← Product Code [HQD](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD) · K991780

# FLUOREX 700 -UV (FLUSILFOCON A), FLUOREX 500-UV (FLUSILFOCON B) AND FLUOREX 300-UV (FLUSILFOCON C) RIGID GAS PERMEABLE C (K991780)

_G.T. Laboratories, Inc. · HQD · Jun 29, 1999 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K991780

## Device Facts

- **Applicant:** G.T. Laboratories, Inc.
- **Product Code:** [HQD](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD.md)
- **Decision Date:** Jun 29, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5916
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The spherical lens is intended to be used in the correction of visual acuity in eyes that are myopic or hyperopic. Bifocal and trifocal designs are available for the correction of visual acuity in eyes that are also presbyopic.

## Device Story

Rigid gas permeable contact lenses composed of fluoro-silicate-acrylic copolymers with an added UV-absorbing compound (Cyasorb UV-416). Lenses function as a refractive medium when placed on the cornea to focus light onto the retina. Designed for daily wear to correct visual acuity in myopic, hyperopic, or presbyopic patients. Prescribed by eye care professionals. UV blocker provides protection by reducing light transmission in the 250-350nm range to 0.0%.

## Clinical Evidence

No clinical data provided; substantial equivalence established via bench testing and material characterization.

## Technological Characteristics

Materials: Fluoro-silicate-acrylic-copolymers with Cyasorb UV-416. Standards: ANZI/ASTM D542-20 (refractive index), ANZI Z80.6 (water content), ANZI/ASTM 280.6.4.5 (light transmission). Properties: Dk 29.3-70.0, Shore D hardness 85.5-87.0, water content <1.0%.

## Regulatory Identification

A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.

## Predicate Devices

- Fluorex® 700 (flusilfocon A) Rigid Gas Permeable Contact Lenses ([P880001](/device/P880001.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 2 9 1999_510(K) SUMMARY FOR PREEDOM OF INFORMATION

510(K) SUMMARY FOR FREEDOM OF INFORMATION

Addition of UV blocker to Fluorex® 700, 500, 300 (flusilfocon A, B, C) Material for Rigid Permeable Contact Lenses

| 1. Submitted by: | G. T. Laboratories, Inc<br>2007 Johns Drive<br>Glenview, IL 60025-1616 |
|------------------|------------------------------------------------------------------------|
|------------------|------------------------------------------------------------------------|

John M. Szabocsik, Ph.D. Contact Person: Szabocsik and Associates 203 N. Wabash, Ste 1200 Chicago, IL 60601 TEL: 312-553-0828

May 18, 1999 2. Date Prepared

Contact Lens 3. Common/Usual

- Fluorex® 700-UV (flusilfocon A), 4. Proprietary Name Fluorex® 500-UV (flusilfocon B), Fluorex® 300-UV (flusilfocon C), Rigid Gas Permeable Contact Lenses
- 5. Device Description The UV blocking materials contain a UV-absorbing compound incorporated into the fluoro-silicate-acryliccopolymers (flusilfocon A, B, C). The Fluorex® UV (flusilfocon) Rigid Gas Permeable Contact Lens when placed on a cornea acts as a refracting medium to focus light rays onto the retina.
- 6. Intended Use The spherical lens is intended to be used in the correction of visual acuity in eyes that are myopic or hyperopic. Bifocal and trifocal designs are available for the correction of visual acuity in eyes that are also presbyopic.
- 7. Equivalence: The sponsor considers these lenses to be substantially equivalent to Fluorex® 700 (flusilfocon A) Rigid Gas Permeable Contact Lenses, approved under PMA #P880001. The attached table summarizes the comparison of the new materials to the predicate device.

{1}------------------------------------------------

|                                                                                | SUBSTANTIAL EQUIVALENCE                                                                                                                                                                                                                            |                    |                                 |                 |              |                 |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|---------------------------------|-----------------|--------------|-----------------|
| Material Type                                                                  | Fluorex® 700                                                                                                                                                                                                                                       | Fluorex® 700-UV    | Fluorex® 500                    | Fluorex® 500-UV | Fluorex® 300 | Fluorex® 300-UV |
| Material<br>Description                                                        | Fluoro-silicate-acrylic-copolymers. Tinted lenses contain one or more of the<br>following dyes: D&C Green dye #6, D&C Yellow dye #10. D&C Violet #2, D&C Red dye<br>#17, CI Solvent Yellow 18. The UV absorbing lenses also contain Cyasorb UV-416 |                    |                                 |                 |              |                 |
| Actions                                                                        | When placed on the human cornea, the hydrated lens acts as a corrective refracting<br>medium to focus light rays on the retina.                                                                                                                    |                    |                                 |                 |              |                 |
| Chord Diameter*                                                                | <                                                                                                                                                                                                                                                  | 7.00 to 10.50 mm   | ->                              |                 |              |                 |
| Center thickness*                                                              | <                                                                                                                                                                                                                                                  | 0.10mm to 1.00mm   | ->                              |                 |              |                 |
| Base Curve*                                                                    | <                                                                                                                                                                                                                                                  | 6.00 to 9.00 mm    | ->                              |                 |              |                 |
| Powers*                                                                        | <                                                                                                                                                                                                                                                  | -20.00 to +12.00 D | ->                              |                 |              |                 |
| Gas permeability<br>(Dk) (polarographic<br>method )                            | 70.0                                                                                                                                                                                                                                               | 70.0               | 50.0                            | 51.5            | 30.0         | 29.3            |
| Hardness<br>(Shore D hardness<br>ANZI / ASTM )                                 | 85.5                                                                                                                                                                                                                                               | 85.5               | 86.5                            | 86.5            | 87.0         | 87.0            |
| Wetting Angle<br>(Wilhelmy plate<br>method )                                   | 14.8°                                                                                                                                                                                                                                              | 14.8°              | 13.8°                           | 13.8°           | 12.9°        | 12.9°           |
| Water content<br>(ANZI Z80.6; 5.6.1)                                           |                                                                                                                                                                                                                                                    |                    | All materials <1.0%             |                 |              |                 |
| Refractive Index<br>(ANZI/ASTM D542-20)                                        | 1.457                                                                                                                                                                                                                                              | 1.457              | 1.460                           | 1.460           | 1.465        | 1.464           |
| Specific Gravity                                                               | 1.097                                                                                                                                                                                                                                              | 1.085              | 1.110                           | 1.110           | 1.118        | 1.119           |
| Light Transmission,<br>Visible (380-780nm)<br>Clear lens (ANZI/ASTM 280.6.4.5) |                                                                                                                                                                                                                                                    |                    | All materials >90%              |                 |              |                 |
| Light Transmission,<br>UV (250-350nm)                                          |                                                                                                                                                                                                                                                    |                    | All UV absorbing materials 0.0% |                 |              |                 |

、ポイン

ﺑﻬ

*Parameters for spherical lenses

## CHEMISTRY

The Fluorex® -UV (flusilfocon) materials are rigid gas permeable materials for making contact lenses, consisting of a fluorine-containing polyacrylate-silicone blend, and identical to the already approved Fluorex® (flusilfocon ) materials in component monomers, with the only difference being the addition of the UV blocker. No monomers were found to be extracted from the material.

### TOXICOLOGY

The Fluorex® UV (flusilfocon) material was found to be noncytotoxic in an agar overlay cytotoxicity test.

### MICROBIOLOGY

No testing was required, based on the similarity to the predicate device

# CLINICAL

No testing was required, based on the similarity to the predicate device

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Public Health Service

JUN 2 9 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

G.T. Laboratories, Inc. C/O John M. Szabocsik, Ph.D. 203 North Wabash Avenue Suite 1200 Chicago, IL 60601

Re: K991780

Trade Name: Fluorex® 700-UV (flusilfocon A), Fluorex® 500-UV (flusilfocon B) and Fluorex® 300-UV (flusilfocon C) Rigid Gas Permeable Contact Lenses

Regulatory Class: II Product Code: 86 HQD Dated: May 21, 1999 Received: May 25, 1999

Dear Dr. Szabocsik:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895, A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{3}------------------------------------------------

Page 2 - John M. Szabocsik, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Roerl lorentbal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

K991780 510 (k) NUMBER (IF KNOWN)

Fluorex® 700-UV (flusilfocon A), DEVICE NAME Fluorex® 500-UV (flusilfocon B) and Fluorex® 300-UV (flusilfocon C) Rigid Gas Permeable Contact Lenses

INDICATIONS FOR USE

The Fluorex® 700-UV (flusilfocon A), Fluorex® 500-UV (flusilfocon B) and Fluorex® 300-UV (flusilfocon C) Spherical Rigid Gas Permeable Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased eyes that are myopic, hyperopic and which may exhibit astiqmatism of 1.50 diopters or less that does not interfere with visual acuity.

The Fluorex® 700-UV (flusilfocon A), Fluorex® 500-UV (flusilfocon B) and Fluorex® 300-UV (flusilfocon C) Translating Bifocal and Trifocal Contact Lenses are indicated for daily wear for the correction of visual acuity in not-aphakic persons with non-diseased presbyopic eyes that are myopic or hyperopic and which may exhibit astigmatism of 3.50 diopters or less that does not interfere with visual acuity.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED )

Concurrence of CDRH, Office of Device Evaluation (ODE)

X Prescription Use (Per 21 CFR 801.109)

Over-The-Counter-Use OR (Optional Format 1-2-96)

Lionel V. C. Brown, Ph.D.

Division of Ophthalmic Devices 510(k) Number --K991780

is

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K991780](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K991780)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
