← Product Code [OTZ](/submissions/OP/subpart-e%E2%80%94surgical-devices/OTZ) · K121579

# EK DELIVERY DEVICE (K121579)

_Tdak Medical, Inc. · OTZ · Oct 3, 2012 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/OTZ/K121579

## Device Facts

- **Applicant:** Tdak Medical, Inc.
- **Product Code:** [OTZ](/submissions/OP/subpart-e%E2%80%94surgical-devices/OTZ.md)
- **Decision Date:** Oct 3, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4300
- **Device Class:** Class 1
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Intended Use

The EK Delivery Device is to be used by ophthalmic surgeons trained in Endothelial Keratoplasty (EK) procedures and trained in the use of the EK Delivery Device as: An aid facilitating the insertion of donor corneal posterior lamellar endothelial graft of 100 – 220 µm of thickness into the anterior chamber of the eye during endothelial keratoplasty. For use in a 5.1 mm or larger incision and a maximum donor tissue diameter of ≤ 8.5 mm) For loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 72 hours.

## Device Story

The EK Delivery Device is a single-use, sterile, disposable instrument designed for ophthalmic surgeons to insert donor corneal posterior lamellar endothelial grafts into the anterior chamber during Endothelial Keratoplasty. The device consists of four components: a trocar to hold and introduce the graft, a trocar holder serving as an ergonomic handle, an injector assembly (plunger) to push the tissue into the eye, and an end plug for shipping. The device delivers the donor tissue in a rolled configuration. It also functions as a storage container for donor tissue during transport by eye bank technicians for up to 72 hours. By facilitating a controlled insertion through a 5.1 mm or larger incision, the device assists the surgeon in graft placement, potentially improving surgical efficiency and graft handling compared to manual techniques.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including sterilization validation, shelf life, mechanical strength, tissue handling and stability, and biocompatibility testing.

## Technological Characteristics

Single-use, disposable, sterile (radiation) device. Components: Trocar, Trocar Holder, Injector Assembly (plunger), and End Plug. Dimensions: Designed for 5.1 mm or larger incisions and donor tissue ≤ 8.5 mm diameter. Material: Not specified. Connectivity: None. Software: None. Energy source: Manual operation.

## Regulatory Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

## Predicate Devices

- EndoSerter ™ ([K090626](/device/K090626.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121579

ОСТ

3 2012

## 510(k) Summary

# TDAK Medical, Inc.

11575 Sorrento Valley Road Suite 214 San Diego, CA 92121

## SUMMARY

Submitter's name:

TDAK Medical Inc. 11575 Sorrento Valley Road Suite 214 San Diego, CA 92121

Name of contact person:

Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821 greg@regulatoryspecialists.com

Date the summary was revised:

September 12, 2012

Name of the device: Trade or proprietary name Common or usual name Classification Panel

EK Delivery Device EK Delivery Device Endothelial Keratoplasty (EK) Injector Ophthalmic

| Product Code | Classification Regulation | Classification Name      |
|--------------|---------------------------|--------------------------|
| OTZ          | 886.4300                  | Lens, Guide, Intraocular |

The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:

{1}------------------------------------------------

| 510(k) Number | Trade or Proprietary or Model Name | Manufacturer         |
|---------------|------------------------------------|----------------------|
| K090626       | EndoSerter ™                       | Ocular Systems, Inc. |

The Endothelial Keratoplasty Injector is a single-use, disposable device that allows an ophthalmic surgeon to insert a previously prepared disc of posterior donor cornea into the eye of a recipient patient through a small incision during a posterior comeal transplant surgical procedure. The function of the device is to insert the flat disc of donor comeal tissue within a cylindrical tube that is sized to fit into 5.1 mm or larger corneal or scleral incision, and to pull or push this donor corneal tissue into the anterior chamber of the eye as part of a Endothelial Keratoplasty procedure. This device was designed to insert donor posterior corneal tissue ≤ 8.5 mm in diameter, 100um to 220um in thickness, through a 5.1 mm or larger corneal incision.

#### Indications:

The EK Delivery Device is to be used by ophthalmic surgeons trained in Endothelial Keratoplasty (EK) procedures and trained in the use of the EK Delivery Device as:

- · An aid facilitating the insertion of donor corneal posterior lamellar endothelial graft of 100 - 220 um into the anterior chamber of the eve during endothelial keratoplasty.
· For use in a 5.1 mm or larger incision and a maximum donor tissue diameter of . ≤ 8.5 mm).

· For loading and storage of donor tissue during transport to the surgeon by trained eve bank technicians, and for storage of donor tissue for up to a maximum of 72 hours.

Comparison to Predicate: .

The EK Delivery Device is similar to its predicate in basic functionality of inserting corneal tissue during endothelial keratoplasty (EK) procedure. Both are intended to be an aid to delivering the corneal tissue in a convenient form for the ophthalmic physician.

The EK Delivery Device is to be used by ophthalmic surgeons trained in Endothelial Keratoplasty (EK) procedures and trained in the use of the EK Delivery Device as:

{2}------------------------------------------------

- An aid facilitating the insertion of donor posterior lamellar endothelial graft of . 100µm to 220µm in thickness into the anterior chamber of the eye during endothelial keratoplasty.
- For loading and storage of donor tissue during transport to the surgeon by . trained eye bank technicians, and for storage of donor tissue for up to a maximum of 72 hours.

{3}------------------------------------------------

.

: 上

|:

| Product Name                 | Proposed:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate:                                                                                                                                                                                                                                                                                        |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | EK Delivery Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K090626                                                                                                                                                                                                                                                                                           |
|                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EndoSerter™                                                                                                                                                                                                                                                                                       |
| Indications for Use          | The EK Delivery Device is to<br>be used by ophthalmic<br>surgeons trained in<br>Endothelial Keratoplasty (EK)<br>procedures and trained in the<br>use of the EK Delivery<br>Device as:<br>An aid facilitating the<br>insertion of donor posterior<br>lamellar endothelial graft of<br>100µm to 220µm in thickness<br>into the anterior chamber of<br>the eye during endothelial<br>keratoplasty. For use in a 5.1 mm or<br>larger incision and a<br>maximum donor tissue<br>diameter of ≤ 8.5 mm). For loading and storage of<br>donor tissue during transport<br>to the surgeon by trained<br>eye bank technicians, and<br>for storage of donor tissue<br>for up to a maximum of 72<br>hours. | The EndoSerter™ is<br>used to insert corneal<br>endothelial allograft<br>tissue measuring less<br>than or equal to 8.5mm in<br>diameter and 175 micron<br>in central thickness into<br>the anterior chamber<br>through a minimum 4mm<br>incision during<br>endothelial keratoplasty<br>procedures |
| Product Code                 | OTZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | OTZ                                                                                                                                                                                                                                                                                               |
| Regulation Number            | 886.4300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 886.4300                                                                                                                                                                                                                                                                                          |
| Class                        | 1, reserved                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 1, reserved                                                                                                                                                                                                                                                                                       |
| Review Advisory<br>Committee | Ophthalmic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Ophthalmic                                                                                                                                                                                                                                                                                        |
| Target Population            | Patients requiring Endothelial<br>Keratoplasty                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Patients requiring<br>Endothelial Keratoplasty                                                                                                                                                                                                                                                    |
| Anatomical Site              | Eye                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Eye                                                                                                                                                                                                                                                                                               |
| Product Name                 | Proposed:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate:                                                                                                                                                                                                                                                                                        |
|                              | EK Delivery Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K090626<br>EndoSerter™                                                                                                                                                                                                                                                                            |
| Design                       | The EK Delivery device<br>consists of four components:<br>(1) the Trocar which is used<br>to hold the graft material and<br>introduce it into the anterior<br>chamber of the recipient eye,<br>(2) the Trocar Holder, which<br>is used to close the proximal<br>end of the Trocar and as an<br>ergonomic handle to aid the<br>surgeon handling the device,<br>(3) the Injector Assembly,<br>which is a plunger to push<br>the endothelial keratoplasty<br>tissue out of the Trocar and<br>into the anterior chamber of<br>the recipient eye, and (4) End<br>Plug used during shipping.                                                                                                         | The EndoSerter®<br>consists of the<br>instrument's body and its<br>internal mechanism and<br>carrier. The EndoSerter®<br>is a sterile, disposable,<br>single use only device.                                                                                                                     |
|                              | Delivers a circular endothelial<br>tissue button in a rolled<br>configuration.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Delivers a circular<br>endothelial tissue button<br>in a rolled configuration.                                                                                                                                                                                                                    |
| Sterility                    | Sterile Radiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Sterile Radiation                                                                                                                                                                                                                                                                                 |
| Storage Capability           | Yes 72 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                |

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

and the comments of the comments of

:

{4}------------------------------------------------

Testing has been completed showing the following:

Sterilization Validation

Shelf Life

:

Mechanical Strength

Tissue Handling and Stability

Biocompatibility

This testing shows that the predicate and TDAK device are equivalent and pose no new safety issues.

---

***_***_ and the other one is the same. I'm sorry.

***_*****_ and**, and the list goes on.

lity

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

OCT

3 2012

TDAK Medical. Inc. c/o Mr. Greg Holland Regulatory Specialist, Inc. 3722 Ave. Sausalito Irvine, CA 92606

Re: K121579

Trade/Device Name: EK Delivery Device Regulation Number: 21 CFR 886.4300 Regulation Name: Graft Insertion Instrument for Endothelial Keratoplasty Regulatory Class: Class I (reserved) Product Code: OTZ Dated: September 12, 2012 Received: September 13, 2012

Dear Mr. Holland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

Page 2 – Mr. Greg Holland

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

## Indications for Use Statement Indications for Use

510(k) Number (if known): K121579

Device Name: EK Delivery Device

Indications for Use:

The EK Delivery Device is to be used by ophthalmic surgeons trained in Endothelial Keratoplasty (EK) procedures and trained in the use of the EK Delivery Device as:

- · An aid facilitating the insertion of donor corneal posterior lamellar endothelial graft of 100 – 220 µm of thickness into the anterior chamber of the eye during endothelial keratoplasty.
· For use in a 5.1 mm or larger incision and a maximum donor tissue diameter of ≤ 8.5 mm)

· For loading and storage of donor tissue during transport to the surgeon by trained eye bank technicians, and for storage of donor tissue for up to a maximum of 72 hours.

AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ==============================================================================================================================================================================

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K121579

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/OTZ/K121579](https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/OTZ/K121579)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
