← Product Code [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL) · K023077

# OCU-FLEX PEDIATRIC APHAKIC (OCUFILCON B) SPHERICAL SOFT CONTACT LENS FOR DAILY WEAR (LATHE-CUT) (K023077)

_Ocu-Ease Optical Products, Inc. · LPL · Dec 10, 2002 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K023077

## Device Facts

- **Applicant:** Ocu-Ease Optical Products, Inc.
- **Product Code:** [LPL](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL.md)
- **Decision Date:** Dec 10, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens are indicated for daily wear for the correction of hyperopia in pediatric aphakic persons. The lens may be disinfected with chemical (not heat) disinfection system.

## Device Story

The OCU-FLEX PEDIATRIC APHAKIC is a lathe-cut, spherical soft contact lens fabricated from ocufilcon B (a copolymer of 2-HEMA and methacrylic acid cross-linked with EGDMA). In its hydrated state (53% water), the lens is soft, pliable, and conforms to the corneal curvature. It is designed specifically for pediatric patients, featuring a smaller lens diameter and higher power range (+10.00 to +40.00 diopters) to address the refractive needs of pediatric aphakia. The device is intended for daily wear and requires chemical disinfection. It is prescribed by eye care professionals to correct hyperopia, providing a transparent optical surface that benefits the patient by restoring visual acuity.

## Clinical Evidence

Bench testing only. Pre-clinical performance data for ocufilcon B (cytotoxicity, systemic injection, ocular irritation) was referenced from PMA P820051. Data for the visibility tint (Reactive Blue 21) was referenced from K960291.

## Technological Characteristics

Material: ocufilcon B (47% polymer, 53% water). Copolymer of 2-HEMA and methacrylic acid cross-linked with EGDMA. Refractive index: 1.41. Oxygen permeability (Dk/t): 18.1 x 10^-11. Specific gravity: 1.18. Light transmission: >95%. Form factor: Lathe-cut spherical soft contact lens. Available in clear or blue visibility tint (Reactive Blue 21). Disinfection: Chemical only (no heat).

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Pediatric/Aphakic lens (methafilcon A) manufactured by Flexlens Products ([K950294](/device/K950294.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## DEC 1 0 2002

# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

K023077 The assigned 510(k) number is:

## Applicant information:

| Date Prepared:                   | September 12, 2002                                                                                           |
|----------------------------------|--------------------------------------------------------------------------------------------------------------|
| Name:<br>Address                 | Ocu-Ease Optical Products, Inc.<br>629 Tennent Avenue<br>Pinole, CA 94564                                    |
| Contact Person:<br>Phone number: | Charles R. Vermette<br>(800) 521-8984                                                                        |
| USA Consultant:                  | Deanna Werber or Martin Dalsing<br>Medvice Consulting, Inc.<br>623 Glacier Drive<br>Grand Junction, CO 81503 |
| Phone number                     | (970) 243-5490<br>Fax (970) 243-5501<br>Email: dwerber@fdapproval.com                                        |

## Device Information:

| Device Classification: | Class II                                                                                            |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Classification Number: | LPL                                                                                                 |
| Classification Name:   | Lenses, Soft Contact, Daily Wear                                                                    |
| Trade Name:            | OCU-FLEX Pediatric Aphakic (ocufilcon B) Spherical Soft<br>Contact Lens for Daily Wear (lathe-cut). |

{1}------------------------------------------------

## Purpose of 510(k) submission:

## NEW DEVICE ~

Ocu-Ease Optical Products, Inc. proposes to manufacture the OCU-FLEX PEDIATRIC APHAKIC, (ocufilcon B) Spherical Soft Contact Lens for Daily Wear (lathe-cut). Data supporting substantial equivalency to the predicate devices, performance, and safety and efficacy of the OCU-FLEX PEDIATRIC APHAKIC, (ocufilcon B) Soft Contact Lens for Daily Wear (lathe-cut) is contained in this submission.

## Equivalent Device:

The OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Soft Contact Lens for Daily Wear (lathecut) is substantially equivalent to the following predicate devices.

- Pediatric/Aphakic lens (methafilcon A) manufactured by Flexlens Products (K950294) .
## Device Description:

The OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Soft Contact Lens for Daily Wear (lathecut) is fabricated from ocufilcon B which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The aphakic lens designs are made for daily wear for the correction of refractive hyperopia in pediatric aphakic persons. The design elements of the OCU-FLEX PEDIATRIC APHAKIC (ocufileon B) Soft Contact Lens for Daily Wear (lathe-cut) is intended to accommodate pediatric patients. The lens design has a higher power range and smaller lens to compensate for the increased power.

In the hydrated state, the lens conforms to the curvature of the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

{2}------------------------------------------------

The non-ionic lens material, ocufilcon B, is a copolymer of 2-hydroxyethyl methacrylate (2-HEMA) methacrylic acid and cross-linked with ethylene glycol dimethacrylate (EGDMA). It consists of 47% ocufilcon B and 53% water by weight when immersed in normal saline solution buffered with sodium bicarbonate. The lenses are available in clear and with a blue visibility-handling tint, Reactive Blue 21.

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 53% water by weight. The physical properties of the lens are:

| Refractive Index            | 1.41                                                                     |
|-----------------------------|--------------------------------------------------------------------------|
| Light Transmission (clear)  | greater than 95% T                                                       |
| Light Transmission (tinted) | greater than 95% T                                                       |
| Water Content               | 53 %                                                                     |
| Specific Gravity            | 1.18                                                                     |
| Oxygen Permeability         | 18.1 X 10-11 (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt method). |

{3}------------------------------------------------

## Intended Use:

The OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens are indicated for daily wear for the correction of hyperopia in pediatric aphakic persons.

The lens may be disinfected with chemical (not heat) disinfection system.

## Pre-Clinical Performance Data:

## ocufilcon B

Pre-clinical performance data addressing the cytotoxicity test, systemic injection test, and ocular eye irritation test for can be referenced for the ocufilcon B in PMA # P820051. Pre-clinical data addressing Reactive Blue 21 can be referenced in Ocu-Ease Optical Products, Inc. 510(k) K960291.

## Substantial Equivalence:

The device will be manufactured according to specified process controls and a quality assurance program already in place at Ocu-Ease Optical Products, Inc. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Ocu-Ease Optical Products, Inc.

The device is similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate device, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above.

{4}------------------------------------------------

The following table illustrates that the production method, lens function and indications for use of the OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens for Daily Wear (lathe-cut) substantially equivalent to the predicate devices.

|     | CHARACTERISTICS  | OCU-FLEX PEDIATRIC APHAKIC<br>(ocufilcon B) Spherical Soft Contact<br>Lens for Daily Wear                                                                                       | X-Cel Contacts<br>Flexlens Pediatric Aphakic                                                                                                                                                                                                                                                                                                                       |
|-----|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | INDICATION       | Daily wear, Soft Contact Lens                                                                                                                                                   | Daily wear, Soft Contact Lens                                                                                                                                                                                                                                                                                                                                      |
| 2.) | INTENDED USE     | The OCU-FLEX PEDIATRIC<br>APHAKIC (ocufilcon B) Spherical Soft<br>Contact Lens are indicated for daily<br>wear for the correction of hyperopia in<br>pediatric aphakic persons. | The Flexlens 55 (methafilcon A) Soft<br>Contact Lens is indicated for Daily Wear<br>use for the correction of visual refractive<br>ametropia and secialized uses such as<br>atypical ametropia, examples include but<br>are not limited to:  adult and pediatric<br>aphakia and irregular astigmatism<br>created by keratoconus or trauma or post<br>keratoplasty. |
| 3.) | MATERIAL         | HYDROPHILIC                                                                                                                                                                     | HYDROPHILIC                                                                                                                                                                                                                                                                                                                                                        |
| a.  | dk/l             | 18.1                                                                                                                                                                            | 18.8                                                                                                                                                                                                                                                                                                                                                               |
| 4.) | DESIGN           | Aphakic Spherical                                                                                                                                                               | Aphakic Spherical                                                                                                                                                                                                                                                                                                                                                  |
| 5.) | PARAMETERS       |                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                    |
| a.  | Base Curve (mm)  | 6.0 mm to 8.5 mm                                                                                                                                                                | 6.0 mm to 10.8 mm                                                                                                                                                                                                                                                                                                                                                  |
| b.  | Dia (mm)         | 8.0 mm to 11.0 mm                                                                                                                                                               | 10.0 mm to 16.0 mm                                                                                                                                                                                                                                                                                                                                                 |
| c.  | Powers Available | +10.00 to +40.00 diopters                                                                                                                                                       | Plano to +30.00 diopters                                                                                                                                                                                                                                                                                                                                           |
| d.  | Thickness        | Varies with power                                                                                                                                                               | Varies with power                                                                                                                                                                                                                                                                                                                                                  |
| e.  | Optical Zone .   | 4.5mm - 8.0mm                                                                                                                                                                   | 4.5mm - 8.0mm                                                                                                                                                                                                                                                                                                                                                      |

## Substantial Equivalence Table

{5}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

### OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Device Name: Lens for Daily Wear (lathe-cut)

## INDICATIONS FOR USE:

The OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens are indicated for daily wear for the correction of hyperopia in pediatric aphakic persons.

The lens may be disinfected with chemical (not heat) disinfection system.

## (PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

or

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

{6}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is enclosed within a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.

### Public Health Service

Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850

Ocu-Ease Optical Products, Inc. c/o Ms. Deanna Werber Medvice Consulting, Inc. 623 Glacier Drive Grand Junction, CO 81503

Re: K023077

Trade/Device Name: Ocu-Flex Pediatric Aphakic (ocufilcon B) Spherical Soft Contact Lens for Daily Wear (lathe-cut clear & visi-tint) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: September 12, 2002 Received: September 16, 2002

Dear Ms. Werber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{7}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address . http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{8}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

OCU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens for Daily Wear (lathe cut)

### INDICATIONS FOR USE:

ic (ICU-FLEX PEDIATRIC APHAKIC (ocufilcon B) Spherical Soft Contact Lens are indicated for Janv wear for the correction of hyperopia in pediatric aphakic persons.

The lens may be disinfected with chemical (not heat) disinfection system.

## PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic Ear, and Thelon of Ophthalmic East
Nose and Throat Devises 510(k) Number ਨ

Prescription Use (Per 21 CFR 801.109)

(Optional Format 1-2-96)

Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K023077](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/LPL/K023077)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
