The VICTUS Femtosecond Laser Platform is indicated for use in: - · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea - · for anterior capsulotomy during cataract surgery - · the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea. - · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
Device Story
VICTUS Femtosecond Laser Platform is a precision ophthalmic surgical laser system. It uses a Yb:KYW femtosecond laser to perform Laser-Induced Optical Breakdown (LIOB) for corneal tissue photodisruption. The system includes a main unit with a laser source, optical unit, video microscope, and Fourier-based OCT for visualization/positioning. A sterile, single-use Patient Interface (contact lens and suction clip) fixes the eye and provides depth control. The surgeon uses a control screen to view live camera and OCT images, plan treatments, and monitor parameters. The system features automated lens apex detection for centration of capsulotomy and lens fragmentation, and a new grid fragmentation pattern. The device is operated by a surgeon in a clinical setting. It assists in surgical procedures by creating precise corneal incisions and lens fragmentation, potentially improving surgical outcomes and procedural efficiency.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing using agarose gel, PETG, and PMMA to verify performance parameters against pre-defined acceptance criteria. Software verification and validation were conducted per FDA guidance for a 'major' level of concern.
Technological Characteristics
Yb:KYW femtosecond laser source; LIOB photodisruption mechanism; Fourier-based OCT for visualization; sterile single-use Patient Interface (contact lens/suction clip); S60 patient bed; software v3.3. Complies with EN ISO 60601-1, 60601-1-2, and 60601-2-22. Connectivity includes GUI PC for treatment planning and data management.
Indications for Use
Indicated for patients undergoing LASIK surgery (corneal flap creation), cataract surgery (anterior capsulotomy, corneal incisions, lens fragmentation of nuclear cataracts), or other ophthalmic treatments requiring corneal incisions or lamellar resection. Contraindicated for fragmentation of posterior subcapsular (PSC) and cortical cataracts.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2017
Ken Nehmer Director, Regulatory Affairs Technolas Perfect Vision GmbH 351 Buena Vista Ave. E. Unit 501E San Francisco, CA 94117
Re: K171014
Trade/Device Name: VICTUS Femtosecond Laser Platform Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: September 19, 2017 Received: September 20, 2017
Dear Ken Nehmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K171014
Device Name VICTUS Femtosecond Laser Platform
#### Indications for Use (Describe)
The VICTUS Femtosecond Laser Platform is indicated for use in:
- · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
- · for anterior capsulotomy during cataract surgery
- · the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
- · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K171014 510(k) Summary
| I. Submitter | Technolas Perfect Vision GmbH<br>Messerschmittstr. 1 +3<br>80992 Munich, Germany<br>Tel: 49 174 3494062 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ken Nehmer<br>Director, Regulatory Affairs<br>Technolas Perfect Vision<br>351 Buena Vista Ave. E<br>Unit 501E<br>San Francisco, CA 94117<br>Ph: 415-297-0408<br>Email: ken.nehmer@bausch.com |
| Date Revised: | October 16, 2017 |
| II. Device | |
| Name of Device: | VICTUS Femtosecond Laser Platform, Model<br>14000 |
| Common Name: | Ophthalmic Laser |
| Classification Name: | Laser, Ophthalmic |
| Regulatory Class: | Class II |
| Product Code | OOE (Ophthalmic Femtosecond Laser) |
| Regulation Number | 21 CFR 886.4390 |
| III. Predicate Device | VICTUS Femtosecond Laser Platform (K151161)<br>This predicate has not been subject to a design-<br>related recall. |
{4}------------------------------------------------
# IV. Device Description
The VICTUS Femtosecond Laser Platform (VICTUS) is a precision ophthalmic surgical laser and is cleared for use in the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for patients undergoing anterior capsulotomy during cataract surgery. The VICTUS has additionally been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea, primary/secondary corneal incisions, and laser-assisted lens fragmentation.
This submission includes the next software iteration of version 3.3 and introduces a new S60 patient bed option for the VICTUS Platform. The v3.3 software introduces an additional pattern (grid pattern) which is a variation of currently cleared patterns and is being added to the already cleared lens fragmentation capability. In addition, the software introduces a lens apex detection feature.
The previously cleared patient contacting lens and previously cleared suction clip assembly are combined in one convenient package for the user as the VERAFIT patient interface kit. With the exception of the introduction of a new S60 patient bed option, the VICTUS Platform hardware remains unchanged from that described and cleared via K151161.
The mode of operation for the VICTUS Platform is using Yb:KYW Femtosecond laser to produce Laser-Induced Optical Breakdown (LIOB) of the corneal tissue. For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.
The indications for use remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K151161.
| Component | Description |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Assistant workstation | Allows to perform the system tests, to enter<br>and manage the patient data, to select and<br>program procedures, to apply vacuum for<br>patient interface and suction clip, to adjust the<br>procedure according to the patient's eye and to<br>export treatment records. |
| Chiller | Helps to ensure that the main components are<br>kept at a steady temperature. |
| Controller | Checks the connectivity among all electronic<br>components. |
| GUI PC | Interface between the user and the laser |
The VICTUS Femtosecond Laser Platform main unit is composed of the following:
{5}------------------------------------------------
| Laser source | Generates the laser beam. |
|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Optical unit | Controls the complete laser beam path. |
| Surgeon control screen | Shows a live camera image, a live OCT image, and all relevant parameters for the selected treatment. |
| Treatment illumination | Contains the laser ring light module. It illuminates the treatment area. |
| Video microscope | Used as part of treatment planning and shows the top view of the eye to be treated. |
| OCT | Fourier based integrated optical coherence tomography unit which is used to visualize and/or position the treatment as an overlay on the video image. The OCT is not used for diagnostic purposes. |
| The core of the main unit is the laser source that generates the laser beam. All other components control and sustain the treatment process. | |
The following components are additional features to the main unit:
| Additional feature | Description |
|-----------------------------------------------------|-------------------------------------------------------------------------|
| Surgical microscope (optional feature) | Allows for centering the suction clip |
| Surgical microscope illumination (optional feature) | Provides a uniform and adjustable illumination |
| Patient Bed | Used to position a supine patient for treatment by the VICTUS Platform. |
The following components are accessories to the main unit:
| Accessory | Description |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Interface Kit (PIK) | The patient interface kit is sterile, single use kit<br>composed of a contact lens component and a<br>suction clip component. The contact lens and<br>suction clip assembly create a reference surface<br>for depth control and fix the eye relative to the<br>delivery of the laser beam. |
# V. Indications for Use
The VICTUS Femtosecond Laser Platform is indicated for use for:
- the creation of a corneal flap in patients undergoing LASIK surgery or other treatment ● requiring initial lamellar resection of the cornea.
- for anterior capsulotomy during cataract surgery. ●
- the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or . other ophthalmic treatment requiring cuts / incisions in the cornea.
{6}------------------------------------------------
- laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for ● fragmentation of posterior subcapsular (PSC) and cortical cataracts
The indications for use remain unchanged from those previously cleared via K151161.
# VI. Comparison of Technological Characteristics with the Predicate
The mode of operation for the VICTUS Platform is using Yb:KYW Femtosecond laser to produce Laser-Induced Optical Breakdown (LIOB) of the corneal tissue and is the same as that previously cleared for the predicate VICTUS under K151161. Both systems deliver femtosecond pulses to produce a pattern of photodisruption to create cuts / separation in ophthalmic tissue. The subject and predicate device contain the same main unit components as described earlier in this document.
This submission includes updated software version 3.3 software will only be installed on those VICTUS Platform systems that have the cleared Fourier based OCT installed. The patient interface remains the same as that previously cleared for the VICTUS Platform. The VICTUS Platform uses vacuum to affix a suction ring to the corneal surface prior to use. There is no change to vacuum suction or force pressure related to the patient interface as compared to that previously described and cleared in K151161.
The VICTUS Platform mode of operation and the technology used to create the cutting action are identical to the predicate VICTUS Platform device (K151161), and therefore substantially equivalent to the legally marketed predicate device.
The following technological differences exist between the subject and predicate devices:
# a) Grid lens fragmentation pattern
The predicate VICTUS Platform includes clearance for several available lens fragmentation patterns including circular, radial, and a combination of circular and radial cuts. The v3.3 software introduces a new grid fragmentation pattern to the already available fragmentation patterns.
# b) Apex detection feature
The predicate VICTUS system includes clearance for detection of pupil and lens surface. The v3.3 software introduces an apex detection feature which marks the apex location and provides an automated centration of the capsulotomy and lens fragmentation on the apex of the crystalline lens during treatment planning. The feature is being introduced as a user convenience and the physician must still review positioning results and confirm the treatment plan prior to initiating treatment.
#### c) S60 Bed option
A new bed configuration is being introduced that provides three locking positions as follows:
| 0°: | Position for treatment |
|--------|-------------------------------------------------------------------|
| 16.5°: | Position for centering the patient interface (PI) or suction ring |
| 60°: | Position for patient entrance/exit |
{7}------------------------------------------------
In order to minimize manufacturing costs, it has been determined to package the predicate VICTUS contacting lens from the patient interface kit (SKU 90000115) with the suction clip from the predicate VICTUS patient interface kit (SKU 90000145) into one convenient package for the user ( this new kit will be known as the VERAFIT patient interface kit -SKU 90000200TPV).
# VII. Performance Data
## Biocompatibility testing
No new biocompatibility testing was performed to support this premarket notification.
## Electrical safety and electromagnetic compatibility (EMC)
The VICTUS Femtosecond Laser Platform has undergone testing and is in compliance with applicable safety standards as listed in the following table.
| Standard | Title |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN ISO 60601-1: 2005 + A1 | Medical electrical equipment – Part I: General<br>requirements for safety |
| EN ISO 60601-1-2: 2007 | Medical electrical equipment – Part 1: General<br>requirements for safety; 2. Collateral standard:<br>electromagnetic compatibility; requirements and<br>tests |
| EN ISO 60601-2-22: 2007 | Medical electrical equipment – Part 2: Particular<br>requirements for the safety of diagnostic and<br>therapeutic laser equipment |
The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate device. The VICTUS Femtosecond Laser Platform and the predicate device therefore have a similar performance profile.
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern since the VICTUS Platform software controls the delivery of treatment such that an error or malfunction could result in death or serious injury.
{8}------------------------------------------------
## Mechanical and acoustic testing
There was no mechanical or acoustical testing performed to support substantial equivalence of this premarket notification.
#### Animal Study
There was no animal study performed to support substantial equivalence of this premarket notification.
## Non-Clinical Performance Data
A variety of test procedures were conducted to demonstrate that the performance of the various indications available provide substantially equivalent results when using the software version 3.3 which is subject of this premarket notification. The collected data were evaluated by comparing the mean values to the specified acceptance criteria and their 95% confidence intervals. Different materials were used for the bench performance testing including: agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA). The acquired test data successfully verified that the various parameters meet their pre-defined acceptance criteria.
## Clinical Studies
There was no clinical study performed to support substantial equivalence of this premarket notification.
#### VIII. Conclusions
A majority of the technological characteristics of the VICTUS Platform are substantially equivalent to the technological characteristics of the predicate VICTUS Platform which was cleared via K151161. The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the VICTUS Platform should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.