDROPHILIC)CONTACT LENS · I-See Vision Technology, Inc. · Jun 16, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# Summary of Safety and Effectiveness
This summary of 510 (k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
K102895 The assigned 510 (k) number is:
Applicant information:
Date Prepared:
March 1, 2011
Menicon Co., Ltd.
Company Name: Address:
Contact Person/ Official Correspondent: Phone number: Fax number:
Nagoya, Japan 460-0006 Mr. Takahiro Ochi
21-19 Aoi, 3-Chome, Naka-Ku
(81)-52-937-5021 (81)-52-935-1121
Medvice Consulting, Inc. U.S. Consultant Martin Dalsing Contact Person: (970) 243-5490 Phone number: (970) 243-5501 Fax number:
#### Device Information:
| Device Classification: | Class II |
|------------------------|-------------------------------------------------------------------------|
| Classification number: | LPL |
| Classification name: | Lenses, Soft Contact, Daily Disposable |
| Trade name: | One Day Flat Pack (hioxifilcon A) Daily Disposable Soft<br>Contact Lens |
Menicon Co., Ltd.
Page SSE 1
APR 2 8 2011
{1}------------------------------------------------
# Purpose of 510(k) notification:
- 1) Change in lens packaging system to new and innovative, 'Flat Pack Technology'.
Was: Traditional, 'Blister' packaging Now: Innovative, 'Flat Pack' packaging
#### Equivalent Predicate Devices:
The One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lenses are substantially equivalent to the following predicate device:
1. Clear 1-Day (hioxifilcon A), K080632, Manufactured by Clearlab SG Pte Ltd.
# Device Description:
The One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens is available as a single vision spherical lens. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The One Day Flat Pack Daily Disposable Contact Lens packaging system is designed to reduce lens handling by always presenting the lens anterior side up upon opening, which ensures correct lens orientation for proper eye insertion. The approximately 1mm flat packaging system is easily opened and reinforces the single-use factor.
The non-ionic lens material, (hioxifilcon A) is a random copolymer of 2- hydroxyethyl methacylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (Glycero) Methacrylate, GMA) cross-linked with ethylene glycol dimethacrylate. It consists of 42% hioxifileon A and 58% water by weight when immersed in a buffered saline solution. The lens is available with a blue visibility-handling tint, color additive 'Reactive Blue # 19', 21 CFR part 73.3121. The United States Adopted Names Council (USAN) has adopted the (hioxifilcon A) name.
In the hydrated state, the lens conforms to the eve covering the corner and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{2}------------------------------------------------
# 510(k) PREMARKET NOTIFICATION One Day Flat Pack (hioxifilcon A) Single Use Contact Lens
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight. The physical properties of the lens are:
| Refractive Index at 21°C: | 1.4011(wet) |
|---------------------------------|----------------------------------------------------------------------------------------------------------|
| Light Transmission: | > 95% |
| Surface Character: | Hydrophilic |
| Water Content at 21°C: | 58.0 % |
| Specific Gravity at 21°C: | 1.086(wet) |
| Oxygen Permeability at 34-36°C: | 25.38 x 10 <sup>-11</sup> (cm <sup>2</sup> /sec) (ml O <sub>2</sub> /ml x mm Hg), (revised Fatt method). |
#### Substantial Equivalence:
Lenses supplied in the Flat Packaging are within established specifications for the lens design and are equivalent to the predicate device lenses supplied in the blister pack packaging. The lens design maintains established physical characteristics, and are stable and biocompatible with the ocular environment.
- . Both the blister and flat package have components made from polypropylene and laminated foil.
- . Traditional blister pack has its base made from polypropylene with a laminated foil sealed on top. Flat package uses the same material with 2 laminated foils and an additional polypropylene disc to ensure correct lens orientation when opened.
The following non-clinical studies demonstrate that lenses supplied in the Flat Pack packaging are substantially equivalent to lenses supplied in the blister pack.
Non-clinical Studies A series of in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the contact lenses packaged in the Flat Packaging system. All non-clinical testing was conducted according to valid scientific protocols.
> Furthermore, a series of packaging integrity tests, stress tests, and simulations tests were performed on the Flat Packaging system to assess overall robustness and demonstrate the integrity of the packaging system.
Test results of the non-clinical testing on the One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens demonstrate that:
- Lenses supplied in the Flat Pack packaging system are sterile, .
- The packaging material and extracts are not toxic and not irritating, and .
{3}------------------------------------------------
# 510(k) PREMARKET NOTIFICATION One Day Flat Pack (hioxifilcon A) Single Use Contact Lens
- The lens physical and material properties are consistent with lenses packaged in . a blister package. See chart below.
| Lens phys/mtl property | One Day Flat Pack | Clear 1-Day -blister pkg. |
|------------------------|-------------------|---------------------------|
| Lens diameter | 14.2 | 14.2 |
| Lens base curve | 8.6 | 8.6 |
| Lens power | +5D ~ -10D | +10D ~ -20D |
| Refractive index | 1.4011 | 1.4011 |
| Light transmittance | >95% | >95% |
| Surface character | Hydrophilic | Hydrophilic |
| Specific gravity | 1.086' | 1.086 |
| Oxygen permeability | 25.38 | 25.38 |
No clinical data is required for this submission, as the safety and efficacy of the contact lens itself has been previously established in 510(k) K080632.
## HOW SUPPLIED/OPENED
Each lens is supplied sterile, in a non-traditional packaging system, the Flat Pack, containing buffered saline solution. Each container of 90 Flat Packs, as well as each Flat Pack is marked with dioptric power, Single Patient Use, Rx Symbol, Sterile Symbol, composition of the lens, manufacturing lot number and expiration date of the lens.
The pack is opened by grasping both top and bottom foil tabs and peeling them apart to fully expose the lens.
Image /page/3/Picture/8 description: The image shows two hands opening a package of disposable contact lenses. The text in the image says "Peel/Pull apart here" and "ONE Day disposable lens". The hands are pulling apart the package to reveal the contact lens inside.
Peel/Pull apart here
Clinical Testing
{4}------------------------------------------------
# Technological Characteristics:
The technological characteristics of the One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens, as compared to the predicate device is illustrated in the following table.
| Pre-Clinical<br>equivalency /Device | One Day Flat Pack<br>(hioxifilcon A)<br>New Device | Clearl-Day®<br>(hioxifilcon A)<br>Predicate Device |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for daily<br>wear for the optical<br>correction of<br>refractive ametropia<br>(myopia and<br>hyperopia) in phakic<br>or aphakic persons<br>with non-diseased<br>eyes. | Indicated for daily<br>wear for the optical<br>correction of<br>refractive ametropia<br>(myopia and<br>hyperopia) in phakic<br>or aphakic persons<br>with non-diseased<br>eyes. |
| Functionality | The contact lenses act<br>as a refractive<br>medium that focus<br>light rays from near<br>and distant objects on<br>the retina. | The contact lenses act<br>as a refractive<br>medium that focus<br>light rays from near<br>and distant objects on<br>the retina. |
| Indications | Daily Disposable,<br>Soft (hydrophilic)<br>contact lens for single<br>use only. | Daily Disposable,<br>Soft (hydrophilic)<br>contact lens for single<br>use only. |
| Production Method | Spun-Cast | Spun-Cast |
| Packaging Method | Flat Pack Technology<br>(FPT) | Blister Pack |
| Lens Orientation<br>upon opening | Anterior side up | Unknown |
| FDA Group # | Group #2 >50%<br>Water, non-ionic<br>Polymer | Group #2 >50%<br>Water, non-ionic<br>Polymer |
| USAN name | Hioxifilcon A | Hioxifilcon A |
| Water Uptake (%) | 58.0% | 59.8% |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Menicon Company, Ltd. C/O Mr. Martin Dalsing Official Correspondent 806 Kimball Avenue Grand Junction, CO 81501
APR 2 8 2011
Re: K102895
Trade/Device Name: One Day Flat Pack Regulation Number: 21 CFR 886.5925 Regulation Name: Lenses, Soft Contact, Daily Wear Regulatory Class: Class II Product Code: LPL Dated: April 12, 2011 Received: April 15, 2011
Dear Mr. Dalsing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Martin Dalsing
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
\$\fr\$
Enclosure
{7}------------------------------------------------
# K102895 INDICATIONS FOR USE STATEMENT
One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lenses. DEVICE NAME:
#### INDICATIONS FOR USE:
The One Day Flat Pack (hioxifilcon A) Daily Disposable Soft Contact Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
The lens is intended to be worn once and then discarded at the end of each wearing period on a daily basis. The patient should be instructed to start the next wearing period with a new lens.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X
or
Over-The-Counter Use
(Per 21 CFR 801.109) (Optional Format 1-2-96)
Denise Hampton
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K102895 510(k) Number
Qne Day Flat Pack (hioxifilcon A) Daily Disposable, Soft Contact Lens
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.