← Product Code [OBO](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/OBO) · K161509

# 3D OCT-1 Maestro (K161509)

_Topcon Corporation · OBO · Jul 28, 2016 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/OBO/K161509

## Device Facts

- **Applicant:** Topcon Corporation
- **Product Code:** [OBO](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/OBO.md)
- **Decision Date:** Jul 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1570
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions. The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures. It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

## Device Story

Non-contact, high-resolution tomographic and biomicroscopic imaging system; merges OCT and fundus camera. Captures OCT images and color fundus images sequentially; supports anterior and posterior segment imaging. Data transferred via LAN to off-the-shelf PC; processed by 'FastMap' or 'OCT Viewer' software. Automated functions include retinal layer segmentation, thickness calculation, optic disc analysis, and comparison against a normative reference database. Used in clinical settings by eye care professionals. Output viewed on PC monitor; aids clinicians in diagnosis, documentation, and management of ocular diseases. Benefits include non-invasive, high-resolution visualization and quantitative assessment of ocular structures.

## Clinical Evidence

Clinical studies evaluated agreement, repeatability, and reproducibility of measurements (retinal thickness, RNFL, ganglion cell complex, optic disc) compared to iVue and TRC-NW300. Included 25 normal subjects, 25 with retinal disease, and 25 with glaucoma. Manual segmentation adjustment rates reported (0-40% depending on scan type). Reference database validated with 399 normal subjects (ages 18-88).

## Technological Characteristics

Spectral domain OCT (SD-OCT) using 840 nm superluminescent diode (SLD). Fundus camera uses Xenon flash and CMOS sensor. Connectivity via LAN to external PC. Software-based analysis. Complies with ISO 15004-1, ISO 10940, ISO 15004-2, AAMI ANSI ES 60601-1, and IEC 60601-1-2.

## Regulatory Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Optovue's iVue with Normative Database ([K121739](/device/K121739.md))
- Topcon's TRC-NW300 Non-Mydriatic Retinal Camera ([K123460](/device/K123460.md))

## Reference Devices

- Topcon's 3D OCT-2000 ([K092470](/device/K092470.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble a ribbon or banner. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2016

Topcon Corporation % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K161509

Trade/Device Name: 3D OCT-1 Maestro Regulation Number: 21 CFR §886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO, HKI Dated: June 1, 2016 Received: June 1, 2016

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K161509

Device Name

3D OCT-1 Maestro

Indications for Use (Describe)

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions.

The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures.

It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

{3}------------------------------------------------

# 510(k) SUMMARY – K161509

# Topcon Corporation's 3D OCT-1 Maestro

# Submitter Information:

| Applicant:           | Topcon Corporation<br>75-1 Hasunuma-cho, Itabashi-ku<br>Tokyo, Japan 174-8580                                                             |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Phone:               | (201) 599-5533                                                                                                                            |
| Fax:                 | (201) 599-5248                                                                                                                            |
| Contact Person:      | Charles Reisman<br>Senior Manager, Global R&D<br>Topcon Medical Systems, Inc.<br>111 Bauer Drive<br>Oakland, NJ 07436<br>1 (201) 599-5283 |
| Date Prepared:       | July 25, 2016                                                                                                                             |
| Subject Device:      |                                                                                                                                           |
| Device Name:         | 3D OCT-1 Maestro                                                                                                                          |
| Common Name:         | Optical Coherence Tomography                                                                                                              |
| Classification Name: | Ophthalmoscope                                                                                                                            |
| Regulation:          | Class II, 21 C.F.R. § 886.1570                                                                                                            |
| Product Code(s):     | OBO, HKI                                                                                                                                  |
| Predicate Device(s): |                                                                                                                                           |

#### Primary Predicate: Optovue's iVue with Normative Database (K121739) Topcon's TRC-NW300 Non-Mydriatic Retinal Camera (K123460) Secondary Predicate:

{4}------------------------------------------------

#### Indications for Use:

The Topcon 3D OCT-1 Maestro is a non-contact, high resolution tomographic and biomicroscopic imaging device that incorporates a digital camera for photographing, displaying and storing the data of the retina and surrounding parts of the eye to be examined under Mydriatic and non-Mydriatic conditions.

The 3D OCT-1 Maestro is indicated for in vivo viewing, axial cross sectional, and three-dimensional imaging and measurement of posterior ocular structures, including retinal nerve fiber layer, macula and optic disc as well as imaging of anterior ocular structures.

It also includes a Reference Database for posterior ocular measurements which provide for the quantitative comparison of retinal nerve fiber layer, optic nerve head, and the human retina to a database of known normal subjects. The 3D OCT-1 Maestro is indicated for use as a diagnostic device to aid in the diagnosis, documentation and management of ocular health and diseases in the adult population.

## Device Description:

The Maestro is a non-contact, high-resolution, tomographic and bio-microscopic imaging system that merges OCT and fundus cameras into a single device. The technological characteristics of the OCT employed are similar to those of already 510(k)-cleared OCT products, such as Topcon's 3D OCT-2000 (K092470), in that it employs conventional spectral domain OCT with widely-used 840 nm light source. The technological characteristics of the fundus camera employed are also similar to those of already cleared fundus cameras, such as Topcon's TRC NW300 (K123460), in terms of field of view (FOV) and camera sensor resolution.

The Maestro captures an OCT image and a color fundus image sequentially. It can take anterior OCT images in addition to fundus OCT images. It also includes a reference database for fundus OCT. Captured images are transferred from the device to an off-the-shelf personal computer (PC) via LAN cable, where the dedicated software for this device is installed. The transferred data is then automatically processed with analysis functions such as the automatic retinal layers segmentation, the automatic thickness calculation with several grids, the optic disc analysis and comparison with a reference database of eyes free of ocular pathology, and is finally automatically saved to the PC.

Two software programs for installation on an off-the-shelf PC are provided with the device. The first PC software program, called "FastMap", captures the images from the device, analyzes them and enables viewing of the data. The second PC software program, called "OCT Viewer", can only analyze and view the data.

Accessories include the following: power cord; chin-rest paper sheet; LAN cable; chin-rest paper pins; external fixation target; dust cover; spare parts case; and stylus pen. An optional Anterior Segment Kit allows the user to activate the anterior segment imaging functionality of the Maestro device.

#### Substantial Equivalence

The Maestro is substantially equivalent to the primary predicate device Optovue's iVue with Normative Database (iVue) cleared under K121739. The Maestro and the iVue with Normative Database have very similar indication for use statements with respect to the reference database and its ability to provide reference analyses. The Maestro and iVue are both OCT devices and each provide cross-sectional images of the fundus and measurement of posterior ocular structures. Both devices provide analysis functions for both retinal disease and glaucoma. Also, both are intended for use as a diagnostic device to aid in the detection and management of various ocular diseases. 

{5}------------------------------------------------

The Maestro and the iVue have similar technological characteristics with respect to the kind of OCT images they capture, the processing methods, and presentation of OCT information. The light source used by both devices is a spectral domain OCT (SD-OCT) using a superluminescent diode (SLD) with a center wavelength of 840 nm. The Maestro has a faster scan speed of 50,000 A-scans per second compared to 26,000 A-scans per second for the iVue. In addition, the Maestro has deeper A-scan depth of 2.6 mm (in tissue) at axial resolution of 6 micron (in tissue) and transverse resolution of 20 micron (in tissue) compared to A-scan depth of 2.0 to 2.3 mm at axial resolution of 5 micron and unspecified transverse resolution for the iVue. Both devices use the 3D Scan pattern and other similar scan patterns for posterior and Radial and Line-like scans for anterior. For the anterior segment feature, the Maestro provides only a visualization while the iVue also provides automated measurement of central corneal thickness.

In addition to the primary predicate device, Topcon's TRC-NW300 Non-Mydriatic Retinal Camera (NW300) (K123460) as a secondary predicate device for functionality. This secondary predicate device serves as a comparator device to the subject device since the scientific methodology and the device performance for the individual components have already been established. Comparative data between the subject and secondary predicate device will be provided in the device labeling.

Based on the intended use/indications for use and technological characteristics described above, the subject device is substantially equivalent to the primary predicate device.

# Performance Data

Performance bench testing has been conducted on the Maestro to demonstrate substantial equivalence to the predicate devices. A list of testing conducted is as follows:

- Axial and Horizontal Resolutions
- . Trueness and Precision for Axial and Horizontal Directions
- . Testing to Consensus Standards for Ophthalmic Devices
	- ISO 15004-1:2006 Ophthalmic instruments Fundamental requirements and test methods-Part 1: o General requirements applicable to all ophthalmic instruments
	- ISO 10940: 2009 Ophthalmic instruments Fundus cameras o
	- ISO 15004-2:2007 Ophthalmic Instruments Fundamental requirements and test methods Part 2: O Light hazard protection
- Electrical Safety ●
	- o AAMI ANSI ES 60601-1:2005/A2:2010/(R)2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (Third Edition)
- Electromagnetic Compatibility
	- IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety O and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests (Third Edition)

Biocompatibility of materials used in the patient contacting components were already evaluated and established in company's predicate device submissions.

{6}------------------------------------------------

# Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could result in an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

#### Clinical Data

Topcon conducted clinical studies to determine agreement, repeatability and reproducibility of the measurement data between the Maestro device and the iVue with normative database. A comparative study between images of the Maestro and the TRC NW300 Non Mydriatic Retinal Camera was also conducted. These studies included data from 25 normal subjects with retinal disease, and 25 subjects with glaucoma. The studies included five main analysis, 2) an agreement analysis, 3) an agreement analysis, 3) a fundus photograph evaluation, 4) an anterior segment image evaluation, and 5) a B scan image quality evaluation.

A reference database study of the Maestro was also conducted.

Consistent with the labeling for the test and control devices, the clinical site was permitted to make manual adjustments to automated segmentation based on the clinician's judgment. The study device manual adjustment rates for macula scans ranged from approximately 0-5% for 6x6 macula and 10-15% for 12x9 wide scans. For RNFL thicknesses and optic disc parameters, the overall manual adjustment rates for the Maestro were 11-32% and 20-40% for 6x6 disc and 12x9 wide scans, respectively.

| Table 1: Full Retinal Thickness - Normal Eyes Repeatability and Reproducibility |    |        |                 |       |       |       |       |       |
|---------------------------------------------------------------------------------|----|--------|-----------------|-------|-------|-------|-------|-------|
| Repeatability                                                                   |    |        | Reproducibility |       |       |       |       |       |
|                                                                                 |    |        |                 |       |       |       |       |       |
| Device                                                                          |    |        |                 |       |       |       |       |       |
| Measurement                                                                     | N  | Mean   | SD              | Limit | CV%   | SD    | Limit | CV%   |
| Maestro 12x9 3D Wide                                                            |    |        |                 |       |       |       |       |       |
| Central Fovea                                                                   | 25 | 233.99 | 2.608           | 7.302 | 1.115 | 3.419 | 9.573 | 1.461 |
| Superior Parafoveal                                                             | 25 | 309.12 | 1.647           | 4.611 | 0.533 | 2.435 | 6.819 | 0.788 |
| Nasal Parafoveal                                                                | 25 | 309.72 | 1.470           | 4.116 | 0.475 | 2.595 | 7.266 | 0.838 |
| Inferior Parafoveal                                                             | 25 | 307.29 | 1.281           | 3.588 | 0.417 | 2.235 | 6.259 | 0.727 |
| Temporal Parafoveal                                                             | 25 | 296.65 | 1.417           | 3.968 | 0.478 | 2.294 | 6.424 | 0.773 |
| Superior Perifoveal                                                             | 25 | 268.84 | 1.442           | 4.039 | 0.536 | 2.140 | 5.991 | 0.796 |
| Nasal Perifoveal                                                                | 25 | 286.14 | 0.818           | 2.289 | 0.286 | 1.681 | 4.707 | 0.587 |
| Inferior Perifoveal                                                             | 25 | 257.54 | 1.059           | 2.965 | 0.411 | 1.356 | 3.796 | 0.526 |
| Temporal Perifoveal                                                             | 25 | 253.97 | 0.948           | 2.655 | 0.373 | 1.519 | 4.252 | 0.598 |
| Maestro 6x6 3D Macula                                                           |    |        |                 |       |       |       |       |       |
| Central Fovea                                                                   | 25 | 231.18 | 1.581           | 4.428 | 0.684 | 2.370 | 6.637 | 1.025 |
| Superior Parafoveal                                                             | 25 | 310.61 | 1.649           | 4.619 | 0.531 | 2.616 | 7.324 | 0.842 |
| Nasal Parafoveal                                                                | 25 | 309.71 | 1.328           | 3.717 | 0.429 | 2.468 | 6.911 | 0.797 |
| Inferior Parafoveal                                                             | 25 | 307.78 | 1.244           | 3.484 | 0.404 | 2.112 | 5.913 | 0.686 |
| Temporal Parafoveal                                                             | 25 | 295.58 | 1.217           | 3.409 | 0.412 | 2.066 | 5.785 | 0.699 |
| Superior Perifoveal                                                             | 25 | 268.73 | 1.468           | 4.111 | 0.546 | 2.017 | 5.648 | 0.751 |

#### Precision Assessment:

# Table 1: Full Retinal Thicknoss - Normal Eves Popoatability and Roproducibili

\\DC - 026811/000013 - 8588672 v1

{7}------------------------------------------------

| Repeatability         |    |        |       | Reproducibility |       |       |       |       |
|-----------------------|----|--------|-------|-----------------|-------|-------|-------|-------|
| Device<br>Measurement | N  | Mean   | SD    | Limit           | CV%   | SD    | Limit | CV%   |
| Nasal Perifoveal      | 25 | 285.91 | 0.873 | 2.445           | 0.305 | 1.912 | 5.354 | 0.669 |
| Inferior Perifoveal   | 25 | 258.27 | 1.285 | 3.599           | 0.498 | 2.143 | 6.002 | 0.830 |
| Temporal Perifoveal   | 25 | 251.83 | 1.088 | 3.045           | 0.432 | 1.963 | 5.496 | 0.779 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{8}------------------------------------------------

Table 2: Full Retinal Thickness - Retinal Disease Eyes Repeatability and Reproducibility

|                       |    |        | Repeatability |        | Reproducibility |       |        |       |
|-----------------------|----|--------|---------------|--------|-----------------|-------|--------|-------|
| Device                |    |        |               |        |                 |       |        |       |
| Measurement           | N  | Mean   | SD            | Limit  | CV%             | SD    | Limit  | CV%   |
| Maestro 12x9 3D Wide  |    |        |               |        |                 |       |        |       |
| Central Fovea         | 26 | 250.15 | 3.697         | 10.351 | 1.478           | 4.745 | 13.285 | 1.897 |
| Superior Parafoveal   | 26 | 317.62 | 1.561         | 4.371  | 0.492           | 2.371 | 6.639  | 0.747 |
| Nasal Parafoveal      | 26 | 315.72 | 1.512         | 4.233  | 0.479           | 2.737 | 7.665  | 0.867 |
| Inferior Parafoveal   | 26 | 313.43 | 2.060         | 5.769  | 0.657           | 2.983 | 8.351  | 0.952 |
| Temporal Parafoveal   | 26 | 303.67 | 2.042         | 5.718  | 0.672           | 3.164 | 8.860  | 1.042 |
| Superior Perifoveal   | 26 | 276.82 | 1.356         | 3.797  | 0.490           | 1.758 | 4.922  | 0.635 |
| Nasal Perifoveal      | 26 | 288.87 | 1.093         | 3.059  | 0.378           | 1.719 | 4.812  | 0.595 |
| Inferior Perifoveal   | 26 | 263.37 | 1.800         | 5.039  | 0.683           | 2.319 | 6.495  | 0.881 |
| Temporal Perifoveal   | 26 | 261.82 | 1.950         | 5.459  | 0.745           | 2.279 | 6.382  | 0.871 |
| Maestro 6x6 3D Macula |    |        |               |        |                 |       |        |       |
| Central Fovea         | 26 | 247.02 | 2.693         | 7.541  | 1.090           | 3.301 | 9.242  | 1.336 |
| Superior Parafoveal   | 26 | 317.54 | 1.754         | 4.910  | 0.552           | 2.714 | 7.599  | 0.855 |
| Nasal Parafoveal      | 26 | 315.98 | 1.515         | 4.241  | 0.479           | 2.575 | 7.210  | 0.815 |
| Inferior Parafoveal   | 26 | 312.70 | 1.555         | 4.354  | 0.497           | 2.488 | 6.966  | 0.796 |
| Temporal Parafoveal   | 26 | 301.80 | 1.484         | 4.156  | 0.492           | 2.704 | 7.570  | 0.896 |
| Superior Perifoveal   | 26 | 276.60 | 1.531         | 4.286  | 0.553           | 2.365 | 6.621  | 0.855 |
| Nasal Perifoveal      | 26 | 288.21 | 1.083         | 3.033  | 0.376           | 1.902 | 5.326  | 0.660 |
| Inferior Perifoveal   | 26 | 263.98 | 1.697         | 4.751  | 0.643           | 2.177 | 6.095  | 0.825 |
| Temporal Perifoveal   | 26 | 258.61 | 1.136         | 3.182  | 0.439           | 1.994 | 5.583  | 0.771 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{9}------------------------------------------------

| Table 3: Full Retinal Thickness - Glaucoma Eyes Repeatability and Reproducibility |  |  |
|-----------------------------------------------------------------------------------|--|--|
|                                                                                   |  |  |

| Table 3: Full Retinal Thickness - Glaucoma Eyes Repeatability and Reproducibility |    |        |       |       |       |       |        |       |
|-----------------------------------------------------------------------------------|----|--------|-------|-------|-------|-------|--------|-------|
| Device<br>Measurement                                                             | N  | Mean   | SD    | Limit | CV%   | SD    | Limit  | CV%   |
| Maestro 12x9 3D Wide                                                              |    |        |       |       |       |       |        |       |
| Central Fovea                                                                     | 25 | 243.09 | 2.914 | 8.159 | 1.199 | 3.985 | 11.158 | 1.639 |
| Superior Parafoveal                                                               | 25 | 304.59 | 1.921 | 5.380 | 0.631 | 2.727 | 7.635  | 0.895 |
| Nasal Parafoveal                                                                  | 25 | 307.97 | 1.520 | 4.255 | 0.493 | 2.676 | 7.492  | 0.869 |
| Inferior Parafoveal                                                               | 25 | 299.65 | 2.012 | 5.633 | 0.671 | 2.723 | 7.624  | 0.909 |
| Temporal Parafoveal                                                               | 25 | 293.57 | 1.587 | 4.444 | 0.541 | 2.687 | 7.524  | 0.915 |
| Superior Perifoveal                                                               | 25 | 261.20 | 1.731 | 4.846 | 0.663 | 2.308 | 6.463  | 0.884 |
| Nasal Perifoveal                                                                  | 25 | 276.95 | 1.393 | 3.901 | 0.503 | 1.831 | 5.127  | 0.661 |
| Inferior Perifoveal                                                               | 25 | 246.04 | 1.726 | 4.832 | 0.701 | 2.225 | 6.230  | 0.904 |
| Temporal Perifoveal                                                               | 25 | 247.98 | 1.685 | 4.718 | 0.679 | 2.229 | 6.241  | 0.899 |
| Maestro 6x6 3D Macula                                                             |    |        |       |       |       |       |        |       |
| Central Fovea                                                                     | 25 | 241.05 | 1.677 | 4.695 | 0.696 | 2.986 | 8.361  | 1.239 |
| Superior Parafoveal                                                               | 25 | 305.20 | 1.462 | 4.095 | 0.479 | 2.555 | 7.154  | 0.837 |
| Nasal Parafoveal                                                                  | 25 | 307.46 | 1.245 | 3.487 | 0.405 | 2.791 | 7.815  | 0.908 |
| Inferior Parafoveal                                                               | 25 | 299.20 | 1.270 | 3.556 | 0.424 | 2.644 | 7.404  | 0.884 |
| Temporal Parafoveal                                                               | 25 | 292.57 | 1.222 | 3.421 | 0.418 | 2.636 | 7.380  | 0.901 |
| Superior Perifoveal                                                               | 25 | 261.20 | 1.379 | 3.860 | 0.528 | 2.142 | 5.998  | 0.820 |
| Nasal Perifoveal                                                                  | 25 | 276.37 | 0.919 | 2.573 | 0.332 | 1.987 | 5.563  | 0.719 |
| Inferior Perifoveal                                                               | 25 | 246.18 | 1.492 | 4.177 | 0.606 | 2.298 | 6.433  | 0.933 |
| Temporal Perifoveal                                                               | 25 | 245.71 | 1.030 | 2.884 | 0.419 | 2.366 | 6.624  | 0.963 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Repeatability limit = 2.8 x Repeatability SD
Reproducibility limit = 2.8 x Reproducibility SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{10}------------------------------------------------

Table 4: Ganglion Cell + IPL - Normal Eves Repeatability and Reproducibility

|                       |                   | Repeatability |       |       | Reproducibility |       |       |       |       |
|-----------------------|-------------------|---------------|-------|-------|-----------------|-------|-------|-------|-------|
| Device                | Measurement       | N             | Mean  | SD    | Limit           | CV%   | SD    | Limit | CV%   |
| Maestro 12x9 3D Wide  | Superior          | 25            | 70.37 | 0.641 | 1.795           | 0.911 | 0.715 | 2.001 | 1.015 |
|                       | Superior Nasal    | 25            | 74.66 | 0.671 | 1.878           | 0.898 | 0.707 | 1.980 | 0.947 |
|                       | Superior Temporal | 25            | 71.46 | 0.590 | 1.651           | 0.825 | 0.831 | 2.326 | 1.162 |
|                       | Inferior          | 25            | 68.11 | 0.634 | 1.776           | 0.931 | 0.663 | 1.856 | 0.973 |
|                       | Inferior Nasal    | 25            | 73.59 | 0.699 | 1.957           | 0.950 | 0.768 | 2.150 | 1.043 |
|                       | Inferior Temporal | 25            | 73.12 | 0.553 | 1.550           | 0.757 | 0.708 | 1.983 | 0.968 |
|                       | Average           | 25            | 71.89 | 0.291 | 0.814           | 0.404 | 0.365 | 1.023 | 0.508 |
| Maestro 6x6 3D Macula | Superior          | 25            | 70.48 | 0.736 | 2.060           | 1.044 | 0.809 | 2.266 | 1.148 |
|                       | Superior Nasal    | 25            | 74.62 | 0.394 | 1.102           | 0.528 | 0.548 | 1.535 | 0.735 |
|                       | Superior Temporal | 25            | 71.93 | 0.611 | 1.712           | 0.850 | 0.712 | 1.993 | 0.989 |
|                       | Inferior          | 25            | 68.33 | 0.453 | 1.268           | 0.663 | 0.657 | 1.841 | 0.962 |
|                       | Inferior Nasal    | 25            | 73.58 | 0.436 | 1.220           | 0.592 | 0.595 | 1.667 | 0.809 |
|                       | Inferior Temporal | 25            | 73.46 | 0.562 | 1.573           | 0.765 | 0.611 | 1.712 | 0.832 |
|                       | Average           | 25            | 72.08 | 0.263 | 0.735           | 0.364 | 0.401 | 1.124 | 0.557 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{11}------------------------------------------------

| Table 5: Ganglion Cell + IPL - Retinal Disease Eyes Repeatability and Reproducibility |    |       |               |       |       |                 |       |       |
|---------------------------------------------------------------------------------------|----|-------|---------------|-------|-------|-----------------|-------|-------|
|                                                                                       |    |       | Repeatability |       |       | Reproducibility |       |       |
| Device                                                                                |    |       |               |       |       |                 |       |       |
| Measurement                                                                           | N  | Mean  | SD            | Limit | CV%   | SD              | Limit | CV%   |
| Maestro 12x9 3D Wide                                                                  |    |       |               |       |       |                 |       |       |
| Superior                                                                              | 26 | 69.57 | 1.234         | 3.456 | 1.774 | 1.779           | 4.982 | 2.558 |
| Superior Nasal                                                                        | 26 | 73.78 | 0.768         | 2.150 | 1.041 | 0.812           | 2.274 | 1.101 |
| Superior Temporal                                                                     | 26 | 70.05 | 0.963         | 2.696 | 1.375 | 1.734           | 4.854 | 2.475 |
| Inferior                                                                              | 26 | 65.84 | 1.760         | 4.928 | 2.673 | 1.826           | 5.113 | 2.774 |
| Inferior Nasal                                                                        | 26 | 70.86 | 1.834         | 5.136 | 2.589 | 1.844           | 5.164 | 2.603 |
| Inferior Temporal                                                                     | 26 | 70.39 | 1.427         | 3.995 | 2.027 | 2.537           | 7.103 | 3.604 |
| Average                                                                               | 26 | 70.08 | 0.762         | 2.134 | 1.088 | 0.819           | 2.294 | 1.169 |
| Maestro 6x6 3D Macula                                                                 |    |       |               |       |       |                 |       |       |
| Superior                                                                              | 26 | 68.87 | 0.687         | 1.923 | 0.997 | 0.695           | 1.945 | 1.008 |
| Superior Nasal                                                                        | 26 | 74.04 | 0.841         | 2.355 | 1.136 | 0.926           | 2.593 | 1.251 |
| Superior Temporal                                                                     | 26 | 70.49 | 1.023         | 2.866 | 1.452 | 1.286           | 3.601 | 1.824 |
| Inferior                                                                              | 26 | 66.77 | 0.810         | 2.268 | 1.213 | 0.940           | 2.633 | 1.408 |
| Inferior Nasal                                                                        | 26 | 71.07 | 0.491         | 1.374 | 0.690 | 0.699           | 1.958 | 0.984 |
| Inferior Temporal                                                                     | 26 | 71.37 | 0.993         | 2.782 | 1.392 | 1.078           | 3.017 | 1.510 |

0.375

Mean = Intercept of the ANOVA model

Average

Repeatability SD = Square root of the residual variance

26

70.45

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

1.049

0.532

0.465

1.302

0.660

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{12}------------------------------------------------

Table 6: Ganglion Cell + IPL - Glaucoma Eves Repeatability and Reproducibility

| Table 6: Ganglion Cell + IPL - Glaucoma Eyes Repeatability and Reproducibility |                   |    |       |               |       |       |                 |       |       |
|--------------------------------------------------------------------------------|-------------------|----|-------|---------------|-------|-------|-----------------|-------|-------|
| Device                                                                         | Measurement       | N  | Mean  | Repeatability |       |       | Reproducibility |       |       |
|                                                                                |                   |    |       | SD            | Limit | CV%   | SD              | Limit | CV%   |
| Maestro 12x9 3D Wide                                                           |                   |    |       |               |       |       |                 |       |       |
|                                                                                | Superior          | 25 | 63.55 | 0.884         | 2.476 | 1.392 | 1.011           | 2.832 | 1.591 |
|                                                                                | Superior Nasal    | 25 | 68.77 | 0.996         | 2.790 | 1.449 | 1.169           | 3.274 | 1.700 |
|                                                                                | Superior Temporal | 25 | 63.93 | 0.883         | 2.472 | 1.381 | 0.908           | 2.542 | 1.420 |
|                                                                                | Inferior          | 25 | 59.61 | 0.932         | 2.608 | 1.563 | 1.063           | 2.977 | 1.784 |
|                                                                                | Inferior Nasal    | 25 | 66.07 | 0.868         | 2.432 | 1.315 | 0.980           | 2.743 | 1.483 |
|                                                                                | Inferior Temporal | 25 | 63.34 | 0.902         | 2.526 | 1.424 | 0.935           | 2.618 | 1.476 |
|                                                                                | Average           | 25 | 64.22 | 0.403         | 1.129 | 0.628 | 0.460           | 1.288 | 0.716 |
| Maestro 6x6 3D Macula                                                          |                   |    |       |               |       |       |                 |       |       |
|                                                                                | Superior          | 25 | 64.27 | 0.592         | 1.658 | 0.921 | 0.708           | 1.983 | 1.102 |
|                                                                                | Superior Nasal    | 25 | 69.54 | 0.480         | 1.345 | 0.691 | 0.559           | 1.566 | 0.804 |
|                                                                                | Superior Temporal | 25 | 66.21 | 0.766         | 2.145 | 1.157 | 0.924           | 2.586 | 1.395 |
|                                                                                | Inferior          | 25 | 60.86 | 0.784         | 2.194 | 1.288 | 0.883           | 2.472 | 1.451 |
|                                                                                | Inferior Nasal    | 25 | 66.11 | 0.503         | 1.407 | 0.760 | 0.689           | 1.928 | 1.042 |
|                                                                                | Inferior Temporal | 25 | 66.00 | 0.765         | 2.142 | 1.159 | 0.891           | 2.494 | 1.350 |
|                                                                                | Average           | 25 | 65.53 | 0.388         | 1.087 | 0.593 | 0.483           | 1.351 | 0.736 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{13}------------------------------------------------

Table 7: Ganglion Cell Complex Thickness - Normal Eyes Repeatability and Reproducibility

| Table 7: Ganglion Cell Complex Thickness - Normal Eyes Repeatability and Reproducibility |             |        |       |               |       |                 |       |       |     |
|------------------------------------------------------------------------------------------|-------------|--------|-------|---------------|-------|-----------------|-------|-------|-----|
| Device                                                                                   | Measurement | N      | Mean  | Repeatability |       | Reproducibility |       |       |     |
|                                                                                          |             |        |       | SD            | Limit | CV%             | SD    | Limit | CV% |
| Maestro 12x9 3D Wide                                                                     |             |        |       |               |       |                 |       |       |     |
| Superior                                                                                 | 25          | 106.36 | 0.731 | 2.046         | 0.687 | 0.872           | 2.441 | 0.819 |     |
| Superior Nasal                                                                           | 25          | 118.01 | 0.969 | 2.713         | 0.821 | 1.070           | 2.995 | 0.906 |     |
| Superior Temporal                                                                        | 25          | 94.11  | 0.595 | 1.665         | 0.632 | 0.755           | 2.115 | 0.803 |     |
| Inferior                                                                                 | 25          | 106.27 | 0.746 | 2.087         | 0.702 | 1.007           | 2.820 | 0.948 |     |
| Inferior Nasal                                                                           | 25          | 120.43 | 0.887 | 2.483         | 0.736 | 1.272           | 3.562 | 1.056 |     |
| Inferior Temporal                                                                        | 25          | 97.48  | 0.640 | 1.791         | 0.656 | 0.821           | 2.299 | 0.842 |     |
| Average                                                                                  | 25          | 107.11 | 0.504 | 1.411         | 0.470 | 0.691           | 1.936 | 0.645 |     |
| Superior Average                                                                         | 25          | 106.16 | 0.608 | 1.702         | 0.573 | 0.729           | 2.042 | 0.687 |     |
| Inferior Average                                                                         | 25          | 108.06 | 0.579 | 1.621         | 0.536 | 0.845           | 2.367 | 0.782 |     |
| Maestro 6x6 3D Macula                                                                    |             |        |       |               |       |                 |       |       |     |
| Superior                                                                                 | 25          | 106.95 | 1.497 | 4.192         | 1.400 | 1.719           | 4.813 | 1.607 |     |
| Superior Nasal                                                                           | 25          | 118.02 | 0.634 | 1.776         | 0.537 | 0.942           | 2.637 | 0.798 |     |
| Superior Temporal                                                                        | 25          | 93.78  | 0.711 | 1.992         | 0.759 | 0.924           | 2.587 | 0.985 |     |
| Inferior                                                                                 | 25          | 107.29 | 0.681 | 1.908         | 0.635 | 1.137           | 3.184 | 1.060 |     |
| Inferior Nasal                                                                           | 25          | 121.27 | 0.687 | 1.925         | 0.567 | 0.901           | 2.522 | 0.743 |     |
| Inferior Temporal                                                                        | 25          | 97.09  | 0.682 | 1.909         | 0.702 | 0.979           | 2.741 | 1.008 |     |
| Average                                                                                  | 25          | 107.43 | 0.535 | 1.497         | 0.498 | 0.783           | 2.193 | 0.729 |     |
| Superior Average                                                                         | 25          | 106.25 | 0.707 | 1.981         | 0.666 | 0.895           | 2.507 | 0.843 |     |
| Inferior Average                                                                         | 25          | 108.55 | 0.541 | 1.515         | 0.498 | 0.845           | 2.365 | 0.778 |     |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{14}------------------------------------------------

Table 8: Ganglion Cell Complex Thickness - Retinal Disease Eyes Repeatability and Reproducibility

| Table 8: Ganglion Cell Complex Thickness - Retinal Disease Eyes Repeatability and Reproducibility |               |        |       |        |                 |       |        |       |
|---------------------------------------------------------------------------------------------------|---------------|--------|-------|--------|-----------------|-------|--------|-------|
|                                                                                                   | Repeatability |        |       |        | Reproducibility |       |        |       |
| Device<br>Measurement                                                                             | N             | Mean   | SD    | Limit  | CV%             | SD    | Limit  | CV%   |
| Maestro 12x9 3D Wide                                                                              |               |        |       |        |                 |       |        |       |
| Superior                                                                                          | 26            | 108.98 | 1.462 | 4.095  | 1.342           | 2.295 | 6.425  | 2.106 |
| Superior Nasal                                                                                    | 26            | 120.00 | 0.926 | 2.593  | 0.772           | 0.985 | 2.758  | 0.821 |
| Superior Temporal                                                                                 | 26            | 94.96  | 1.024 | 2.867  | 1.078           | 1.262 | 3.534  | 1.329 |
| Inferior                                                                                          | 26            | 105.20 | 2.236 | 6.259  | 2.125           | 2.236 | 6.259  | 2.125 |
| Inferior Nasal                                                                                    | 26            | 116.09 | 3.729 | 10.442 | 3.213           | 3.752 | 10.506 | 3.232 |
| Inferior Temporal                                                                                 | 26            | 97.86  | 2.231 | 6.248  | 2.280           | 2.420 | 6.775  | 2.473 |
| Average                                                                                           | 26            | 107.20 | 1.192 | 3.337  | 1.112           | 1.192 | 3.337  | 1.112 |
| Superior Average                                                                                  | 26            | 107.98 | 0.756 | 2.118  | 0.700           | 1.157 | 3.240  | 1.072 |
| Inferior Average                                                                                  | 26            | 106.39 | 2.280 | 6.383  | 2.143           | 2.288 | 6.407  | 2.151 |
| Maestro 6x6 3D Macula                                                                             |               |        |       |        |                 |       |        |       |
| Superior                                                                                          | 26            | 107.96 | 0.899 | 2.518  | 0.833           | 1.081 | 3.027  | 1.001 |
| Superior Nasal                                                                                    | 26            | 120.10 | 0.793 | 2.221  | 0.660           | 0.952 | 2.665  | 0.792 |
| Superior Temporal                                                                                 | 26            | 93.87  | 0.741 | 2.076  | 0.790           | 0.972 | 2.721  | 1.035 |
| Inferior                                                                                          | 26            | 106.13 | 1.160 | 3.248  | 1.093           | 1.160 | 3.248  | 1.093 |
| Inferior Nasal                                                                                    | 26            | 117.24 | 0.721 | 2.019  | 0.615           | 0.862 | 2.414  | 0.736 |
| Inferior Temporal                                                                                 | 26            | 95.85  | 0.780 | 2.184  | 0.814           | 0.901 | 2.524  | 0.940 |
| Average                                                                                           | 26            | 106.88 | 0.518 | 1.452  | 0.485           | 0.643 | 1.799  | 0.601 |
| Superior Average                                                                                  | 26            | 107.31 | 0.599 | 1.677  | 0.558           | 0.802 | 2.246  | 0.747 |
| Inferior Average                                                                                  | 26            | 106.41 | 0.612 | 1.712  | 0.575           | 0.686 | 1.920  | 0.644 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{15}------------------------------------------------

Table 9: Ganglion Cell Complex Thickness - Glaucoma Eyes Repeatability and Reproducibility

|                       |             |        | Repeatability |       |        | Reproducibility |       |        |       |
|-----------------------|-------------|--------|---------------|-------|--------|-----------------|-------|--------|-------|
|                       |             |        |               |       |        |                 |       |        |       |
| Device<br>Measurement | N           | Mean   | SD            | Limit | CV%    | SD              | Limit | CV%    |       |
| Maestro 12x9 3D Wide  |             |        |               |       |        |                 |       |        |       |
| Superior              | 25          | 95.85  | 1.053         | 2.947 | 1.098  | 1.188           | 3.327 | 1.240  |       |
| Superior Nasal        | 25          | 108.78 | 0.850         | 2.379 | 0.781  | 0.897           | 2.513 | 0.825  |       |
| Superior Temporal     | 25          | 85.28  | 1.014         | 2.840 | 1.189  | 1.033           | 2.893 | 1.211  |       |
| Inferior              | 25          | 91.59  | 1.037         | 2.903 | 1.132  | 1.071           | 2.998 | 1.169  |       |
| Inferior Nasal        | 25          | 105.43 | 0.777         | 2.175 | 0.737  | 0.974           | 2.728 | 0.924  |       |
| Inferior Temporal     | 25          | 85.08  | 0.756         | 2.116 | 0.888  | 0.818           | 2.289 | 0.961  |       |
| Average               | 25          | 95.34  | 0.608         | 1.703 | 0.638  | 0.655           | 1.833 | 0.687  |       |
| Superior Average      | 25          | 96.64  | 0.805         | 2.255 | 0.833  | 0.857           | 2.400 | 0.887  |       |
| Inferior Average      | 25          | 94.03  | 0.651         | 1.823 | 0.692  | 0.724           | 2.028 | 0.770  |       |
| Maestro 6x6 3D Macula |             |        |               |       |        |                 |       |        |       |
| Superior              | 25          | 96.85  | 0.700         | 1.960 | 0.723  | 0.908           | 2.542 | 0.937  |       |
| Superior Nasal        | 25          | 108.81 | 0.691         | 1.935 | 0.635  | 0.892           | 2.498 | 0.820  |       |
| Superior Temporal     | 25          | 84.93  | 0.597         | 1.671 | 0.703  | 0.809           | 2.264 | 0.952  |       |
| Inferior              | 25          | 92.15  | 1.042         | 2.917 | 1.131  | 1.166           | 3.264 | 1.265  |       |
| Inferior Nasal        | 25          | 105.99 | 0.706         | 1.977 | 0.666  | 0.914           | 2.558 | 0.862  |       |
| Inferior Temporal     | 25          | 84.51  | 0.742         | 2.079 | 0.878  | 0.939           | 2.629 | 1.111  |       |
| Average               | 25          | 95.59  | 0.486         | 1.360 | 0.508  | 0.648           | 1.814 | 0.678  |       |
| Superior Average      | 25          | 96.86  | 0.509         | 1.425 | 0.526  | 0.676           | 1.893 | 0.698  |       |
| Inferior Average      | 25          | 94.22  | 0.642         | 1.798 | 0.682  | 0.782           | 2.189 | 0.830  |       |
| Device                | Measurement | N      | Mean          | SD    | Limit  | CV%             | SD    | Limit  | CV%   |
| Maestro 12x9 3D Wide  |             |        |               |       |        |                 |       |        |       |
| Average               |             | 25     | 105.30        | 1.388 | 3.887  | 1.318           | 1.598 | 4.474  | 1.517 |
| Superior              |             | 25     | 127.89        | 3.754 | 10.512 | 2.936           | 3.893 | 10.901 | 3.044 |
| Nasal                 |             | 25     | 79.37         | 2.765 | 7.743  | 3.484           | 3.044 | 8.522  | 3.835 |
| Inferior              |             | 25     | 135.23        | 3.375 | 9.449  | 2.495           | 4.094 | 11.464 | 3.028 |
| Temporal              |             | 25     | 78.68         | 1.356 | 3.796  | 1.723           | 2.087 | 5.843  | 2.652 |
| 12-hour - T           |             | 25     | 65.26         | 1.606 | 4.496  | 2.461           | 1.984 | 5.554  | 3.040 |
| 12-hour - TS          |             | 25     | 93.36         | 2.636 | 7.381  | 2.824           | 3.024 | 8.467  | 3.239 |
| 12-hour - ST          |             | 25     | 138.21        | 4.533 | 12.691 | 3.280           | 4.933 | 13.812 | 3.569 |
| 12-hour - S           |             | 25     | 125.15        | 5.968 | 16.711 | 4.769           | 6.355 | 17.795 | 5.078 |
| 12-hour - SN          |             | 25     | 120.31        | 5.282 | 14.789 | 4.390           | 5.534 | 15.496 | 4.600 |
| 12-hour - NS          |             | 25     | 98.29         | 5.169 | 14.474 | 5.259           | 5.483 | 15.353 | 5.579 |
| 12-hour - N           |             | 25     | 62.75         | 2.490 | 6.972  | 3.968           | 3.514 | 9.840  | 5.600 |
| 12-hour - NI          |             | 25     | 77.11         | 4.042 | 11.318 | 5.242           | 4.072 | 11.402 | 5.281 |
| 12-hour - IN          |             | 25     | 113.67        | 5.013 | 14.035 | 4.410           | 5.880 | 16.463 | 5.173 |
| 12-hour - I           |             | 25     | 145.75        | 5.964 | 16.700 | 4.092           | 6.621 | 18.538 | 4.543 |
| 12-hour - IT          |             | 25     | 146.36        | 4.057 | 11.359 | 2.772           | 4.904 | 13.732 | 3.351 |
| 12-hour - TI          |             | 25     | 77.41         | 2.295 | 6.426  | 2.965           | 3.490 | 9.771  | 4.508 |
| Maestro 6x6 3D Disc   |             |        |               |       |        |                 |       |        |       |
| Average               |             | 25     | 104.32        | 0.973 | 2.724  | 0.933           | 1.147 | 3.211  | 1.099 |
| Superior              |             | 25     | 128.41        | 2.627 | 7.356  | 2.046           | 2.917 | 8.167  | 2.272 |
| Nasal                 |             | 25     | 78.14         | 2.007 | 5.619  | 2.569           | 2.302 | 6.445  | 2.946 |
| Inferior              |             | 25     | 135.51        | 2.003 | 5.610  | 1.478           | 2.503 | 7.008  | 1.847 |
| Temporal              |             | 25     | 75.22         | 1.335 | 3.738  | 1.775           | 1.859 | 5.206  | 2.471 |
| 12-hour - T           |             | 25     | 61.65         | 1.396 | 3.908  | 2.264           | 1.816 | 5.086  | 2.946 |
| 12-hour - TS          |             | 25     | 90.01         | 2.161 | 6.050  | 2.401           | 2.792 | 7.817  | 3.101 |
| 12-hour - ST          |             | 25     | 139.48        | 3.271 | 9.157  | 2.345           | 4.344 | 12.163 | 3.114 |
| 12-hour - S           |             | 25     | 125.52        | 4.964 | 13.898 | 3.954           | 5.145 | 14.407 | 4.099 |
| 12-hour - SN          |             | 25     | 120.20        | 4.160 | 11.647 | 3.461           | 4.294 | 12.024 | 3.573 |
| 12-hour - NS          |             | 25     | 97.39         | 3.622 | 10.143 | 3.720           | 4.295 | 12.025 | 4.410 |
| 12-hour - N           |             | 25     | 61.11         | 1.526 | 4.273  | 2.497           | 1.759 | 4.926  | 2.879 |
| 12-hour - NI          |             | 25     | 75.87         | 2.995 | 8.385  | 3.947           | 3.423 | 9.585  | 4.511 |
| 12-hour - IN          |             | 25     | 113.57        | 3.298 | 9.234  | 2.904           | 3.590 | 10.052 | 3.161 |
| 12-hour - I           |             | 25     | 147.83        | 3.633 | 10.172 | 2.457           | 4.388 | 12.287 | 2.968 |
| 12-hour - IT          |             | 25     | 145.02        | 3.980 | 11.144 | 2.745           | 4.366 | 12.224 | 3.010 |
| 12-hour - TI          |             | 25     | 73.98         | 2.212 | 6.194  | 2.990           | 3.259 | 9.125  | 4.405 |

N = Number of subjects

Mean = Intercept of the ANOVA model

Repeatability SD = Square root of the residual variance

Reproducibility SD = Square root of the operator/device variance, the interaction variance and the residual variance

Repeatability limit = 2.8 x Repeatability SD

Reproducibility limit = 2.8 x Reproducibility SD

Repeatability CV% = (Repeatability SD)/Intercept x 100%

{16}------------------------------------------------

Table 10: Retinal Nerve Fiber Layer - Normal Eyes Repeatability and Reproducibility

{17}------------------------------------------------

|                                                                                                                                                                                                                 |   |      | Repeatability |       |     | Reproducibility |       |     |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|------|---------------|-------|-----|-----------------|-------|-----|
| Device<br>Measurement                                                                                                                                                                                           | N | Mean | SD            | Limit | CV% | SD              | Limit | CV% |
| T=Temporal; S=Superior; N=Nasal; I=Inferior<br>All statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interaction<br>between operator/device and eye. |   |      |               |       |     |                 |       |     |
| N = Number of subjects                                                                                                                                                                                          |   |      |               |       |     |                 |       |     |
| Mean = Intercept of the ANOVA model                                                                                                                                                                             |   |      |               |       |     |                 |       |     |
| Repeatability SD = Square root of the residual variance                                                                                                                                                         |   |      |               |       |     |                 |       |     |
| Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance                                                                                 |   |      |               |       |     |                 |       |     |
| Repeatability limit = 2.8 x Repeatability SD                                                                                                                                                                    |   |      |               |       |     |                 |       |     |
| Reproducibility limit = 2.8 x Reproducibility SD                                                                                                                                                                |   |      |               |       |     |                 |       |     |
| Repeatability CV% = (Repeatability SD)/Intercept x 100%                                                                                                                                                         |   |      |               |       |     |                 |       |     |
| Reproducibility CV% = (Reproducibility SD)/Intercept x 100%                                                                                                                                                     |   |      |               |       |     |                 |       |     |

{18}------------------------------------------------

Table 11: Retinal Nerve Fiber Layer - Retinal Disease Eyes Repeatability and Reproducibility

| Device               | Measurement | N  | Mean   | SD    | Repeatability |       |       | Reproducibility |       |     |
|----------------------|-------------|----|--------|-------|---------------|-------|-------|-----------------|-------|-----|
|                      |             |    |        |       | SD            | Limit | CV%   | SD              | Limit | CV% |
| Maestro 12x9 3D Wide |             |    |        |       |               |       |       |                 |       |     |
| Average              |             | 26 | 101.46 | 1.618 | 4.529         | 1.594 | 1.916 | 5.364           | 1.888 |     |
| Superior             |             | 26 | 123.56 | 4.099 | 11.476        | 3.317 | 4.517 | 12.648          | 3.656 |     |
| Nasal                |             | 26 | 76.76  | 3.346 | 9.370         | 4.359 | 3.948 | 11.055          | 5.143 |     |
| Inferior             |             | 26 | 128.81 | 3.343 | 9.360         | 2.595 | 5.427 | 15.194          | 4.213 |     |
| Temporal             |             | 26 | 76.73  | 1.708 | 4.783         | 2.226 | 2.166 | 6.064           | 2.823 |     |
| 12-hour - T          |             | 26 | 67.02  | 1.942 | 5.438         | 2.898 | 2.510 | 7.028           | 3.745 |     |
| 12-hour - TS         |             | 26 | 91.79  | 2.691 | 7.535         | 2.932 | 3.143 | 8.799           | 3.424 |     |
| 12-hour - ST         |             | 26 | 133.09 | 4.608 | 12.901        | 3.462 | 5.515 | 15.443          | 4.144 |     |
| 12-hour - S          |             | 26 | 122.77 | 6.633 | 18.572        | 5.403 | 7.191 | 20.135          | 5.857 |     |
| 12-hour - SN         |             | 26 | 114.84 | 6.720 | 18.816        | 5.852 | 6.720 | 18.816          | 5.852 |     |
| 12-hour - NS         |             | 26 | 91.74  | 6.559 | 18.365        | 7.150 | 6.839 | 19.149          | 7.455 |     |
| 12-hour - N          |             | 26 | 64.72  | 3.133 | 8.772         | 4.840 | 3.512 | 9.833           | 5.426 |     |
| 12-hour - NI         |             | 26 | 73.86  | 6.014 | 16.839        | 8.143 | 6.407 | 17.939          | 8.675 |     |
| 12-hour - IN         |             | 26 | 111.63 | 5.625 | 15.749        | 5.039 | 9.004 | 25.210          | 8.066 |     |
| 12-hour - I          |             | 26 | 146.30 | 5.483 | 15.353        | 3.748 | 6.994 | 19.583          | 4.781 |     |
| 12-hour - IT         |             | 26 | 128.52 | 5.212 | 14.593        | 4.055 | 5.932 | 16.610          | 4.616 |     |
| 12-hour - TI         |             | 26 | 71.39  | 2.549 | 7.136         | 3.570 | 2.791 | 7.815           | 3.910 |     |
| Maestro 6x6 3D Disc  |             |    |        |       |               |       |       |                 |       |     |
| Average              |             | 26 | 101.72 | 1.102 | 3.086         | 1.084 | 1.505 | 4.214           | 1.480 |     |
| Superior             |             | 26 | 125.25 | 2.512 | 7.034         | 2.006 | 2.609 | 7.305           | 2.083 |     |
| Nasal                |             | 26 | 76.24  | 2.175 | 6.090         | 2.853 | 3.523 | 9.863           | 4.620 |     |
| Inferior             |             | 26 | 131.15 | 2.293 | 6.419         | 1.748 | 2.830 | 7.924           | 2.158 |     |
| Temporal             |             | 26 | 74.25  | 1.287 | 3.603         | 1.733 | 1.446 | 4.048           | 1.947 |     |
| 12-hour - T          |             | 26 | 64.47  | 1.600 | 4.480         | 2.482 | 1.916 | 5.365           | 2.972 |     |
| 12-hour - TS         |             | 26 | 89.31  | 2.218 | 6.210         | 2.483 | 2.644 | 7.404           | 2.961 |     |
| 12-hour - ST         |             | 26 | 134.93 | 4.220 | 11.815        | 3.127 | 5.015 | 14.042          | 3.717 |     |
| 12-hour - S          |             | 26 | 123.86 | 5.350 | 14.980        | 4.320 | 5.865 | 16.423          | 4.736 |     |
| 12-hour - SN         |             | 26 | 116.90 | 3.694 | 10.343        | 3.160 | 4.234 | 11.854          | 3.622 |     |
| 12-hour - NS         |             | 26 | 91.83  | 4.177 | 11.694        | 4.548 | 6.026 | 16.873          | 6.562 |     |
| 12-hour - N          |             | 26 | 63.16  | 3.237 | 9.065         | 5.125 | 3.583 | 10.031          | 5.672 |     |
| 12-hour - NI         |             | 26 | 73.70  | 4.219 | 11.814        | 5.725 | 5.444 | 15.242          | 7.387 |     |
| 12-hour - IN         |             | 26 | 114.30 | 4.239 | 11.869        | 3.709 | 5.469 | 15.314          | 4.785 |     |
| 12-hour - I          |             | 26 | 150.51 | 4.359 | 12.206        | 2.896 | 4.874 | 13.648          | 3.239 |     |
| 12-hour - IT         |             | 26 | 128.61 | 3.770 | 10.556        | 2.932 | 4.834 | 13.535          | 3.759 |     |
| 12-hour - TI         |             | 26 | 68.95  | 2.020 | 5.656         | 2.930 | 2.473 | 6.925           | 3.587 |     |

{19}------------------------------------------------

|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |    |        |       | Repeatability |       |                 | Reproducibility |       |     |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|--------|-------|---------------|-------|-----------------|-----------------|-------|-----|
| Device<br>Measurement                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | N  | Mean   | SD    | SD            | Limit | CV%             | SD              | Limit | CV% |
| T=Temporal; S=Superior; N=Nasal; I=Inferior<br>All statistics are estimated from two-way random-effect ANOVA model with random effects operator/device, eye and interaction<br>between operator/device and eye.                                                                                                                                                                                                                                                                           |    |        |       |               |       |                 |                 |       |     |
| N = Number of subjects<br>Mean = Intercept of the ANOVA model<br>Repeatability SD = Square root of the residual variance<br>Reproducibility SD = Square root of the sum of the operator/device variance, the interaction variance and the residual variance<br>Repeatability limit = 2.8 x Repeatability SD<br>Reproducibility limit = 2.8 x Reproducibility SD<br>Repeatability CV% = (Repeatability SD)/Intercept x 100%<br>Reproducibility CV% = (Reproducibility SD)/Intercept x 100% |    |        |       |               |       |                 |                 |       |     |
| Table 12: Retinal Nerve Fiber Layer - Glaucoma Eyes Repeatability and Reproducibility                                                                                                                                                                                                                                                                                                                                                                                                     |    |        |       |               |       |                 |                 |       |     |
| Device<br>Measurement                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | N  | Mean   | SD    | Repeatability |       | Reproducibility |                 |       |     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |    |        |       | Limit         | CV%   | SD              | Limit (Ratio)   | CV%   |     |
| Maestro 12x9 3D Wide                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |    |        |       |               |       |                 |                 |       |     |
| Average                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 25 | 83.55  | 1.646 | 4.609         | 1.970 | 1.752           | 4.906           | 2.097 |     |
| Superior                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 25 | 97.32  | 3.814 | 10.679        | 3.919 | 3.948           | 11.055          | 4.057 |     |
| Nasal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 25 | 70.63  | 3.805 | 10.653        | 5.387 | 4.036           | 11.301          | 5.715 |     |
| Inferior                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 25 | 102.90 | 3.271 | 9.160         | 3.179 | 3.402           | 9.526           | 3.306 |     |
| Temporal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 25 | 63.29  | 1.370 | 3.837         | 2.165 | 1.941           | 5.434           | 3.066 |     |
| 12-hour - T                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 25 | 55.91  | 1.779 | 4.980         | 3.181 | 2.534           | 7.094           | 4.532 |     |
| 12-hour - TS                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 25 | 73.53  | 2.471 | 6.918         | 3.360 | 2.882           | 8.071           | 3.920 |     |
| 12-hour - ST                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 25 | 101.84 | 4.691 | 13.134        | 4.606 | 4.731           | 13.246          | 4.645 |     |
| 12-hour - S                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 25 | 99.85  | 6.176 | 17.294        | 6.185 | 7.648           | 21.415          | 7.659 |     |
| 12-hour - SN                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 25 | 90.18  | 5.725 | 16.031        | 6.349 | 5.828           | 16.318          | 6.463 |     |
| 12-hour - NS                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 25 | 80.96  | 6.688 | 18.727        | 8.261 | 6.719           | 18.813          | 8.299 |     |
| 12-hour - N                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 25 | 58.29  | 4.138 | 11.586        | 7.099 | 4.436           | 12.422          | 7.611 |     |
| 12-hour - NI                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 25 | 72.63  | 4.578 | 12.819        | 6.304 | 4.903           | 13.727          | 6.750 |     |
| 12-hour - IN…

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/OBO/K161509](https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/OBO/K161509)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
