K203586 · Corneagen, Inc. · OTZ · Feb 2, 2022 · Ophthalmic
Device Facts
Record ID
K203586
Device Name
EndoSerter-PL
Applicant
Corneagen, Inc.
Product Code
OTZ · Ophthalmic
Decision Date
Feb 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The EndoSerter®-PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4.0 mm incision during endothelial keratoplasty procedures and for loading and storage of donor transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
Device Story
EndoSerter®-PL is a sterile, single-use, manual ophthalmic surgical instrument. It facilitates loading, storage, and transport of donor corneal endothelial allograft tissue by eye bank technicians; subsequently used by ophthalmic surgeons to deliver the tissue into the anterior chamber during endothelial keratoplasty. The device holds the graft in a rolled configuration. By enabling pre-loading and transport, it streamlines surgical workflows. The device is purely mechanical; no electronic or software components are involved. It benefits patients by providing a standardized, sterile method for graft delivery, reducing potential for tissue damage during insertion.
Clinical Evidence
Bench testing only. Studies included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity), tissue handling/stability (endothelial cell loss assessment), sterilization validation (SAL 10^-6), shelf life/transportation (leak testing), aseptic handling (USP <71>), and dimensional/mechanical functional testing. No clinical data presented.
Technological Characteristics
Handheld, manual, single-use ophthalmic instrument. Materials: polycarbonate, isoplast, polyimide, stainless steel, silicone, ABS, and colorants. Biocompatibility per ISO 10993-1. Sterilization via radiation. Dimensions support ≤8.0 mm graft and ≥4.0 mm incision. No software or energy source.
Indications for Use
Indicated for patients requiring endothelial keratoplasty (EK) for corneal endothelial disease (e.g., Fuchs' dystrophy, bullous keratopathy, transplant failure). Used for insertion of corneal endothelial allograft tissue (≤8.0 mm diameter, ≤100 microns thickness) through ≥4.0 mm incision.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
K242389 — EyeGility Inserter for Preloaded enVista IOLs · Bausch & Lomb, Incorporated · Oct 10, 2024
K123736 — CRYSTALSERT DELIVERY SYSTEM · Bausch & Lomb · May 16, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 2, 2022
CorneaGen, Inc % Omid Kodai President Khodai Consulting, Inc. 23 Bellflower Lake Forest, California 92630
Re: K203586
Trade/Device Name: EndoSerter®-PL Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I, reserved Product Code: OTZ Dated: December 22, 2021 Received: December 27, 2021
Dear Omid Kodai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203586
Device Name EndoSerter®-PL
### Indications for Use (Describe)
The EndoSerter® PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4.0 mm incision during endothelial keratoplasty procedures and for loading and storage of donor transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### l. SUBMITTER
CorneaGen, Inc. 101 N Chestnut St Suite 303 Winston Salem NC 27101 Phone: 336-516-9600 Contact Person: Thomas Miller Date Prepared: February 1, 2022
#### DEVICE II.
| Name of Device: | EndoSerter®-PL |
|----------------------|------------------------------------------------------------------------------|
| Common Name: | Graft Insertion Device |
| Classification Name: | Graft Insertion Instrument for Endothelial Keratoplasty (21<br>CFR 886.4300) |
| Regulatory Class: | 1, Reserved |
| Product Code: | OTZ |
#### lll. PREDICATE DEVICE
## Predicate Device:
| 510(k)<br>Number | Trade or Proprietary or<br>Model Name | Manufacturer | Clearance Date |
|------------------|---------------------------------------|-------------------|-----------------|
| K121579 | EK Delivery Device | TDAK Medical Inc. | October 3, 2012 |
This predicate has not been subject to a design-related recall.
## Reference Device:
| 510(k)<br>Number | Trade or Proprietary or<br>Model Name | Manufacturer | Clearance Date |
|------------------|---------------------------------------|--------------|------------------|
| K090626 | EndoSerter® | CorneaGen | January 21, 2011 |
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Endothelial keratoplasty (EK), also referred to as posterior lamellar keratoplasty, is a form of corneal transplantation during which the diseased inner layer of the cornea, the endothelium, is replaced with healthy donor tissue. Diseases that affect the endothelial layer include Fuchs' endothelial dystrophy, aphakic and pseudophakic bullous keratopathy (corneal edema following cataract extraction), and failure or rejection of a previous corneal transplant.
The EndoSerter®-PL is a sterile, single use, handheld, manual ophthalmic surgical instrument. It is used to preload a processed donor corneal endothelial allograft for storage and transportation to the ophthalmic surgeon and to deliver the allograft into the anterior chamber of the eye during corneal surgery. It is designed to deliver a corneal endothelial allograft into the eye during corneal endothelial keratoplasty. The loading and storage of donor tissue for transport to the surgeon is performed by trained technician at the eye bank. Users of this graft insertion instrument for EK are trained ophthalmic surqeons in a healthcare facility or hospital.
#### V. INDICATIONS FOR USE
The indications for use statement for the EndoSerter®-PL by CorneaGen is:
The EndoSerter®-PL is used to insert corneal endothelial allograft tissue measuring less than or equal to 8.0 mm in diameter and 100 microns in central thickness into the anterior chamber through a minimum 4 mm incision during endothelial keratoplasty procedures and for loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 48 hours.
The Indications for Use statement for the EndoSerter®-PL is not identical to the predicate or reference devices; however, the differences do not alter the intended therapeutic use of the device, nor do they affect the safety and/or effectiveness of the device relative to the predicate or reference devices. The subject, predicate, and reference devices have the same intended use for the insertion of corneal endothelial allograft tissue during endothelial keratoplasty procedures.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE AND REFERENCE DEVICES
Graft Insertion for Endothelial Keratoplasty (EK) is the technological principle for both the subject and predicate as well as the reference devices. It is based on the use of insertion instrumentation for corneal tissue graft EK procedures.
At a high level, the subject and predicate devices and when appropriate the reference device are based on the following same technological elements:
- . Same mode of operation in the insertion of corneal endothelial graft for patients requiring Endothelial Keratoplasty (EK)
- Used for loading and storage of donor tissue during transport .
- Delivers a circular endothelial tissue button in a rolled configuration .
- . Provided as a sterile, single use, disposable device
- Loaded by trained technicians at the eye bank .
- . Donor posterior lamellar endothelial graft of central corneal thickness of 100 microns
The following technological differences exist between the subject and predicate devices:
- . Minimum 4 mm incision for the subjective device and 5.1 mm or larger incision for the predicate device. However, the subject device and the reference device are the same in terms of incision size of minimum 4 mm incision.
- Insertion of corneal endothelial allograft tissue measuring less than or equal to . 8.0 mm and the predicate and reference device requiring the insertion of donor tissue diameter of less than or equal to 8.5 mm. This difference of 0.50 mm graft size is negligible and does not introduce any potential impact to safety and/or effectiveness.
- Subject device is manufactured using polycarbonate, isoplast, polyimide, ● stainless steel, silicone, acrylonitrile butadiene styrene (ABS), white, blue, and black colorant. Manufacturing materials for the predicate device is not commercially available or available in public records. The reference device is manufactured using polycarbonate, isoplast, polyimide, stainless steel, white, blue, and black colorant. The addition of silicone and ABS does not introduce any
{6}------------------------------------------------
impact to safety and/or effectiveness based upon biocompatibility testing performed in accordance with ISO 10993-1.
- . Subjective device has tissue storage capability up to 48 hours and the predicate device has tissue storage capability up to 72 hours. This difference does not introduce any potential impact to the safety and/or effectiveness of the subject device in delivering the allograft, and relevant testing has confirmed the stability of the subjective device during tissue handling and stability testing.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for the EndoSerter®-PL device was conducted in accordance with the FDA Guidance issued on September 4, 2020 "Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process"." The battery of testing included the following tests which demonstrated biocompatibility of the device with direct contact to the human tissue:
- Cytotoxicity ●
- Sensitization
- Irritation ●
- Acute Systemic toxicity ●
- . Pyrogenicity
The EndoSerter®-PL device is considered tissue contacting for a duration of less than 24 hours, while the allograft tissue is considered a permanent implant.
## Tissue Handling and Stability Testing
The purpose of this study was to quantitatively determine the endothelial cell damage (endothelial cell loss) when using the EndoSerter®-PL to load, store, transport, and deliver donor allograft tissue. The test results demonstrated that the device did not have any leaking, damage, or shifting of the tissue grafts during the transportation and storage.
{7}------------------------------------------------
# Sterilization Validation
The purpose of this study was to adopt, by equivalency, the EndoSerter®-PL into the existing radiation sterilization process used for the EndoSerter®. The radiation sterilization dose range provided a Sterility Assurance Level (SAL) of 10° in accordance with FDA Guidance issued on January 21, 2016 "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
# Shelf Life and Transportation Testing
The purpose of this study was to perform leak test to demonstrate that the EndoSerter® PL does not leak corneal storage media during storage and transportation over a 48hour period prior to and after accelerated aging. The results of this leak test validated that storage media did not leak from the EndoSerter®-PL during storage and transportation over a 48-hour period prior to and after accelerated aging.
# Aseptic Handling Testing
This testing, performed in accordance with the USP<71>. Method Suitability Test procedures, was conducted to demonstrate the ability to maintain aseptic handling of tissue without increased risk of contamination on transfer to and storage in the EndoSerter®-PL. The testing results demonstrated that the aseptic handling utilized by CorneaGen to load and transport cornea tissue in the EndoSerter®-PL did not increase the risk of contamination of the tissue.
# Dimensional, Functional, and Mechanical Testing
The purpose of this study was to perform dimensional, and mechanical testing utilizing measurement of critical dimensions including additional dimensions relating to mating components, visual inspection, and fit of those components. The study confirmed that the EndoSerter®-PL components perform mechanically and functionally as designed. Therefore, supporting the acceptable use of the device components and materials required in the design and assembly of the EndoSerter®-PL.
#### VIII. CONCLUSION
The EndoSerter®-PL, the subject device, is substantially equivalent to both the predicate and the reference devices' intended use which includes:
- same mode of operation in the insertion of corneal endothelial graft for patients ● requiring Endothelial Keratoplasty (EK)
{8}------------------------------------------------
- . used for loading and storage of donor tissue during transport
- delivering a circular endothelial tissue button in a rolled configuration .
- provided as a sterile, single use, disposable device ●
- . loaded by trained technicians at the eye bank
- . posterior lamellar endothelial graft of central corneal thickness of 100 microns corneal thickness, as well as the duration of time that tissue can be stored within the device.
The minimal technological differences of EndoSerter®-PL, the subject device, to the predicate and the reference devices, do not pose any potential impact to safety and/or effectiveness as demonstrated by the performance testing conducted.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.