← Product Code [MSS](/submissions/OP/subpart-e%E2%80%94surgical-devices/MSS) · K231838

# RxSight® Insertion Device (63002) (K231838)

_Rxsight, Inc. · MSS · Aug 15, 2023 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/MSS/K231838

## Device Facts

- **Applicant:** Rxsight, Inc.
- **Product Code:** [MSS](/submissions/OP/subpart-e%E2%80%94surgical-devices/MSS.md)
- **Decision Date:** Aug 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4300
- **Device Class:** Class 1
- **Review Panel:** Ophthalmic

## Intended Use

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3-piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device (Model 63002) is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

## Device Story

Sterile, single-use IOL inserter and haptic puller; used during cataract surgery. Surgeon places IOL into loading cartridge; screw plunger advances IOL through cartridge, folding lens and delivering it into eye through small incision. Device facilitates IOL implantation; benefits patient by enabling minimally invasive cataract surgery. Used in clinical/surgical settings by ophthalmologists.

## Clinical Evidence

Bench testing only. Simulated surgical manipulation and recovery properties evaluated per ISO 11979-3:2012 (mechanical dimensions, sagitta) and ISO 11979-2:2014 (optical properties, surface/bulk homogeneity) pre- and post-injection.

## Technological Characteristics

Single-use, sterile IOL injector. Materials: Polycarbonate (injector), Polybutylene terephthalate polymer (cartridge), Polypropylene (cartridge). Features LubriMATRIX™ cartridge coating. Sterilization: Ethylene Oxide (SAL 10^-6). Mechanical operation via screw plunger.

## Regulatory Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

## Predicate Devices

- RxSight Insertion Device ([K231466](/device/K231466.md))
- pioli IOL Delivery System ([K172228](/device/K172228.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 15, 2023

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration".

RxSight, Inc. Maureen OConnell Senior Vice President, Clinical and Regulatory Affairs 100 Columbia Aliso Viejo, California 92656

## Re: K231838

Trade/Device Name: RxSight® Insertion Device (63002) Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I, reserved Product Code: MSS Dated: June 22, 2023 Received: June 22, 2023

## Dear Ms. Maureen OConnell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Elvin Y. I

for Bennett Walker, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231838

Device Name

RxSight® Insertion Device (63002)

#### Indications for Use (Describe)

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3-piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device (Model 63002) is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                             |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY RxSight® Insertion Device (Model 63002)

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(a).

| APPLICANT:                | RxSight, Inc.<br>100 Columbia<br>Aliso Viejo, CA 92656                                                                  |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON:           | Maureen O'Connell<br>Sr. Vice President Clinical and Regulatory Affairs<br>moconnell@rxsight.com<br>Tel: (978) 207-1245 |
| DATE SUMMARY<br>PREPARED: | August 15, 2023                                                                                                         |
| TRADE NAME:               | RxSight Insertion Device (Model 63002)                                                                                  |
| COMMON NAME:              | IOL Injector                                                                                                            |
| CLASSIFICATION NAME:      | Intraocular Lens (IOL) Guide                                                                                            |
| DEVICE CLASSIFICATION:    | Class I; 21 CFR 886.4300                                                                                                |
| PRODUCT CODE:             | MSS                                                                                                                     |
| PREDICATE DEVICE(S):      | Primary: RxSight Insertion Device, K231466<br>Secondary: pioli IOL Delivery System, K172228                             |

#### 1.1 DEVICE DESCRIPTION

The RxSight Insertion Device (Model 63002) is a sterile, single-use device to be used to fold and insert intraocular lenses (IOL) into the eye through a small incision during cataract surgery. The RxSight Insertion Device (Model 63002) consists of:

- A single-use, disposable, sterile IOL inserter with a non-pyrogenic cartridge ●
- A single-use, disposable, sterile haptic puller. ●

#### 1.2 INDICATIONS FOR USE

The RxSight® Insertion Device (Model 63002) is indicated for the folding and insertion of a 3piece silicone, intraocular lens into the human eye through a surgical incision. The RxSight Insertion Device is intended for the insertion of the RxSight Light Adjustable Lens®, RxSight

{4}------------------------------------------------

Light Adjustable Lens+, and IOL models validated for use with this device in IOL approved labeling.

#### 1.3 TECHNOLOGICAL CHARACTERISTICS COMPARISON

RxSight believes that the RxSight Insertion Device (Model 63002) described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to legally marketed predicate devices that are Class I medical devices. The following table compares the RxSight Insertion Device (Model 63002) with the primary predicate device which is the RxSight Insertion Device (Model 63000/63001) cleared most recently in K231466 and the secondary predicate device, the AST Products pioli IOL Delivery System cleared in K172228.

|                     | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                         | Primary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                   | Secondary Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                     | RxSight Insertion Device<br>(Model 63002)                                                                                                                                                                                                                                                                                                                                                                                                               | RxSight Insertion Device<br>(Model 63000/63001)                                                                                                                                                                                                                                                                                                                                                                                                               | AST Products<br>pioli IOL Delivery System                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 510(k) Number       | -                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K231466                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K172228                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Code        | MSS                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MSS                                                                                                                                                                                                                                                                                                                                                                                                                                                           | MSS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended Use        | Intraocular Lens Delivery                                                                                                                                                                                                                                                                                                                                                                                                                               | Intraocular Lens Delivery                                                                                                                                                                                                                                                                                                                                                                                                                                     | Intraocular Lens Delivery                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use | The RxSight® Insertion<br>Device (Model 63002) is<br>indicated for the folding and<br>insertion of a 3-piece<br>silicone, intraocular lens into<br>the human eye through a<br>surgical incision. The<br>RxSight Insertion Device<br>(Model 63002) is intended<br>for the insertion of the<br>RxSight Light Adjustable<br>Lens®, RxSight Light<br>Adjustable Lens+, and IOL<br>models validated for use with<br>this device in IOL approved<br>labeling. | The RxSight® Insertion<br>Device is indicated for the<br>folding and insertion of a 3-<br>piece silicone, intraocular<br>lens into the human eye<br>through a surgical incision.<br>The RxSight Insertion Device<br>is intended for the insertion<br>of the RxSight Light<br>Adjustable Lens®, the<br>RxSight Light Adjustable<br>Lens+, the Bausch & Lomb<br>LI61A0 IOL and IOL models<br>validated for use with this<br>device in IOL approved<br>labeling. | The pioli IOL Delivery system<br>is a single-use, sterile device<br>intended to insert a single-<br>piece foldable intraocular lens<br>(IOL) into the human eye<br>through a surgical procedure.<br>The system provides a tubular<br>pathway for lens implantation<br>through an incision. The pioli<br>IOL Delivery System is only<br>for the insertion of the Lenstec<br>Softec I IOL and IOL models<br>validated for use with this<br>device as indicated in the IOL<br>approved labeling. |
| Operating Principle | An IOL is placed in a loading<br>cartridge. Screw plunger<br>advances the IOL through the<br>cartridge which folds the IOL<br>and advances it into the eye.                                                                                                                                                                                                                                                                                             | An IOL is placed in a loading<br>cartridge. Cartridge snapped<br>into the handpiece. Screw<br>plunger advances the IOL<br>through the cartridge which<br>folds the IOL and advances it<br>into the eye.                                                                                                                                                                                                                                                       | Cartridge is back loaded into<br>the injection system with a<br>plunger that advances the IOL<br>through the cartridge and into<br>the eye                                                                                                                                                                                                                                                                                                                                                    |
| Pre-loaded IOL      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                                                            | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Material (Injector) | Polycarbonate                                                                                                                                                                                                                                                                                                                                                                                                                                           | Titanium                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

## COMPARISON OF THE RXSIGHT INSERTION DEVICE (MODEL 63002) TO THE PREDICATE DEVICES

{5}------------------------------------------------

|                                          | Proposed Device                           | Primary Predicate<br>Device                                         | Secondary Predicate<br>Device             |
|------------------------------------------|-------------------------------------------|---------------------------------------------------------------------|-------------------------------------------|
|                                          | RxSight Insertion Device<br>(Model 63002) | RxSight Insertion Device<br>(Model 63000/63001)                     | AST Products<br>pioli IOL Delivery System |
|                                          | Polybutylene terephthalate<br>polymer     |                                                                     |                                           |
| Material (cartridge)                     | Polypropylene                             | Polypropylene                                                       | Polypropylene                             |
| Cartridge Coating                        | LubriMATRIX™                              | LubriMATRIX™                                                        | LubriMATRIX™                              |
| How Supplied<br>(Reusable/Single<br>Use) | Single Use, supplied sterile              | Handpiece - Reusable<br>Cartridge - Single Use,<br>supplied sterile | Single Use, supplied sterile              |
| Method of<br>Sterilization               | Ethylene Oxide to SAL 10-6                | Handpiece - Autoclave<br>Cartridge - Ethylene Oxide to<br>SAL 10-6  | Ethylene Oxide to SAL 10-6                |

#### 1.4 SUMMARY OF PERFORMANCE TEST RESULTS

The descriptive characteristics are well-defined and adequate to ensure equivalence of the RxSight Insertion Device with the predicate devices.

Non-clinical performance testing included simulated surgical manipulation and recovery of properties including both pre- and post-injection evaluation of the IOLs in accordance with ISO 11979-3:2012. Specifically, mechanical dimensions and sagitta were verified following lens delivery and compared to the measurements performed prior to the lens delivery per ISO 11979-3:2012. Additionally, optical properties and overall surface and bulk homogeneity were tested pre and post lens injection per ISO 11979-2:2014.

# Conclusion

The RxSight Insertion Device (Model 63002) has the same intended use as the legally marketed predicate devices identified in this 510(k) premarket notification and other IOL injectors regulated under 21 CFR 886.4300. This device meets all product design requirements and applicable standards and embodies technological characteristics similar to the predicate devices. Non-clinical data establish that the device is safe and effective in the delivery of the RxSight Light Adjustable Lens and Light Adjustable Lens+ and is shown to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/MSS/K231838](https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/MSS/K231838)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
