The Keeler Slit Lamp is an AC-powered Slit lamp biomicroscope and is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment This device is intended to be used only by suitably trained and authorised healthcare professionals.
Device Story
AC-powered slit lamp biomicroscope; projects thin, intense light beam into patient eye via control diaphragm. System includes XYZ translation base, illumination unit, and observation system (Galilean converging binoculars). Patient positioned with chin and forehead rests; clinician uses control lever to focus slit on cornea. Illumination system features adjustable filters, graticules, and rheostat for intensity control. Magnification body allows selection of multiple magnification levels. Used in clinical settings by trained healthcare professionals to visualize anterior eye segment structures; aids diagnosis of ocular disease or trauma.
Clinical Evidence
Bench testing only. Device evaluated for compliance with BS EN ISO 15004-2 (radiation hazards), BS EN ISO 10939, IEC 60601-1 (electrical safety), and IEC 60601-1-2 (electromagnetic compatibility).
Technological Characteristics
AC-powered biomicroscope; Galilean converging binoculars; 6V 20W halogen or LED light source; XYZ translation base; adjustable slit width/length (0-12mm); magnification range 6x-40x. Complies with ISO 10939, ISO 15004-2, IEC 60601-1, and IEC 60601-1-2. Permanently installed IR heat-absorbing filter.
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties. For use by trained healthcare professionals.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K131589
## Ophthalmic Instruments
# 3. 510(k) Summary
# 3.1 Submitter's information
The submitter of this pre-market notification is:
Mr. Nickie Power (Quality Manager). Keeler limited Clewer hill Road, Windsor, Berkshire, SL4 4AA, UK Company Phone No: +44 (0) 1753 827153 Company Fax No: +44 (0) 1753 827145 Contact Person: Mr. Nickie Power
Date summary prepared: January 14, 2013
### 3.2 Device Identification
| Device Trade Name: | Keeler Slit Lamp - H Series |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | AC Powered Slit lamp Biomicroscope |
| Class: | ============================================================================================================================================================================== |
| Classification Panel: | 86 |
| Product Code: | HJO |
| Regulation Number: | 886.1850 |
# 3.3 Predicate Device
| Predicate Device: | Slit Lamp BQ 900 |
|-----------------------|------------------------------------|
| 510(k) No: | K100202 |
| Manufacturer: | Haag-Streit AG |
| Common Name: | AC Powered Slit lamp Biomicroscope |
| Class: | II |
| Classification Panel: | 86 |
| Product Code: | HJO |
| Regulation Number: | 886.1850 |
510K Keeler H-Series Slit Lamp
Project 210
Issue B
JUN 2 7 2013
Page 5 of 53
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the words "Keeler Ophthalmic Instruments". The word "Keeler" is in a larger, bold font, and the words "Ophthalmic Instruments" are in a smaller font below it. The text is black and the background is white.
### 3.4 Device description
An AC-powered slit lamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into the patient's eye through a control diaphragm a thin, intense beam of light.
This Keeler Slit Lamp has an XYZ translation base that is either mounted onto a custom table top supplied by Keeler or can be mounted on a third parties table top (refraction unit) by suitably trained technicians.
Fitted to the XYZ base is the illumination and observation system; fitted to the table top is the chinrest assembly with fixation target.
The illumination system rotates around its mounting axis and has a number of filters and graticules allowing the user to control precisely the characteristics of the examination light. The magnification body has a magnification change systems and eyepieces adjustable for the user's individual pupillary distance and distance correction.
The patient is seated in front of the slit lamp with his/her chin in the chin rest and forehead against the forehead rest. The chin rest can be adjusted in height to align the eyes of the patient with the optics/light beam.
With the control lever the instrument can be moved back and forward until the slit appears in focus on the cornea. The image can be observed through the microscope. The base unit has a rheostat to control the illumination level.
Various magnifications can be selected on the microscope. For different observations the slit width can be changed, the slit can be tilted horizontally, and the angle between the illumination unit and the microscope can also be varied horizontally.
### 3.5 Indications for Use
The Keeler Slit Lamp is an AC-powered Slit lamp biomicroscope and is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment
This device is intended to be used only by suitably trained and authorised healthcare professionals.
Project 210
Issue B
{2}------------------------------------------------
Keeler
### Ophthalmic Instruments
# 3.6 Comparison of the device with the Predicate
The Haag-Streit BQ900 has been chosen as the substantially equivalent device, it has the same intended use and classification.
The Keeler Slit Lamp has the following similarities to those of the predicate device:
-has the same indication for use,
-uses the same operating principle,
-incorporates the same basic optical design.
Refer to the table below for comparison between Keeler and Haag Streit BQ 900 Slit Lamps (K100202).
| Device | Predicate Device<br>Haag-Streit Slit Lamp<br>BQ 900 | Keeler Slit Lamp<br>H- Series | Discussion |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | An ac-Powered slit lamp<br>biomicroscope is an AC-<br>powered device that is a<br>microscope intended for use<br>in eye examination that<br>projects into the patient's eye<br>through a control diaphragm<br>a thin, intense beam of light | An ac-Powered slit lamp<br>biomicroscope is an AC-<br>powered device that is a<br>microscope intended for use<br>in eye examination that<br>projects into the patient's eye<br>through a control diaphragm<br>a thin, intense beam of light | Same |
| Microscope | Galilean converging<br>binoculars | Galilean converging<br>binoculars | Same |
| Type | AC Powered Biomicroscope<br>Slit Lamp | AC Powered Biomicroscope<br>Slit Lamp | Same |
| Angle of<br>Convergence | 13° | 13° | Same |
| Magnification | 6.3x / 10x / 16x / 25x / 40x | x6, x10, x16, x25 & x40 | The difference are not<br>significant and does not<br>affect the safety or<br>effectiveness of the<br>device |
| Objective lens<br>working distance | 105mm | 107mm | The difference are not<br>significant and does not<br>affect the safety or<br>effectiveness of the<br>device |
| Field of view | 32.0 @ 6.3x / 20.0 @ 10x /<br>12.7@ 16x / 8.0@ 25x /<br>5.1@ 40x | 34.0 @ 6x / 22.0 @ 10x / 14<br>@ 16x / 8.5 @ 25x / 5.5 @<br>40x | The difference are not<br>significant and does not<br>affect the safety or<br>effectiveness of the<br>device |
| Range of PD<br>adjustment | 52 - 78 mm | 49.0 to 77mm | There is a difference in<br>the range of adjustment<br>however this does not<br>affect safe or<br>effectiveness on the<br>devices as both<br>conform to the<br>requirement of ISO<br>10939 |
| Eyepiece dioptric<br>adjustment<br>range | + 8 to - 8 diopters | + 8 to - 8 diopters | same |
| Slit Projection<br>Unit | | | |
| Slit Length | 1-8 mm continuous | 12mm (1mm - 12mm<br>continuously variable) | The difference is not<br>significant<br>An Iris usually has a<br>diameter of 6mm max,<br>therefore a slit length of<br>12mm is more than<br>sufficient, meeting the<br>requirement specified<br>in ISO 10939 and does<br>not affect the safety or<br>effectiveness of the<br>device |
| Slit Width | 0 - 8 mm continuous | 0-12mm continuously<br>variable | The difference is not<br>significant; this does<br>not affect safety or<br>effectiveness of the<br>devices as both<br>conform to the<br>requirement of ISO<br>10939.<br>Slit lamps are used to<br>examine the inner part<br>of the eye and usually<br>a narrow slit of less<br>than 1mm is needed.<br>For this reason the<br>maximum slit width is<br>not critical as long as it<br>is ≥ 8mm |
| Light source | 7.5V, 38 W Halogen Lamp<br>24VDC 1A LED | 6V 20W Halogen lamp and<br>LED - As the slit lamp<br>conform to the relevant<br>standard<br>ISO 10939, there are no<br>issues of safety and<br>effectiveness. | There is a difference in<br>light sources used<br>however all light<br>sources must conform<br>to the requirements of<br>15004-2 Photo-toxicity<br>therefore there is no<br>affect on safety or<br>effectiveness |
| Duration of<br>illumination | Maximum examination times<br>according to ISO 15004-2<br>and ISO 10939 | Maximum examination times<br>according to ISO 15004-2<br>and ISO 10939 | Same |
| Light intensity<br>selection | Three selections: Full, 1/4<br>and 1/16 via neutral density<br>filter. | 0-100% continuous<br>potentiometer | The difference is not<br>significant as they allow<br>the user to select the<br>necessary light<br>intensity to perform the<br>relevant task |
| Filter | Blue, red free (green), grey<br>(10%)<br>The UV filter and the heat<br>absorption filter are<br>permanently mounted | Clear; red free; neutral<br>density; diffuser; blue; IR<br>heat absorbing filter<br>permanently installed | The difference is not<br>significant and does not<br>affect safety or<br>effectiveness, both<br>devices have a<br>permanently installed<br>heat absorbing filter |
| Electrical<br>ratings | | | |
| Input voltage | AC 100-240 V, 50/60Hz | AC 100-240 V, 50/60Hz | Same |
| Power output | 30VA | 30VA | Same |
| Output Voltage | 12 V DC | 12 V DC | Same |
| Compliance<br>with Safety<br>standards | IEC60601-1-2<br>IEC60601-1<br>ISO 15004-2 | IEC 60601-1-2<br>IEC 60601-1<br>ISO15004-2 | Same |
510K Keeler H-Series Slit Lamp
Issue B
{3}------------------------------------------------
# Keeler
# Ophthalmic Instruments
510K Keeler H-Series Slit Lamp
,
・
:
Project 210
.
100 - 100
Issue B
Page 8 of 53
{4}------------------------------------------------
### Ophthalmic Instruments
## 3.7 Summary of Performance, EMC and Safety Testing
The Keeler Slit Lamp (H - Series) was evaluated against the requirements of BS EN ISO 15004-2 and BS EN ISO 10939 for radiation hazards, to IEC 60601-1 for electrical safety and to IEC 60601-1-2 for electromagnetic compatibility.
In all tests the slit lamp was in compliance with these FDA recognized standards.
# 3.8 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this Premarket Notification, Keeler limited concludes that the H- Series Slit Lamp is safe and effective, and substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
June 27, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Keeler Limited % Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories Inc. I 2 Laboratory Drive Research Triangle Park, NC 27709
Re: K131589
Trade/Device Name: Keeler Slit Lamp - H Series Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slit Lamp Regulatory Class: Class II Product Code: HJO Dated: June 4, 2013 Received: June 13, 2013
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includions for use stated in the enclosure) to legally marketed predicated vices marked in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drig. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbrandi, nsand adulteration. Please note: CDRH does not evaluate information related to contractling allie warranties. We remind you, however, that device labeling must be truthful and not miolesaling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDA coul publish further announcements concerning your device in the Federal Register
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You most comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Mr. Jeff D. Rongero
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor Yow/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D.
Director
Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Keeler
Ophthalmic Instruments
### 2. Indications for Use Form
510(k) Number (if known):
Device Name: Keeler Slit Lamp H-Series.
Indications for Use:
The Keeler Slit Lamp is an AC-powered Slit lamp biomicroscope and is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment
This device is intended to be used only by suitably trained and authorised healthcare professionals.
Prescription Use _v (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------|-------------------------------------------------------------------------------------------|
| | <img alt="Signature" src="signature.png"/> Andrew Yang -5 2013.06.27 16:08:34 -04'00' |
| 510K Keeler H-Series Slit Lamp | Project 210 |
| | Division of Ophthalmic and Ear, Nose and Throat Devices |
| | 510(k) Number K131589 |
| | Issue B |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.