← Product Code [HQF](/submissions/OP/subpart-e%E2%80%94surgical-devices/HQF) · K220974

# R:GEN (K220974)

_Lutronic Vision, Inc. · HQF · Dec 27, 2022 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQF/K220974

## Device Facts

- **Applicant:** Lutronic Vision, Inc.
- **Product Code:** [HQF](/submissions/OP/subpart-e%E2%80%94surgical-devices/HQF.md)
- **Decision Date:** Dec 27, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4390
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Intended Use

The R:GEN is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment of Clinically Significant Macular Edema (CSME).

## Device Story

R:GEN is a surgical laser system for focal laser treatment (selective retinal therapy) of retinal diseases; specifically CSME. System components include main body, LCD monitor, slit lamp set, footswitch, and Real Time Feedback (RTF) contact lens. Device uses Q-switched Nd:YLF laser (527 nm) to wound RPE cells. Physician operates system via slit lamp delivery; laser beam passes through RTF contact lens to retina. RTF sensor provides feedback during procedure. Device offers multiple laser patterns (single, square, circle, 2x2 combo) to improve treatment precision. Output is controlled laser energy (0.03-0.35 mJ) limited to 15 pulses per session. Benefits include targeted RPE treatment while minimizing collateral damage to surrounding retinal tissue.

## Clinical Evidence

Evidence includes published clinical performance testing confirming device efficacy for CSME treatment. Supporting data includes animal studies demonstrating RPE cell wounding and safety. Bench testing covers biocompatibility (ISO 10993), sterilization (ISO 11135), software V&V (IEC 62304), and electrical/laser safety (IEC 60601-2-22, IEC 60825-1).

## Technological Characteristics

Q-switched Nd:YLF laser; 527 nm wavelength; 1.7 µs pulse duration; 0.03-0.35 mJ energy. Delivery via slit lamp adapter with optical fiber and RTF contact lens. Connectivity: Footswitch, LCD touchscreen. Power: 100-240V, 50/60 Hz. Software: Major level of concern. Sterilization: RTF sensor validated per ISO 11135.

## Regulatory Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

## Predicate Devices

- R:GEN Laser System ([K153769](/device/K153769.md))

## Reference Devices

- Ellex 2RT ([K122202](/device/K122202.md))
- TxCell Scanning Laser Deliver System ([K121475](/device/K121475.md))
- Pascal Synthesis Ophthalmic Laser ([K123542](/device/K123542.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

December 27, 2022

Lutronic Vision, Inc. % Dulciana Chan Principal Consultant Ram+ 2251 San Diego Avenue, Suite B-257 San Diego, California 92110

Re: K220974

Trade/Device Name: R:gen Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: November 23, 2022 Received: November 25, 2022

#### Dear Dulciana Chan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Alexander Date: 2022.12.27 Beylin -S 17:58:40 -05'00'

> for Elvin Ng, Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220974

Device Name R:GEN

Indications for Use (Describe)

The R:GEN is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment of Clinically Significant Macular Edema (CSME).

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K220974

### DATE PREPARED

December 22, 2022

## 510(k) OWNER

Lutronic Vision Inc. 19 Fortune Drive Billerica, MA 01821 Telephone: Official Contact: Haelyung Hwang, CEO

## REPRESENTATIVE/CONSULTANT

Dulciana Chan, MSE Allison C. Komiyama, Ph.D., R.A.C. RQM+ Telephone: +1 (412) 816-8253 Email: dchan@rqmplus.com, akomiyama@rqmplus.com

### DEVICE INFORMATION

Proprietary Name/Trade Name: R:GEN Common Name: Laser, Ophthalmic Regulation Number: 21 CFR § 886.4390 Class: II Product Code: HQF Premarket Review: OPEQ/OHT1/DHT1A/Retinal and Diagnostic Devices Review Panel: Ophthalmic

#### PREDICATE DEVICE IDENTIFICATION

The R:GEN is substantially equivalent to the following predicate:

| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------|-------------------|
| K153769       | R:GEN Laser System                   | ✓                 |

{4}------------------------------------------------

Lutronic submits the following information to demonstrate that the R:GEN has similar technology to the following legally marketed reference devices:

| 510(k) Number | Reference Device Name / Manufacturer                     | Reference Device |
|---------------|----------------------------------------------------------|------------------|
| K122202       | Ellex 2RT /Ellex Medical                                 | ✓                |
| K121475       | TxCell Scanning Laser Deliver System /IRIDEX Corporation | ✓                |
| K123542       | Pascal Synthesis Ophthalmic Laser/ IRIDEX Corporation    | ✓                |

## DEVICE DESCRIPTION

The R:GEN is a surgical laser system for use by ophthalmic physicians for performing focal laser treatment, also referred to as selective retinal therapy for the treatment of various retinal diseases by wounding Retinal Pigmented Epithelium (RPE) cells. It is the product of continual technological evolution of the R:GEN technology described in the predicate device (K153769). It consists of the following main components: Main Body, LCD Monitor, Slit Lamp Set, Footswitch, and Real Time Feedback (RTF) Contact Lens which consists of the RTF Body (Real Time Feedback Body) and RTF Sensor (Real Time Feedback Sensor). The Main Body allows the proper operation of the entire system. It contains the laser and is attached to the Footswitch and optical fiber. The laser is a Q-switched Nd:YLF (Neodymium-doped Yttrium Lithium Fluoride) laser with an emission wavelength of 527 nm (produced by second-harmonic generation). The integrated software provides all the functions which are necessary to use the device. The Slit Lamp Set contains the laser delivery system. The laser beam emitted through the Slip Lamp Set is transmitted through the RTF Contact Lens (first through the RTF body and then through the RTF Sensor), and finally irradiated to the patient's retina.

## INDICATIONS FOR USE

The R:GEN is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment of Clinically Significant Macular Edema (CSME).

## INTENDED USE

The subject device has the same intended use as the predicate device which is to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye.

{5}------------------------------------------------

# SUBSTANIAL EQUIVALENCE

|                                | Subject Device                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                             | Reference Device                                                                                                                                                                                                                                                                                       | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |              | 2 x 2 Combo Pattern |                    |                    | Rotation               | Triple Ring                          |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|---------------------|--------------------|--------------------|------------------------|--------------------------------------|
| Intended<br>Use                | Lutronic Vision<br>Inc.<br>R:GEN<br>K220974                                                                                                                                                                                                                                            | Lutronic<br>Corporation<br>R:GEN Laser<br>System<br>K153769                                                                                                                                                                                                                                  | Ellex Medical<br>Ellex 2RT<br>K122202                                                                                                                                                                                                                                                                  | IRIDEX Corporation<br>TxCell Scanning Laser<br>Delivery System<br>K121475                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | IRIDEX Corporation<br>Pascal Synthesis<br>Ophthalmic Laser<br>K123542                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |              |                     |                    |                    | Arc                    | Arc                                  |
| Indications<br>for Use         | To produce a wound<br>to the Retinal<br>Pigmented<br>Epithelium (RPE) of<br>the retina of the eye                                                                                                                                                                                      | To produce a wound<br>to the Retinal<br>Pigmented Epithelium<br>(RPE) of the retina of<br>the eye                                                                                                                                                                                            | To produce a<br>wound to the<br>Retinal Pigmented<br>Epithelium (RPE) of<br>the retina of the<br>eye                                                                                                                                                                                                   | To deliver laser energy to<br>various ocular targets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | To deliver laser energy to<br>various ocular targets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |              |                     |                    |                    | Triple Arc             | Line (1-10 spots in a line)          |
| Indications<br>for Use         | The R:GEN is<br>indicated for use by<br>an ophthalmic<br>physician to<br>produce a wound to<br>the Retinal<br>Pigmented<br>Epithelium (RPE) of<br>the retina of the<br>eye, via focal laser<br>treatment, in the<br>treatment of<br>Clinically Significant<br>Macular Edema<br>(CSME). | The R:GEN Laser<br>System is indicated<br>for use by an<br>ophthalmic physician<br>to produce a wound<br>to the Retinal<br>Pigmented Epithelium<br>(RPE) of the retina of<br>the eye, via focal laser<br>treatment, in the<br>treatment of Clinically<br>Significant Macular<br>Edema (CSME) | The 2RT (LR1532) is<br>indicated for use by<br>a trained<br>ophthalmic<br>physician to<br>produce a wound<br>to the Retinal<br>Pigmented<br>Epithelium (RPE) of<br>the retina of the<br>eye, via focal laser<br>treatment, in the<br>treatment of<br>Clinically Significant<br>Macular Edema<br>(CSME) | When the TxCell Scanning<br>Laser Delivery System is<br>connected to the IQ 532<br>(532 nm), the IQ 577 (577<br>nm) or the IQ 810 (810<br>nm) Laser Console, from<br>the IRIDEX<br>Family of IQ Laser Systems<br>and used to deliver laser<br>energy in CW-Pulse,<br>MicroPulse or LongPulse<br>mode, it is intended to be<br>used by a trained<br>ophthalmologist for the<br>treatment of ocular<br>pathology of both the<br>anterior and posterior<br>segments of the eye.<br>532 nm<br>Indicated for retinal<br>photocoagulation laser                                                                                                                                        | The PASCAL® Synthesis<br>Ophthalmic Scanning Laser<br>System is intended for use in<br>the treatment of ocular<br>pathology in both the posterior<br>and anterior segments.<br>Intended for use in the<br>posterior segment to perform<br>retinal photocoagulation,<br>panretinal photocoagulation.<br>focal photocoagulation and<br>grid photocoagulation for<br>vascular and structural<br>abnormalities of the retina and<br>choroid including:<br>(532 nm)<br>• proliferative and non-<br>proliferative diabetic<br>retinopathy<br>• macular edema                                                                          |              |                     |                    |                    | Circle                 | Hexagon (6-sided array)              |
| Patient<br>Population          | Adult                                                                                                                                                                                                                                                                                  | Adult                                                                                                                                                                                                                                                                                        | Adult                                                                                                                                                                                                                                                                                                  | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |              |                     |                    |                    |                        | Octant (up to 4 rings divided        |
| Product<br>Code /              | HQF/ 21 CFR<br>886.4390                                                                                                                                                                                                                                                                | HQF/ 21 CFR<br>886.4390                                                                                                                                                                                                                                                                      | HQF/ 21 CFR<br>886.4390                                                                                                                                                                                                                                                                                | GEX/ 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | HQF/ 21 CFR 886.4390                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |              |                     |                    |                    |                        | into 8 radial segments)              |
|                                |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                        | trabeculoplasty,<br>iridotomy, iridoplasty<br>including:<br>Retinal photocoagulation<br>(RPC) for the treatment<br>of:<br>Diabetic retinopathy,<br>including:<br>Nonproliferative<br>retinopathy<br>Macular edema<br>Proliferative retinopathy<br>Retinal tears and<br>detachments<br>Lattice degeneration<br>Age-related macular<br>degeneration (AMD) with<br>choroidal<br>neovascularization (CNV)<br>Sub-retinal (choroidal)<br>neovascularization<br>Central and branch retinal<br>vein occlusion<br>Laser trabeculoplasty for<br>the treatment of:<br>Primary open angle<br>glaucoma<br>Laser iridotomy,<br>iridoplasty for the<br>treatment of:<br>Angle closure glaucoma | choroidal<br>neovascularization<br>associated with wet age-<br>related macular<br>degeneration age-related macular<br>degeneration lattice degeneration retinal tears and<br>detachments<br>(577nm) proliferative and non-<br>proliferative diabetic<br>retinopathy macular edema choroidal<br>neovascularization<br>associated with wet age-<br>related macular<br>degeneration age-related macular<br>degeneration lattice degeneration retinal tears and<br>detachments Intended for use in the<br>treatment of ocular pathology<br>in the anterior segment<br>including:<br>(532 nm and 577nm) iridotomy<br>trabeculoplasty |              |                     |                    |                    |                        | Enhanced Octant (multiple            |
| Regulation<br>Number           |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |              |                     |                    |                    |                        | rings from 500-3000 μm               |
| Regulation<br>Description      | Ophthalmic Laser                                                                                                                                                                                                                                                                       | Ophthalmic Laser                                                                                                                                                                                                                                                                             | Ophthalmic Laser                                                                                                                                                                                                                                                                                       | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Ophthalmic Laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |              |                     |                    |                    |                        | radius)                              |
| Laser Type                     | Q-switched Nd:YLF<br>Laser                                                                                                                                                                                                                                                             | Q-switched Nd:YLF<br>Laser                                                                                                                                                                                                                                                                   | Q-switched Nd:YAG<br>Laser, frequency<br>doubled                                                                                                                                                                                                                                                       | Frequency doubled solid-<br>state                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Pumped semiconductor<br>(OPSL); solid-state                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |              |                     |                    |                    |                        | Circle (1-3 rings in full or partial |
| Laser<br>Wavelength            | 527 nm (green)                                                                                                                                                                                                                                                                         | 527 nm (green)                                                                                                                                                                                                                                                                               | 532 nm                                                                                                                                                                                                                                                                                                 | 532 nm, 577 nm, 810 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 532 nm or 577 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |              |                     |                    |                    |                        | circle)                              |
| Laser Energy                   | 0.03 mJ to 0.35 mJ                                                                                                                                                                                                                                                                     | 0.03 mJ to 0.4 mJ                                                                                                                                                                                                                                                                            | Minimum: 0.1 mJ<br>Maximum: 0.6 mJ<br>0.10-0.30 in 0.02<br>steps<br>0.30-0.60 in 0.05<br>steps                                                                                                                                                                                                         | 2W power                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 0.3mW-2000 mW power                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Screen       | Touch LCD Module    | Touch LCD Module   | Tablet             | Unknown                | Touchscreen LCD Module               |
| Pulse<br>Duration              | 1.7 µs                                                                                                                                                                                                                                                                                 | 1.7 µs                                                                                                                                                                                                                                                                                       | 0.003 μs<br>No less than 500 ms<br>for 20 continuous<br>shots                                                                                                                                                                                                                                          | Variable (≤100ms)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 5-1000 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Emission     | Footswitch          | Footswitch         | Footswitch         | Footswitch             | Footswitch                           |
| Aiming<br>Beam                 | 635 nm                                                                                                                                                                                                                                                                                 | 635 nm                                                                                                                                                                                                                                                                                       | 635 nm                                                                                                                                                                                                                                                                                                 | 635 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 635 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Control      |                     |                    |                    |                        |                                      |
| Type of<br>Delivery<br>System  | Slit lamp adapter<br>with optical fiber<br>connected to a<br>separate laser<br>source                                                                                                                                                                                                  | Slit lamp adapter with<br>optical fiber<br>connected to a<br>separate laser source                                                                                                                                                                                                           | Slit lamp adapter<br>with optical fiber<br>connected to a<br>separate laser<br>source                                                                                                                                                                                                                  | Slit lamp adapter with<br>optical fiber connected to<br>a separate laser source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Slit lamp adapter with optical<br>fiber connected to a separate<br>laser source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Laser        | RTF Contact Lens    | Laser Contact Lens | Laser Contact Lens | Laser Contact Lens not | Laser Contact Lens not               |
| Single Spot<br>Size            | 200 μm                                                                                                                                                                                                                                                                                 | 200 μm                                                                                                                                                                                                                                                                                       | 400 μm                                                                                                                                                                                                                                                                                                 | 50-500 μm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 50, 100, 200, 400 μm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Contact Lens | provided            | not provided       | not provided       | provided               | provided                             |
| Multi-spot<br>Deliver<br>Modes | Yes<br>Square Pattern (2x2,<br>3x3, or 4x4)<br>Circle Pattern<br>(radius 500-2000<br>μm)                                                                                                                                                                                               | No                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                                     | Yes<br>Multi-spot patterns (100-<br>500 μm spot sizes)<br>Radius<br>Grid (2x2, 3x3, 4x4, 5x5,<br>6x6,7x7)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes Multi-spot patterns (40-<br>400 µm spot sizes)<br>Array (1x1 to 5x5 grid)<br>Triple Arc<br>Wedge (3-sided 3x3x3 to<br>6x6x6 segment)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Electrical   | AC 100-240V, 50/60  | AC 100-240V, 50/60 | 100-240 VAC,       | 100-240 VAC, 50/60 Hz  | 100 - 230 VAC,                       |
| Rating                         | Hz, Power                                                                                                                                                                                                                                                                              | Hz, Power                                                                                                                                                                                                                                                                                    | 50/60 Hz                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 50/60 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |              |                     |                    |                    |                        |                                      |
|                                | consumption 300VA                                                                                                                                                                                                                                                                      | consumption 300VA                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |              |                     |                    |                    |                        |                                      |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The R:GEN that is the subject of this 510(k) employs the same mechanism of action as the predicate device and achieves its intended use with very similar technological characteristics when compared with the legally marketed predicate and reference devices. Differences in technological characteristics, do not raise difference safety or effectiveness questions and data demonstrates that the R:GEN is as safe and effective as the predicate device for the same intended use. The R:GEN is substantially equivalent to the predicate device based on the information summarized here:

The subject device has a similar design and dimensions, and identical materials as the device cleared in K153769.

The differences in technological characteristics from the predicate include using the same laser technology to provide additional treatment patterns and the use of the Real Time Feedback (RTF) Contact Lens. The subject device offers additional laser treatment patterns including single, square, circle and 2x2 Combo while the predicate device provided a single spot treatment pattern.

The maximum laser energy per pulse for the subject device is lower than the predicate device. The predicate device was designed to deliver a single pulse to the patient with no limitations on the number of pulses to be delivered during a treatment session. The subject device delivers a maximum of 15 pulses per treatment.

The R:GEN subject to this 510(k) has undergone testing to ensure the device is as safe and effective as the predicate for the intended use.

## PERFORMANCE DATA

The R:GEN substantially conforms to the performance standards for light emitting products (21 CFR 1040). Bench testing was performed in accordance with IEC 60825-1 Safety of laser products - Part 1: Equipment classification, and requirement and IEC 60601-2-22 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

## Bench Testing

The following tests were performed to demonstrate safety and performance based on current industry standards:

- . Biocompatibility testing of patient contacting material in compliance to ISO 10993-1, ISO 10993-5, and ISO 10993-10
- Sterilization testing and shelf life validation of the RTF Sensor in compliance to ISO . 11737-1, ISO 10993-7, ISO 11135, and ASTM F1980-16

{10}------------------------------------------------

- Software Verification and Validation were conducted in compliance to IEC 60601-1 and ● IEC 62304. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Major" level of concern, since a failure or latent flaw could indirectly result in death or serious injury of the patient or operator through incorrect or delayed information or through the action of a care provider.
- Electromagnetic Compatibility (EMC) and Electrical Safety testing in compliance to IEC 60601-1, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1, and IEC/EN 60601-1-2

## Animal Testing

An animal studies were performed with the R:GEN that established that the use of subject device wounds RPE cells. Additional published animal studies showing the safety and effectiveness of the subject were provided.

#### Clinical Evidence

Published clinical performance testing established that the subject device performs as intended in the treatment of CSME.

## CONCLUSION

Based on the testing performed, including biocompatibility, sterilization, shelf life validation, software verification and validation, EMC and electrical safety, bench performance testing, animal testing, and published clinical evidence, it can be concluded that the subject device is able to perform equivalently to the predicate device. The R:GEN has the same indications for use, and technological characteristics, that do not raise different questions of safety and effectiveness and therefore is assessed to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQF/K220974](https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQF/K220974)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
