K963079 · Canon USA, Inc. · HKX · Mar 3, 1997 · Ophthalmic
Device Facts
Record ID
K963079
Device Name
CANON TX-10 TONOMETER
Applicant
Canon USA, Inc.
Product Code
HKX · Ophthalmic
Decision Date
Mar 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.
Device Story
TX-10 is a non-contact tonometer using air-puff technology to measure intraocular pressure. Device features auto-alignment system for measurement; includes auto and manual alignment modes. Operated by clinician in clinical setting. System includes power-assisted chin rest and 4-inch color LCD display. Output consists of intraocular pressure measurements, date, time, and patient ID. Device provides automated measurement, reducing need for special operator skills. Safety mechanism includes software-controlled stopper.
Clinical Evidence
Bench testing comparing TX-10 and T-2 performance. TX-10 demonstrated regression slope of 0.996, Y-intercept of 0.462 mmHg, and standard deviation of difference (Sd) of 1.56 mmHg. Results meet proposed ISO specifications (slope 0.95-1.05; intercept -1.0 to +1.0; Sd < 2.5 mmHg).
Technological Characteristics
Non-contact air-puff tonometer. Features auto-alignment system, power-assisted chin rest, and 4-inch color LCD. Software-controlled safety stopper. Measuring range 0-60 mmHg; 1 mmHg increment. Measuring time 3 milliseconds. LED fixation target.
Indications for Use
Indicated for measurement of intraocular pressure in human eyes.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
K023816 — CANON FULL AUTO TONOMETER TX-F · Canon, Inc.-Utsunomiya Optical Products Operation · Nov 26, 2002
K073087 — NON-CONTACT TONOMETER, MODEL FT-1000 · Tomey Corporation · Nov 18, 2008
K111710 — FULL AUTO TONOMETER MODEL TX-20 · Canon Inc. -Medical Equipment Group · Sep 20, 2011
K031733 — NON-CONTACT TONOMETER, MODEL NT-4000 · Nidek, Inc. · Dec 4, 2003
K223739 — VS Tabletop Tonometer · Medimaging Integrated Solution, Inc. · Nov 8, 2023
Submission Summary (Full Text)
{0}
03/03/97 16:58 FAX 316 328 5.52
104 10000000
Canon
K963079
MAR - 3 1997
CANON U.S.A., INC.
ONE CANON PLAZA
LAKE SUCCESS, NY 11042-1198
TELEPHONE: (516) 328-5000
GENERAL FAX TELEPHONE
(516) 328-5069
# 510(k) Summary
1. Date Prepared: July 23, 1996
2. Applicant's Name: Canon USA, Inc.
One Canon Plaza
Lake Success, NY 11042
U.S.A.
Mr. Glenn Impal / Mr. Ken Shadoff
Phone: (516) 328-5600
Fax: (516) 328-5169
3. Model Name: Canon TX-10 Tonometer
86HKX, Tonometer, AC Powered
Document Control Number: ?
4. Predicate Model: Canon T-2 Tonometer
86HKX, Tonometer, AC Powered
Document Control Number: K943939
5. Device Description: The TX-10 is a tonometer designed using a non-contact measurement system. Air puff gently measures the intraocular pressure with the help of an auto-alignment system. Features quick and easy measurement that does not require any special skills.
6. Intended Use: The TX-10 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.
7. Summary of Technological Characteristics: See the attached comparison chart.
8. Assessment of Performance Data: See Attachment 1
{1}
Comparison in the measurement accuracy among Canon TX-10 and Canon T-2
| | TX-10 | T-2 | ISO Specifications (proposed) |
| --- | --- | --- | --- |
| Regression Slope | 0.996 | 0.907 | 0.95 to 1.05 |
| Regression Y intercept | 0.462mmHg | 1.08mmHg | -1.0 to +1.0 |
| Standard Deviation of Difference, Sd | 1.56mmHg | 2.12mmHg | 2.5mmHg |
| Coefficient of Correlation | 0.926 | 0.858 | Unspecified |
{2}
K963079
# CANON TONOMETERS COMPARISON CHART
| MODEL | | CANON NON CONTACT TONOMETER T-2 | CANON TONOMETER TX-10 |
| --- | --- | --- | --- |
| Specification | Measuring Range Setting | 0~30/0~60mmHg
(Automatic) | Same as T-2 |
| | Alignment Method | One bright dot with alignment level display | **Auto Mode**
Monitor and auto fine alignment |
| | | | **Manual Mode**
Monitor with two dots |
| | Increment | 1 mmHg | Same as T-2 |
| | Measuring Time | 3 millisecond
(0.003 second) | Same as T-2 |
| | Fixation Target | LED (green) | Same as T-2 |
| | Chin Rest | Manual | Power assisted
(driving at controller) |
| Safety Mechanism | | Mechanical Stopper | Software Controlled Stopper |
| Printout Format | | R/L, date, time, avarage, ID, 3 or 10 data for each eye | Same as T-2 |
| Display | | 5" (inch) B/W CRT | 4" (inch) Color LCD |
| Dimentions
(width x depth x height) mm | 286 x 445 x 472 | **Body**
299 x 340 x 177 | |
| | | **Controller**
170 x 183 x 177 | |
| Weight (kg) | 22.0 | **Body**
16.0 | |
| | | **Controller**
1.5 | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches