The Storz D4600 is intended to be used as an accessory to the Storz PREMIERE® (K894278, K921460, K946227) or Storz DAISY® (K854508) Microsurgical Systems for use in air and fluid exchange during posterior segment ophthalmic surgical procedures. The D4600 Air Exchange Line provides the initial pathway through which air generated by the microsurgical system is routed into the posterior chamber of the eye.
Device Story
Accessory tubing pack for ophthalmic microsurgical systems; routes air from system to posterior chamber of eye during surgery. Single-patient use; sterile; gamma-irradiated. Used by ophthalmic surgeons in clinical settings to facilitate air/fluid exchange. Device provides physical pathway for air delivery; no active electronic or software components.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Tubing set for air delivery; sterile; gamma irradiation sterilization; packaging includes Tyvek® and PET/PE pouch. No electronic components, software, or active sensing.
Indications for Use
Indicated for use in posterior segment ophthalmic surgical procedures requiring air and fluid exchange.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
K091441 — SYNERGETICS ADJUSTABLE GAS PRESSURIZED INFUSION (AGPI) TUBE SET · Synergetics, Inc. · Oct 30, 2009
K974561 — MOISTAIR FLUID AIR TUBING SET · American Medical Devices, Inc. · Feb 19, 1998
Submission Summary (Full Text)
{0}
K962131
storz®
AUG 23 1996
Premarket Notification
Storz Instrument Company
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
The following information is submitted in accordance with the requirements of 21 CFR 807.92:
Contact Person: Regulatory Affairs Associate
Date Prepared: May 31, 1996
Proprietary Name: Storz D4600 Air Exchange Line Pack.
Common/Usual Name: Ophthalmic surgical system accessory pack for posterior segment procedures.
Classification Name: System, Phacofragmentation (86 HQC).
Device Description/Intended Use: The Storz D4600 is intended to be used as an accessory to the Storz PREMIERE® (K894278, K921460, K946227) or Storz DAISY® (K854508) Microsurgical Systems for use in air and fluid exchange during posterior segment ophthalmic surgical procedures. The D4600 Air Exchange Line provides the initial pathway through which air generated by the microsurgical system is routed into the posterior chamber of the eye.
Predicate Device: The proposed Storz D4600 Air Exchange Line Pack is substantially equivalent to the existing Storz D4600 Air Exchange Line Pack (K894278), and to the Intek International, Inc. TR-9006-11 Air Delivery Tubing System (K832761).
Predicate Comparison: A chart comparing the Storz D4600 to the predicate devices, demonstrating substantial equivalence, is attached.
Submitted by:
Audrey Swearingen
Storz Instrument Company
3365 Tree Court Industrial Blvd.
St. Louis, MO 63122
Ph. 314/225-5051
STORZ
58
{1}
59
# DEVICE COMPARISON CHART
## Storz D4600 Air Exchange Line Pack
| Device | Storz D4600, modified | Storz D4600, existing | Intek International TR9006-11 |
| --- | --- | --- | --- |
| Intended Use | To deliver air to the eye during posterior ophthalmic surgical procedures | To deliver air to the eye during posterior ophthalmic surgical procedures | To deliver air to the eye during ophthalmic surgical procedures |
| Recommended System | Storz PREMIERE® or Storz Daisy™ Microsurgical Systems | Storz PREMIERE® or Storz Daisy™ Microsurgical Systems | Escalon™ CFAE-40 Air/Fluid Exchange Pump |
| Instructions For Use Included? | Yes | No | Unknown |
| Single Patient Use? | Yes | Yes | Yes |
| Provided Sterile? | Yes | Yes | Yes |
| Sterilization Method | Gamma Irradiation | Gamma Irradiation | Unknown |
| Packaging | Singly in Tyvek® and PET/PE Pouch, Ten pouches per Box | Singly in Tyvek® and PET/PE Pouch, Ten pouches per Box | Unknown package type, Ten units per Box |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.