K103023 · Abbott Medical Optics, Inc. · HQC · Jun 1, 2011 · Ophthalmic
Device Facts
Record ID
K103023
Device Name
UNIT, PHACOFRAGMENTATION
Applicant
Abbott Medical Optics, Inc.
Product Code
HQC · Ophthalmic
Decision Date
Jun 1, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.4670
Device Class
Class 2
Intended Use
The proposed irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Device Story
AMO Irrigation Sleeve is a molded silicone component placed over a phacoemulsification tip; used during ocular surgery. Device directs irrigation solution across the tip shaft into the eye. Operated by ophthalmic surgeons in clinical settings. Performance characteristics include flow rate >= 18cc and length of 0.75-1.50 inches. Device ensures proper fluid delivery during phacoemulsification procedures.
Clinical Evidence
Bench testing only. Compliance with ISO 10993-1 (biocompatibility), ISO 11137-1/2/3 (sterilization), and ISO 14971 (risk management). Testing validated durometer, tensile strength, flow rate, and post-sterilization product integrity.
Technological Characteristics
Material: Silicone (80+/-5 durometer). Dimensions: 0.75-1.50 inches. Flow rate: >= 18cc. Sterilization: Validated per ISO 11137. Non-electronic, mechanical component.
Indications for Use
Indicated for use during ocular surgery to direct irrigation solution across the shaft of a phacoemulsification tip into the eye.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
K981109 — QUITETIP · Paradigm Medical Industries, Inc. · Jun 9, 1998
K192418 — Phacoemulsification Needle (Tip) and Irrigation Sleeve · Raico, LLC · May 27, 2020
K971173 — ML DISPOSABLE ALK TUBING · Med-Logics, Inc. · May 22, 1997
K081681 — VITRECTOMY CUTTER AND ACCESSORIES · Advanced Medical Optics, Inc. · Oct 8, 2008
K961831 — STORZ DP5200 IRRIGATION ADDMINISTRATION SET PACK · Storz Instrument Co. · Aug 6, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
KIO3023
JUN - 1 2011
# 510(K) SUMMARY
This 510(k) summary is being submitted in accordance with the Medical Device Amendments of 1976, the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
| Applicant Name: | Abbott Medical Optics, Inc.<br>1700 E. St. Andrew Place<br>Santa Ana CA 92705, USA<br>Establishment Registration Number: 2020664 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Evelyn De La Vega<br>Manager, Regulatory Affairs<br>Tel: (714) 247-8487<br>Fax: (714) 247-8784 |
| Date of Summary Preparation: | March 22, 2011 |
| Device Trade Name: | AMO Irrigation Sleeve |
| Device Classification: | CDRH Device Class II |
| Classification Code: | HQC<br>21 CFR 886.4670<br>Unit, Phacofragmentation |
Reviewing Panel: Ophthalmic Devices
Devices to Which Substantial Equivalence is Claimed:
| 510(k) Number | Trade Name | Manufacturer |
|---------------|--------------------------|---------------|
| K844373 | Ultra Phaco | OMS* |
| K840695 | Phacoemulsification Kits | Allergan ISP* |
* AMO currently owns the rights to these devices.
## DEVICE DESCRIPTION
The Abbott Medical Optics (AMO) irrigation sleeve is a device that is intended to direct irrigation solution across the shaft of the phacoemulsification tip, allowing the irrigation solution to enter the eye during ocular surgery. It is a small, molded component that is placed over the phacoemulsification tip. The proposed irrigation sleeve is made of silicone, with a hardness specification of 80+/-5 durometer units. The flow rate of the sleeve is greater than or equal to 18cc, with a length of 0.75 -1.50 inches.
Abbott Medical Optics Inc.
{1}------------------------------------------------
### INDICATIONS FOR USE
The proposed irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
#### STATEMENT OF COMPARISON TO PREDICATE DEVICES
The physical and performance characteristics of the proposed irrigation sleeve are substantially equivalent to those of the predicate devices cleared under K844373, and K840695.
#### BRIEF SUMMARY OF NONCLINICAL TESTS AND RESULTS
The proposed AMO irrigation sleeve has undergone testing and is in compliance with the following standards; ISO 10993-1:2003 Biological Evaluation of Medical Devices Part 1 Evaluation and Testing, ISO 11137-1:2006 Sterilization of Health Care Product Requirements for Validation and Routine Control- Part 1 Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices, ISO 11137-2:2006 Sterilization of Health Care Product Requirements for Validation and Routine Control- Part 2 Establishing Dose, ISO 11137-3:2006 Sterilization of Health Care Product Requirements for Validation and Routine Control- Part 3 Guidance on Dosimetric Aspects, and ISO 14971:2007 Medical Devices- Application of Risk Management to Medical Devices. Testing of the sleeve included validation of the durometer and tensile strength, flow rate, and product integrity after sterilization. The AMO irrigation sleeve was found to perform equivalently to the predicate irrigation sleeves. Thus, the proposed AMO irrigation sleeve and the predicate devices have similar effectiveness and performance profiles and are considered substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Abbott Medical Optics, Inc. % Ms. Evelyn De La Vega Regulatory Affairs Manager 1700 St. Andrew Place Santa Ana, CA 92705
JUN -1 201
Re: K103023
Trade/Device Name: Phacofragmentation Unit Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HOC Dated: May 23, 2011 Received: May 24, 2011
Dear Ms. De La Vega:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
1 :: 21
1. 1 :11 :
・・
Sincerely yours,
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: AMO Irrigation Sleeve
Indication for Use:
The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
# : (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Use
(Per 21 CFR 801.109)
96)
OR
Over-The-Counter
(Optional Format 1-2-
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
510(k) Number. K103023
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.