← Product Code [HQC](/submissions/OP/subpart-e%E2%80%94surgical-devices/HQC) · K031404

# DIGNITY OPHTHALMIC SURGICAL SYSTEM (K031404)

_Phacor Inc., A Santen Company · HQC · May 22, 2003 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQC/K031404

## Device Facts

- **Applicant:** Phacor Inc., A Santen Company
- **Product Code:** [HQC](/submissions/OP/subpart-e%E2%80%94surgical-devices/HQC.md)
- **Decision Date:** May 22, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4670
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The Dignity™ Ophthalmic Surgical System is a Phacofragmentation device intended for use in cataract surgery to emulsify, and extract the cataract.

## Device Story

Dignity™ Ophthalmic Surgical System is a phacofragmentation unit used by ophthalmic surgeons during cataract surgery. The system console connects to various surgical accessories to perform phacoemulsification, irrigation/aspiration, diathermy, and vitrectomy. It functions by emulsifying the cataractous lens and extracting the material from the eye. The device features a graphical user interface (GUI) for navigation and controls for footpedal operation and cassette/cartridge loading. It is intended for clinical use in an operating room setting. The system's primary benefit is facilitating the surgical removal of cataracts through controlled emulsification and aspiration.

## Clinical Evidence

Bench testing only. In vitro performance testing demonstrated substantial equivalence to predicate devices regarding phacoemulsification, irrigation/aspiration, diathermy, and vitrectomy capabilities.

## Technological Characteristics

Phacofragmentation system (21 CFR 886.4670). Console-based system with integrated GUI, footpedal, and cassette/cartridge loading mechanism. Performs phacoemulsification, irrigation/aspiration, diathermy, and vitrectomy. Software facilitates user interface and hardware control.

## Regulatory Identification

A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.

## Predicate Devices

- Sovereign™ ([K981116](/device/K981116.md))
- Legacy® Series 20000™ ([K955789](/device/K955789.md))
- Millenium™ ([K980488](/device/K980488.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K031404

Page 1 of 2

# Premarket Notification, 510(k) Summary

Date: 5/2/03

Submitter Information: Jeff B. Paul Director. Quality Assurance and Requlatory Affairs Phacor Inc., A Santen Company 775 Fiero Lane San Luis Obispo, CA 93401 805-546-1815 805-546-1826 fax

Device Name:

Trade or Proprietary Name:

Dignity™ Ophthalmic Surgical System

Common or Usual Name:

Phacofragmentation unit

Classification Number:

21 CFR section 886.4670

### Predicate Device Comparison:

The Dignity™ Ophthalmic Surgical System device that is the subject of this 510(k) is substantially equivalent to predicate devices marketed by Advanced Medical Optics Inc., Alcon Laboratories Inc. and Bausch & Lomb. The Advanced Medical Optics device most closely equivalent to the Phacor Inc. device is the Sovereign™, cleared for commercial distribution under 510(k) number K981116. The Alcon Laboratories device most closely equivalent to the Phacor Inc. device is the Legacy® Series 20000™, cleared for commercial distribution under 510(k) number K955789. The Bausch & Lomb device most closely equivalent to the Phacor Inc. device is the Millenium™, cleared for commercial distribution under 510(k) number K980488.

The significant differences between the Dignity™ Ophthalmic Surgical System and the three predicate devices, Sovereign™, Legacy™ and Millenium™ are listed below:

- t External and internal structural differences. The external differences are in size, weight, and configuration. Also the Graphical User Interface has a different navigational format.
- Software will facilitate a more user friendly format in coordination with the . footbedal and cassette/cartridge loading and unloading.

{1}------------------------------------------------

K031404

Page 2 of 2

# Device Description:

The Dignity™ Ophthalmic Surgical System is used by ophthalmic surgeons to perform cataract surgery. The system utilizes accessories connected to the console to perform the cataract surgery.

#### Intended Use:

The Dignity™ Ophthalmic Surgical System is a Phacofragmentation device intended for use in cataract surgery to emulsify, and extract the cataract.

## Summary of non-clinical tests:

Performance tests were conducted on the Dignity™ Ophthalmic Surgical System. The performance in an in vitro model was substantially equivalent to the predicate devices with respect to Phacoemulsification, Irrigation/Aspiration, Diathermy and Vitrectomy.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines extending from its head, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

MAY 2 2 2003

Food and Drug Administration 9200 Corporate Boulevard · Rockville MD 20850

Phacor Inc., A Santen Company c/o Jeff B. Paul Director, Quality Assurance & Regulatory Affairs 775 Fiero Lane San Luis Obispo, CA 93401

Re: K031404

Trade/Device Name: Dignity™ Ophthalmic Surgical System Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation system Regulatory Class: Class II Product Code: HQC Dated: May 2, 2003 Received: May 8, 2003

Dear Mr. Paul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

# Page 2 - Jeff B. Paul

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

A képi korentbol

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

S10(k) Number (if known): K031404

Device Name: DignityTM Ophthalmic Surgical System

Indications for Use:

The Dignity™ Ophthalmic Surgical System is a Phacofragmentation device intended for use in cataract surgery to emulsify, and extract the cataract.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

Denis L. McCarthy

(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises

510(k) Number K031404

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQC/K031404](https://fda.innolitics.com/submissions/OP/subpart-e%E2%80%94surgical-devices/HQC/K031404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
