← Product Code [HKI](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HKI) · K170527

# Phoenix Clinical ICON (K170527)

_Phoenix Technology Group, Inc. Dba Phoenix Clinical · HKI · Sep 15, 2017 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HKI/K170527

## Device Facts

- **Applicant:** Phoenix Technology Group, Inc. Dba Phoenix Clinical
- **Product Code:** [HKI](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HKI.md)
- **Decision Date:** Sep 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1120
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

General ophthalmic imaging including retinal, corneal, and external structures of the eye.

## Device Story

Phoenix Clinical ICON is a cart-based, handheld, wide-field, high-resolution ophthalmic imaging system. It captures retinal, corneal, and external eye images using an interchangeable LED-based light source (white or blue) integrated into the handpiece. The device requires direct contact between the glass lens tip and the patient's cornea, typically using coupling agents like Goniosol or GenTeal. Operated by clinicians, nurses, or technicians in hospitals, clinics, or offices, the system captures video or still images, which are stored on the cart-based computer for review. The device facilitates clinical assessment of ocular structures; output is viewed by healthcare providers to support diagnostic decision-making. Connectivity to IT networks is supported under supervision.

## Clinical Evidence

Bench testing only. Evaluated per ISO 10940:2009 for image quality and ISO 15004-2:2007 for light safety (Group 2 compliance). Retinal imaging performance compared to predicate using vascular edges as resolution targets via ImageJ analysis. A panel of five clinicians performed a qualitative comparison of image quality across four parameters, confirming equivalence.

## Technological Characteristics

Cart-based system with handheld camera; 10W white (450-675nm) or blue (450-460nm) LED light sources. Contact lens interface; uses Goniosol/GenTeal. 100-degree field of view; 12-micron resolution. Connectivity: IT network capable. Power: Battery and AC. Software-based image capture and storage.

## Regulatory Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- RetCam 3 ([K102859](/device/K102859.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a way that they also resemble a bird-like shape.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2017

Phoenix Technology Group, Inc. Dba Phoenix Clinical Christopher A. Henderson Director of Regulatory Affairs and Ouality Assurance 6630 Owens Drive Pleasanton, CA 94588

Re: K170527

Trade/Device Name: Phoenix Clinical Icon Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: August 15, 2017 Received: August 16, 2017

Dear Christopher A. Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Bradley S. Cunningham -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K170527

Device Name Phoenix Clinical ICON

Indications for Use (Describe)

General ophthalmic imaging including retinal, corneal, and external structures of the eye.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## TECHNOLOGY GROUP

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92 510(k).

| Applicant:              | Phoenix Technology Group, Inc.<br>6630 Owens Drive<br>Pleasanton, CA 94588                                                                                 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person          | Christopher Henderson<br>Director of QA and RA<br>Phone: (925) 462-5000 x204<br>Facsimile: (925) 485-1155<br>E-Mail: chenderson@phoenixtechnologygroup.com |
| Date Prepared:          | September 13, 2017                                                                                                                                         |
| Trade Name(s):          | Phoenix Clinical ICON, Phoenix ICON                                                                                                                        |
| Common Name:            | Ophthalmic Imaging System, Retinal Camera, Fundus Camera                                                                                                   |
| Classification:         | II                                                                                                                                                         |
| Classification Name(s): | Ophthalmic Camera                                                                                                                                          |
| Regulation Number:      | 886.1120                                                                                                                                                   |
| Product Code:           | HKI                                                                                                                                                        |
| Classification Panel:   | Ophthalmic                                                                                                                                                 |
| Predicate Device(s):    | RetCam 3 K102859                                                                                                                                           |

#### General Information

#### Device Description:

The Phoenix Clinical ICON is a battery and AC-powered, cart-based, non-sterile, wide-field, handheld, high resolution, real-time retinal imaging device. It is intended to be used for general ophthalmic imaging including retinal, corneal, and external structures of the eye. The intended users of the Phoenix Clinical ICON are clinical imaging technicians, ophthalmic technicians, nurses, and physicians. The Phoenix Clinical ICON may be used in hospitals, medical clinics, and physician's offices. The Phoenix Clinical ICON is a system comprised of a hand-held camera with an interchangeable LED based light source incorporated into the hand piece that is attached to the cart contains the battery, the keyboard controls, monitor, attached foot pedal, and computer with software. The hand piece is designed such that the glass lens in the tip will come into contact with the cornea. LED light is emitted into the eye to illuminate the retina for image capture. The Phoenix ICON is operated by attaching the appropriate module (white or blue LED) to the handpiece and, after the software is running, placing the camera lens in direct contact with the patient's cornea. The Phoenix Clinical ICON software can capture either video or still images and store them on the cart computer for later review. The Phoenix Clinical ICON may be connected to, under IT supervision, IT networks.

{4}------------------------------------------------

## PHOENIX TECHNOLOGY GROUP

### Substantial Equivalence Summary of the ICON and the predicate device:

The design of the Phoenix ICON is substantially equivalent to the design of the predicate device. The methods of action and use are identical and the user functions (handpiece, software, procedure of image obtaining) are all identical in functionality between the ICON and the predicate. The intended use of both devices being "General Ophthalmic Imaging of retinal, corneal, and external structures" is identical. The proposed devices are substantially equivalent to the predicate devices in regards to its intended use, design, technology, and materials.

| Device                            | K102859(Predicate)                                 |                                                                                                        | K170527                        |                                                                                                              |
|-----------------------------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------------------------------------------------------------------------------|
| Features                          | White Light Module                                 | FA Module                                                                                              | White Light Module             | FA Module                                                                                                    |
| Observation Light Source          |                                                    |                                                                                                        |                                |                                                                                                              |
| Light source type, wavelength     | White light LED,                                   | Blue light LED                                                                                         | 10W White light LED, 450-675nm | 10W Blue light LED, 450-460nm                                                                                |
| Maximum Light source output power | 21 mW/cm^2                                         | 85 mW/cm^2                                                                                             | 6.1 mW/cm^2                    | 28 mW/cm^2                                                                                                   |
| Light intensity control           | Zero to maximum                                    | Zero to maximum                                                                                        | Zero to maximum                | Zero to maximum                                                                                              |
| External fixation light (if any)  | none                                               | None                                                                                                   | none                           | none                                                                                                         |
| Camera                            |                                                    |                                                                                                        |                                |                                                                                                              |
| Field of View                     | 30 to 130 degrees                                  | 30 to 130 degrees                                                                                      | 100 degrees                    | 100 degrees                                                                                                  |
| Resolution                        | Varies with lens between 13 microns and 47 microns | Varies with lens between 13 microns and 47 microns                                                     | Single lens at 12 microns      | Single lens at 12 microns                                                                                    |
| Frame Rate                        | 30 frames per second                               | 30 frames per second                                                                                   | 30 frames per second           | 30 frames per second                                                                                         |
| Other Accessories                 |                                                    |                                                                                                        |                                |                                                                                                              |
| Insert filter                     | No insert filter needed.                           | Change light source at cart and remove objective lens and insert yellow barrier filter, filters >510nm | No insert filter needed.       | Use blue LED module and flip filter on hand piece, barrier filter blocking band 500nm, edge wavelength 515nm |
| Imaging lens                      | Flat field inserted                                | N/A                                                                                                    | Flat field external camera     | N/A                                                                                                          |
| Eye Contact Materials             | Goniosol or GenTeal                                | Goniosol or GenTeal                                                                                    | Goniosol or GenTeal            | Goniosol or GenTeal                                                                                          |

### Discussion of key technological differences between the ICON and the predicate:

{5}------------------------------------------------

## TECHNOLOGY GROUP

| Performance           |                                                          |                                                          |                              |                              |
|-----------------------|----------------------------------------------------------|----------------------------------------------------------|------------------------------|------------------------------|
| Imaging<br>Format     | .TIF/.JPEG/.AVI/.BMP                                     | .TIF/.JPEG/.AVI/.BMP                                     | .TIF/.JPEG/.AVI              | .TIF/.JPEG/.AVI              |
| Imaging<br>Resolution | Varies with lens<br>between 13 microns<br>and 47 microns | Varies with lens<br>between 13 microns<br>and 47 microns | Single lens at 12<br>microns | Single lens at 12<br>microns |

There are several key technological differences noted above in the Phoenix Clinical ICON system:

- Changes in light sources: Optical radiation hazard evaluation was performed that these changes will not affect the device safety.
- Changes in optical system, accessories and imaging: The image quality testing results were supportive of substantial equivalence to the predicate device.
- -The predicate device has an LED source that is delivered to the hand piece that is generated on the cart where the application device has the LED source located in the hand piece that allows for the swapping of modules from white light to blue light for fluorescein angiography. This change has no affect in regards to safety or effectiveness and is merely a function of usability.

### Indications for Use:

General ophthalmic imaging including retinal, corneal, and external structures of the eye.

### Bench Testing Summary:

### Testing per ISO 10940:2009:

ISO 10940:2009 - Ophthalmic instruments: fundus cameras was utilized as the first step for validation of substantial equivalence of image quality. The results were supportive of substantial equivalence to the predicate device.

### Light safety testing per ISO 15004-2:

The Phoenix Clinical ICON was tested according to 15004-2:2007 to determine acceptable light safety limits for both the white and blue light. The test results demonstrate the ICON is in compliance with the of Group 2 instrument requirements provided by the standard.

### Retinal imaging testing compared to predicate device:

Using vascular edges as a resolution target and the NIH analytical tool ImageJ. The images from the predicate device and the ICON were presented to a panel of five clinicians for comparison in four parameters of image quality.

The entire retina can be imaged as a set of multiple images with both cameras. In the figure below it can be seen that the ICON can very adequately cover the retina with six images from the RetCam.

{6}------------------------------------------------

# PHOENIX TECHNOLOGY GROUP

Image /page/6/Figure/2 description: The image shows two diagrams comparing the field of view of two different camera systems. The diagram on the left, labeled "ICON six images", shows six overlapping blue circles within a square grid. The diagram on the right, labeled "RetCam 130 five images", shows five overlapping gray circles within a similar grid. Both diagrams include a dashed red circle encompassing the overlapping circles, and the grids are labeled with degree measurements from -90 to +90.

Angle measured from the center of the eye

The resolution results were supportive of substantial equivalence to the predicate device.

### Conclusion:

The bench performance tests and qualitative comparison of the retinal images support our conclusion of substantial equivalence of Phoenix Clinical ICON to the RetCam predicate.

### Company Contact:

Christopher Henderson Director of Quality Assurance and Regulatory Affairs Phone: (925) 485-1100 x204 E-Mail: chenderson@phoenixtechnologygroup.com

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HKI/K170527](https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HKI/K170527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
