K971679 · Ophthalmic Technologies, Inc. · MPA · Jun 25, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K971679
Device Name
I-SCOPE OPHTHALMIC ENDOSCOPE
Applicant
Ophthalmic Technologies, Inc.
Product Code
MPA · Gastroenterology, Urology
Decision Date
Jun 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Ophthalmic Technologies i-Scope™ is intended to be used for visualization of intra-ocular structures, as a means to introduce micro instruments into the eye, or as a means to introduce laser fibers into the eye for photocoagulation.
Device Story
The OTI i-Scope is a rigid fiber-optic microendoscope used for intra-ocular visualization and surgical access. The device features an imaging fiber bundle (10,000 pixels), an illumination fiber, and a working channel for micro-instruments or disposable laser fibers. The illumination bundle connects to a Xenon light source (e.g., OTI XE-LITE or other FDA-cleared sources). The imaging bundle connects to an orientation ring, video adapter, and camera. Used by clinicians in ophthalmic procedures to visualize structures and facilitate photocoagulation. The device provides direct optical visualization via video output, assisting the surgeon in navigating the eye and performing therapeutic interventions.
Clinical Evidence
No clinical data required; substantial equivalence supported by non-clinical bench testing including resolution and light output comparisons.
Technological Characteristics
Rigid fiber-optic microendoscope; 10,000-pixel imaging fiber bundle; illumination fiber; working channel for micro-instruments/laser fibers. Connects to external Xenon light source and video camera system. Materials are identical to those used in predicate devices.
Indications for Use
Indicated for visualization of intra-ocular structures, introduction of micro-instruments into the eye, and introduction of laser fibers for photocoagulation.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
ElectroFiber Optics Fiber Optic Endoscope
EndoOptiks Microprobe
Related Devices
K032598 — SYNERGETICS SYNERLIGHT FIBER OPTIC LIGHTSOURCE · Synergetics, Inc. · Nov 24, 2003
K971338 — SYNTEC, INC. DISPOSABLE BARE END FIBER · Syntec, Inc. · May 6, 1997
K973293 — SYNTEC TRUE LIGHT END IRRIGATING ENDOILLUMINATOR · Vaughan Weeks · Nov 18, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
# K971679
Ophthalmic Technologies i-scope 510(k) submission p. A - 3
JUN 2 5 1997
Ophthalmic Technologies Inc. 37 Kodiak Crescent Unit 16 Downsview, Ontario M3J 3E5 Canada 416-631-6932 Revised 6/16/96 510 (k) Summary
- 1. Submitter Information
a. Name: Ophthalmic Technologies Inc.
b. Address: Ophthalmic Technologies Inc. 30 Kodiak Crescent, Unit 16 Downsview, Ontario M3J 3E5, CANADA
- c. Telephone Number: 416-631-6932
- d. Contact Person: Dr. George Myers, 201-727-1703
- e. Date Summary Prepared: March 31, 1997
- 2. Device Information
- a. Trade or proprietary name: OTI i-scope ophthalmic endoscope .
- b. Common name: Ophthalmic Endoscope
- c. Classification name: Endoscope, Ophthalmic
### 3. Predicate Devices:
- a. ElectroFiber Optics Fiber Optic Endoscope
- b. EndoOptiks Microprobe
- 4. Description of the device
The Ophthalmic Technologies i-Scope™ is a rigid ophthalmic fiber-optic microendoscope. The i-Scope has an imaging fiber bundle with 10,000 pixels, an illumination fiber, and a working channel for an optional laser fiber or microinstruments. The laser fiber is disposable and can be
{1}------------------------------------------------
Ophthalmic Technologies i-scope 510(k) submission p. A - 4
replaced between cases. The fiber is not manufactured by Ophthalmic Technologies. The image fiber bundle terminal is connected to an orientation ring, video adapter and video camera. The illuminating bundle fiber can be connected to the OTI "XE-LITE" Xenon Light Source, or other light sources which have been cleared by the FDA. The laser fiber is disposable and can be replaced between cases.
#### 5. Intended Use
The Ophthalmic Technologies i-Scope™ is intended to be for visualization of intra-ocular structures, as a used means to introduce micro instruments into the eye, or as a means to introduce laser fibers for photocoagulation.
## 6. Technological Characteristics
a. Comparisons
The i-scope is comparable to the EndoOptiks Microprobe, an ophthalmic endoscope, and the ElectroFiber Optics Fiber Optic Endoscope, a general-purpose endoscope. Its intended use and specifcations are very close to that of the EndoOptiks device. Materials and construction are also similar to that used in the ElectroFiber Optics device.
b. Data
- (1) Non-Clinical Tests presented:
- (a) Resolution with test pattern
- (b) Light output compared to other devices
- (2) Clinical tests: Not required
(3) Materials: All materials used in this device have been used in predicate devices.
c. Conclusion:
The tests provided show that the OTI i-scope is as safe and effective, and perform as well as or better than the legally marketed devices identified above.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three wavy lines that resemble a stylized bird or human figure.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1100 - 5 1097
Ophthalmic Technologies, Inc. c/o Dr. George Myers Official Correspondent Medsys, Inc. 377 Route 17 South Hasbrouck Heights, NJ 07604
Re: K971679 Trade Name: i-SCOPE Ophthalmic Endoscope Regulatory Class: II Product Code: 86 MPA Dated: April 4, 1997 Received: May 7, 1997
Dear Dr. Myers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS)inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Dr. George Myers
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K971679
Ophthalmic Technologies i-scope 510(k) submission p. i - 4
> Page of
510(k) Number (if known):
- -
Device Name: i-scope Ophthalmic Endoscope_____________________________________________________________________________________________________________________________________
Indications for Use:
.
The Ophthalmic Technologies i-Scope™ is intended to be used for visualization of intra-ocular structures, as a means to introduce micro instruments into the eye, or as a means to introduce laser fibers into the eye for photocoagulation.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------|---------|
| Division of Ophthalmic Devices | |
| 510(k) Number | K971679 |
| Prescription Use (Per 21 CFR 810.109) | <span style="text-decoration: underline;">✓</span> |
|---------------------------------------|----------------------------------------------------|
|---------------------------------------|----------------------------------------------------|
OR
| Over-the-Counter Use | <span style="text-decoration: underline;"></span> |
|----------------------|---------------------------------------------------|
|----------------------|---------------------------------------------------|
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.