The ENFLU 18 (enflufocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The ENFLU 18 (enflufocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters. The ENFLU 18 (enflufocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters. The ENFLU 18 (enflufocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Device Story
ENFLU 18 is a lathe-cut, rigid gas permeable (RGP) contact lens made of siloxanyl fluoroitaconate copolymer (enflufocon A). Lenses are tinted for handling and optionally include a benzophenone UV-blocking monomer. Lenses may be optionally treated with Tangible Hydra-PEG, a covalently bonded polyethylene glycol (PEG) polymer layer (~30nm thick) to improve wettability. Lenses are prescribed by eye care practitioners for daily wear to correct refractive errors or manage irregular corneas. Patients clean and disinfect lenses using chemical (non-heat) systems. The device acts as a refractive medium, focusing light on the retina. Benefits include vision correction and management of corneal irregularities; risks are consistent with standard RGP lens use.
Clinical Evidence
No clinical data. Safety and effectiveness established via non-clinical testing, including GLP-compliant toxicology, biocompatibility, microbiology, and shelf-life stability studies. Physicochemical, mechanical, and optical properties were compared to predicates.
Indicated for daily wear for correction of refractive error (myopia, hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. Also indicated for refractive astigmatism (up to 10.00D for toric, up to 4.00D for multifocal/bifocal), presbyopia (add power up to +4.00D), and management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, or post-refractive surgery).
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Boston ES (enflufocon A) Daily Wear Contact Lens (K943177)
Optimum GP with HPT (roflufocon C, D, and E) Daily Wear Contact Lenses (K161100)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 24, 2017
CONTAMAC Ltd. % Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln West Linn, OR 97068
Re: K171575
Trade/Device Name: ENFLU 18 (enflufocon A) Daily Wear Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: September 11, 2017 Received: September 14, 2017
Dear Bret Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue
Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Denise L. Hampton -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171575
Device Name
ENFLU 18 (enflufocon A) Daily Wear Contact Lens
Indications for Use (Describe)
The ENFLU 18 (enflufocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic persons with non-diseased eyes with myopia or hyperopia.
The ENFLU 18 (enflufocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters.
The ENFLU 18 (enflufocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The ENFLU 18 (enflufocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise nondiseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;"><b></b></span> | <span style="font-size:100%;">Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|----------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:100%;"><b></b></span> | <span style="font-size:100%;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K171575 |
|--------------------------------|--------------------------------------------------------------------------------|
| Applicant information: | |
| Date Prepared: | May 22nd, 2017 |
| Name: | Contamac Ltd. |
| Address | Carlton House<br>Shire Hill<br>Saffron Walden<br>Essex CB11 3AU |
| Contact Person: | Robert McGregor<br>Managing Director |
| Phone number: | 44-1799 542 000 |
| Consultant: | Bret Andre<br>EyeReg Consulting, Inc<br>6119 Canter Ln.<br>West Linn, OR 97068 |
| Phone number | (503) 372-5226 |
## Device Information:
| Device Classification: | Class II |
|------------------------|------------------------------------------------------------------|
| Product Code: | HQD |
| Classification Name: | Daily Wear Rigid Gas Permeable Contact Lens<br>(21 CFR 886.5916) |
| Common Name: | Rigid Gas Permeable Contact Lens |
| Trade Name: | ENFLU 18 (enflufocon A) Daily Wear Contact Lens |
{4}------------------------------------------------
Purpose of 510(k) Submission: ~ New Device ~
## Predicate Devices:
The ENFLU 18 (enflufocon A) Daily Wear Contact Lenses are substantially equivalent to the following predicate devices:
"Boston ES (enflufocon A) Daily Wear Contact Lens" By Bausch & Lomb, Inc. 510(k) number; K943177 -primary predicate
"Optimum GP with HPT (roflufocon C, D, and E) Daily Wear Contact Lenses" by Contamac Ltd. 510(k) number; K161100 -reference predicate
## Device Description:
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens is manufactured from a machine latheable rigid gas permeable material composed of siloxanyl fluoroitaconate copolymer that is tinted for visibility and available with or without an ultraviolet (UV) light absorber.
In the ENFLU 18 (enflufocon A) Daily Wear Contact Lens with UV Blocker, a Benzophenone UV blocking monomer is used to block UV radiation. The UV blocking for ENFLU 18 averages > 98% in the UVB range of 280nm - 315nm and >83% in the UVA range of 316 - 380mm.
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens incorporates a visibility tint to make the lens more visible for handling. The tinted lenses contain one or more of the following color additives: D&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2 and D&C red No. 17.
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens may be optionally treated to incorporate Tangible "M Hydra-PEG-which is a thin polyethylene glycol (PEG)-based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible™ Hydra-PEG, the underlying material (enflufocon A) is encapsulated in a thin layer of polymer that results in measurable improvement of wettability (sessile drop contact angle) compared to untreated lenses. The resulting layer is hydrophilic and approximately 30nm in thickness. The following table depicts the contact angle of the coated vs. uncoated ENFLU 18 (enflufocon A) Daily Wear Contact Lens:
{5}------------------------------------------------
| | ENFLU 18 - Enflufocon A | |
|------------------------------------------------------|-------------------------|-------------------------------|
| | Uncoated* | Tangible TM Hydra-PEG Coated* |
| Average Sessile Drop Contact Angle<br>(degrees) n=30 | 86.5 | 48.0 |
| Standard Deviation | 2.85 | 4.20 |
| % Standard Deviation | 3.30 | 8.75 |
*lenses were manufactured and soaked in borate buffered saline for 24 hours prior to contact angle measurement
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens may be packaged and shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the ENFLU 18 (enflufocon A) Daily Wear Contact Lenses is the PolyVial Contact Lens Case. When shipped "wet", the ENFLU 18 (enflufocon A) Daily Wear Contact Lenses may be packaged and shipped in the Unique pH contact lens care system by Menicon Co., Ltd. The active ingredients in Unique pH solution are Edetate Disodium 0.01% and Polyquaternium 10.0011%.
The physical properties of the ENFLU 18 (enflufocon A) Daily Wear Contact Lens are as follows:
| | ENFLU 18 (enflufocon A) | ENFLU 18 (enflufocon A) with<br>Tangible HydraPEG |
|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Refractive Index (dry) | 1.4436 | 1.4432 |
| Light Transmission (@380-<br>780nm) | >90% | >90% |
| UVA Transmission (@316-<br>380nm) | <17% | <14% |
| UVB Transmission (@280-<br>315nm) | <2% | <1% |
| Specific Gravity (g/cm³) | 1.221 | 1.221 |
| Oxygen Permeability<br>(Dk)<br>ISO/FATT Method | $21.03 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35 °C) | $23.96 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35 °C) |
| Visitint lenses contain one or<br>more of the following color<br>additives conforming to:<br>21 CFR Part 73 & 74, Subpart D | D&C Green No. 6, C.I. Solvent Yellow<br>No. 18, D&C Violet No. 2 and D&C red<br>No. 17 | D&C Green No. 6, C.I. Solvent Yellow<br>No. 18, D&C Violet No. 2 and D&C red<br>No. 17 |
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens is available in the Spherical, Toric, Multifocal/Bifocal, Irregular Cornea (Scleral) design configurations, within the following lens parameters:
- Chord Diameter: 7.0mm to 22.0mm ●
- Center Thickness: Varies
- Base Curve:
●
- Spherical Powers:
- -20.00 Diopters to +20.00 Diopters up to -10.00 Diopters
5.0mm to 8.0mm
- o Toric Powers: Add Powers: up to +4.00 Diopters ●
{6}------------------------------------------------
## Intended Use:
The ENFLU 18 (enflufocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia.
The ENFLU 18 (enflufocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters.
The ENFLU 18 (enflufocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The ENFLU 18 (enflufocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eyes that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
Eye care practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
## Testing:
Non-clinical Testing A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the ENFLU 18 (enflufocon A) Daily Wear Contact Lens. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols. Test results of the nonclinical testing on the ENFLU 18 (enflufocon A) Daily Wear Contact Lens demonstrate that:
- The finished lenses are not toxic and not irritating, ●
- The physicochemical, mechanical, and optical properties of the lenses are ● substantially equivalent to the predicate device.
- . Bioburden levels are below the acceptance criteria (<100 cfu/lens) initially (dry) and following 30 days of storage in Unique pH at ambient temperatures.
- . Physical parameters are stable following 30 days of storage in Menicon Unique pH at ambient temperatures,
- . The surface properties of the lens are stable following 24 days of accelerated aging in Menicon Unique pH.
{7}------------------------------------------------
Clinical Testing The clinical safety and effectiveness has been previously established for contact lenses manufactured from (enflufocon A) and contact lenses treated with Tangible 114 Hydra-PEG.
## Conclusions Drawn from Studies
## Validity of Scientific Data
Laboratories under Good Laboratory Practice regulations conducted toxicology, microbiology, and shelf-life stability studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7
#### Substantial Equivalence
Information presented in this Premarket Notification establishes that the ENFLU 18 (enflufocon A) Daily Wear Contact Lens is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.
#### Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
#### Substantial Equivalence:
Comparison to Predicate Device(s):
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens is substantially equivalent to the Boston ES (enflufocon A) Daily Wear Contact Lens (cleared under K943177) in terms of the following:
- contact lens material (enflufocon A)
- production method: lathe cut manufacturing process
- functionality
- indications for use (sphere, toric, multifocal, irregular cornea) o
The ENFLU 18 (enflufocon A) Daily Wear Contact Lens is substantially equivalent to the Optimum GP with HPT (roflufocon C, D. and E) Daily Wear Contact Lens (cleared under K161100) in terms of the following:
- may optionally incorporate the Tangible™ Hydra-PEG surface coating ●
The following matrix illustrates the production method, lens function and material characteristics of the ENFLU 18 (enflufocon A) Daily Wear Contact Lens as well as the predicate devices.
{8}------------------------------------------------
# Substantial Equivalence Matrix
| | ENFLU 18<br>(enflufocon A)<br>New Device | Boston ES (enflufocon A)<br>Predicate Device | Optimum GP with HPT<br>(roflufocon C, D, E)<br>Predicate Device |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia and/or presbyopia. The<br>lens may also be prescribed for<br>management of irregular corneal<br>conditions such as keratoconus and<br>post graft fitting. | Indicated for daily wear for the<br>correction of visual acuity in<br>aphakic and not aphakic persons<br>with non-diseased eyes with myopia<br>or hyperopia and/or presbyopia.<br>The lens may also be prescribed for<br>management of irregular corneal<br>conditions such as keratoconus and<br>post graft fitting. | Indicated for daily wear for the<br>correction of visual acuity in<br>aphakic and not aphakic persons<br>with non-diseased eyes with<br>myopia or hyperopia and/or<br>presbyopia. The lens may also be<br>prescribed for management of<br>irregular corneal conditions such<br>as keratoconus and post graft<br>fitting. |
| Functionality | The contact lenses act as a<br>refractive medium that focus<br>light rays from near and<br>distant objects on the retina | The contact lenses act as a<br>refractive medium that focus<br>light rays from near and<br>distant objects on the retina | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina |
| Indications | Daily Wear | Daily Wear | Daily Wear |
| Production Method | Lathe-Cut, custom<br>manufactured | Lathe-Cut, custom<br>manufactured | Lathe-Cut, custom<br>manufactured |
| USAN name | enflufocon A | enflufocon A | roflufocon C, D, E |
| Water Content (%) | <1% | <1% | <1% |
| Specific Gravity | 1.221 | 1.220 | Roflufocon C: 1.178<br>Roflufocon D: 1.166<br>Roflufocon E: 1.155 |
| Oxygen Permeability<br>(cm2/sec) (ml O2/ml x mm Hg @<br>35°C) | Uncoated: 21 x 10-11<br>Coated: 24 x 10-11 | 22 x 10-11 | Roflufocon C: 65x10-11<br>Roflufocon D: 100x10-11<br>Roflufocon E: 125x10-11 |
| Wettability<br>(sessile drop contact<br>angle) | Uncoated: 86.5°<br>Coated: 48.0° | 87.5° | Roflufocon C: 40.28°<br>Roflufocon D: 40.40°<br>Roflufocon E: 36.90°<br>(measured by captive bubble<br>advancing contact angle) |
| UV Absorber/Blocker<br>Available | Yes, Optional | Yes | Yes, Optional |
| Includes Hydra-PEG<br>Surface Coating | Yes, Optional | No | Yes |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.