K243664 · Ai Optics · HKI · Dec 17, 2024 · Ophthalmic
Device Facts
Record ID
K243664
Device Name
Sentinel Camera
Applicant
Ai Optics
Product Code
HKI · Ophthalmic
Decision Date
Dec 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Sentinel Camera is a non-mydriatic medical digital camera that is intended to capture digital images of the human eye.
Device Story
Sentinel Camera is a non-mydriatic digital ophthalmic camera used in medical environments for documentation and consultation. Device captures retinal images using infrared light for targeting and white light for image acquisition; features three internal fixation targets (middle for macula-centered, left/right for disc-centered images). Operated by healthcare professionals; images stored in internal memory and transferred via Wi-Fi to AI Optics Server. Powered by rechargeable Li-ion battery. Benefits include convenient examination without pupil dilation; provides high-resolution documentation for clinical decision-making.
Clinical Evidence
Bench testing only. Performance data demonstrates compliance with recognized standards including IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 15004-1, ANSI Z80.36, ISO 10940, and ISO 10993-1 (biocompatibility). No clinical data presented.
Technological Characteristics
Non-mydriatic ophthalmic camera; 5 MP color CMOS sensor; 4.0" TFT-LCD display (720x720); illumination via white and IR LEDs; diopter compensation -20 D to +15 D; 50x40 degree FOV. Connectivity: Wi-Fi. Power: 3.65V 5200 mAh Li-ion battery. Standards: IEC 60601-1:2005+A2:2020, IEC 60601-1-2:2014+A1:2020, ISO 15004-1:2020, ANSI Z80.36:2016, ISO 10940:2009, NEMA PS3.1-30.20:2021, IEC 62304:2015, ISO 10993-1:2018.
Indications for Use
Indicated for capturing digital images of the human eye in patients of any age or gender. Intended for use without mydriasis, though mydriatic drops may be used if necessary.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2024
AI Optics % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K243664
Trade/Device Name: Sentinel Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: November 27, 2024 Received: November 27, 2024
## Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243664
Device Name Sentinel Camera
Indications for Use (Describe)
Sentinel Camera is a non-mydriatic medical digital camera that is intended to capture digital images of the human eye.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for AI Optics. The logo consists of a circular icon to the left of the text "AI Optics". The icon is a partial circle with lines radiating outward from the open side of the circle. The text "AI Optics" is in a sans-serif font and is aligned horizontally with the icon.
Page 1 of 5
# Sentinel Camera 510(k) Summary (K243664)
| 510(k) Submitter: | AI Optics |
|----------------------|------------------------------------------------------------------------------------------|
| Address: | 145 Lexington Ave<br>Floor 10<br>New York, New York 10016 |
| Company Phone: | 914-874-3256 |
| Contact Person: | Andrew DiGiore |
| Date: | December 16, 2024 |
| Subject Device: | |
| Trade Name: | Sentinel Camera |
| 510(k) Number: | K243664 |
| Common Name: | Ophthalmic Camera |
| Classification Name: | Camera, Ophthalmic, AC-Powered<br>(21 CFR 886.1120) |
| Regulatory Class: | II |
| Product Code: | HKI |
| Predicate Device: | |
| Trade Name: | Optomed Aurora Camera<br>Optomed Aurora Retinal Module<br>Optomed Aurora Anterior Module |
| 510(k) Number: | K180378 |
| Regulatory Class: | II |
| Product Code: | HKI |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AI Optics. The logo consists of a blue circle that is partially filled with blue lines radiating from the center. To the right of the circle are the words "AI Optics" in blue sans-serif font.
# Description of the Device
The Sentinel Camera is accompanied by the following accessories: a charging base, eye cup, and power cable.
The Sentinel Camera is designed for use in a medical environment. Captured images are used for documentation and consultation. Sentinel Camera has internal memory where captured images are stored.
The Sentinel Camera is designed for non-mydriatic retinal imaging. In non-mydriatic imaging no mydriasis is needed because infrared light is used for targeting the retina and white light is flashed when an image is taken. The pupil does not respond to the infrared light, so examination is convenient for the patient. With small pupils, it is recommended to use mydriatic drops. The Sentinel Camera has three internal fixation targets for the patient to fixate on during imaging. The middle fixation target provides a macula-centered image. The left and right fixation targets provide disc-centered images.
The transfer of images to the AI Optics Server is carried out via Wi-Fi communication.
The Sentinel Camera has a rechargeable Li-ion battery that is charged when the camera is placed on charging base, which is connected to the mains by the power cable.
#### Indications for Use
Sentinel Camera is a non-mydriatic medical digital camera that is intended to capture digital images of the human eye.
#### Comparison of Technological Characteristics
The technological characteristics of the Sentinel Camera closely match those of the Optomed Aurora Camera Optomed Aurora Retinal Module Optomed Aurora Anterior Module (K180378). Both devices utilize similar imaging technology, including a high-resolution camera and light sources for illumination. The slight variations in specifications does not raise new questions of safety and effectiveness.
Table 1 below includes a summary of the technical information used in the substantial equivalence discussion.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for AI Optics. The logo consists of a circular graphic on the left and the text "AI Optics" on the right. The circular graphic is a partial circle made up of several lines, and the text is in a sans-serif font.
# Sentinel Camera Traditional 510(k) Premarket Submission Page 3 of 5
# Table 1: Substantial Equivalence
| Characteristic | Subject Device | Predicate Device |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sentinel Camera | Optomed Aurora Camera<br>Optomed Aurora Retinal Module<br>Optomed Aurora Anterior Module |
| 510(k) Number | K243664 | K180378 |
| Classification | HKI | HKI |
| Device Class | Class II | Class II |
| CFR Section | 21 CFR 886.1120 | 21 CFR 886.1120 |
| Common Name | Ophthalmic camera | Ophthalmic camera |
| Indication for Use | Sentinel Camera is a medical digital<br>camera that is intended to capture<br>images of the eye. | Optomed Aurora Camera is a medical<br>digital camera that is used with<br>dedicated optics modules intended to<br>capture images and videos of fundus of<br>the eye and surface of the eye.<br>Optomed Aurora Camera with Optomed<br>Aurora Retinal Module is intended to<br>capture digital images and videos of the<br>human eye.<br>Optomed Aurora Camera with Optomed<br>Aurora Anterior Module is intended to<br>capture digital images and video of the<br>surface of the human eye and<br>surrounding areas. |
| Usage | Prescription Use | Prescription Use |
| Use Condition | Intended to be used without mydriasis<br>but can be used also with mydriatic<br>drops. | Intended to be used without mydriasis<br>but can be used also with mydriatic<br>drops. |
| Patient Population | System is not patient population<br>specific. | System is not patient population<br>specific. |
| Illumination Source | • White: OSRAM Oslon Compact PL<br>• IR: Roithner Lasertechnik SMB1N<br>• Target LEDs: Kingbright | Aurora Retinal Module:<br>• White: OSRAM Oslon LUWH9GP<br>• NIR: OSRAM Oslon SFH-4716<br>• Target LEDs: Vishay VLMS1500-GS08<br><br>Aurora Anterior Module:<br>• White: OSRAM Advanced Power Topled LW G6SP-EAFS-JKQL-1<br>• Blue: OSRAM Advanced Power Topled LB G6SP-V2BB-35-1 |
| Display System | 4.0", TFT-LCD, 720x720 pixels | 4.0", TFT-LCD, 800x480 pixels |
| Camera Sensor | Color CMOS camera maximum resolution 5 MP | Color CMOS camera maximum resolution 5 MP |
| Diopter<br>Compensation | From -20 D to +15 D | From -20 D to +20 D |
| Field of View | 50 x 40 degrees | 50 x 40 degrees |
| Storage Media | Internal storage | MicroSDHC memory card |
| Image Data Format | PNG | JPEG, MPEG-4 |
| Weight | 900 g | Aurora Camera: 514g<br>Aurora Retinal Module: 310g<br>Aurora Anterior Module: 105g |
| Battery | Rechargeable Li-Ion battery, 3.65 V, 5200 mAh | Rechargeable Li-Ion battery, 5000065, 3.63 V, 2600 mAh |
| Data Transfer | Data transfer via Wi-Fi | Data transfer via USB (1.1) |
| Standards | • IEC 60601-1:2005+A2:2020 (Edition 3.2)<br>• IEC 60601-1-2: 2014 +A1: 2020 (Edition 4.1)<br>• ISO 15004-1:2020<br>• ANSI Z80.36:2016<br>• ISO 10940:2009<br>• NEMA PS3.1-30.20: 2021<br>• IEC 62304:2015<br>• ISO 10993-1:2018<br>• NEMA PS 3.1 - 3.20 2022 | • IEC 60601- 1:2005+A1:2012 (Edition 3.1)<br>• IEC 60601-1-2:2014 (Edition 4.0)<br>• IEC 60601-1- 6:2010+A1:2013 (Edition 3.1)<br>• IEC 62471:2006<br>• ISO 15004-1:2006<br>• ISO 15004-2:2007<br>• ISO 10940:2009<br>• IEC 62304:2006 +A1:2015 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for AI Optics. The logo consists of a circular icon on the left and the text "AI Optics" on the right. The circular icon is a stylized representation of a lens or aperture, with a series of lines radiating outward from the center. The text "AI Optics" is written in a sans-serif font, with the letters "AI" slightly larger than the word "Optics". The logo is a simple and modern design, and the use of the color blue gives it a professional and trustworthy feel.
Page 4 of 5
## Performance Data
Performance data for the Sentinel Camera, including extensive testing against recognized standards, supports the substantial equivalence of the Sentinel Camera to the predicate device. Both devices underwent similar testing to external standards. The testing demonstrates that the Sentinel Camera meets
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for AI Optics. The logo consists of a circular icon on the left and the text "AI Optics" on the right. The circular icon is a partial circle with lines radiating outward from the center, resembling a stylized lens or aperture. The text "AI Optics" is written in a sans-serif font and is aligned horizontally with the circular icon.
Page 5 of 5
the necessary safety and performance criteria, with no new questions of safety or effectiveness raised by the differences in technological characteristics, therefore supporting the conclusion of substantial equivalence.
### Conclusion
Based on the provided performance data and the comparison, Sentinel Camera is as safe, as effective, and performs as well as or better than the predicate device. Sentinel Camera is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.