← Product Code [HQD](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD) · K000795

# RIGID GAS PERMEABLE CONTACT LENS (K000795)

_Polymer Technology · HQD · May 26, 2000 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K000795

## Device Facts

- **Applicant:** Polymer Technology
- **Product Code:** [HQD](/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD.md)
- **Decision Date:** May 26, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5916
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

BOSTON XO (hexafocon A) Contact Lens Material is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and not-aphakic persons with nondiseased eyes. The lens may be disinfected using a chemical disinfecting system only.

## Device Story

BOSTON XO is a rigid gas permeable (RGP) contact lens material composed of hexafocon A (aliphatic fluoroitaconate siloxanyl methacrylate copolymer). Designed for daily wear, the lens corrects refractive errors including myopia, hyperopia, astigmatism, and presbyopia. It is intended for use in aphakic and non-aphakic patients with healthy eyes. The material features high oxygen permeability (DK 100) and is hydrophobic with <1% water content. The device is supplied dry and requires chemical disinfection by the user. Clinical performance is evaluated through visual acuity, refractive changes, and slit lamp findings. The lens provides a vision correction benefit to patients by compensating for refractive ametropia.

## Clinical Evidence

Clinical study included 128 eyes (64 patients) with 102 eyes (51 patients) completing the study. Safety endpoints included adverse events, slit lamp findings, symptoms, and keratometry changes. Efficacy endpoints included refractive changes, visual acuity, lens deposits, and wettability. Results concluded the material is equivalent in safety and efficacy to the predicate device. Bench testing included USP Agar Diffusion Cytotoxicity, acute ocular irritation, systemic injection, and 30-cycle solution compatibility testing, showing no significant changes to lens parameters.

## Technological Characteristics

Material: hexafocon A (aliphatic fluoroitaconate siloxanyl methacrylate copolymer). DK: 100. Specific gravity: 1.26. Visible light transmittance: >=70%. Water content: <1%. Form factor: Rigid gas permeable contact lens. Sterilization: Shipped dry; requires chemical disinfection. Color additives conform to 21 CFR 74.3206.

## Regulatory Identification

A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.

## Predicate Devices

- BOSTON ES (enflufocon A) ([K943177](/device/K943177.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

### FOR

## BOSTON XO

#### 1. SUBMITTER INFORMATION:

Polymer Technology Global Vision Care 1400 N. Goodman Street Rochester, New York 14603-0450

#### 2. CONTACT PERSON:

|                 | Manager, Regulatory Affairs    |
|-----------------|--------------------------------|
| Address:        | 1400 North Goodman Street      |
|                 | P.O. Box 450                   |
|                 | Rochester, New York 14603-0450 |
| Telephone No.:  | (716) 338-8638                 |
| Fax No.:        | (716) 338-0702                 |
| E-mail Address: | dketchum@bausch.com            |

Debra Ketchum

#### 3. DEVICE IDENTIFICATION:

| Classification Name: | Rigid Gas Permeable (hydrophobic) Contact<br>Lens Material            |
|----------------------|-----------------------------------------------------------------------|
| Proprietary Name:    | BOSTON XO (hexafocon A) Contact Lens<br>Material                      |
| Common Name:         | fluoro silicone acrylate rigid gas permeable<br>contact lens material |

#### 4. PREDICATE DEVICE:

BOSTON ES (enflufocon A) has been selected as the predicate device for BOSTON XO (hexafocon A).

#### 5. DESCRIPTION OF THE DEVICE:

The BOSTON XO Contact Lens Material, hexafocon A, is composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer.

The color additives conform to 21 CFR part 74.3206. The hexafocon A material has an oxygen permeability, DK of 100, a specific gravity of 1.26, and the lens visible light transmittance of at least 70%. The hexafocon A name has been adopted by the United States Adopted Names Council (USAN).

{1}------------------------------------------------

#### 6. INDICATIONS FOR USE:

The BOSTON XO contact lens material is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and not-aphakic persons with nondiseased eyes. The lens may be disinfected using a chemical disinfection system only.

### Description of Safety and Substantial Equivalence: 7.

A series of preclinical testing were performed to demonstrate the safety and effectiveness of the BOSTON XO contact lens material. The results of all testing demonstrated that the safety and effectiveness of the BOSTON XO is equivalent to the currently marketed BOSTON ES contact lens material. A summary of these results from the preclinical studies is presented below.

### Toxicology:

### In-Vitro Cytotoxicity:

USP Agar Diffusion Cytotoxicity was completed in accordance with USP XXII. The test article meets the requirements of the Agar Diffusion Test.

### Acute Ocular Irritation:

Acute Ocular Irritation test was performed and produced no ocular irritation.

### Systemic Injection

The lens material meets the requirements of the Systemic Injection Test and is considered non-toxic.

### Shelf Life:

The BOSTON XO (hexafocon A) is a hydrophobic rigid gas permeable contact lens material with <1% water content. This material will be shipped dry. The data presented supports substantial equivalence of this BOSTON XO (hexafocon A) contact lens material to the already marketed BOSTON ES (enflufocon A) contact lens material. Based on the Premarket Notification Guidance Document for Daily Wear Contact Lenses, May 12, 1994, shelf-life studies are not required for clearance of this material.

{2}------------------------------------------------

# Solution Compatibility:

Studies were conducted on blue tinted lens material with the ultraviolet light absorber. Lenses were run through 30 cycles of cleaning and light uboorben - Establish the compatibility of the lens material with the conditioning to octabligion. The parameters of ultraviolet and visible light (UV/vis) spectra, base curve, lens diameter, power and surface inguality were recorded prior to and upon completion of 30 cycles. Initial quality were rooorded phot to . There were no significant changes to lens parameters after 30 complete cycles.

# Clinical Testing

Below is a summary of the clinical study carried out to evaluate the safety and efficacy of the BOSTON XO (hexafocon A) contact lens material and onloady of a daily wear contact lens for the correction of visual acuity.

A total of 128 eyes (64 patients) were entered into the study by 3 Investigators. Prior to entry into this study each patient was required to read and sign a Statement of Informed Consent. All patients who signed a Statement of Informed Consent are accounted for in this report. Of the 128 eyes (64 patients enrolled), 102 eyes (51 patients) completed the study.

The safety and efficacy measures for this study were:

Safety: Adverse Events, Positive Slit Lamp Findings, Symptoms/Complaints and Keratometry Changes

Efficacy: Refractive Changes, Lens Visual Acuity, Lens VA Line Changes, Lens Deposits, and Lens Wettability.

The sponsor concludes that BOSTON XO (hexafocon A) contact lens material is equivalent in safety and efficacy to the predicate device, Boston ES (enflufocon A).

#### SUBSTANTIAL EQUIVALENCE 8.

The BOSTON XO contact lens material is substantially equivalent to the currently marketed BOSTON ES contact lens material, which was cleared in 510(k) Premarket Notification No. K943177 on August 25, 1994. The difference between the two devices is a change in the components.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 6 2000

Ms. Debra L.B. Ketchum Manager. Regulatory Affairs Polymer Technology Global Vision Care 1400 N Goodman Street P.O. Box 450 Rochester, NY 14603-0450

Re: K000795 Trade Name: BOSTON XO Contact Lens Regulatory Class: II Product Code: 86 HQD Dated: March 10, 2000 Received: March 13, 2000

Dear Ms. Ketchum:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{4}------------------------------------------------

Page 2 - Ms. Debra L.B. Ketchum

This letter will allow you to begin marketing your device as described in your 510(k) premarket This icter will anow you to begin massantial equivalence of your device to a legally marketed notification. The PDAT micing of substantial vqur device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and IT you desire specific active for your avicedevices), please contact the Office of Compliance at (301) additionally 807.10 for in neve angilosas on the promotion and advertising of your device, please 571-0119. Tructionally, rol question (301) 594-4639. Also, please note the regulation entitled, contact the Office of Complantee as to should (21CFR 807.97). Other general information Wisoranding by relettine to premained nombailing in the Division of Small Manufacturers On your responsibilities and in 1000 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon

Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Polymer Technology 1400 North Goodman Street P.O. Box 450 Rochester, NY 14603-0450

### Indications for Use Statement

510(k) Number (if known): k 000795

Device Name:__________________________________________________________________________________________________________________________________________________________________ BOSTON XO

Indications for Use:

BOSTON XO (hexafocon A) Contact Lens Material is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and nresbyopia) in aphakic and not-aphakic persons with nondiseased eyes. The lens may be disinfected using a chemical disinfecting system only.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter-Use_

BnJuss

(Division Sign-Off) Division of Ophthalmic Devices 510(k) Number_k_000 795

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K000795](https://fda.innolitics.com/submissions/OP/subpart-f%E2%80%94therapeutic-devices/HQD/K000795)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
