bioli lOL Delivery System

K200057 · Ast Products, Inc. · MSS · Jun 4, 2020 · Ophthalmic

Device Facts

Record IDK200057
Device Namebioli lOL Delivery System
ApplicantAst Products, Inc.
Product CodeMSS · Ophthalmic
Decision DateJun 4, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4300
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The bioli™ IOL Delivery System is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through an incision. The bioli™ IOL Delivery System is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling.

Device Story

The bioli™ IOL Delivery System is a single-use, sterile device used by surgeons during cataract surgery to fold and implant foldable acrylic intraocular lenses (IOLs) into the human eye. The system consists of a cartridge and an injector with a silicone-tipped plunger. The surgeon loads the IOL into the cartridge, which features internal geometry to progressively fold the lens, and then uses the injector to advance the lens through a nozzle into the capsular bag. The cartridge surfaces are treated with LubriMATRIX™, a hydrophilic and lubricious polymer coating, to facilitate smooth delivery. The device is operated manually by a physician. By enabling controlled delivery of the IOL through a small incision, the device assists in the replacement of the human crystalline lens, potentially improving patient visual outcomes following cataract surgery.

Clinical Evidence

No clinical data. Bench testing only. Performance was validated through simulated surgical manipulation, biocompatibility testing (ISO 10993-5, -10, -11), sterilization validation (ISO 11135), and shelf-life/transport stability testing (ASTM F1980, ASTM D4169). IOLs delivered via the system were evaluated for optical and dimensional properties per ISO 11979-3:2012, showing no damage or performance degradation.

Technological Characteristics

Single-use, sterile, manual injector system. Materials: Polycarbonate (injector barrel), ABS (plunger), Silicone (tip), Polypropylene (cartridge), Stainless Steel (spring). Surface treatment: LubriMATRIX™ (hydrophilic/lubricious polymer). Sterilization: EtO (SAL 10^-6). Standards: ISO 11979-3:2012, ISO 10993-5:2009, ISO 10993-10:2010.

Indications for Use

Indicated for patients undergoing cataract surgery requiring implantation of Lenstec Softec I IOL or other IOL models validated for use with this device.

Regulatory Classification

Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. AST Products, Inc. William Lee VP, R&D and Regulatory Affairs 9 Linnell Circle Billerica, Massachusetts 01821 Re: K200057 Trade/Device Name: bioli IOL Delivery System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I, reserved Product Code: MSS Dated: April 26, 2020 Received: April 28, 2020 Dear William Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Tieuvi Nguyen, Ph.D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K200057 Device Name bioli™ IOL Delivery System #### Indications for Use (Describe) The bioli™ IOL Delivery System is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through an incision. The bioli™ IOL Delivery System is for implantation of qualified Lenstec Softec 1 IOL and IOL models validated for use with this device as indicated in the IOL approved labeling. Type of Use (Select one or both, as applicable) | <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |---------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR § 807.92, 510(k) summary is provided. ### 1. Submitter AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778 ## 2. Official Correspondence/Contact Person William Lee, Ph.D. VP, R&D and Regulatory Affairs AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778 Email: wlee@astp.com ## 3. 510(k) Preparer William Lee, Ph.D. VP, R&D and Requlatory Affairs AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778 Email: wlee@astp.com ### 4. Product Name - Trade Name: bioli™ IOL Delivery System (Models: BIOLI-C, BIOLI-rC, BIOLI-D) - Common Name: Folders and Injectors, Intraocular Lens (IOL) ## 5. Device Classification - | Classification Name: Class I (21 CFR § 886.4300 Intraocular lens guide) - Classification Panel: Ophthalmic (86) - Product Code: MSS ## 6. Legally Marketed Predicates - K182965 BL-Cart™ IOL Delivery Cartridge {4}------------------------------------------------ #### 7. Device Description The bioli™ IOL Delivery System is an intraocular lens delivery system used for progressive folding and delivering intraocular lenses into the eye for replacement of the human crystalline lens. The bioli™ IOL Delivery System is intended for use during cataract surgery. The bioli™ IOL Delivery System is composed of two parts: - (1) a cartridge that holds the intraocular lens in position in preparation for loading and folding, and - (2) a single-use injector that houses a silicone-tipped plunger used to advance the lens into the capsular bag The cartridge incorporates a smooth, internal geometry that progressively folds the lens for delivery through conventional nozzle geometry. A coating is applied to the cartridge including the inner surfaces to increase lubricity. The cartridge is loaded by positioning the intraocular lens into the cartridge, folding the trailing haptic onto the anterior side of the optic, and finally pushing the optic edge of the lens to position it as far into the cartridge as the forceps will permit. After the cartridge is attached onto the injector, the system is now ready for the advancement of the plunger to deliver the lens. The bioli™ IOL Delivery System and the referenced legally marketed predicate, BL-Cart™ IOL Delivery Cartridge, share the same principal method to implant an intraocular lens into the eye. Both utilize the same lens folding system design in which IOLs are folded and/or delivered from an injector system and the resulting optical and physical properties of the IOL remain unchanged as a result of folding/delivery. Note: LubriMATRIX™ is a proprietary treatment composed of a hydrophilic polymer and a lubricious polymer engrafted onto the surface of the cartridge via a chemical polymerization process. Refer to MAF-1963 for further information. LubriMATRIX™ is applied to AST Products' current line of IOL Delivery Systems: lioli™ IOL Delivery System (K142056), pioli™ IOL Delivery System (K172228), and BL-Cart™ IOL Delivery Cartridge (K182965). #### 8. Intended Use and Indications for Use The bioli™ IOL Delivery System is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through an incision. The bioli™ IOL Delivery System is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling. The bioli™ IOL Delivery System and the referenced legally marketed predicate device share the same principal folding method to implant an intraocular lens into the eye and similar intended use and indications for use for lens insertion into the eye, and thus does {5}------------------------------------------------ not raise any new questions or concerns regarding safety and effectiveness. Refer to Table 1. | Trade Name | Manufacturer | Regulation<br>No. | Product<br>Code | Regulation<br>Description | 510(k) No. | Intended Use &<br>Indications for Use | |------------------------------------------------------------------------------|-----------------------|-------------------|-----------------|---------------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | bioli™ IOL<br>Delivery System<br>(Models: BIOLI-<br>C, BIOLI-rC,<br>BIOLI-D) | AST Products,<br>Inc. | 886.4300 | MSS | Intraocular<br>Lens Guide | New device | The bioli™ IOL Delivery System is a<br>single-use, sterile device intended to<br>insert an approved single-piece foldable<br>acrylic intraocular lens (IOL) into the<br>human eye through an incision.<br>The bioli™ IOL Delivery System is only for<br>the insertion of Lenstec Softec I IOL and<br>IOL models validated for use with this<br>device as indicated in the approved IOL<br>labeling. | | BL-Cart™ IOL<br>Delivery<br>Cartridge<br>(Model: Type D) | AST Products,<br>Inc. | 886.4300 | MSS | Intraocular<br>Lens Guide | K182965 | The BL-Cart™ IOL Delivery Cartridge is a<br>single-use, sterile device intended to<br>insert an approved single-piece foldable<br>acrylic intraocular lens (IOL) into the<br>human eye through a surgical procedure.<br>The cartridge is intended to be used in<br>conjunction with Alcon Monarch® III IOL<br>Delivery System injector.<br>The BL-Cart™ IOL Delivery Cartridge is<br>only for the insertion of Lenstec Softec I<br>IOL and IOL models validated for use with<br>this device as indicated in the approved<br>IOL labeling. | Table 1. Device Comparisons – Intended Use and Indications for Use {6}------------------------------------------------ #### 9. Technological Characteristics The bioli™ IOL Delivery System is an intraocular lens delivery system used for progressive folding and delivering of an intraocular lens into the eye for replacement of the human crystalline lens during cataract surgery. The device consists of a lubricious coated cartridge and a single-use injector with a silicone-tipped plunger used to advance the lens into the capsular bag. The bioli™ IOL Delivery System shares similar technological characteristics for lens insertion into the eye with the predicate device referenced in this 510(k) premarket notification application. Refer to Tables 2 and 3. {7}------------------------------------------------ | Trade Name | Design | Operating Principle | Advancement<br>Mechanism | Biocomp. | Sterilization<br>Method | Setting Used | Patient<br>Contact | Recommended<br>for Use With | Human<br>Factors | Performance<br>Criteria Met | |--------------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|----------|---------------------------------------|--------------------------------|-------------------------------------------------------|------------------------------|----------------------------------|-----------------------------| | bioli™ IOL<br>Delivery System<br>(Models: BIOLI-C,<br>BIOLI-rC, BIOLI-D) | A sterile,<br>single-use<br>disposable<br>device | The cartridge is loaded<br>into the injection system<br>and the IOL is pushed<br>through the cartridge<br>into the eye. | Plunger /<br>syringe type | Yes | ETO sterilization<br>SAL level 1x10-6 | Hospital/<br>Surgery<br>center | Distal end of<br>cartridge to<br>place lens in<br>eye | FDA-approved<br>viscoelastic | For use<br>by<br>doctors<br>only | Yes | | BL-Cart™ IOL<br>Delivery Cartridge<br>(Model: Type D) | A sterile,<br>single-use<br>disposable<br>coated<br>cartridge | The one-piece<br>disposable cartridge is<br>loaded into a reusable<br>handpiece injection<br>system and the IOL is<br>pushed through the<br>cartridge into the eye | Plunger-like,<br>non-screw type | Yes | ETO sterilization<br>SAL level 1x10-6 | Hospital/<br>Surgery<br>center | Distal end of<br>cartridge to<br>place lens in<br>eye | FDA-approved<br>viscoelastic | For use<br>by<br>doctors<br>only | Yes | Table 2. Device Comparisons – Technological Characteristics # Table 3. Device Comparisons – Materials | Trade Name | Device Materials | Standards Met | |--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | bioli™ IOL Delivery System<br>(Models: BIOLI-C, BIOLI-rC, BIOLI-D) | Injector Barrel - Polycarbonate (144R)<br>Plunger - ABS (PA-758)<br>Silicone Tip - Silicone (R 401)<br>Cartridge - Polypropylene (RJ880MO-12)<br>Spring - Stainless Steel (SS304) | ISO 11979-3:2012<br>ISO10993-5: 2009<br>ISO 10993-10: 2010 | | BL-Cart™ IOL Delivery Cartridge (Model: Type D) | Cartridge - Polypropylene (RJ880MO-12) | ISO 11979-3:2012<br>ISO10993-5: 2009<br>ISO 10993-10: 2010 | {8}------------------------------------------------ #### 10. Performance Data (Nonclinical and/or Clinical) - Non-clinical Tests: The bioli™ IOL Delivery System contacts the patient's eye directly for a limited (≤24hr) time. The device also has limited contact with the IOL. Based on the contact duration, cytotoxicity, sensitization and ocular irritation tests were performed. In addition, acute systemic toxicity and material-mediated pyrogenicity tests were also performed. The biocompatibility testing was performed in accordance with International Standard Orqanization (ISO) Biocompatibility evaluation of medical devices- parts 5, 10 and 11. The biocompatibility testing was found acceptable. All lenses were delivered through the bioli™ IOL Delivery System according to the loading and delivery instructions. The IOLs were evaluated for optical properties, dimensional properties and overall surface and bulk homoqeneity before and after being surgically manipulated using the bioli™ IOL Delivery System, as well as lens opening time after folding. The bioli™ IOL Delivery System was also evaluated for its cartridge performance, such as overall cartridge surface and bulk homogeneity. IOL optical properties and overall surface and bulk homogeneity inspection were conducted in accordance with ISO 11979-3:2012, Ophthalmic implants – Intraocular lenses - Part 3: Mechanical properties and test methods. After delivery, lenses were observed for possible damages or scratches using a 10x microscope and showed passing results, and were within dimensional specifications. The bioli™ IOL Delivery System also showed no damage after lens delivery. The resulting data from simulated surgical manipulation of bioli™ IOL Delivery System (Models: BIOLI-C, BIOLI-IC, BIOLI-D) to deliver intraocular lenses showed that the device can successfully deliver Lenstec's hydrophilic Softec 1 IQLs of low to high diopters and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens. Package stability and performance stability test results met the defined acceptance criteria and requirements for a shelf-life claim of 5 years. Visual inspection, package seal and package integrity of the sterile barrier system of bioli™ IOL Delivery System met the requirements in accordance with ASTM F1980-2007(2011 reapproved). The post aging tests were carried out in accordance with ASTM F88-09, ASTM F1929-12, ISO 11737-2:2009, USP 35 and ASTM F1140/F1140M-2013. Results from sterilization validation testing conducted in accordance with ISO 11135:2014 demonstrated that bioli™ IOL Delivery System is able to be sterilized using EtO sterilization parameters with residual EO found to be within safe levels post-aeration period. A transport stability study was conducted according to requirements in ISO 2233:2001, ASTM D4169-16, ASTM F88/F88M-15, ASTM F2096-11, and ASTM F1929-15. The results demonstrated that bioli™ IOL Delivery System was able to {9}------------------------------------------------ resist the physical damage caused by drop, vibration and short extreme temperature impacts during transportation. - Clinical Tests: Not required for this device. #### 11. Conclusion The resulting data from simulated surgical manipulation of bioli™ IOL Delivery System to deliver intraocular lenses showed that bioli™ (Models: BIOLI-C, BIOLI-rC, BIOLI-D) can successfully deliver Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens. Based on the assessment of these findings, along with the results of Biocompatibility, Sterilization and Shelf-life testing, we conclude that the bioli™ IOL Delivery System is substantially equivalent to the referenced legally marketed predicate device provided in this 510(k) premarket notification submission.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...