← Product Code [HJO](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HJO) · K072861

# CEEMAX SLIT LAMP (HALOGEN LAMP SERIES/SOLID STATE ILLUMINATION SERIES) (K072861)

_Acheve Technology, Inc. · HJO · Jun 24, 2008 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HJO/K072861

## Device Facts

- **Applicant:** Acheve Technology, Inc.
- **Product Code:** [HJO](/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HJO.md)
- **Decision Date:** Jun 24, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1850
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Intended Use

The CeeMax™ Slit Lamp (Halogen Lamp Series/LED Illumination series) is an AC powered microscope and accessories intended for the use in the examination of the anterior eye segment, from cornea, endothelium to posterior capsule. Slit Lamp which is intended for that is used to aid the diagnosis of diseases or traumas which affect the structure or properties of the eye.

## Device Story

CeeMax Slit Lamp is an AC-powered biomicroscope for anterior eye segment examination; used by clinicians to aid diagnosis of ocular disease or trauma. Device utilizes halogen or LED illumination sources; provides variable brightness control (zero to full power) and standard slit beam geometry. Operator adjusts optics to visualize eye structures; output is direct visual observation by the clinician. Device is used in clinical settings. Benefits include enhanced visualization of anterior segment for diagnostic assessment.

## Clinical Evidence

Bench testing only. Evidence includes electrical safety certifications (EN/IEC 60601-1), EMC testing (EN/IEC 60601-1-2, EN6100-3-2, EN6100-3-3), flammability testing of materials, and bench verification of radiation outputs and light source parameters.

## Technological Characteristics

AC-powered slit lamp biomicroscope. Materials: Thermosetting phenol formaldehyde resin. Illumination: Halogen (12V, 30W; 460-700nm) or LED (12V, 10W; 450-650nm). Brightness control: 0-100%. Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC).

## Regulatory Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

## Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Reichert PSL Slit Lamp ([K061330](/device/K061330.md))
- VISION-TECH YZ SLIT LAMP ([K033190](/device/K033190.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 072861

### JUN 2 4 2008

## 510(k) Summary of Safety and Effectiveness for the CeeMax Slit Lamp™ (Halogen Lamp Series/LED Illumination series) (per 21CFR807.92)

#### 1. SPONSOR

CeeMax Ophthalmic Division of Achevé Technology, Inc 17100 Gillette Ave. #113 Irvine, CA 92614 Telephone: 949-885-8875 949-854-2388 Fax: E Mail: choujs@sbcglobal.com Contact Person: Jim-Son Chou, Ph.D.

#### 2. DEVICE NAME and Classification

| Proprietary Name:    | CceMax™ Slit Lamp (Halogen Lamp Series/LED<br>Illumination series) |
|----------------------|--------------------------------------------------------------------|
| Common/Usual Name:   | Slit Lamp                                                          |
| Classification Name: | AC-powered slit lamp biomicroscope                                 |
| Regulation Number:   | 886.1850                                                           |
| Product Code:        | HJO                                                                |
| Class:               | II                                                                 |

#### PREDICATE DEVICES 3.

(1) Reichert PSL Slit Lamp ----- FDA 510K K061330 (2) VISION-TECH YZ SLIT LAMP ---- FDA 510K K033190

#### 4. DEVICE DESCRIPTION

The CceMax™ Slit Lamp's (Halogen Lamp Series/LED Illumination series) with models of 801, 802, 901, 902, 8015S, 802SS, 901SS, 902SS are similar in design, function, materials, operation, and electro-optical electro-optical performance to other conventional AC-powered slit lamp biomicroscope.

#### INTENDED USE 5.

{1}------------------------------------------------

The CeeMax™ Slit Lamp (Halogen Lamp Series/LED Illumination series) is an AC powered microscope and accessories intended for the use in the examination of the anterior eye segment, from cornea, endothelium to posterior capsule. Slit Lamp which is intended for that is used to aid the diagnosis of diseases or traumas which affect the structure or properties of the eye.

6.

### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

# Side by Side Comparison of the CeeMax™ Slit Lamp (Halogen Lamp Series/LED Illumination series) with Predicate Devices

|                                | Submission Device                                                                                                                                                                                     | Predicate Device                                                                                                                                                                                              |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | CeeMaxTM Slit Lamp<br>(Halogen Lamp Series/LED<br>Illumination series)                                                                                                                                | Slit Lamp's<br>YZ5F (K033190)                                                                                                                                                                                 |
|                                |                                                                                                                                                                                                       | SL-15 (K061330)                                                                                                                                                                                               |
| Indications<br>for Use         | Same.                                                                                                                                                                                                 | AC powered microscope and<br>accessories intended for the<br>use in the examination of the                                                                                                                    |
|                                |                                                                                                                                                                                                       | anterior eye segment, from<br>cornea, endothelium to<br>posterior capsule. Slit Lamp<br>which is intended for that is<br>used to aid the diagnosis of<br>diseases or traumas which<br>affect the structure or |
|                                |                                                                                                                                                                                                       | properties of the eye                                                                                                                                                                                         |
| Method of<br>operation         | Same as the predicate device<br>YZ5F (K033190) table top                                                                                                                                              | YZ5F (K033190) Table Top<br>SL-15 (K061330) Hand held                                                                                                                                                         |
| Exposure<br>parameters         | The Halogen lamp exposure<br>parameters are the same as the<br>predicate device.<br>The LED exposure parameters<br>are similar to the predicate                                                       | Halogen lamp:<br>YZ5F (K033190)<br>LED: Reichert<br>SL-15(K061330)                                                                                                                                            |
|                                | device, detailed description<br>was also provided.                                                                                                                                                    |                                                                                                                                                                                                               |
| Flammability                   | Thermosetting phenol<br>formaldehyde resin.                                                                                                                                                           | Thermosetting phenol<br>formaldehyde resin for the                                                                                                                                                            |
|                                |                                                                                                                                                                                                       | lamp cap YZ5F (K033190)                                                                                                                                                                                       |
| Brightness<br>control          | From zero to full power<br>illumination control                                                                                                                                                       | Same                                                                                                                                                                                                          |
| Light source                   | Halogen lamp and white LED                                                                                                                                                                            | Halogen lamp for the YZ5F<br>(K033190)<br>LED for the SL-15(K061330)                                                                                                                                          |
| Beam                           | The same as the YZ5F                                                                                                                                                                                  | The standard slit opening                                                                                                                                                                                     |
| geometry                       | (K033190)                                                                                                                                                                                             | range YZ5F (K033190)                                                                                                                                                                                          |
| Radiation<br>safety            | The Halogen lamp exposure<br>parameters are the same as the<br>predicate device.<br>The LED exposure parameters<br>are similar to the predicate<br>device, detailed description<br>was also provided. | Halogen lamp:<br>YZ5F (K033190)<br>LED: Reichert<br>SL-15(K061330)                                                                                                                                            |
| Optical<br>radiation<br>hazard | No, warning in the manual<br>regarding the caution and<br>particularly for the aphakic<br>diagnostic                                                                                                  | No                                                                                                                                                                                                            |
| Electrical                     | Safety: EN(IEC) 60601-1                                                                                                                                                                               | Safety: IEC 60601-1                                                                                                                                                                                           |
| Electrical                     | EMC: EN(IEC) 60601-1-2<br>EN6100-3-2<br>EN6100-3-3                                                                                                                                                    | EMC: EN(IEC) 60601-1-2                                                                                                                                                                                        |
| Light Output<br>Wavelength     | The Halogen Lamp emits<br>Visible light covers 460nm to<br>700nm. The LED light source<br>covers 450nm to 650nm.                                                                                      | The Halogen Lamp emits<br>Visible light covers 460nm to<br>650nm. The LED light source<br>covers 450nm to 650nm.                                                                                              |
| Light Source<br>Power          | Halogen: 12V, 30W<br>LED: 12V, 10W                                                                                                                                                                    | Halogen(Vision Tech):<br>12V, 30W<br>LED(Reichert): 12V, 6W                                                                                                                                                   |

{2}------------------------------------------------

### PERFORMANCE TESTING 7.

Testing provided in this premarket notification includes certifications of electrical safety, flammability, and bench test of radiation outputs.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in all capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 4 2008

Jim-Son Chou, Ph.D. President CeeMax Ophthalmic Division of Achevé Technology, Inc 17100 Gillette Ave. #113 Irvine, CA 92614

Re: K072861

Trade/Device Name: CceMax Slit Lamp (Halogen Lamp Series/LED Illumination Series) Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: May 22, 2008 Received: May 23, 2008

Dear Dr. Chou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Jim-Son Chou, Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mahina Eggleston, und

Malvina B. Eydelman, M. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K 072861

Device Name: CeeMax™ Slit Lamp (Halogen Lamp Series/LED Illumination series)

Indications For Use:

The CeeMax™ slit-lamp series is an AC-powered slit-lamp biomicroscope intended for use in eye examination of the anterior segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affect the structural properties of the anterior eye segment.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use

*x* 6/18/2008

510(k) Number K072861

---

**Source:** [https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HJO/K072861](https://fda.innolitics.com/submissions/OP/subpart-b%E2%80%94diagnostic-devices/HJO/K072861)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
