EZVIEW (POLYMACON) SOFT (HYDROPHILLIC) CONTACT LENS FOR DAILY WEAR
K082883 · I-See Vision Technology, Inc. · LPL · Jun 16, 2009 · Ophthalmic
Device Facts
Record ID
K082883
Device Name
EZVIEW (POLYMACON) SOFT (HYDROPHILLIC) CONTACT LENS FOR DAILY WEAR
Applicant
I-See Vision Technology, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Jun 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity. Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.
Device Story
EZvue is a soft, hydrophilic contact lens manufactured via cast molding. Composed of Polymacon (HEMA cross-linked with EGDMA) with 38% water content. Tinted blue with Reactive Blue Dye #19 for handling visibility. Supplied sterile in blister packs with isotonic phosphate buffered saline. Intended for daily wear; requires chemical (non-heat) disinfection. Prescribed by eyecare practitioners for refractive error correction. Benefits include vision correction for ametropic patients.
Clinical Evidence
Clinical study conducted on 60 human eyes over 6 months. Results showed nearly 100% vision correction and high patient satisfaction with lens wear and care. Nonclinical testing included biocompatibility (cytotoxicity, irritation, systemic toxicity), leachability, and steam sterilization validation.
Technological Characteristics
Material: Polymacon (HEMA/EGDMA copolymer). Water content: 38%. Refractive index: 1.430. Light transmittance: >93%. Oxygen permeability: 9x10^-11 (cm2/sec)(mlO2/ml-mmHg) @ 35C. Diameter: 13.90-14.30mm. Power: -0.50D to -12.00D. Center thickness: 0.080-0.150mm. Sterilization: Steam.
Indications for Use
Indicated for daily wear correction of myopia and hyperopia in aphakic/non-aphakic persons with non-diseased eyes and refractive astigmatism up to 2.00 diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
SuperView (polymacon) Contact Lens for Daily Wear (K033136)
Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA number N16895/S087)
Related Devices
K082879 — EZVUE (OCUFILCON D) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR · I-See Vision Technology, Inc. · Jun 16, 2009
K082880 — EZVUE (OCUFILCON D) 1-DAY SOFT (HYDROPHILIC)CONTACT LENS · I-See Vision Technology, Inc. · Jun 16, 2009
K083288 — DISCON PLUS (ETAFILCON A) CONTACT LENS VISIBILITY TINT WITH UV BLOCKER · Innova Vision, Inc. · Sep 3, 2009
K051129 — DISCON (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS FOR DAILY WEAR · Innova Vision, Inc. · Aug 12, 2005
K232839 — Eye Secret 38 UV Aspheric (polymacon) and Air Light 38 UV Aspheric (polymacon) Soft (hydrophilic) Contact Lens for Daily Wear · Yung Sheng Optical Co., Ltd. · Dec 16, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
K082883
# JUN 1 6 2009
## 510 (K) Summary
JUN 16 2009
SUMMARY OF SAFETY AND EFFECTIVENESS FOR EZvue (Polymacon) Soft (Hydrophilic) Contact Lens for Daily Wear
| Submitter Information: | |
|------------------------|---------------------------------------------------------------|
| Company: | I-SEE VISION TECHNOLOGY INC. |
| | 1F, No. 45, Lane 2, Section 2, Kuang-Fu Rd., Hisnchu, Taiwan, |
| | 30071 |
| Contact Person: | Mr. Tom Lin, |
| | Special Assistant of President Office |
| Phone: | 886-3-5752822 |
| FAX: | 886-3-5752823 |
| E-Mail: | tomlin@iseevision.com |
#### Identification of Device:
| Classification Name: | Lens, Soft Contact, Daily Wear, per 21 CFR. 886.5925 |
|-----------------------|-------------------------------------------------------|
| Trade Name: | EZvue (Polymacon) Soft (Hydrophilic) Contact Lens for |
| | Daily Wear |
| Common or usual Name: | Soft (hydrophilic) Contact lens (daily wear) |
| FDA Classification: | Class II |
### Predicate Device:
- BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted છે Contact Lens cleared via K020927 BAUSCH & LOMB®
- SuperView (polymacon) Contact Lens for Daily Wear (K033136) made � from INNOVA VISION INC.
- Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA 0 number N16895/S087) by BAUSCH & LOMB, INC.
#### Indications for Use
EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity..
Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.
{1}------------------------------------------------
#### Description of Device
EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 38% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photo- polymerization. The EZvue (Polymacon) Contact lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.
#### Summary of Clinical Study:
The EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear were tested over 60 human eyes separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.
#### Nonclinical Studies:
A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:
a) Technological characteristics studies
EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:
Diameter range; 13.90~14.30mm
- Power range: -0.50D~-12.00D
Center thickness: varies with power (0.080~0.150mm)
Lenses have the following properties:
Refractive index: 1.430 (hydrated)
Light transmittance: >93%
Water content: 38 %
Oxygen permeability (edged corrected): 9×10~15(cm²/sec)(mlO2/ml-mmHg)]@ 35℃
There characterizations of EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear s are equivalent and comparable to those of predicate lenses.
{2}------------------------------------------------
#### b) Biocompatibility
In accordance with the May 1994 Guidance Document for daily wear contact lenses, toxicity studies have been conducted on the model: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.
#### c) Microbiology
Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.
#### d) Leachability
Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.
#### Substantial equivalence Statement:
Testing performed on the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927), SuperView (polymacon) Contact Lens for Daily Wear (K033136) made from INNOVA VISION INC. , & Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA number N16895/S087) by BAUSCH & LOMB, INC. , when used in accordance with the instructions for use. It is due to the facts that The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.
In conclusion, it is I-SEE's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-- BAUSH &LOMB® SofLens™ MultiFocal (Polymacon) Visibility Tinted Contact Lens (K020927) , SuperView (polymacon) Contact Lens for Daily Wear (K033136) made from INNOVA VISION INC. , & Optima FW™ (polymacon) Visibility Tinted Contact Lenses (NDA number N16895/S087) by BAUSCH & LOMB, INC ..
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three bars representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
I-See Vision Technology Inc. c/o Ms. Jennifer Reich Senior Consultant Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001
Jun 1 6 2009
Re: K082883
Trade/Device Name: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL Dated: June 9, 2009 Received: June 11, 2009
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K081883
Device Name: EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear I-SEE Vision Technnology Inc.
Indications For Use:
EZvue (Polymacon) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
Eyecare practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
MD
Division Sim Division of Onthalm Nose and Throat Dev
Page 1 of 1
Nose and Throat Devices,
510(k) Number
K082883
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.