M2 SINGLE USE MICROKERTOME

K022560 · Moria SA · HMY · Sep 27, 2002 · Ophthalmic

Device Facts

Record IDK022560
Device NameM2 SINGLE USE MICROKERTOME
ApplicantMoria SA
Product CodeHMY · Ophthalmic
Decision DateSep 27, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4370
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The M2 SINGLE USE microkeratome is intended for use in making of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.

Device Story

The M2 Single Use microkeratome is a surgical instrument used by ophthalmologists to create corneal flaps during LASIK or lamellar keratoplasty. The system consists of a power unit (providing vacuum and motor control), a motor (electric or gas turbine), a keratome head with a pre-inserted blade, suction rings for eye fixation, and applanator lenses for diameter control. The device is operated in a clinical setting. The surgeon uses the power unit to control vacuum pressure and motor speed to drive the blade oscillation and advancement. The device creates a circular lamellar resection of predetermined diameter and thickness. By automating the resection process, the device provides consistent flap thickness and bed smoothness, facilitating subsequent refractive surgery steps and potentially improving patient outcomes.

Clinical Evidence

Clinical evidence includes in-vitro studies on porcine eyes demonstrating flap thickness consistency, safety, and quality of corneal resections. In-vivo studies on 72 human eyes confirmed the device is safe and capable of creating circular lamellar resections of predetermined diameter and thickness with bed smoothness equivalent to the predicate device.

Technological Characteristics

System includes power unit (vacuum/motor control), electric or gas turbine motor, keratome head with low carbon steel blade, methylmethacrylate applanator lenses, and stainless steel handles. Operates via vacuum fixation and mechanical oscillation/advancement. Power unit features vacuum/gas pressure gauges and battery indicators. Designed for clinical use; no software-based automated analysis or ML algorithms.

Indications for Use

Indicated for patients undergoing LASIK surgery or other treatments requiring initial lamellar resection of the cornea.

Regulatory Classification

Identification

A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### MORIA S.A. M2 SINGLE USE microkeratome July 2nd, 2002 Premarket Notification KOZZ560 # SEP 2 7 2002 Section 4 page 1 ## 510(k) SUMMARY ### 1. Submitter's identification - a. MORIA S.A. 15, rue Georges Besse 92160 ANTONY France Tel: (33-1) 46 74 46 36 Fax: (33-1) 46 74 46 70 - Contact person: Mélanie RENAUD-SAMIRI QA & Regulatory Affairs Manager E mail: mrenaud@moria-int. com July 2nd, 2002 C. Date Summary Prepared: ### 2. Device name b. M2 SINGLE USE microkeratome Trade Name: ### 3. Classification name Keratome (per CFR 886.4370) ### 4. Substantial equivalence Substantial equivalence is being claimed to the following legally marketed devices: Device name: CARRIAZO BARRAQUER SINGLE USE Microkeratome Company name: MORIA S.A. Document control number: K003594 ### And Device name: CARRIAZO BARRAQUER II Microkeratome Company name: MORIA S.A. Document control number: K002191 {1}------------------------------------------------ ### Section 4 page 2 ### 5. Device description List of components - a) Power unit - b) Motor - c) Suction rings - d) Applanator lenses - e) Footswitches - f) Keratome head with pre-inserted Keratome blade ### Power unit a) The power unit used for the M2 SINGLE USE microkeratome is the same as the power unit used for the predicate devices CARRIAZO BARRAQUER SINGLE USE Microkeratome (K003594) & CARRIAZO BARRAQUER II Microkeratome (K002191) already legally marketed in the USA by our company. The power unit includes pumps for producing vacuum. The power unit has been designed to operate the Keratome by means of electric motor or by means of a gas turbine motor. Only one of the above power options can be selected at the time by means of a 2 position switches in the front panel. The front panel has several displays and features: - Vacuum pressure gauge, - - Gas pressure gauge, - - -Battery level indicator, - Battery charge indicator, - - -Connectors: - . DC motor outlet, - . Gas outlet. - Vacuum outlet, ● The back panel has several displays and features: - -Connectors: - Gas inlet, - . Foot pedals, - . Battery charger. All connectors are of different type for preventing connecting mistakes. {2}------------------------------------------------ Section 4 page 3 #### b) Keratome motor ### Option 1: Turbine motor The turbine motor is gas powered. The recommended gas is medical grade nitrogen. The turbine has been marketed by MORIA since six years (see predicate device PLANCON MICROLAMELLAR KERATOME K970377). ### Option 2: The electrical motor The drive system has two built-in electrical motors (one motor for the blade oscillation and one motor for the advance of the microkeratome). ### Keratome head C) The Keratome head includes the blade that is moved by the motor. Different heads are available in order to adjust the thickness of the cut. #### d) Suction rings The suction rings are used to fixate and pressurize the eye and provide a base for the microkeratome. #### e) Applanator lenses The applanator lenses are made of clear methylmethacrylate with a stainless steel handle. They are used with the rings to control disk diameter before the cut. The upper face is convex for magnification. The base face (contact face) is plane, with an engraved and calibrated reticule diameter. {3}------------------------------------------------ Section 4 page 4 ### f) Keratome blade The blade is made of two parts: the blade edge in low carbon steel, and the blade holder, which is not in contact with the patient's eye. ### 6. Statement of intended use The M2 SINGLE USE microkeratome is intended for use in making of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. ### 7. Discussion of tests and results Keratomes have been used for lamellar keratoplasty for more than 30 years. In-vitro studies on porcine eyes demonstrated: - -The flap thickness consistency, - The safety of corneal resections, - - The good quality of corneal resections. - In-vivo studies on 72 human eyes showed that the M2 SINGLE USE microkeratome is a safe Keratome able to create, equivalently to the predicate device, circular lamellar resection of a predetermined diameter and thickness and bed smoothness. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with three heads. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MORIA S.A. c/o Mrs. Melanie Renaud-Samiri Ouality Assurance & Regulatory Affairs Manager 15, Rue Georges Besse 92160 Antony France Re: K022560 Trade/Device Name: M2 Single Use microkeratome Regulation Number: 886.4370 Regulation Name: Keratome Regulatory Class: I Product Code: HMY Dated: July 2, 2002 Received: August 2, 2002 Dear Mrs. Renaud-Samiri: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, A. Ralph Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health {6}------------------------------------------------ : Section 5 page 1 510(k) Number (if known): Device Name: : M2 SINGLE USE microkeratome ### Indications for use: The M2 SINGLE USE microkeratome is intended for use in the making of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Qevice Eyaluation (ODE) (Division Sign-Off) Division of Ophthalmic Ear. Nose and Throat Devises 510(k) Number_ ( 0 2 2 5 6 0 2 2 5 6 0 - Prescription Use V (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...