The Discam® Digital Imaging System is intended to be used as an ophthalmic diagnostic tool, which is complimentary to other accepted diagnostic technologies, to track the progress of glaucoma, relative to monitoring changes in the appearance (size and shape) of the Optic Disc.
Device Story
Discam Digital Imaging System captures stereo images of the optic disc using a CCD stereo camera mounted on a table with an adjustable chin rest. Input consists of optical images of the fundus; operator initiates capture via joystick in 0.2 seconds. System interfaces with a PC (Windows 95) for image display, measurement, and archiving. Software enables basic analysis, including measurement of cup and disc size and cup-to-disc ratio. Used in clinical settings by eye care professionals to track glaucoma progression by monitoring optic nerve head changes. Output allows clinicians to visualize and quantify structural changes over time, aiding in clinical decision-making regarding glaucoma management.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing confirming power and luminance levels are within specifications and reliance on published literature regarding the safety and effectiveness of digital imaging systems for fundus and optic nerve head evaluation.
Technological Characteristics
CCD stereo camera; 12V 55W halogen illumination; 10° field of view; PC-based image capture and analysis; switch mode power supply (100-220V, 50-60Hz). Software provides measurement of cup and disc dimensions.
Indications for Use
Indicated for evaluation of the Optic Disc (Optic Nerve Head) to monitor progress in patients with Ocular Hypertension and Glaucoma.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
KOWA PROFESSIONAL FUNDUS CAMERA MODELS FX-500 (K954780)
{0}------------------------------------------------
#### AUG 17 1998
K974535
510(k) Summary of Safety and Effectiveness Information Marcher Enterprises, Ltd.
# DISCAM® DIGITAL IMAGING SYSTEM
## Applicant/Manufacturer
Marcher Enterprises, Ltd. Twyford Road, Unit 449 Rotherwas Industrial Estate Hereford HR2 6JR United Kingdom
Contact Name: Patrick Kirby
Tel: (44) 1432 278 400 Fax: (44) 1432 354218
Date of Summary Preparation: October 23, 1997
#### Device Name
Proprietary Name: Discam® Digital Imaging System Common Name: Ophthalmic Digital Camera Classification Name: Camera, Ophthalmic, AC-Powered
### Predicate Device
- · KOWA OPTIMED, INC .: KOWA PROFESSIONAL FUNDUS CAMERA MODELS FX-500 (K954780, cleared 12/01/95)
- · OPHTHALMIC IMAGING SYSTEMS, INC .: DFC-512 DIGITAL IMAGING SYSTEM (K913929, cleared 11/27/91)
- · LASER DIAGNOSTIC TECHNOLOGIES, INC .: TOPOGRAPHIC SCANNING SYSTEM (K923742, cleared 03/26/93)
- · HEIDELBERG ENGINEERING: HEIDELBERG RETINA TOMOGRAPH (K912891, cleared 09/27/91)
510(k) SUMMARY OF SAFETY AND EFF DISCAM® DIGITAL IMAGI
{1}------------------------------------------------
## Device Description
The Discam® Digital Imaging System consists of a CCD Stereo Camera that captures stereo images through an image capture board interfaced with a PC with the minimum requirements mages arough an RAM, 1 GB Hard Disc, 133 MHz CPU, Windows® 95 OS. The CCD stereo camera and housing is mounted on a standard rise and fall table, along with a standard adjustable patient chin rest.
The patient fixates on fixation lights, (fixed LEDs), and the operator views the image on the PC's monitor. Image capture, which is accomplished in 0.2 seconds, is initiated by pushing a iov stick.
Windows® based software allows basic analysis, (measurement of cup and Disc, cup to Disc ratio), of the captured images, using either our optional analysis software, or third party software with user developed algorithms, and data archiving.
The camera measures 370 mm Depth x 360 mm Height x 108 mm Width. It is powered by a switch mode power supply 100 V to 220 V, 50 Hz to 60 Hz.
Illumination is from a 12 Volt 55 Watt Halogen Lamp. There is no flash.
# Intended Use
The Discam® Digital Imaging System is intended to be used as an ophthalmic diagnostic tool, which is complimentary to other accepted diagnostic technologies, to track the progress of glaucoma, relative to monitoring changes in the appearance (size and shape) of the Optic Disc.
## Comparison of the Technological Characteristics of the Device to the Predicate.
The determination of the substantial equivalence of the Discam® Digital Imaging System is based upon the following:
- · Intended Use
- · Operational Characteristics
- · Technological Characteristics
A side by side comparison of the characteristics of the Discam® Digital Imaging System to the predicate devices listed above is provided in Table 3.
{2}------------------------------------------------
# TABLE 3: COMPARISON OF DEVICE
| | Marcher<br>Enterprises, Ltd<br>Discam® | Kowa Optimed<br>Fundus Camera<br>FX-500 | |
|---------------------------------------------------------------------------|----------------------------------------|-----------------------------------------|--|
| Intended Use: To view, capture and<br>archive images of the human fundus. | YES | YES | |
| Ability to measure changes in<br>Optic Disc/Cup size and shape | YES | YES (1) | |
| Halogen Illumination Source | YES | YES | |
| Confocal Laser Scanning | NO | NO | |
| PC Interface | YES | YES (1) | |
| CCD Camera | YES | YES (2) | |
| Maximum Field of View | 10° | 60° | |
Note (1): With Optional PC Interface and Software Note (2): With Optional 35 mm Film Based or Digital CCD
:
:
{3}------------------------------------------------
# CHARACTERISTICS
а сталогія страдае пасадае парадаецца парасміна такцыя выяванн
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
:
| ser Diagnostic<br>Technologies<br>TopSS | Heidelberg Eng.<br>Retina<br>Tomograph | Ophthalmic<br>Imaging Systems<br>DFC-512 |
|-----------------------------------------|----------------------------------------|------------------------------------------|
| YES | YES | YES |
| YES | YES | YES |
| NO | NO | YES |
| YES | YES | NO |
| YES | YES | YES |
| NO | NO | YES |
| 30° | 30° | 60° |
r
Table 3
e la personal programments on the comments of the comments of the summer a AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AN
{4}------------------------------------------------
## Testing
- · Non-Clinical Testing
The Discam® Digital Imaging System was tested to assure that the power and luminance levels delivered in every mode of operation were within specification.
- · Clinical Testing
Published data indicate that viewing and capturing images of the Fundus and Optic Nerve Head/Optic Disc with Digital Imaging Systems is safe and effective.
- Conclusions Drawn
Marcher Enterprises Ltd. believes that the Discam® Digital Imaging System is substantially equivalent to those predicate devices cited above, that any differences are minor, and that these differences do not raise any issue of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 4 1998
Tim Dawson Development Director Marcher Enterprises LTD Unit 2, Clearview Court, Twyford Road Rotherwas, Hereford HR2 6JR UK
Re: K974535 Trade Name: Discam Digital Imaging System Regulatory Class: II Product Code: 86 HKI Dated: July 20, 1998 Received: July 22, 1998
Dear Mr. Dawson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{6}------------------------------------------------
Page 2 - Tim Dawson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Discam® Digital Imaging System
#### Indications For Use:
The Discam® Digital Imaging System is intended to be used to evaluate the Optic Disc, (Optic Nerve Head). It is used in the same fashion as a Fundus camera, except that it's field of view is narrower, in order to focus exclusively on the Optic Disc, (Optic Nerve Head). Evaluation of changes in the size and shape of the Optic Disc, (Optic Nerve Head), is useful in monitoring the progress of patients with Ocular Hypertension, and Glaucoma.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY ) ャー・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Everett Been
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K974535