K142837 · Heine Optotechnik GmbH & Co. KG · HLJ · Jan 29, 2015 · Ophthalmic
Device Facts
Record ID
K142837
Device Name
HEINE SIGMA 250 & HEINE SIGMA 250 M2
Applicant
Heine Optotechnik GmbH & Co. KG
Product Code
HLJ · Ophthalmic
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 886.1570
Device Class
Class 2
Intended Use
The HEINE SIGMA® 250 (M2) are battery powered indirect ophthalmoscopes for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
Device Story
Battery-powered indirect ophthalmoscope; worn on user's head (headband or spectacle-mounted). Provides illumination and viewing optics for examination of ocular media and retina. Operated by medical professionals in clinical settings. Device uses white LED illumination; brightness controlled via dial. Output allows visualization of patient's fundus; aids clinical diagnosis of ocular conditions. No data collection or display system.
Clinical Evidence
Bench testing only. Testing performed per Ophthalmoscope Guidance regarding optical radiation hazard (ISO 10943) and fundamental requirements for ophthalmic instruments (ISO 15004-2).
Technological Characteristics
Battery-powered binocular indirect ophthalmoscope; headband or spectacle-mounted. Illumination: 3V white LED. Viewing optics: +2 diopter. Materials: self-extinguishing (UL94 HB or V-0). Safety: Complies with IEC 60601-1, ISO 10943, and ISO 15004-2. No data collection/display. Manual brightness control.
Indications for Use
Indicated for medical professionals to examine the ocular media (cornea, aqueous, lens, vitreous) and retina of the eye.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
K123587 — HEINE MINI 3000 LED OPHTHALMOSCOPE · Heine Optotechnik GmbH & Co. KG · Mar 22, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the figure.
Public Health Service
January 29, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
HEINE Optotechnik GmbH & Co. KG Mrs. Bettina Seim Director, Regulatory Affairs Kientalstr. 7 Herrsching, Germany 82211
Re: K142837
Trade/Device Name: HEINE SIGMA® 250 and 250 M2 Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLJ Dated: December 17, 2014 Received: December 23, 2014
Dear Mrs. Seim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) number (if known): Device Name:
HEINE SIGMA ® 250 HEINE SIGMA ® 250 M2
Indications For Use:
The HEINE SIGMA® 250 (M2) are battery powered indirect ophthalmoscopes for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
xl Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for Heine Optotechnik. The logo is an orange triangle with the word "HEINE" written in white, stylized to look like a snake wrapped around a lightning bolt. The snake and lightning bolt are also white. The logo is simple and recognizable.
## 510(k) Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act, HEINE Optotechnik GmbH & Co. KG herewith submits a Summary of Safety and Effectiveness.
| Submitter Information: | HEINE Optotechnik GmbH & Co. KG<br>Kientalstr. 7<br>82211 Herrsching<br>Germany<br>Registration Number: 1000379039<br>Owner/Operator Number: 9003020 |
|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact Person: | Mrs. Bettina Seim<br>Director Regulatory Affairs<br>HEINE Optotechnik GmbH & Co. KG<br>Phone: +49 8152 38 0<br>E-mail: <a href="mailto:bseim@heine.com">bseim@heine.com</a> |
| US Agent (Contact): | Belinda Labourdette<br>HEINE USA, Itd.<br>10 Innovation Way<br>Dover, NH 03820 USA<br>Phone: +1 603 842-6605<br>E-mail: <a href="mailto:blab@heine-na.com">blab@heine-na.com</a> |
| Date Prepared: | September 22, 2014 |
| Device(s) Identification:<br>Device Trade Name: | HEINE SIGMA® 250 (M2) |
| Common Name: | (indirect) Ophthalmoscope |
| Classification of the device:<br>Device Classification Name:<br>Product Code:<br>Device Classification No.: | Ophthalmoscope<br>HLJ<br>Part 886.1570 |
{4}------------------------------------------------
## HEINE
Image /page/4/Picture/1 description: The image is a logo for Heine Optotechnik, a German manufacturer of medical diagnostic instruments. The logo features an orange triangle with the word "HEINE" written in white, stylized letters. A white caduceus symbol, commonly associated with medicine, is overlaid on the text. The caduceus has a lightning bolt running through it.
Panel: Requlatory Status:
Ophthalmic Devices (86) Class II
#### Device Description:
The HEINE SIGMA® 250 (M2) are indirect ophthalmoscopes, worn on the user's head to provide illumination and viewing optics in order to examine the media and the retina of a patient`s eye. The indirect ophthalmoscopes are battery operated.
#### Intended Use:
The HEINE SIGMA® 250 (M2) are battery powered indirect ophthalmoscopes for medical professionals, containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, vitreous) and the retina of the eye.
#### Predicate Device:
| Device Trade Name: | HEINE OMEGA® 500 |
|--------------------|---------------------------------|
| Applicant: | HEINE Optotechnik GmbH & Co. KG |
| 510(k) No.: | K123316 |
The indirect ophthalmoscopes HEINE SIGMA® 250 (M2) are considered substantial equivalent to the HEINE OMEGA® 500 Ophthalmoscope (K123316).
There is no significant difference in intended use or technology.
| | SIGMA 250 | SIGMA 250 M2 | HEINE OMEGA® 500 | Assessment |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Intended Use | The HEINE SIGMA®<br>250 (M2) is a battery<br>powered indirect<br>ophthalmoscope for<br>medical professionals,<br>containing illumination<br>and viewing optics<br>intended to examine<br>the media (cornea,<br>aqueous, lens,<br>vitreous) and the<br>retina of the eye. | The HEINE SIGMA®<br>250 (M2) is a battery<br>powered indirect<br>ophthalmoscope for<br>medical professionals,<br>containing illumination<br>and viewing optics<br>intended to examine<br>the media (cornea,<br>aqueous, lens,<br>vitreous) and the<br>retina of the eye. | The indirect<br>ophthalmoscope HEINE<br>OMEGA® 500 is an AC-<br>powered or battery<br>powered device for<br>medical professionals,<br>containing illumination<br>and viewing optics<br>intended to examine the<br>media (cornea,<br>aqueous, lens, vitreous)<br>and the retina of the<br>eye. | Equivalent |
| Type | Binocular (Headband<br>and Spectacles<br>mounted) | Binocular (Headband<br>and Spectacles<br>mounted) | Binocular (Headband<br>mounted) | Similar |
| | SIGMA 250 | SIGMA 250 M2 | HEINE OMEGA® 500 | Assessment |
| Method of<br>operation | Used to examine the<br>retina by an examiner<br>in a specified distance<br>to the eye | Used to examine the<br>retina by an examiner<br>in a specified distance<br>to the eye | Used to examine the<br>retina by an examiner in<br>a specified distance to<br>the eye | Equivalent |
| Illumination | White LED, 3V | White LED, 3V | White 6V, 5 Watt<br>LED,<br>Xenon<br>Halogen bulb | Equivalent |
| Selectable<br>filter | Red-free | Red-free | Blue, yellow, red-free,<br>diffuser | Similar |
| Safety filter IR<br>blocker | LED light source, no<br>IR radiation | LED light source, no<br>IR radiation | Permanent | Similar |
| Light Output¹ | 444lx (@ 210mA)<br>615lx (@ 350mA) | 444lx (@ 210mA)<br>615lx (@ 350mA) | 507 lx 258 lx (max.)<br>(max.) | Similar |
| Light<br>apertures¹ | Small circle:<br>n/a | Small circle:<br>22,5mm | Small circle:<br>18 mm | Similar |
| | Middle circle: 35mm | Middle circle:<br>n/a | Middle circle:<br>39 mm | Similar |
| | Large circle: 80mm | Large circle:<br>83 mm | Large circle: 74 mm | Similar |
| Inter pupillary<br>distance<br>adjustment | 47mm - 72mm | 47mm - 72mm | 46 - 74 mm | Similar |
| Data collection<br>and/or display<br>system | None | None | None | Equivalent |
| Lens power<br>viewing optics | +2 diopter | +2 diopter | +2 diopter | Equivalent |
| Power sources | n/a | n/a | Wireless battery pack | Not existing |
| | n/a | n/a | Wall mounted unit | Not existing |
| | Belt battery pack | Belt battery pack | Belt battery pack | Equivalent |
| Brightness<br>controls | Control dial | Control dial | Control dial | Equivalent |
| Maximum<br>temperature of<br>parts of the<br>device held by<br>the operator or<br>accessible to<br>the patient | Complies with IEC<br>60601-1 | Complies with IEC<br>60601-1 | Complies with IEC<br>60601-1 | Equivalent |
| | SIGMA 250 | SIGMA 250 M2 | HEINE OMEGA® 500 | Assessment |
| Flammability<br>of materials | Low probability. All<br>measures have been<br>taken to use self-<br>extinguishing<br>materials which are<br>either of flame<br>classification HB or V-<br>O (UL94). The system<br>is illuminated using a<br>LED and all materials<br>used in the vicinity are<br>specially designed to<br>safely operate in high<br>temperature<br>environments. | Low probability. All<br>measures have been<br>taken to use self-<br>extinguishing<br>materials which are<br>either of flame<br>classification HB or V-<br>O (UL94). The system<br>is illuminated using a<br>LED and all materials<br>used in the vicinity are<br>specially designed to<br>safely operate in high<br>temperature<br>environments. | Low probability. All<br>measures have been<br>taken to use self-<br>extinguishing materials<br>which are either of<br>flame classification HB<br>or V-0 (UL94). The<br>system is illuminated<br>using a LED or 5W XHL<br>Xenon Halogen lamp<br>and all materials used in<br>the vicinity are specially<br>designed to safely<br>operate in high<br>temperature<br>environments. | Equivalent |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Heine Optotechnik. The logo is an orange triangle with the word "HEINE" written in white, stylized letters. A white medical symbol, resembling a snake wrapped around a lightning bolt, is placed in front of the text.
# HEINE
{6}------------------------------------------------
# HEINE
Image /page/6/Picture/1 description: The image shows the logo for Heine Optotechnik, a German manufacturer of medical diagnostic instruments. The logo features an orange triangle with the word "HEINE" written in a stylized, cursive font. A symbol resembling a snake wrapped around a lightning bolt is integrated into the text, adding a medical and technological element to the design. The logo is simple, recognizable, and reflects the company's focus on medical technology.
Note 1:
The measurements are taken from 500 mm in front of the instrument.
#### Summary of Non-Clinical Performance Testing:
The HEINE SIGMA® 250 (M2) indirect ophthalmoscopes were tested according to the "Ophthalmoscope Guidance" in respect to optical radiation hazard with ophthalmoscopes (ISO 10943). Additionally testing in accordance with applicable requirements of ISO 15004-2 "Ophthalmic instruments - Fundamental requirements and test methods" has been performed.
#### Conclusion:
HEINE Optotechnik believes that the HEINE SIGMA® 250 (M2) indirect ophthalmoscopes are substantially equivalent to the currently legally marketed devices. They do not introduce new indications for use, have the same technological characteristics, and do not introduce new potential hazards or safety risks.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.