AHMED ClearPath Glaucoma Drainage Device

K182518 · New World Medical, Inc. · KYF · Jan 18, 2019 · Ophthalmic

Device Facts

Record IDK182518
Device NameAHMED ClearPath Glaucoma Drainage Device
ApplicantNew World Medical, Inc.
Product CodeKYF · Ophthalmic
Decision DateJan 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.3920
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K182518 · Jan 18, 2019AHMED ClearPath Glaucoma Drainage DeviceNew World Medical, Inc.Published clinical literature (Ophthalmology, American Journal of Ophthalmology, British Journal of Ophthalmology, Investigative Ophthalmology & Visual Science)The sponsor utilized peer-reviewed clinical literature to support the substantial equivalence of the device's design features (e.g., lack of ridge, ripcord use for tube occlusion) and surgical placement techniques by referencing outcomes from studies of predicate and similar glaucoma drainage devices.Literature review; Clinical outcomes; Surgical technique; Real-world clinical practice

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Ahmed Baerveldt Comparison (ABC) Study; Comparative clinical study (referenced in literature); Follow-up/Duration: 5 yearsPatients with refractory glaucomaBaerveldt Glaucoma ImplantTreatment outcomes and postoperative complications
Location of glaucoma drainage devices relative to the optic nerve; Observational/Anatomical studyPatients undergoing glaucoma drainage device implantationNot applicable for this studyOptimal positioning of glaucoma drainage devices

Indications for Use

The AHMED® ClearPath glaucoma drainage device is indicated for the management of refractory glaucoma where previous surgical treatment has failed or is not expected to provide satisfactory results. Such refractory glaucoma may include but is not limited to: neovascular glaucoma, congenital or infantile glaucoma, and refractory glaucoma resulting from aphakia or uveitis.

Device Story

Non-valved glaucoma drainage device; shunts aqueous humor from anterior chamber to subconjunctival space to reduce intraocular pressure. Consists of medical-grade silicone tube attached to barium-impregnated silicone episcleral plate. Implanted by ophthalmologists in surgical setting. Optional 23-gauge needle and polypropylene ripcord provided for tube occlusion to prevent early postoperative hypotony. Plate conforms to globe equator; facilitates bleb formation. Single-use, sterile, permanent implant.

Clinical Evidence

Bench testing only. No clinical data provided for the subject device. Performance verified via ANSI Z80.27 testing (pressure/flow, structural integrity, dimensional verification) and biocompatibility (ISO 10993-1). Sterilization validated per ISO 11137-1/2. Literature cited for predicate performance and clinical context of design features (e.g., ripcord use).

Technological Characteristics

Materials: Medical-grade silicone, barium sulfate (impregnated in plate). Energy: None (passive). Dimensions: 250mm² and 350mm² plate surface areas; 32mm tube length. Connectivity: None. Sterilization: Gamma irradiation (15 kGy). Design: Non-valved aqueous shunt.

Indications for Use

Indicated for patients with refractory glaucoma (e.g., neovascular, congenital/infantile, aphakic, or uveitic) where prior surgical treatment failed or is expected to provide unsatisfactory results.

Regulatory Classification

Identification

An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.

Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and (3) “Aqueous Shunts—510(k) Submissions.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 18, 2019 New World Medical, Inc. Mukesh Sabarad Manager, Regulatory Affairs 10763 Edison Ct Rancho Cucamonga, California 91730 Re: K182518 Trade/Device Name: AHMED® ClearPath Glaucoma Drainage Device Regulation Number: 21 CFR 886.3920 Regulation Name: Aqueous Shunt Regulatory Class: Class II Product Code: KYF Dated: December 14, 2018 Received: December 17, 2018 #### Dear Mukesh Sabarad: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Tieuvi H. Nguyen -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K182518 Device Name AHMED® ClearPath Glaucoma Drainage Device Indications for Use (Describe) The AHMED® ClearPath glaucoma drainage device is indicated for the management of refractory glaucoma where previous surgical treatment has failed or is not expected to provide satisfactory results. Such refractory glaucoma may include but is not limited to: neovascular glaucoma, congenital or infantile glaucoma, and refractory glaucoma resulting from aphakia or uveitis. Type of Use (Select one or both, as applicable): | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. | I. SUBMITTER | New World Medical, Inc<br>10763 Edison Ct<br>Rancho Cucamonga, CA 91730<br>USA<br>(909) 466-4304 | |-----------------|---------------------------------------------------------------------------------------------------| | Contact Person: | Mukesh Sabarad<br>Manager, Regulatory Affairs<br>New World Medical, Inc.<br>(909) 466-4304 Ext146 | January 17, 2019 Date Summary Prepared: #### II. DEVICE | Trade Name: | AHMED® ClearPath Glaucoma Drainage Device | |----------------------|-------------------------------------------| | Common Name: | Glaucoma Implant | | Classification Name: | Aqueous shunt (21 CFR 886.3920) | | Device Class: | Class II | | Device Product Code: | KYF | #### III. PREDICATE DEVICES | Predicate devices information | | |------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | 510(k) Number: K905129<br>Device Name: Baerveldt Glaucoma Implant (BGI)<br>Decision Date: 02/11/1991 | 510(k) Number: K955455<br>Device Name: Baerveldt Pars Plana Glaucoma Implant (BGI)<br>Decision Date: 02/18/1997 | | Reference device information | | | 510(k) Number: K162060<br>Device Name: Ahmed Glaucoma Valve<br>Decision Date: 10/24/2016 | | {4}------------------------------------------------ #### IV. DEVICE DESCRIPTION The AHMED® ClearPath Glaucoma Drainage Device is a non-valved drainage device designed to shunt aqueous in eyes suffering from refractory glaucoma. Two models CP250 are available covering surface areas of approximately 250mm². The implant consists of a medical grade silicone tube secured to a medical grade silicone episcleral plate near the anterior suture points. The anterior suture points are located on the anterior side of the plate, flanking each side of the tubing track. The silicone plate is barium impregnated to increase ultrasound resolution and identification with CT scan, MRI and plain skull films. The plate conforms to the shape of the globe at its equator and provides a surface from which fluid can be dispersed. Each AHMED® ClearPath is supplied with a 23-gauge hypodermic needle and a 2-inch polypropylene ripcord (pre-loaded in the lumen of the tube) in a sterile, sealed double- pouch. The supplied needle and ripcord are for optional use and are available to be incorporated into the implantation procedure per the surgeon's usual routine. The device is for single use only. #### V. INTENDED USE The AHMED® ClearPath glaucoma drainage device has the same intended use as the predicates and all other devices regulated within the generic type of device known as aqueous shunts in accordance with 21 CFR 886.3920. The device is a prescription (Rx) device that is intended to be permanently implanted to reduce intraocular pressure for the management of refractory glaucoma. The AHMED® ClearPath glaucoma drainage device will bear the following indications for use statement: The AHMED® ClearPath glaucoma drainage device is indicated for the management of refractory glaucoma where previous surgical treatment has failed or is not expected to provide satisfactory results. Such refractory glaucoma may include but is not limited to: neovascular glaucoma, congenital or infantile glaucoma, and refractory glaucoma resulting from aphakia or uveitis. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITHTHE VI. PREDICATE DEVICES The technological characteristics of the AHMED® ClearPath glaucoma drainage device and the Baerveldt glaucoma implant are similar. Furthermore, the differences in technological characteristics that exist between the AHMED® ClearPath glaucoma drainage device and both legally marketed predicate devices do not raise different questions of safety and effectiveness. {5}------------------------------------------------ ## Comparison of the AHMED® ClearPath glaucoma drainage device and the Predicate Devices | Feature | AHMED® ClearPath (ACP) | Baerveldt Glaucoma Implant (BGI) | |--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | To be permanently implanted to reduce<br>intraocular pressure for the management of<br>Glaucoma | To be permanently implanted to reduce<br>intraocular pressure for the management of<br>Glaucoma | | Indication for Use | The AHMED® ClearPath Glaucoma<br>Drainage Device is indicated for the<br>management of refractory glaucoma where<br>previous surgical treatment has failed or is<br>not expected to provide satisfactory results.<br>Such refractory glaucoma may include but is<br>not limited to neovascular glaucoma,<br>congenital or infantile glaucoma, and<br>refractory glaucoma resulting from aphakia<br>or uveitis. | The Baerveldt Glaucoma Implant is<br>indicated for use in patients (with prior<br>vitrectomy for Pars Plana) with medically<br>uncontrollable glaucoma and poor surgical<br>prognosis, such as, but not limited to:<br>neovascular glaucoma, aphakic/pseudophakic<br>glaucomas, patients who have failed<br>conventional surgery, congenital glaucomas<br>and secondary glaucomas due to uveitis,<br>epithelial downgrowth, etc. | | Regulation<br>Number/Product Code | 886.3920, KYF | 886.3920, KYF | | Rx or OTC | Rx | Rx | | Permanent Implant | Yes | Yes | | Anatomical Site | The tube is implanted in the anterior chamber<br>and the endplate is implanted under the<br>conjunctiva | The tube is implanted in the anterior chamber<br>and the endplate is implanted under the<br>conjunctiva | | Material (endplate) | Medical grade silicone, barium impregnated | Medical grade silicone, barium impregnated | | Material (tube) | Medical grade silicone | Medical grade silicone | | Location of anterior<br>suturing holes<br>(relative to the limbus) | 6-8mm | 8-10 mm for model BG101-350 and BG103-<br>250. Model BG102-350 with Pars Plana Clip it<br>is 2-7mm | | Mechanism of action | Creates a channel via a tube for aqueous flow<br>from the anterior chamber to the<br>subconjunctival space resulting in a bleb<br>formed within the encapsulated<br>polypropylene. | Creates a channel via a tube for aqueous flow<br>from the anterior chamber to the<br>subconjunctival space resulting in a bleb<br>formed within the encapsulated polypropylene | | Design | Silicone tube attached to a silicone endplate | Silicone tube attached to a silicone endplate | | Plate Footprint Area<br>(Surface area) | 250 mm2 and 350 mm2 | 250 mm2 and 350 mm2 | | Tube Dimensions | Length: 32 mm<br>Inner diameter: 0.30 mm<br>Outer Diameter: 0.63 mm | Length: 32 mm<br>Inner diameter: 0.305 mm<br>Outer Diameter: 0.635 mm | | Feature | AHMED® ClearPath (ACP) | Baerveldt Glaucoma Implant (BGI) | | Plate Thickness | 0.81mm | 0.84mm (1.5mm for the Pars-Plana Clip) | | Sterilization | Gamma | Gamma | | How Supplied | Sterile | Sterile | | Single Use | Yes | Yes | {6}------------------------------------------------ ## VII. PERFORMANCE DATA #### Bench Testing: The nonclinical bench testing conducted on the AHMED® ClearPath glaucoma drainage device included design verification and functional product testing, sterilization validation, packaging and shelf life testing, and biocompatibility testing. Results of the nonclinical testing demonstrate that the AHMED® ClearPath glaucoma drainage device meets the defined specifications. #### Design Verification and Functional Product Testing: The AHMED® ClearPath glaucoma drainage device was evaluated to verify that the design output met the original design input and intent. This testing was based on tests described in ANSI Z80.27, Implantable Glaucoma Devices, which includes: pressure/flow characterization, structural integrity, and dimensional verification. The results establish that the AHMED® ClearPath glaucoma drainage device meets the design intent and complies with the applicable requirements. #### Device Performance Comparison to Predicate Device Baerveldt Glaucoma Implant The AHMED® ClearPath glaucoma drainage device was evaluated against the predicate Baerveldt Glaucoma Implant for pressure/flow characterization and effectiveness of tube occlusion utilizing a ripcord. The results establish that the AHMED® ClearPath glaucoma drainage device and Baerveldt Glaucoma Implant are equivalent. #### Sterilization Validation: The gamma irradiation sterilization method was validated using the VDmax 15 method described in BS EN ISO 11137-1:2015 and BS EN ISO 11137-2:2015. Validation results demonstrate that a minimum exposure dose of 15 kGy has been substantiated for the routine sterilization of the AHMED® ClearPath glaucoma drainage device to provide a 10th sterility assurance level (SAL). {7}------------------------------------------------ ## Packaging and Shelf Life Testing: The AHMED® ClearPath glaucoma drainage device is labeled with an expiration date of 2 (two) years. The shelf life study evaluated the functional performance of the AHMED® ClearPath glaucoma drainage device, as well as the packaging integrity. Additional testing was completed to evaluate the impact of environmental conditioning and distribution factors. Test results confirm that the AHMED® ClearPath glaucoma drainage device meet their functional requirements and the sterile barrier (package integrity) remains intact after simulated distribution and aging. This testing provides the justification for the 2-year shelf life and the maintenance of the sterile barrier. #### Biocompatibility Testing Biocompatibility of the final finished AHMED® ClearPath glaucoma drainage device was demonstrated through testing in accordance with BS EN ISO 10993-1 "Biological evaluation of medical devices. Part 1: Evaluation and testing within a risk management process" (refer to the table below for a listing of all tests performed). Testing was performed for the Implant, Ripcord, and Hypodermic Needle. The testing demonstrated that the device materials have an acceptable biocompatibility profile. | Test | Results | |----------------------------------------|-----------------------------------------------| | Cytotoxicity (MEM Elution) | Non-cytotoxic | | Guinea Pig Maximization Sensitization | Non-irritant compared to negative control | | Acute Systemic Toxicity (Mouse) | Non-toxic | | Intraocular Irritation (Rabbit) | No evidence of irritation | | Bacterial Mutagenicity – Ames Assay | Non-mutagenic | | 4 Week Intramuscular Implant (Rabbit) | Non-irritant compared to USP negative control | | 13 Week Intramuscular Implant (Rabbit) | Non-irritant compared to USP negative control | | Material mediated pyrogenecity | Non-pyrogenic | | Direct Contact Cytotoxicity | Non-cytotoxic | #### Listing of All Biocompatibility Tests Performed on All Patient Contacting Materials ## Clinical Evidence Supporting Substantial Equivalence Although the AHMED® ClearPath has not been clinically evaluated for performance, the following reference in the published clinical literature summarizes the performance of the predicate Baerveldt Glaucoma Implant. {8}------------------------------------------------ - Budenz, Donald L., et al. "Five-year treatment outcomes in the Ahmed Baerveldt comparison o study." Ophthalmology 122.2 (2015): 308-316. The following additional clinical literature is referenced in support of differences in design of the AHMED® ClearPath as compared to the predicate Baerveldt Glaucoma Implant. - Placement of Device: This study was conducted to determine optimal positioning of various glaucoma drainage devices posterior to the limbus while maintaining a safe distance relative to the optic nerve. - Kahook, Malik Y., et al. "Location of glaucoma drainage devices relative to the optic o nerve." British Journal of Ophthalmology 90.8 (2006): 1010-1013. - Implant Design Ridge: The following studies evaluated the safety and effectiveness of the AGV-FP7 ● reference device to the predicate Baerveldt Glaucoma Implant and indicate that a lack of a ridge does not significantly impact implant performance. - Budenz. Donald L., et al. "Five-vear treatment outcomes in the Ahmed Baerveldt comparison O study." Ophthalmology 122.2 (2015): 308-316. - Budenz. Donald L., et al. "Postoperative complications in the Ahmed Baerveldt comparison study o during five years of follow-up." American Journal of Ophthalmology 163 (2016): 75-82. - Use of Ripcord for Tube Occlusion: A ripcord for optional use is provided within the tube lumen of ● the Ahmed ClearPath device. Tube occlusion via intraluminal ripcord is a common technique to prevent early postoperative hypotony in non-valved glaucoma drainage devices. Below are examples of medical publications that discuss this technique, its effectiveness and its impact on implant performance. - Sherwood, Mark B., and M. Fran Smith. "Prevention of early hypotony associated with Molteno O implants by a new occluding stent technique." Ophthalmology 100.1 (1993): 85-90. - O An, Selena J., et al. "Scheduled postoperative ripcord removal in Baerveldt 350 implants corresponds with increased complications: a prospective, randomized trial." Investigative Ophthalmology & Visual Science 59.9 (2018): 2073-2073. ## VIII. SUBSTANTIAL EQUIVALENCE The AHMED® ClearPath glaucoma drainage device has the same intended use as the legally marketed predicate devices identified in this 510(k) submission and all other aqueous shunts regulated by FDA under 21 CFR § 886.3920. The provision of the 23-gauge hypodermic needle and the 2-inch polypropylene ripcord for use (optional) with the AHMED® ClearPath glaucoma drainage device during the implantation procedure does not represent a departure from the standard of care because their use is common with non-valved shunt designs including the Baerveldt the predicate device. Results of the nonclinical testing demonstrate that the AHMED® ClearPath glaucoma drainage {9}------------------------------------------------ device meets the defined specifications. In summary, the AHMED® ClearPath glaucoma drainage device is substantially equivalent to the Baerveldt glaucoma implant when used in the management of refractory glaucoma.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...