The VISTAKON® (senofilcon A) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The VISTAKON® (senofilcon A) Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, phakic or aphakit persons with non-diseased eyes who may have 0.75 D of astigmatism or less. The VISTAKON® (senofilcon A) Toric Contact Lens is indicated for daily wear for the correction of visual acuity in phakic or aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less. The VISTAKON® (senofilcon A) Multifocal-Toric Contact Lens is indicated for daily wear for the correction of distance and near vision in presby opic phakic or apakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less. VISTAKON® (senofilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. Eye Care Practitioners may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only.
Device Story
VISTAKON (senofilcon A) contact lens; hemispherical or hemitoric shell; silicone hydrogel material with internal wetting agent; contains benzotriazole UV-absorbing monomer. Available as spherical, multifocal, toric, and multifocal-toric designs. Packaged in buffered saline with 0.005% methyl ether cellulose. Prescribed by eye care practitioners for daily wear; single-use or frequent/planned replacement. Provides refractive correction for myopia, hyperopia, presbyopia, and astigmatism; blocks UV radiation. Patient benefits include vision correction and UV protection. Used by patients; handled by practitioners for fitting.
Clinical Evidence
Three-month clinical study; at least 50 patients; 2:1 ratio of subject device to predicate. Evaluated safety and efficacy for daily wear myopic correction. Measured adverse reactions, keratometry changes, discontinuations, and unscheduled replacements. Results demonstrated similar performance in vision, health, comfort, and fit compared to concurrent controls.
Technological Characteristics
Silicone hydrogel (senofilcon A); 62% polymer, 38% water. FDA Group I material. Includes internal wetting agent and benzotriazole UV-absorbing monomer. Transmittance: <1% UVB (280–315nm), <10% UVA (316–380nm). Hemispherical/hemitoric form factor. Sterile, buffered saline packaging.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, and astigmatism (up to 10.00D) in phakic or aphakic persons with non-diseased eyes; includes multifocal options for presbyopia. Contraindicated for patients with diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Focus® NIGHT AND DAY™ (lotrafilcon A)
Related Devices
K073485 — VISTAKON (NARAFILCON) A SOFT CONTACT LENS · Johnson & Johnson Vision Care, Inc. · Mar 3, 2008
K100349 — VISTAKON (NARAFILCON B) CONTACT LENS VISIBILITY TINTED WITH UV BLOCKER · Johnson & Johnson Vision Care, Inc. · May 21, 2010
K130342 — SAUFLON CLARITI (SOMOFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER · Sauflon Pharmaceuticals, Ltd. · Aug 5, 2013
K032340 — VISTAKON (GALYFILCON A) SOFT CONTACT LENS · Johnson & Johnson Vision Care, Inc. · Oct 16, 2003
K210930 — ACUVUE® OASYS MAX 1-DAY Contact Lenses (senofilcon A) · Johnson & Johnson Vision Care, Inc. · Nov 29, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
042275
i
510(k) PREMARKET NOTIFICATION VISTAKON®® (senofilcon A) Contact Lens
## Summary of Safety and Effectiveness
| Submitter<br>Information | Company: | VISTAKON®<br>Division of Johnson & Johnson Vision Care, Inc.<br>7500 Centurion Parkway<br>Suite 100<br>Jacksonville, FL 32256 |
|---------------------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Annette M. Hillring<br>President<br>Hillring & Associates, Inc. |
| | Telephone: | 813-269-8246 |
| | FAX: | 813-908-8706 |
| | Date Prepared: | August 20, 2004 |
| Identification of<br>the Device | Common Name: | Soft Contact Lens |
| | | Device Name: |
| | | Classification Name: |
| | | Device Classification: |
| Predicate<br>Device | | The predicate device was selected to address both intended use (daily<br>wear) and material type (FDA Group I; low water, nonionic polymer):<br>• Focus® NIGHT AND DAY™ (lotrafilcon A), FDA Group I. |
Continued on next page
and the same of the same of the same of the same of the same of the same of the same of the seat the seat the seat the seat of the seat of the seat of the seat of the seat of
{1}------------------------------------------------
| Description of<br>Device | • The VISTAKON® (senofilcon A) Contact Lens Clear and Visibility<br>Tint with UV Blocker is available as a spherical lens, a multifocal<br>lens, a toric lens and a multifocal-toric lens. |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • The lenses are made of a silicone hydrogel material containing an<br>internal wetting agent. |
| | • The VISTAKON® (senofilcon A) Contact Lens may be tinted blue<br>using Reactive Blue Dye #4 to make the lens more visible for<br>handling. |
| | • A benzotriazole UV absorbing monomer is used to block UV<br>radiation. The transmittance characteristics are less than 1% in the<br>UVB range of 280 – 315nm and less than 10% in the UVA range of<br>316 – 380nm. |
| | • The VISTAKON® (senofilcon A) Contact Lens is a hemispherical or<br>hemitoric shell. |
| | • The lens is supplied in a sterile state, packaged in a buffered saline<br>solution with 0.005% methyl ether cellulose. |
| | • The composition of the lens is 62% senofilcon A and 38% water by<br>weight when hydrated and stored in the buffered saline solution. |
and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contra
{2}------------------------------------------------
#### Indications for Use
| Lens Design | Indication |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Spherical | The VISTAKON® (senofilcon A) Contact Lens is indicated<br>for daily wear for the correction of refractive ametropia<br>(myopia and hyperopia) in phakic or aphakic persons with<br>non-diseased eyes who may have 1.00D of astigmatism or<br>less. |
| Multifocal | The VISTAKON® (senofilcon A) Multifocal Contact Lens<br>is indicated for daily wear for the correction of distance and<br>near vision in presbyopic, phakic or aphakic persons with<br>non-diseased eyes who may have 0.75D of astigmatism or<br>less |
| Toric | The VISTAKON® (senofilcon A) Toric Contact Lens is<br>indicated for daily wear for the correction of visual acuity in<br>phakic or aphakic persons with non-diseased eyes that are<br>hyperopic or myopic and may have 10.00D of astigmatism<br>or less. |
| Multifocal Toric | The VISTAKON® (senofilcon A) Multifocal-Toric Contact<br>Lens is indicated for daily wear for the correction of<br>distance and near vision in presbyopic phakic or aphakic<br>persons with non-diseased eyes who may have 10.00D of<br>astigmatism or less. |
- VISTAKON® (senofilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
- Eye care practitioners may prescribe the lens for single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement, the lens may be disinfected using a chemical disinfection system only.
{3}------------------------------------------------
Technological Characteristics The characteristics of the VISTAKON® (senofilcon A) Contact Lens are compared to the characteristics of the predicate device, Focus® NIGHT & DAY™ (lotrafilcon A) Contact Lens, in the following tables.
| Material Comparison | | |
|---------------------|----------------------------------------------------|---------------------------------------|
| | Focus® NIGHT AND DAY™ (lotrafilcon A) Contact Lens | VISTAKON® (senofilcon A) Contact Lens |
| Material | lotrafilcon A | senofilcon A |
| Type | Group I | Group I |
| Parameter Comparison | | | | |
|----------------------------|----------------------------------------------------|---------|---------------------------------------|---------|
| | Focus® NIGHT AND DAY™ (lotrafilcon A) Contact Lens | | VISTAKON® (senofilcon A) Contact Lens | |
| | Measured | Labeled | Measured | Labeled |
| Water Content, % | 23 | 24 | 38 | 38 |
| Refractive Index @<br>20°C | 1.43 | 1.43 | 1.42 | 1.42 |
| Dk*, edge corrected | 141 | 140 | 107 | 103 |
| Dk*, non-edge<br>corrected | 168 | NA | 126 | 122 |
| Base Curve, mm | 8.47 | 8.4 | 8.81 | 8.8 |
| Diameter, mm | 13.80 | 13.8 | 14.62 | 14.6 |
| Power, D | -1.11 | -1.00 | -0.35 | -0.50 |
*Dk units: 10 -11(cm2/sec)(ml O2/ml x mmHg)
{4}------------------------------------------------
A series of in vitro and in vivo preclinical toxicology and Non-clinical Testing biocompatibility tests were performed to assess the safety and effectiveness of the contact lens. All non-clinical toxicology tests were conducted in accordance with the GLP regulation (21 CFR Part 56). All other testing was conducted according to valid scientific protocols.
> The results of the non-clinical testing on the VISTAKON® (senofilcon A) Contact Lens demonstrate that:
- · the lens material and extracts are not toxic and not irritating, and
- · lens physical and material properties are consistent with currently marketed lenses.
- Clinical Testing A three-month clinical study was completed to evaluate the safety and efficacy of the VISTAKON® (senofilcon A) Contact Lens when worn by myopic patients on a daily wear basis. The clinical study was conducted in accordance with FDA's 1994 Guidance Document for Daily Wear Contact Lenses.
The study evaluated at least 50 patients with a 2:1 ratio of subject device to predicate device for three (3) months. Parameters measured included adverse reactions, keratometry changes, reasons for discontinuations, and the number of reasons for unscheduled lens replacements.
Clinical evaluation demonstrated similar overall performance in the clinically relevant areas of vision, health, comfort and fit as compared to concurrent controls when used under daily wear conditions.
{5}------------------------------------------------
#### Conclusions Drawn from Studies
| Validity of<br>Scientific Data | Toxicology studies were conducted under the Good<br>Laboratory Practices Regulations by a contract laboratory.<br>Microbiology, chemistry, shelf-life stability, and<br>leachability studies were conducted by VISTAKON®<br>laboratories and followed scientific protocols. The data<br>were determined to be scientifically valid under 21 CFR<br>860.7 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence | Information presented in this Premarket Notification<br>establishes that the VISTAKON® (senofilcon A) Contact<br>Lens is as safe and effective as the predicate device when<br>used in accordance with the labeled directions for use and<br>for the requested indication. |
| Risk and<br>Benefits | The risks of the subject device as the same as those<br>normally attributed to the wearing of soft (hydrophilic)<br>contact lenses on a daily wear basis. The benefits to the<br>patient are the same as those for other soft (hydrophilic)<br>contact lenses. |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 8 2004
Vistakon c/o Ms. Annette M. Hillring President, Hillring & Associates, Inc. 3012 St. Charles Drive Tampa, FL 33618
Re: K042275
Ro-ZZ75
Trade/Device Name: Vistakon8 (senofilcon A) Contact Lens for Daily Wear (Clear and Visibility Tint with UV Blocker) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL; MVN Dated: August 20, 2004 Received: August 23, 2004
Dear Ms. Hillring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. kelpi Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use
510(k) Number:
Device Name:
VISTAKON® (senofilcon A) Contact Lens Clear and Visibility Tint, with UV blocker
Indications for Use:
The VISTAKON® (senofilcon A) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
The VISTAKON® (senofilcon A) Multifocal Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, phakic or aphakit persons with non-diseased eyes who may have 0.75 D of astigmatism or less.
The VISTAKON® (senofilcon A) Toric Contact Lens is indicated for daily wear for the correction of visual acuity in phakic or aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less.
The VISTAKON® (senofilcon A) Multifocal-Toric Contact Lens is indicated for daily wear for the correction of distance and near vision in presby opic phakic or apakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.
VISTAKON® (senofilcon A) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Eye Care Practitioners may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------|---------|
| Division of Ophthalmic Ear, Nose and Throat Devises | |
| 510(k) Number | K042275 |
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.