CIBA VISION (NELFILCON A) ONE-DAY SOFT CONTACT LENSES

K083216 · Ciba Vision Corporation · LPL · Apr 14, 2009 · Ophthalmic

Device Facts

Record IDK083216
Device NameCIBA VISION (NELFILCON A) ONE-DAY SOFT CONTACT LENSES
ApplicantCiba Vision Corporation
Product CodeLPL · Ophthalmic
Decision DateApr 14, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 886.5925
Device ClassClass 2

Indications for Use

Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.

Device Story

Soft contact lenses composed of nelfilcon A polymer; 69% water content. Designed for daily wear; single-use disposable. Corrects refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia. Manufactured via Lightstream Technology (full mold cast); includes integrated printing step for visibility tint or color additives (phthalocyanine blue/green). Supplied sterile in blister packs with buffered saline. Prescribed by eye care professionals; patient self-applied. Output is optical correction; improves visual acuity for patients. No cleaning or disinfection required; discarded after single use.

Clinical Evidence

No clinical testing performed. Safety and effectiveness established via non-clinical stability testing and comparison to predicate devices with identical material and design characteristics.

Technological Characteristics

Material: 31% nelfilcon A polymer, 69% water. FDA Group 2 (high water, nonionic). Oxygen permeability: 26 barrer. Refractive index: 1.38. Manufacturing: Lightstream Technology (full mold cast). Sterilization: Steam (autoclave). Packaging: Blister pack with phosphate-acetate buffered saline and Poloxamer 108.

Indications for Use

Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism) in non-aphakic persons with non-diseased eyes. Progressives indicated for presbyopia with reading add ≤ +3.00D and astigmatism ≤ 2.00D. Single-use daily disposable.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | <div> <img alt="CIBA VISION Logo" src="..."/> <br/> Shared Passion for Healthy Vision and Better Life </div> | | CIBA VISION Corporation<br>11460 Johns Creek Parkway<br>Duluth, Georgia USA 30097 | 04 February 2009<br>Page 1 of 4<br>v03 | |------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------|----------------------------------------| |------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------|----------------------------------------| CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence 510(k) Summary K 083216 1. Submitter Information: | Company: | CIBA VISION® Corporation<br>11460 Johns Creek Parkway<br>Duluth, Georgia USA 30097 | APR 14 2009 | |-----------------|------------------------------------------------------------------------------------------------------------------|-------------| | Contact Person: | Martina Heim, PhD, RAC<br>Senior Regulatory Specialist, Global Regulatory Affairs<br>martina.heim@cibavision.com | | | Telephone: | 678-415-3565 | | | FAX: | 678-415-3454 | | | Date Prepared: | 27 October 2008 | | 2. Device Name: | Common Name: | Soft Contact Lens | |-------------------------|-------------------------------------------------------------------------| | Trade/Proprietary Name: | Focus® DAILIES®, Focus® DAILIES® Toric,<br>Focus® DAILIES® Progressives | | Classification Name: | Daily Wear Soft Contact Lens | | Device Classification: | Class II [21 CFR 886.5925 (b) (1)] | #### 3. Predicate Device: Predicate devices are: 1) Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses, as cleared under K050065, 2) FreshLook®, FreshLook® Toric, and FreshLook® Progressives (nelfilcon A) One-Day Color Contact Lens, as cleared under K050213. #### 4. Description of Device: The lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). For VISITINT® lenses, the color additive phthalocyanine blue (also known as copper phthalocyanine) is added to the lens material to create a light blue edge to edge color to make them easier to see when handling. The lenses may be printed with inks containing one or more of the following color additives: phthalocyanine blue, phthalocyanine green. {1}------------------------------------------------ | CIBAOVISION. | CIBA VISION Corporation | |---------------------------------------------------|---------------------------| | Shared Passion for Healthy Vision and Better Life | 11460 Johns Creek Parkway | | | Duluth, Georgia USA 30097 | ## CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives One-Day contact lenses include spherical, toric, and multifocal lens designs in the following parameter ranges: - Power Range: -20.00D to +20.00D . - 0.010 mm for -3.00D spherical (varies with power) Center Thickness: . Lenses have the following properties: - 1.38 Refractive index: . - Light transmittance: approximately 96 %T . - Water content : 69% by weight . - Oxygen permeability 26 barrer - measured at 35°C (single point Dk-Polarographic method) Lenses are supplied sterile in sealed blister-packs containing buffered saline. The compatibility and package integrity of the blister-pack packaging system has been demonstrated and successfully used for other CIBA VISION marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blisterpack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date. #### 5. Indications for Use: ◆ Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eves who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the CIBA Vision logo. The logo is black and white and features the words "CIBA VISION" in a bold, sans-serif font. The "O" in "VISION" is stylized to look like an eye. Below the logo is the tagline "Shared Passion for Healthy Vision and Better Life" in a smaller font. CIBA VISION Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097 ## CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence #### દ. Substantial Equivalence: The following matrix summarizes the characteristics of the modified device as compared to the predicate devices: | | Modified Device | K050065 | K050213 | |------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------| | | Focus DAILIES Family<br>(nelfilcon A) One-Day Soft<br>Contact Lens | Focus DAILIES Family<br>(nelfilcon A) One-Day<br>Contact Lens | FreshLook Family<br>(nelfilcon A) One-Day Color<br>Contact Lens | | Lens Material: | nelfilcon A | nelfilcon A | nelfilcon A | | Material<br>Classification: | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) | | Water Content: | 69% | 69% | 69% | | Power Range: | +20.00 to -20.00D | +20.00 to -20.00D | +20.00 to -20.00D | | Visibility Tint: | With or without copper<br>phthalocyanine | With or without copper<br>phthalocyanine | With or without copper<br>phthalocyanine | | Manufacturing<br>Method: | Lightstream Technology™:<br>Full mold cast,<br>integrated print step | Lightstream Technology™:<br>Full mold cast | Lightstream Technology™:<br>Full mold cast,<br>integrated print step | | Lens Designs: | Spherical, toric, multifocal | Spherical, toric, multifocal | Spherical, toric, multifocal | | Sterilization: | Steam sterilization, validated<br>autoclave | Steam sterilization, validated<br>autoclave | Steam sterilization, validated<br>autoclave | | Packaging: | Blister pack | Blister pack | Blister pack | | Package Storage<br>saline solution | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 | #### Non-clinical Testing: Successful stability testing supports the labeled expiration date. Additional chemical, physical, and toxicological testing results can be referenced from the predicate devices. #### Clinical Testing: The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device. ### Substantial Equivalence: The Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses are substantially equivalent to the predicate {3}------------------------------------------------ | 110100<br>CIBAC<br>Colli | | |---------------------------------------------------|--| | Shared Passion for Healthy Vision and Better Life | | CIBA VISION Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097 # CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence lenses and similar to other daily wear soft contact lenses in terms of water content (69% water) and ionic characteristics (FDA Group II: high water, nonionic), clinical performance, and indications for use. Any differences which may exist between the Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses and other Group II soft hydrophilic contact lenses do not adversely affect the safety and effectiveness of the device. :) {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in black. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 CIBA VISION Corporation c/o Martina Heim, Ph.D., RAC Senior Regulatory Specialist 11460 Johns Creek Parkway Duluth, Georgia 30097 APR 1 4-2009 Re: K083216 Trade/Device Name: Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 24, 2009 Received: March 25, 2009 Dear Dr. Heim; We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, M.B. Egeland, MD Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT PART II. : : : : : : : : : : : : : : : K083216 510(k) Number: Device Name: Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives Indications For Use: Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. Section and States and 122 Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: Over the Counter Use I ing-chun Shi (Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices P or 510(k) Number K083216 CONFIDENTIAL
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...