The EyeGility Inserter for enVista preloaded is indicated for folding and inserting of enVista IOL models approved for use with this IOL insertion device as indicated in the IOL approved labeling.
Device Story
Sterile, single-use manual surgical instrument; used by ophthalmologists in clinical settings. Device consists of ABS body, Pebax cartridge, PPA plunger, and silicone soft tip. Operation: preloaded IOL shuttle snapped into loading chamber; push-type plunger advances IOL through shuttle into cartridge; cartridge folds IOL and delivers it through incision into capsular bag. Facilitates IOL implantation; reduces manual handling of lens; ensures proper folding and delivery. No electronic components or software.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including fold and recovery, dioptric power, image quality, lens dimensions, surface/bulk homogeneity, biocompatibility, EO sterilization validation, and transport stability.
Technological Characteristics
Materials: ABS body, Pebax cartridge (Medicoat A), PPA plunger, silicone soft tip. Manual push-type actuation. Sterile, single-use, ethylene oxide sterilized. Complies with ISO 11979-2, ISO 11979-3, ISO 10993-1/7/18, ISO 11135, ISO 11607-1, and ANSI/AAMI ST72.
Indications for Use
Indicated for folding and inserting of enVista IOL models approved for use with this device into the human eye during surgical procedures.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Bausch + Lomb PreVue Inserter for enVista Preloaded (K192005)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2024
Bausch & Lomb Incorporated Renee Stoffel Associate Director, Regulatory Affairs 1400 North Goodman Street Rochester, New York 14609
## Re: K242389
Trade/Device Name: EyeGility™ Inserter for Preloaded enVista® IOLs Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I, reserved Product Code: MSS Dated: August 12, 2024 Received: August 12, 2024
### Dear Renee Stoffel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ginny Liu
For Bennett Walker, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242389
Device Name
EyeGility™ Inserter for Preloaded enVista® IOLs
Indications for Use (Describe)
The EyeGility Inserter for en Vista preloaded is indicated for folding and inserting of en Vista IOL models approved for use with this IOL insertion device as indicated in the IOL approved labeling.
| Type of Use (Select <i>one or both</i> , as applicable) | |
|---------------------------------------------------------|--|
|---------------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Submitter: | Bausch & Lomb Incorporated<br>1400 North Goodman Street<br>Rochester, NY 14609 |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Renee Stoffel<br>Associate Director, Regulatory Affairs<br>(651) 392-4619 |
| Date Prepared: | September 20, 2024 |
| Trade Name: | EyeGility™ Inserter for Preloaded enVista® IOLs |
| Classification Name: | Intraocular lens guide (21 CFR 886.4300) |
| FDA Product Code: | MSS |
| Predicate Devices: | Primary Predicate Device<br>Bausch + Lomb PreVue Inserter for enVista Preloaded (K192005) |
| | Predicate Device<br>Bausch + Lomb IOL Injector, INJ 100 (K113852) |
| Device Description: | The EyeGility Inserter is a sterile, single-use device used to fold and<br>insert an intraocular lens through surgical procedure into a human eye.<br>The Inserter provides a tubular pathway through an incision over the<br>iris, allowing delivery of an IOL into the capsular bag. |
| Indications for Use: | The EyeGility Inserter for enVista preloaded is indicated for folding<br>and inserting of enVista IOL models approved for use with this IOL<br>insertion device as indicated in the IOL approved labeling. |
| Summary of Performance<br>and Non-clinical Testing: | The EyeGility Inserter has successfully undergone functional testing in<br>conformance with the requirements set forth in ISO 11979-3.<br>Additional nonclinical testing information is provided in Table 1. All<br>acceptance criteria were met, demonstrating the subject device is<br>substantially equivalent to the predicate devices. |
| Clinical Testing: | No clinical testing was completed in support of the subject device. |
| Comparative Analysis: | The EyeGility Inserter has been demonstrated to be equivalent to the<br>predicate devices for its intended use. Table 2 includes a detailed<br>comparison of the subject and predicate devices. |
| Conclusion: | The EyeGility Inserter is substantially equivalent to the predicate<br>devices. |
{5}------------------------------------------------
| Testing Category | Test | Standards | Results |
|-----------------------------------------------------------|---------------------------------------------------|------------------------------------------|---------|
| Delivery Verification<br>Testing and Stability<br>Testing | Fold and Recovery Test | ISO 11979-3 | Pass |
| | Dioptric Power and Image Quality Post<br>Delivery | ISO 11979-2 | Pass |
| | Lens Dimensions Post Delivery | ISO 11979-3 | Pass |
| | Surface and Bulk Homogeneity | ISO 11979-3 | Pass |
| | Cosmetic Inspection | N/A | Pass |
| Biocompatibility | Biocompatibility Assessment | ISO 10993-1 | Pass |
| | Particulate Study | ISO 10993-1 | Pass |
| | Coating Transfer Study | ISO 10993-1 | Pass |
| | Leachable Extractions | ISO 10993-18 | Pass |
| EO Sterilization | Validation Product Adoption Analysis | ISO 11135 | Pass |
| EO Residuals | ETO/ECH Transfer Engineering Study | ISO 10993-7 | Pass |
| Bacterial Endotoxin | Bacterial Endotoxin Test | ANSI/AAMI ST72 | Pass |
| Packaging | Transport Stability Testing | ISO 11607-1<br>ISO 11979-6<br>ASTM F2096 | Pass |
Table 1. Nonclinical Testing Summary
Table 2. Comparison of the subject and predicate device.
| Technological<br>Similarities | Subject Device – EyeGility<br>Inserter | Primary Predicate Device -<br>Bausch + Lomb PreVue<br>Inserter (K192005) | Predicate Device - Bausch +<br>Lomb IOL Injector, INJ100<br>(K113852) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | The EyeGility Inserter for<br>enVista preloaded is indicated<br>for folding and inserting of<br>enVista IOL models approved<br>for use with this IOL insertion<br>device as indicated in the IOL<br>approved labeling. | The Bausch +Lomb PreVue<br>inserter for en Vista preloaded<br>is indicated for folding and<br>inserting of enVista IOLs<br>(Model MX60PL) and IOL<br>models approved for use with<br>this IOL insertion device as<br>indicated in the IOL approved<br>labeling. | The Bausch + Lomb IOL<br>Injector is indicated for<br>folding and injection of<br>Bausch + Lomb IOLs<br>approved for use with this<br>injector |
| How device is<br>used | The preloaded IOL shuttle is<br>snapped into loading chamber<br>of the enVista inserter body.<br>The push-type plunger<br>advances the IOL through the<br>shuttle into the cartridge which<br>folds the IOL and advances it<br>into the eye. | The preloaded IOL shuttle is<br>snapped into loading chamber<br>of the PreVue inserter. The<br>screw plunger advances the<br>IOL through the shuttle into<br>the cartridge which folds the<br>IOL and advances it into the<br>eye. | The IOL is placed in the<br>loading chamber. A plunger<br>pushes the IOL into the tip,<br>which folds the IOL.<br>Pushing the plunger further<br>advances<br>the IOL out through the tip<br>into the eye. |
{6}------------------------------------------------
| Technological<br>Similarities | Subject Device - EyeGility<br>Inserter | Primary Predicate Device -<br>Bausch + Lomb PreVue<br>Inserter (K192005) | Predicate Device - Bausch +<br>Lomb IOL Injector, INJ100<br>(K113852) |
|-------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Inserter<br>Components | ABS Body<br>Pebax Cartridge (Medicoat A)<br>PPA Plunger<br>Silicone Soft tip | ABS Body<br>Pebax Cartridge (Medicoat A)<br>PPA Plunger<br>ABS Thread Bushing and<br>Screw Spindle | ABS Body<br>ABS Plunger<br>Polyamide Cartridge<br>Medicoat A Coating |
| Single Use | Injector: Yes<br>Cartridge: Yes | Inserter: Yes<br>Cartridge: Yes | Inserter: Yes<br>Cartridge: Yes |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Shelf Life | 22 months | 36 months | 24 months |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.