The Cetus Laser System is indicated for use in for phacofragmentation of the cataractous crystalline lens.
Device Story
Cetus System is 1064nm pulsed Nd:YAG Q-switched laser; used with sterile single-use laser probe and external Phacoemulsification machine. Operated by physician in operating room. Laser light delivered via 300um quartz optical fiber; probe converts laser energy into acoustic shockwaves to fragment crystalline lens. Probe includes integrated aspiration and irrigation. System console features digital touch screen for control. Output assists surgeon in lens fragmentation during cataract procedures; benefits patient through mechanical lens breakdown. System uses air cooling and integrated software.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (AAMI ANSI ES60601-1), laser safety (IEC 60825-1), electromagnetic compatibility (IEC 60601-1-2), sterilization (DIN EN ISO 11135-1), packaging integrity (ASTM F1886/F1886M-09, ASTM F1980-07, DIN EN 868-5, ASTM F1608-00), and biological evaluation (DIN ISO 10993-7).
Technological Characteristics
1064nm pulsed Nd:YAG Q-switched laser; pulse length < 10 ns; max repetition rate 20Hz. Quartz optical fiber (300um) delivery. Air-cooled. Digital touch screen interface. Integrated software. Sterilization via EO (DIN EN ISO 11135-1). Electrical safety: AAMI ANSI ES60601-1:2005(R) 2012/A1:2012. Laser safety: IEC 60825-1:2007. EMC: IEC 60601-1-2:2007.
Indications for Use
Indicated for phacofragmentation of the cataractous crystalline lens in patients requiring cataract surgery.
Regulatory Classification
Identification
A vitreous aspiration and cutting instrument is an electrically powered device, which may use ultrasound, intended to remove vitreous matter from the vitreous cavity or remove a crystalline lens.
Special Controls
*Classification.* Class II (special controls). The device, when it is phacofragmentation unit replacement tubing, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 27, 2017
A.R.C. Laser GmbH % Angela Thyzel General Manager Bessemer St. 14 Nurnberg, 90411 DE
Re: K170183
Trade/Device Name: Cetus System, Cetus Probe Regulation Number: 21 CFR 886.4150 Regulation Name: Vitreous Aspiration and Cutting Instrument Regulatory Class: Class II Product Code: HQE Dated: August 24, 2017 Received: August 24, 2017
Dear Angela Thyzel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Bradley S. Cunningham -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170183
Device Name Cetus System
Indications for Use (Describe)
The Cetus Laser System is indicated for use in phacofragmentation of the cataractous crystalline lens.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for A.R.C. Laser. The logo is set against a blue background. The text "A.R.C." is in large, bold, white letters. Below that, the word "LASER" is in a white box. Underneath the box, the phrase "enlighten your surgery" is written in a smaller, cursive font.
K170183
# 510(k) SUMMARY
| Title: | A.R.C. Cetus System |
|--------------------------|--------------------------------------------------------------------|
| Submitter: | A.R.C. Laser GmbH<br>Bessemer St. 14<br>90411, Nurnberg<br>Germany |
| Contact : | Angela Thyzel |
| Tittle : | General Manager |
| Phone : | 0911-21779-0 |
| Fax : | 0911-21779-99 |
| E-Mail: | a.thyzel@arclaser.de |
| Address: | Bessemer St. 14<br>90411, Nurnberg<br>Germany |
| Date<br>Prepared: | September 26, 2017 |
| Device Trade<br>Name: | Cetus System |
| Common<br>Name: | Phacofragmentation System |
| Classification<br>Name: | Vitreous aspiration and cutting instrument, ophthalmic |
| Device product<br>code: | HQE |
| Device<br>Classification | 21 CFR 886.4150 |
| Predicate<br>Device: | Laser Photolysis System, K993154 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image features a logo for A.R.C. Laser, set against a vibrant blue background. The acronym "A.R.C." is prominently displayed in large, bold white letters, with the word "LASER" directly beneath it in a smaller, yet still noticeable, white font. Below the word "LASER" is a tagline in a cursive font that reads "enjoyed in your surgery".
# Device Description:
The Cetus system is a 1064nm. pulsed, Nd: YAG Q-switched laser system used together with a sterile, single use laser probe and a Phaco machine. The system console has a digital touch screen and it is used in conjunction with the Cetus probe and a Phacoemulsification machine in the operating room. The probe converts the laser light energy into acoustic energy which is used to perform the fragmentation of the the crystalline lens and has integrated aspiration and irrigation functions.
### Intended Use:
The Cetus Laser System is indicated for use in for phacofragmentation of the cataractous crystalline lens.
## Substantial Equivalence:
The Cetus system have the same mechanism of action, same intended use and the same technological characteristics as the predicate device, the Laser Photolysis System (K993154). and that the minor differences should not raise concerns regarding safety and/ or effectiveness.
| Specification laser<br>system | Subject Device | Predicate |
|-------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| 510(k) # | K170183 | K993154 |
| Manufacturer | A.R.C. Laser | A.R.C. Laser |
| Indications for Use | Intended for use in<br>phacofragmentation of the<br>cataractous crystalline lens | Intended for use in<br>phacofragmentation of the<br>cataractous crystalline lens |
| Submission includes | Cetus laser system & Cetus probe<br>only | Laser photolysis system, probe &<br>phaco machine |
| Laser media | Nd: YAG | Nd: YAG |
| Mode of operation | Pulsed, Q-Switch | Pulsed, Q-Switch |
| Wavelength | 1064 nm | 1064 nm |
| Used in combination | with a probe and a Phaco machine | with a probe and a Phaco machine |
| Laser mechanism | Shockwave | Shockwave |
| Repetition rate | Max. 20Hz | Max. 20Hz |
| Laser Class | I | I |
| Pulse length | < 10 ns | < 10 ns |
| Mains supply | 100V to 240V – 50/60 Hz | 100V to 240V – 50/60 Hz |
| Cooling | Air, internal | Water, internal |
| Software | Integrated in system | From Phaco machine |
Comparison Table - Cetus system & Laser Photolysis System
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for A.R.C. Laser. The logo is set against a blue background. The text "A.R.C." is in large, bold, white letters, with the word "LASER" directly underneath in a smaller font. Below that, the tagline "enjoy laser surgery" is written in a cursive font.
### K170183
| Specification | Subject Device | Predicate |
|---------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------|
| Laser beam delivery | Quartz optical fiber 300um with laser probe | Quartz optical fiber 300um with laser probe |
| Dimensions system | H 13,5 cm. (5,31 in.) X W 47,8 cm.<br>(18,8 in.) X L 42,1 cm. (16,5 in.) | H 16 cm. (6,3 in.) X W 47 cm.<br>(18,5 in.) X L 56 cm. (22 in.) |
| Display | Digital touch screen | External in Phaco machine |
| Weight | 12 Kg. (26.4 pounds) | 18 Kg. (39.6 pounds) |
Comparison Table - Cetus system & Laser Photolysis System (Continuation)
The Cetus system is a further development of its predicate (A.R.C. Laser Photolysis System, K993154). The predicate device (K993154) includes a laser system, a probe and a phaco machine. The current submission includes the laser system and the probe only. Additional minor differences between the subject and predicate systems include the cooling system (air vs. water), software (integrated vs. external), display, dimensions and weight as shown in the table above.
Non clinical Performance Data: Performance testing was conducted in order to demonstrate that the A.R.C. Cetus System is as safe and as effective as the cleared predicate device.
The Cetus system has been tested for Electrical Safety compliance to AAMI ANSI ES60601-1:2005(R) 2012 and A1:2012, for Laser Safety to IEC 60825-1:2007, for Electromagnetic Compatibility to IEC 60601-1-2:2007, Sterilization to DIN EN ISO 11135-1:2014, Packaging Integrity to ASTM F1886/ F1886M-09:2013, Accelerated ageing of sterile barrier to ASTM F1980-07:2011, Sterile Packaging to DIN EN 868-5:2009, Microbial ranking of porous packaging materials to ASTM F1608-00:2009, Biological evaluation of EO residuals to DIN ISO 10993-7:2009 and Risk Management been conducted according to ISO 14971: 2007 + Am 2010.
According to the test results provided in this submission, we conclude that the Cetus system is as safe and effective as the predicate device.
Clinical Performance Data: None