Leo Colors and Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses 1. are indicated for daily wear to enhance or alter the apparent color of the eye, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for the management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear to 2. enhance or alter the apparent color of the eye, including ocular masking, either in sighted or nonsighted eyes that require a prosthetic contact lens for the management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by persons who have astigmatism of 12.00 diopters or less or for occlusive therapy for conditions such as diplopia, amblyopia an extreme photophobia. When prescribed for Frequent/Planned Replacement Wear, the Leo Colors, Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are to be cleaned, rinsed and disinfected each time they are removed from the eye and discarded after the recommended wearing period prescribed by the eye care practitioner. When prescribed for Disposable Wear, the wearing time prescribed by the eye care practitioner is for daily wear (single use). Patients should be instructed to discard the lenses at each removal.
Device Story
Soft hydrophilic contact lenses composed of methafilcon A; random co-polymer of hydroxyethylmethacrylate and methacrylic acid cross-linked with ethyleneglycol dimethacrylate. Lenses modified by affixing colorants (Titanium Dioxide, C.I. Reactive Blue 21, C.I. Reactive Red 180, C.I. Reactive Yellow 15, C.I. Reactive Black 5) to the front surface corresponding to the iris. Manufactured via molding or FIPS II; tinting via post-forming ink jetting. Prescribed by eye care practitioners for daily wear; frequent/planned replacement or disposable (single-use) modalities. Patients clean/disinfect lenses or discard after single use. Provides refractive correction or prosthetic/occlusive therapy. Benefits include improved visual acuity, cosmetic enhancement, or management of ocular abnormalities.
Clinical Evidence
Bench testing only. No clinical studies were conducted. Safety and efficacy established through substantial equivalence to predicate devices, including toxicology testing (ocular irritation, cytotoxicity, systemic toxicity) and leachability studies showing <1ppm dye extraction.
Technological Characteristics
Material: Methafilcon A (hydroxyethylmethacrylate/methacrylic acid co-polymer). Water content: 55%. Lens Group 4 (high water, ionic). Dk: 15.04. Index of Refraction: 1.41. Powers: -20.00 to +20.00D. Tinting: Ink jetting post-lens forming. Colorants: Titanium Dioxide, C.I. Reactive Blue 21, C.I. Reactive Red 180, C.I. Reactive Yellow 15, C.I. Reactive Black 5.
Indications for Use
Indicated for daily wear to enhance/alter eye color or provide ocular masking for corneal, iris, or lens abnormalities in sighted/non-sighted eyes. Also for correction of myopia, hyperopia, and astigmatism (up to 2.00D for non-toric, up to 12.00D for toric) in aphakic/not-aphakic persons with non-diseased eyes, or for occlusive therapy for diplopia, amblyopia, or extreme photophobia.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Frequency 55 (methafilcon A) Soft (hydrophilic) Contact Lens (K971164)
Frequency Colors and Expressions (methafilcon A) Soft (hydrophilic) Contact Lenses (K001090)
Submission Summary (Full Text)
{1}------------------------------------------------
## Goper Vision P
Leo Colors Methafilcon A
510(k) Summary
Premarket Notification
# Ko41138
- 1. SUBMITTER:
## Submitted on Behalf of:
- Company Name:
- I Address:
- I Phone:
- Fax:
## 2. CONTACT PERSON:
- Company Name:
- E Address:
- 트 Phone:
- 트 Fax:
## Bonnie Tsymbal
April 28, 2004
Unit 2. South Point
011 44 2380 605200 011 44 2380 605299
Hamble SO3 4RF Southampton UK
CooperVision, Inc. 711 North Road Scottsville, NY 14546 (585) 264-3210 (585) 889-5688
CooperVision Manufacturing, Ltd.
- 3. DATE SUMMARY PREPARED:
## 4. DEVICE IDENTIFICATION:
- 프 Trade Name:
Leo Colors, Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lens
- I Common Name: Hydrophilic Soft Contact Lens
- I Classification
- 트 Device Classification:
Lenses, Soft Contact, Daily Wear 86LPL Class II (21 CFR 886.5925)
## 5. DEVICE DESCRIPTION:
The Leo Colors and Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are available as spherical lenses. Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are available as astigmatic lenses. The lens material, methafilcon A, is a random co-polymer of hydroxyethylmethacrylate and methacrylic acid, which is cross-linked with ethyleneglycol dimethacrylate. The lenses are made by modifying the uncolored methafilcon A base lens by affixing colorants on that portion of the front surface that corresponds to the iris. The colorants Titanium Dioxide, C.I. Reactive Blue 21, C.I. Reactive Red 180, C.I. Reactive Yellow 15, and C.I Reactive Black 5 are listed in 21 CFR Part 73.
{2}------------------------------------------------
| Chemical Name | CI Number | CFR# |
|---------------------------------------------------|-------------------------|---------|
| Titanium dioxide | 77891 | 73.3126 |
| Poly (hydroxyethyl methacrylate)-dye<br>copolymer | C.I. Reactive Blue 21 | 73.3121 |
| Poly (hydroxyethyl methacrylate)-dye<br>copolymer | C.I. Reactive Red 180 | 73.3121 |
| Poly (hydroxyethyl methacrylate)-dye<br>copolymer | C.I. Reactive Yellow 15 | 73.3121 |
| Poly (hydroxyethyl methacrylate)-dye<br>copolymer | C.I. Reactive Black 5 | 73.3121 |
### 6. INTENDED USE
- 1. Leo Colors and Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear to enhance or alter the apparent color of the eve, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for the management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia and hyperopia) in aphakic and notaphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia.
- 2. Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear to enhance or alter the apparent color of the eye, including ocular masking, either in sighted or nonsighted eyes that require a prosthetic contact lens for the management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with nondiseased eyes. The lenses may be worn by persons who have astigmatism of 12.00 diopters or less or for occlusive therapy for conditions such as diplopia an extreme photophobia.
#### Frequent/Planned Replacement Wear
When prescribed for Frequent/Planned Replacement Wear, the Leo Colors, Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are to be cleaned, rinsed and disinfected each time they are removed from the eye and discarded after the recommended wearing period prescribed by the eye care practitioner.
### Disposable Wear
When prescribed for Disposable Wear, the wearing time prescribed by the eve care practitioner is for daily wear (single use). Patients should be instructed to discard the lenses at each removal.
{3}------------------------------------------------
#### 7. PREDICATE DEVICE:
The Leo Colors and Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses for daily wear manufactured by Cooper\ision Manufacturing Ltd. Is equivalent to the Frequency 55 (methafilcon A) Soft (hydrophilic) Contact Lens (K971164) and Frequency Colors and Expressions (methafilcon A) Soft (hydrophilic) Contact Lenses for daily wear (K001090)
As the methafilcon A base lens is produced at the same manufacturing facility as the Frequency 55, Frequency Colors and Expressions, the physical, optical and chemical properties are the same. They are all in Lens Group 4, high water ionic polymer as established by FDA and located in the May 1994 Guidance Document for Daily Wear Contact Lenses.
| | Leo Colors<br>Subject Device | Frequency 55<br>Predicate Device<br>K971164 | Frequency Colors<br>Predicate Device<br>K001090 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Methafilcon A | Equivalent | Equivalent |
| Material<br>Classification | Hydrophilic lens, Group 4 | Equivalent | Equivalent |
| Indication for use | Daily Wear<br>Myopia, hyperopia and<br>astigmatism | Daily Wear<br>Myopia and hyperopia | Daily Wear<br>Myopia, hyperopia and<br>astigmatism |
| Water Content | 55% | Equivalent | Equivalent |
| Light transmittance | >90.0% | Equivalent | Equivalent |
| Dk (35°) | 15.04 | Equivalent | Equivalent |
| Index of Refraction | 1.41 | Equivalent | Equivalent |
| Powers | -20.00 to +20.00D | Equivalent | Equivalent |
| Colorants | Titanium Dioxide<br>CI Reactive Blue 21<br>CI Reactive Red 180<br>CI Reactive Yellow 15<br>CI Reactive Black 5. | CI Reactive Blue 4 | Carbozale violet<br>chromim oxide green<br>dihydrodinaphto brown<br>dihydrodioxo yellow<br>phthalocyanine green<br>iron oxide red<br>iron oxide brown<br>iron oxide black<br>phthalocyanine blue<br>and titanium dioxide |
| Tint Process | Ink jetting<br>post lens forming. | In-monomer tint | Pad printing<br>post lens forming |
| Manufacturing<br>Method | Molded or FIPS II | Molded | Molded or FIPS II |
{4}------------------------------------------------
#### PRECLINICAL INFORMATION 8.
The results of toxicology testing, including Ocular Irritation, Cytoxicity and Systemic Toxicity have demonstrated that the subject lens is non-toxic.
A leachability study was conducted to assess the color fastness of the listed dyes used to tint the Leo Colors. Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses. The study demonstrates that after two weeks of extraction at 37°C in saline, undetectable levels (<1ppm) of dye were observed in the extraction solution.
The physical, optical and chemical properties of the subject lens are equivalent to the predicate device.
#### ഗ CLINICAL DATA
It was determined that Clinical Studies were not necessary to establish the safety and efficacy of the Leo Colors, Lens Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses. This determination was based on the following:
- . The Leo Colors, Lens Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses were demonstrated to be substantially equivalent to the predicate Frequency 55 (methafilcon A) Soft (hydrophilic) Contact Lens (K971164) and (K973063)
- . The Leo Colors, Lens Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses were demonstrated to be substantially equivalent to the predicate Frequency Colors and Expressions (methafilcon A) Soft (hydrophilic) Contact Lens (K001090).
## 10. CONCLUSION
The information provided in this 510(k) established that the Leo Colors. Lens Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses is equivalent in optical, chemical and physical properties of the predicate device and does not raise any questions of safety and effectiveness. Therefore, the device is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 3 2004
Cooper Vision c/o Ms. Bonnie Tsymbal Manager, Regulatory Affairs 711 North Road Scottsville, NY 14546
Re: K041138
Trade/Device Name: Leo Colors, Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL; MVN Dated: April 29, 2004 Received: May 3, 2004
Dear Ms. Tsymbal:
We have reviewed your Section 510(k) premarket notification of intent to market the device w oneve roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass bated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Ms. Bonnie Tsymbal
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Levy Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for CooperVision. The logo is in black and white and features the company name in a stylized font. To the right of the name is a square with a circle inside.
Regulatory Affairs 711 North Road Scottsville, NY 14546 (585) 385-6810 Fax: (585) 889-5688
#### Indication for Use Statement
Device Name:
Leo Colors (methafilcon A) Soft (hydrophilic) Contact Lenses Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses
#### Indication for Use:
- Leo Colors and Leo Aspheric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses l. are indicated for daily wear to enhance or alter the apparent color of the eye, including ocular masking, either in sighted or non-sighted eves that require a prosthetic contact lens for the management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity or for occlusive therapy for conditions such as diplopia. amblyopia or extreme photophobia.
- Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear to 2. enhance or alter the apparent color of the eye, including ocular masking, either in sighted or nonsighted eyes that require a prosthetic contact lens for the management of conditions such as corneal. iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eves. The lenses may be worn by persons who have astigmatism of 12.00 diopters or less or for occlusive therapy for conditions such as diplopia, amblyopia an extreme photophobia.
#### Frequent/Planned Replacement Wear
When prescribed for Frequent/Planned Replacement Wear, the Leo Colors, Leo Aspheric Colors and Leo Toric Colors (methafilcon A) Soft (hydrophilic) Contact Lenses are to be cleaned, rinsed and disinfected each time they are removed from the eye and discarded after the recommended wearing period prescribed by the eye care practitioner.
#### Disposable Wear
When prescribed for Disposable Wear, the wearing time prescribed by the eye care practitioner is for daily wear (single use). Patients should be instructed to discard the lenses at each removal.
#### PLEASE DO NO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Fis 6
510(k) Number K041138
Prescription Use
(Per 21 CFR 801.109) ✓
AND/OR
Over-The-Counter
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.