MAIA, MODEL 1

K092187 · Centervue S.P.A. · HPT · May 27, 2010 · Ophthalmic

Device Facts

Record IDK092187
Device NameMAIA, MODEL 1
ApplicantCentervue S.P.A.
Product CodeHPT · Ophthalmic
Decision DateMay 27, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1605
Device ClassClass 1

Intended Use

The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference database that is a quantitative tool for the comparison of macular sensitivity to a database of known normal subjects.

Device Story

MAIA™ is a non-mydriatic, confocal, line-scanning, infrared ophthalmoscope combined with an automated perimeter for microperimetry. It captures 36° x 36° infrared retinal images; tracks eye movements to quantify fixation characteristics; and measures differential light sensitivity (threshold sensitivity) at macular locations via patient response to projected light stimuli. The device integrates a computer, touch-screen, and custom Linux OS. Operated by clinicians in a clinical setting, it compares patient threshold data against a normative reference database. Output includes retinal images, fixation analysis, and sensitivity maps, aiding clinicians in diagnosing and monitoring retinal pathologies like age-related macular degeneration and diabetic retinopathy. Benefits include objective, quantitative assessment of macular function and fixation stability.

Clinical Evidence

Clinical evaluation included a normative database study (n=494 eyes, 270 subjects, age 21-86) and a precision study. Precision study involved 24 subjects (12 normal, 12 with retinal pathology) tested across 3 devices and 3 operators. Results for normal eyes: mean 29.7 dB, repeatability SD 0.42 dB, reproducibility SD 0.96 dB. Results for pathology eyes: mean 23.5 dB, repeatability SD 0.75 dB, reproducibility SD 0.75 dB. Data supports the device's ability to measure macular sensitivity and fixation stability.

Technological Characteristics

Confocal, line-scanning, infrared ophthalmoscope; automated perimeter; non-mydriatic. Features include infrared illumination, visible light projection for perimetry, touch-screen interface, and integrated computer. Software runs on custom Linux OS. Connectivity includes power cord and push-button input. Dimensions/materials not specified. Class II device.

Indications for Use

Indicated for patients requiring assessment of macular sensitivity, fixation stability, and locus of fixation, and infrared retinal imaging. Applicable to patients with retinal pathologies, including age-related macular degeneration and diabetic retinopathy, as well as normal subjects.

Regulatory Classification

Identification

A perimeter is an AC-powered or manual device intended to determine the extent of the peripheral visual field of a patient. The device projects light on various points of a curved surface, and the patient indicates whether he or she sees the light.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white graphic that appears to be an abstract representation of a circular or spiral structure. The central part of the image features a solid black circle, surrounded by a white ring. The outer area of the image is filled with a pattern of irregular black and white shapes, which could be interpreted as buildings or other architectural elements. The overall impression is that of a stylized map or diagram. mmary of safety and effectiveness Image /page/0/Picture/3 description: The image contains the word "centervue" in a bold, sans-serif font. The letters are closely spaced together, creating a compact and unified appearance. The word is written in lowercase. **SUBMITTER INFORMATION** | A. Company name: | CenterVue SpA | |---------------------------|---------------------------------------| | B. Company address: | via Tommaseo 77, 35131, Padova, ITALY | | C. Company phone: | +39 049 781 1828 | | Company fax: | +39 049 781 1899 | | D. Contact person: | Giuliano Barbaro<br>Managing Director | | E. Date summary prepared: | 4 April 2010 | #### DEVICE IDENTIFICATION - Ophthalmoscope, AC-powered; Automated Perimeter, AC-powered Generic Device Name: A. - MAIA™, Macular Integrity Assessment Trade/proprietary name: в. - C. Classification: Class II - HLI, HPT Product Code: D. ### DEVICE DESCRIPTION The MAIA™ is a confocal, line scanning, infrared, ophthalmoscope, combined with a system for visible light projection to obtain perimetric measurements, using "fundus perimetry" (also "microperimetry"). MAIA™ integrates in one device an automated perimeter and an ophthalmoscope, providing: - images of the central retina over a field of view of 36° x 36°, acquired under infrared . illumination; - recordings of eye movements obtained by "tracking" retinal details in the live retinal images and . providing a quantitative analysis of fixation characteristics; - measurements of differential light sensitivity (or threshold sensitivity) at multiple locations in . the macula, obtained by recording a patient's subjective response (see / don't see) to a light stimulus projected at a certain location over the retina; - comparison of measured threshold sensitivity with a reference database obtained from normal . subjects, indicating whether measured thresholds are above or below certain percentiles. MAIA 100 works with no pupil dilation (non-mydriatic). MAIA™ integrates a computer for control and data processing and a touch-screen display and it is provided with a power cord and a push-button. MAIA™ works with a dedicated software application running on a custom Linux O.S. #### INDICATIONS FOR USE The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference CenterVue SpA ia Tommaseo. 7 35131 Padova, Italy www.centervue.com 0 +30 Mg 781 1829 fax +39 049 781 1899 e-mail: info@centervue.com 378243 CCIAA PD oc .: € 480.000,00 i.v. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image is a black and white graphic design. It features a circular shape in the center, surrounded by a more complex, abstract pattern. The pattern consists of various geometric shapes and lines, creating a visually intricate design around the central circle. centervue t the center of vision that is a quantitative tool for the comparison of macular sensitivity to a database of known JIVALENCE device is substantially equivalent to the combination of the following predicate devices with regard to intended use, operating principle, function: | Predicate Device | 510(k) Holder | 510(k) No. | Date cleared | |---------------------|-------------------------|------------|---------------------| | MP-1 MICROPERIMETER | Nidek Technologies Srl | K023719 | December 23rd 2002 | | MP-1 MICROPERIMETER | Nidek Technologies Srl | K061768 | September 28th 2006 | | CIRRUS HD-OCT(*) | Carl Zeiss Meditec Inc. | K063378 | January 26th 2007 | (*) only with regard to its retinal imaging sub-system #### CLINICAL EVALUATION Clinical data was collected and evaluated to support the indications for use statement for MAIA™ and to demonstrate substantial equivalence to the above predicate devices. In particular, the reference database was developed by obtaining threshold sensitivity data from 494 eyes of 270 normal subjects, enrolled at 4 different clinical sites. The age range of the measured population was 21-86 (mean 43, std. dev. 15). Subjects were recruited among the clinics' personnel and among the relatives of the clinics' regular patients. The following MAIA perimetric grid and other device settings were used for the measurements: - Number of stimuli: 61; . - Macular coverage: 10°; . - Size of stimulus: Goldmann III; . - Duration of stimulus: 200 ms; ● - . Background luminance: 4 asb; - Maximum luminance: 1000 asb; . - Threshold algorithm: standard 4-2 staircase; ● - . Measurement scale: 0 to 36 dB; - . Fixation target: central circle of 1° diameter. #### PRECISION A precision study was conducted which measured both individual location thresholds (n = 37 per eye) and average threshold over all locations. Three devices were used, each operated by a different operator. The study included 24 subjects, each tested on one eye only: 12 without pathology and 12 with bathology. Each subject/eye was tested 3 times within a session (3 repeated measures for each subject). Subjects were nested within each configuration, that is, each subject appears in one, unique device-operator configuration only. The design was repeated twice: once for healthy eyes and once for eyes with pathologies. CenterVue SpA ia Tommasoo, 77 35131 Padova, Italy www.centervue.com tel +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com A 478243 COLA PO CF /P VA 04296580287 Cap. soc .: € 480.000.00 i.v. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a circular design with a black circle in the center. Surrounding the central circle is a pattern of black and white shapes that resemble buildings or structures. The overall design has a somewhat abstract and urban feel, with the contrasting colors creating a visually interesting composition. t the center of vision #### ubiect selection criteria Twelve subjects with normal eyes and 12 subjects with retinal pathologies were enrolled at two different clinical sites based on the following criteria: - 4 normal subjects and 4 subjects with retinal pathologies enrolled at site 1 and measured with device 1 by operator 1; - 4 normal subjects (different from above) and 4 subjects with retinal pathologies (different from above)enrolled at site 1 and measured with device 2 by operator 2; - 4 normal subjects and 4 subjects with retinal pathologies enrolled at site 2 and measured with device 3 by operator 3; Within each pathology category, subjects with varying parameters including age and visual acuity were enrolled. For the pathology group, subjects with early and intermediate age-related macular degeneration as well as with mild, moderate, severe and proliferative diabetic retinopathy were enrolled. Diagnosis of retinal pathology was made by a complete eye examination by an ophthalmologist, including dilated funduscopic examination and pertinent history. Subjects met defined inclusion criteria, but none of the exclusion criteria. #### Precision results | | Normal | Pathology | |----------------------------|---------|-----------| | # subjects | 12 | 12 | | Overall Mean | 29.7 dB | 23.5 dB | | Overall Standard deviation | 1.14 dB | 4.23 dB | | Repeatability SD* | 0.42 dB | 0.75 dB | | Reproducibility SD** | 0.96 dB | 0.75 dB | * estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning. ** estimate of the standard deviation among measurements taken on the same subject using different operators and devices, including repeatability. #### Individual Grid Point Results | Group/Parameter | Repeatability SD | Reproducibility SD | | |-----------------|------------------|--------------------|------| | Normal | Minimum | 0.94 | 1.06 | | | Median | 1.40 | 1.80 | | | Maximum | 2.43 | 2.70 | | Pathology | Minimum | 1.33 | 1.33 | | | Median | 2.36 | 2.43 | | | Maximum | 3.16 | 3.24 | #### CONCLUSION As described in this 510(k) Summary, all testing deemed necessary was conducted on the MAIA™ to ensure that the device is safe and effective for its intended use when used in accordance with its Instructions for Use. CenterVue SpA via Tommaseo. 77 35131 Padova, Italy www.centervue.com tel. +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com REA: 378243 CCIAA PD CFJP, NA: 04296580287 Cap. soc.: € 480.000,00 i.v. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 CenterVue SpA c/o Mr. Giuliano Barbaro Managing Director Via Tommaseo 77 Padova, Italy 35131 MAY -2 7 2010 Re: K092187 Trade/Device Name: Macular Integrity Assessment (MAIA ") Device Regulation Number: 21 CFR 886.1605 Regulation Name: Perimeter, Automatic, AC-powered Regulatory Class: Class II Product Code: HPT, HLI Dated: April 22, 2010 Received: April 26, 2010 Dear Mr. Barbaro: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Alexander for Ma Dir Div Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image is a black and white graphic that resembles an eye. The pupil is a solid black circle in the center. The iris is a white ring around the pupil, and the rest of the eye is made up of a pattern of black and white shapes that create a spiral effect. centervue at the center of vision # Indications for use statement 510(k) Number: K092187 Device Name: Macular Integrity Assessment (MAIATM) ## Indications for use: The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference database that is a quantitative tool for the comparison of macular sensitivity to a database of known normal subjects. Prescription Use M AND/OR Over-The-Counter Use Callaway (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K092187 via Tommaseo, 77 35131 Padova, Italy www.centervue.com tel. +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com REA: 378243 CCIAA PD C.F./P. IVA: 04296580287 Cap. soc.; € 480,000,00 i.v.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...