← Product Code [NFJ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NFJ) · K182376

# Harmony (K182376)

_Topcon Medical Systems, Inc. · NFJ · Sep 28, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/NFJ/K182376

## Device Facts

- **Applicant:** Topcon Medical Systems, Inc.
- **Product Code:** [NFJ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NFJ.md)
- **Decision Date:** Sep 28, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.

## Device Story

Harmony is a software-only ophthalmic image management platform. It imports clinical images, videos, and diagnostic data from ophthalmic devices (OCT, retinal cameras, slit lamps) via direct connection or network. It provides tools for viewing (zoom, pan, enhancement), measurement (line, area, Cup-to-Disc ratio, MPS), and data management. Used by clinicians in clinical settings to organize and analyze patient diagnostic information. The system facilitates clinical decision-making by centralizing data and reports from various diagnostic instruments. It supports web-based access and DICOM communication with PACS. The platform is a modification of the Synergy ODM, featuring a responsive GUI and a transition from Microsoft Silverlight to HTML5, allowing execution on Windows and iPad platforms.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by software verification and validation testing confirming the device performs as intended and meets specifications.

## Technological Characteristics

Software-only image management system. Web-based architecture; front-end migrated from Microsoft Silverlight to HTML5. Supports DICOM communication, web-based data access, and responsive GUI. Compatible with Windows XP or above and iPad. Functionality includes image import, viewing (zoom, pan, enhancement), measurement (line, area, Cup-to-Disc, MPS), and archive/backup.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Topcon Corporation Synergy ODM ([K151952](/device/K151952.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 28, 2018

Topcon Medical Systems. Inc. % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864

Re: K182376

Trade/Device Name: Harmony Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: NFJ Dated: August 29, 2018 Received: August 31, 2018

Dear Maureen O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

# Bradley S. Cunningham -A

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K182376

Device Name Harmony

Indications for Use (Describe)

Harmony is a comprehensive software platform intended for use in importing, processing, measurement, and storage of clinical images and videos of the eye as well as in management of patient data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY

#### Topcon Medical Systems, Inc. Harmony

#### 510(k) Owner

Topcon Medical Systems, Inc. 111 Bauer Drive Oakland, NJ 07436 Phone: (201) 599-5208 Contact Person: James Lorkowski

#### Submission Correspondent

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 (978) 207-1245 Phone: Email: Maureen@oconnellregulatory.com

Date Prepared:

# Trade Name of Device

Harmony

#### Common or Usual Name

System, image management, ophthalmic

#### Classification Name

21 C.F.R. 892.2050 Picture and Archiving communications system

#### Predicate Device

Topcon Corporation Synergy ODM (K151952)

#### Intended Use / Indications for Use

Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye as well as in management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments

{4}------------------------------------------------

through either a direct connection with the instruments or through computerized networks.

## Device Description

Harmony is a modification to Synergy ODM cleared in K151952. The differences between the new version and the currently cleared version are modifications to the GUI using a responsive design and a change to the front end platform from Microsoft Silverlight to HTML5.

Harmony is a comprehensive software platform intended for use in importing, processing, measurement, analysis and storage of clinical images and videos of the eye, as well as for management of patient data, diagnostic data, clinical information, reports from ophthalmic diagnostic instruments through either a direct connection with the instruments or through computerized networks.

Harmony is used together with a number of computerized digital imaging devices, including:

- . Optical Coherence Tomography devices
- Mydriatic retinal cameras .
- Non-mydriatic retinal cameras .
- Biomicroscopes (slit lamps) ●

## Performance Data

No performance data was required or provided. Software validation and verification demonstrate that Harmony performs as intended and meets its' specifications.

## Substantial Equivalence

Harmony is substantially equivalent to Topcon's Synergy ODM cleared in K151952. Harmony has the same intended use and indications for use, similar principles of operation, and similar technological characteristics as the previously cleared predicate device. The intended use of the Harmony and intended use of the Synergy ODM cleared in K151952 are identical and both systems also have the identical indications for use statement.

Harmony has the same technological characteristics as the Synergy ODM (K151952). Both devices are software only image management systems which have the same acquisition, importing, viewing, measurement and analysis; network and security; print, archive and backup functionality.

Regarding acquisition. Harmony and the Synergy ODM (K151952) do not offer capture components. Both Harmony and Synergy ODM import digital 

{5}------------------------------------------------

images, patient data, diagnostic data and clinical information from other software capture systems and directly from ophthalmic devices.

Regarding importing, Harmony and the Synergy ODM (K151952) allow importing and management of files and images from a range of ophthalmic diagnostic devices including DICOM files, image files of known format, video images, and printer files.

Regarding viewing, Harmony and the Synergy ODM (K151952) allow standard viewing operations such as zoom in/out, panning, etc. and standard image enhancements such as contrast and brightness adjustment, and drawing tools

Regarding measurement and analysis. Harmony and the Synergy ODM (K151952) provide line and area measurement capabilities and provide Cup to Disc ratio and MPS (Macular Photocoagulation Study) measurements.

Regarding network and security, both Harmony and the Synergy ODM (K151952) provide web-based access to all data files. Harmony and the Synergy ODM (K151952) provide DICOM communication with other PACS.

Regarding printing, Harmony and the Synergy ODM (K151952) provide similar customizable print templates. Regarding archiving and backup, Harmony and the Synergy ODM (K151952) include archive and backup functionality. The operating system for review stations running each of the systems is Windows XP or above. Harmony can also be executed on an iPad while Synergy ODM (K151952) can be used on a review station with Mac OS X operating system. This minor difference does not impact substantial equivalence.

Modifications were made as the front end platform was changed from Microsoft Silverlight to HTML5 and the design of the GUI was changed to a responsive design. Software documentation is provided in Section VIII in support of these changes. It is not believed that any of these changes impact the substantial equivalence of the Harmony to the prior Synergy ODM.

In conclusion. Harmony shares similar technological characteristics with the predicate device, both in terms of the manner in which images are imported, analyzed, and stored, as well as the operation of the device by the intended user.

## Conclusion

Topcon's Harmony has the same intended use and indications for use as the previously cleared predicate device. In addition, Harmony has the same technological characteristics as its predicate. The modifications to the

{6}------------------------------------------------

cleared version of Synergy ODM are minor and do not impact the substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/NFJ/K182376](https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/NFJ/K182376)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
