Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis OCT Blue Peak, Spectralis OCT with Multicolor

K172649 · Heidelberg Engineering GmbH · OBO · Nov 1, 2017 · Ophthalmic

Device Facts

Record IDK172649
Device NameSpectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis OCT Blue Peak, Spectralis OCT with Multicolor
ApplicantHeidelberg Engineering GmbH
Product CodeOBO · Ophthalmic
Decision DateNov 1, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 886.1570
Device ClassClass 2

Indications for Use

The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging, cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak) and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include reference databases for retinal nerve fiber layer thickness and optic nerve head neuroretinal rim parameter measurements, which are used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values found in normal subjects.

Device Story

Non-contact ophthalmic diagnostic imaging device; acquires infrared, blue reflectance, and composite MultiColor images; performs OCT cross-sectional, 2D, and 3D imaging; utilizes fluorescein/indocyanine green angiography and autofluorescence. Used in office settings by ophthalmologists. Device captures ocular anatomy/lesion data; processes via OCT2 module and Heyex 2 management system; outputs images and quantitative measurements (retinal nerve fiber layer thickness, optic nerve head parameters). Comparison against normal subject reference databases aids clinical assessment of disease progression and management. Modifications include OCT2 module, Wide Field Objective 2 (WFO2) with optimized anti-reflective coating, Heyex 2 software upgrade, and Windows 10 support.

Clinical Evidence

Bench testing only. Device evaluated against consensus standards including ISO 14971, AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 10993 series, IEC 60825-1, and AAMI/ANSI/IEC 62304. Design controls and risk management procedures per 21 CFR 820 and ISO 13485 were applied to verify modifications.

Technological Characteristics

Non-contact ophthalmic imaging; laser scanning ophthalmoscopy and OCT; OCT2 module; WFO2 objective lens (55° field of view) with optimized AR-coating for 486nm, 518nm, 815nm; Windows 10 platform; Heyex 2 database management. Standards: ISO 14971, AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 10993, IEC 60825-1, AAMI/ANSI/IEC 62304.

Indications for Use

Indicated for patients requiring ophthalmic diagnostic imaging of the posterior segment of the eye to aid in the detection and management of ocular diseases including age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, geographic atrophy, and neurodegenerative eye changes.

Regulatory Classification

Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" and the words "U.S. Food & Drug Administration". November 1, 2017 Heidelberg Engineering GmbH % Lena Sattler Consultant Orasi Consulting, LLC. 1655 Forest Dr. Medina, OH 44256 Re: K172649 Trade/Device Name: Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis OCT Blue Peak, Spectralis OCT with Multicolor Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO, MYC Dated: October 20, 2017 Received: October 23, 2017 Dear Lena Sattler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Denise L. Hampton -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172649 #### Device Name Spectralis HRA+OCT and variants #### Indications for Use (Describe) The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak) and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including agerelated macular degeneration, macular edema, diabetic retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include reference databases for retinal nerve fiber layer thickness and optic nerve head neuroretinal rim parameter measurements, which are used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values found in normal subjects. Type of Use (Select one or both, as applicable): | | Prescription Use (Part 21 CFR 801 Subpart D) | |--|----------------------------------------------| | | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the first line and to the right of the second line. ## 510(K) SUMMARY - K172649 Date Prepared October 29, 2017 #### SPONSOR/510(K) OWNER/ MANUFACTURER Heidelberg Engineering GmbH Max-Jarecki-Strasse 8 69115 Heidelberg Germany Telephone: Facsimile: Email: Establishment Registration No .: +49 6221 64630 +49 6221 / 64 63 62 Gerhard.Zinser@HeidelbergEngineering.com 8043762 ## OFFICIAL CONTACT PERSON Lena Sattler Orasi Consulting, LLC. 1655 Forest Dr. Medina, OH 44256 Telephone: Facsimile: E-mail: (440) 554-3706 (866) 904-4315 lena@orasiconsulting.com #### COMMON/USUAL NAME Optical Coherence Tomography #### PROPRIETARY OR TRADE NAMES #### SPECTRALIS HRA+OCT and variants #### CLASSIFICATION INFORMATION | Classification Name: | Ophthalmoscope, AC Powered<br>Ophthalmoscope, Laser, Scanning | |-----------------------|---------------------------------------------------------------| | Medical Specialty: | Ophthalmic | | Device Class: | II | | Classification Panel: | Ophthalmic Device Panel | | Product Codes: | OBO, MYC | Heidelberg Engineering GmbH Proprietary and Confidential Information 510(k) Summary: SPECTRALIS Page 1 of 7 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text, with a red square on either side. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". ## PRODUCT CODE: CLASSIFICATION / CFR TITLE OBO, MYC: Class II § 21 CFR 886.1570 ## LEGALLY MARKETED UNMODIFIED PREDICATE DEVICE | Trade/Device Name: | SPECTRALIS HRA+OCT and variants | |---------------------------------------|---------------------------------------------------------------| | Applicant: | Heidelberg Engineering GmbH | | 510(k) Premarket Notification number: | K152205 | | Classification: | Class II | | CFR Title: | 21 CFR 886.1570 | | FDA Product Code(s): | OBO, MYC | | Classification Name: | Ophthalmoscope, AC Powered<br>Ophthalmoscope, Laser, Scanning | | Common Name: | Optical Coherence Tomography | | Medical Specialty: | Ophthalmic | | Classification Panel: | Ophthalmic Device Panel | The device name and classification information for the modified device is identical to the cleared device. ## GENERAL DEVICE DESCRIPTION The Heidelberg Engineering SPECTRALIS HRA+OCT is a device used to image the anterior and posterior segments of the human eye and to aid in the assessment and management of various diseases of the posterior segment, such as age-related macular degeneration, diabetic retinopathy, and glaucoma. The device is capable of acquiring infrared and blue reflectance images (K101223), as well as a composite color image referred to as "MultiColor image" due to addition of a green laser light source (K121993). New reference data for peripapillary retinal nerve fiber layer thickness and optic nerve head parameters have been added in K152205, as well as two (2) new accessory objective lenses (Wide Field, Ultra-Widefield), and OCT Enhanced Depth Imaging mode. The modifications to the SPECTRALIS are: - 1. OCT2 Module (OCT2) replacement of the Line Scan Camera and PC interface. OCT2 was developed due to the discontinuation of components. - 2. Wide Field Objective 2 (WFO2) The WFO provides a 55° diameter circular field of view. The field of view can be reduced by the software to 35° and 25°. The AR-coating of the WFO was optimized in order to reduce the remaining {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The top line reads "HEIDELBERG" and the bottom line reads "ENGINEERING". There is a red square on either side of the text. reflections at the MultiColor wavelengths 486 nm. 518 nm and 815 nm. The AR-coating also allows for MultiColor imaging. - 3. An upgrade to the data and patient management system (Heyex 2). Heyex 2 has the same basic functionality as the data and patient management system of the cleared SPECTRALIS, Heyex 1. The main improvements of Heyex 2 compared to Heyex 1 are an up-to-date and more scalable underlying image database, and a modern User Interface. - 4. Windows 10 Operating System support. - 5. Labeling The user manual has been updated to (1) improve clarity, (2) improve grammar and (3) include the modifications made to the SPECTRALIS HRA+OCT and variants. The Intended Use/Indications for Use of the modified device, as described in the labeling has not changed as a result of the modifications. ## INDICATIONS FOR USE The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and threedimensional imaging, cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak) and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include reference databases for retinal nerve fiber layer thickness and optic nerve head neuroretinal rim parameter measurements, which are used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values found in normal subjects. #### SUBSTANTIAL EQUIVALENCE The modified SPECTRALIS HRA+OCT and variants is a device modification to the cleared SPECTRALIS HRA+OCT and variants (K152205) predicate device. The modified SPECTRALIS has the same Intended Use/Indications for Use and the same fundamental scientific technology as the predicate device. Heidelberg Engineering GmbH Proprietary and Confidential Information {6}------------------------------------------------ The Substantial Equivalence Summary tables below illustrate the comparisons of the modified SPECTRALIS to the predicate device. The modified SPECTRALIS HRA+OCT and variants measures the same ophthalmic features and parameters as the cleared SPECTRALIS HRA+OCT and variants in K152205. The changes applied to the SPECTRALIS since the clearance in K152205 do not change the subject population for whom the device is used, the type of images to be acquired, or the usefulness of the SPECTRALIS as a quantitative aid to an ophthalmologic examination of the eye. Technological detail characteristics of the device are unchanged except for the modifications as stated in the General Device Description. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the first line and to the right of the second line. ## INTENDED USE/INDICATIONS FOR USE STATEMENT CHART | K152205<br>PREDICATE DEVICE | SUBJECT DEVICE | Same or<br>Different | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | The SPECTRALIS is a non-contact<br>ophthalmic diagnostic imaging device. It is<br>intended for viewing the posterior segment<br>of the eye, including two- and three-<br>dimensional imaging, cross-sectional<br>imaging (SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT), fundus photography,<br>fluorescence imaging (fluorescein<br>angiography, indocyanine green<br>angiography; SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA), autofluorescence<br>imaging (SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA and SPECTRALIS<br>OCT with BluePeak) and to perform<br>measurements of ocular anatomy and ocular<br>lesions. The device is indicated as an aid in<br>the detection and management of various<br>ocular diseases, including age-related<br>macular degeneration, macular edema,<br>diabetic retinopathy, retinal and choroidal<br>vascular diseases, glaucoma, and for<br>viewing geographic atrophy as well as<br>changes in the eye that result from<br>neurodegenerative diseases. The<br>SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT include reference<br>databases for retinal nerve fiber layer<br>thickness and optic nerve head neuroretinal<br>rim parameter measurements, which are<br>used to quantitatively compare the retinal<br>nerve fiber layer and neuroretinal rim in the<br>human retina to values found in normal<br>subjects. | The SPECTRALIS is a non-contact<br>ophthalmic diagnostic imaging device. It is<br>intended for viewing the posterior segment<br>of the eye, including two- and three-<br>dimensional imaging, cross-sectional<br>imaging (SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT), fundus photography,<br>fluorescence imaging (fluorescein<br>angiography, indocyanine green<br>angiography; SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA), autofluorescence<br>imaging (SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA and SPECTRALIS<br>OCT with BluePeak) and to perform<br>measurements of ocular anatomy and ocular<br>lesions. The device is indicated as an aid in<br>the detection and management of various<br>ocular diseases, including age-related<br>macular degeneration, macular edema,<br>diabetic retinopathy, retinal and choroidal<br>vascular diseases, glaucoma, and for<br>viewing geographic atrophy as well as<br>changes in the eye that result from<br>neurodegenerative diseases. The<br>SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT include reference<br>databases for retinal nerve fiber layer<br>thickness and optic nerve head neuroretinal<br>rim parameter measurements, which are<br>used to quantitatively compare the retinal<br>nerve fiber layer and neuroretinal rim in the<br>human retina to values found in normal<br>subjects. | Same | | | | PREDICATE DEVICE<br>K152205<br>SPECTRALIS<br>HRA+OCT | SUBJECT DEVICE | Same or Different | | User | Ophthalmologist, office<br>setting | Ophthalmologist, office<br>setting | Same | | Eye contact<br>required? | No | No | Same | | Dilation of<br>pupil<br>required? | No | No | Same | | Working<br>position | The patient is sitting<br>straight in front of the<br>device. The examiner is<br>sitting opposite the patient. | The patient is sitting<br>straight in front of the<br>device. The examiner is<br>sitting opposite the patient. | Same | | Images of<br>posterior<br>segment of<br>eye | Yes, using standard, WFO,<br>or UWF objective lens | Yes, using standard, WFO<br>/ WFO2, or UWF<br>objective lens | Same; WFO2 was<br>introduced, which has an<br>enhanced anti-reflective<br>coating, but offers the same<br>field of view and has the<br>same optical design. | | Images of<br>anterior<br>segment of<br>eye | With Anterior Segment<br>Module (K113129) | With Anterior Segment<br>Module (K113129) | Same | | RNFL<br>Reference<br>Database | RNFL thickness in<br>reference database for 12°<br>circle scans (K101223)<br>RNFL thickness in<br>reference database for<br>fixed diameter circle scans<br>(K152205) | RNFL thickness in<br>reference database for 12°<br>circle scans (K101223)<br>RNFL thickness in<br>reference database for<br>fixed diameter circle scans<br>(K152205) | Same | | ONH<br>Reference<br>Database | ONH parameters (BMO-<br>MRW, BMO-MRA)<br>reference database<br>(K152205) | ONH parameters (BMO-<br>MRW, BMO-MRA)<br>reference database<br>(K152205) | Same | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the word "HEIDELBERG" and to the right of the word "ENGINEERING". ## INTENDED USE CHARACTERISTICS CHART {9}------------------------------------------------ ## NON-CLINICAL PERFORMANCE SUMMARY The modified SPECTRALIS was evaluated according to the requirements of FDA recognized consensus standards ISO 14971, AAMI / ANSI ES60601-1:2005, IEC 60601-1-2 Edition 3: 2007-03, ISO 10993-1 Fourth Edition 2009-10-15, AAMI/ANSI/ISO 10993-5:2009/(R) 2014, ISO 10993-12 Fourth Edition 2012-07-01, ISO 10993-18:2005, IEC 60825-1 Edition 2.0 2007-03, and AAMI / ANSI / IEC 62304:2006 and was found to meet the requirements of the applicable parts. #### DESIGN CONTROL Heidelberg Engineering designed and developed the modified SPECTRALIS per the company's Design Control procedure, which complies with the FDA Quality System Regulations CFR Part 820 and ISO 13485:2012. The Design Control procedure also incorporates Risk Management procedures, which comply with ISO 14971:2007. Risk assessment was conducted on the modified SPECTRALIS, and the impact of the design modifications were assessed on the predicate 510(k) cleared device. The modified SPECTRALIS is manufactured and tested in the exact manner as the predicate 510(k) cleared device. #### CONCLUSION The modified SPECTRALIS is a device modification to the cleared SPECTRALIS predicate device. The modified SPECTRALIS has the same Intended Use/Indications for Use and the same fundamental scientific technology as the cleared SPECTRALIS predicate device. In summary, Heidelberg Engineering GmbH is of the opinion that the modified SPECTRALIS HRA+OCT and variants, does not introduce any new potential safety risks, is as effective as the cleared SPECTRALIS HRA+OCT and variants and concludes that the modified SPECTRALIS HRA+OCT and variants, is substantially equivalent to the predicate device. This 510(k) summary for the Heidelberg Engineering SPECTRALIS HRA+OCT and variants is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR§807.92.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...