K082850 · Product and Educational Services, LLC · HPZ · Feb 27, 2009 · Ophthalmic
Device Facts
Record ID
K082850
Device Name
EYE SPHERES
Applicant
Product and Educational Services, LLC
Product Code
HPZ · Ophthalmic
Decision Date
Feb 27, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3320
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Eye Spheres are permanent implants that occupy the eye cavity when it has become necessary to surgically remove the eye (enucleation) or the contents of the eye sac (evisceration) Conformers are temporary devices which maintain the shape of the eye and prevent closure or adhesion during the healing process
Device Story
Eye spheres and conformers are ocular implants/devices used following enucleation or evisceration. Eye spheres are permanent implants occupying the eye cavity; conformers are temporary devices maintaining eye shape and preventing adhesion during healing. Both are manufactured from polymethylmethacrylate (PMMA). Eye spheres are available in 10mm-22mm diameters; conformers are available in small, medium, and large sizes, with or without holes. Devices are used in clinical settings by ophthalmologists/surgeons. The device provides structural support to the eye socket post-surgery, facilitating proper healing and preparation for future prosthetic fitting.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Polymethylmethacrylate (PMMA). Form factor: Spheres (10-22mm) and conformers (small, medium, large). Sterilization: Gamma sterilization. Connectivity: None. Software: None.
Indications for Use
Indicated for pediatric and adult patients requiring ocular implants following enucleation or evisceration, or temporary conformers to maintain eye shape and prevent adhesion during post-surgical healing.
Regulatory Classification
Identification
An eye sphere implant is a device intended to be implanted in the eyeball to occupy space following the removal of the contents of the eyeball with the sclera left intact.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ocular peg which is supplied sterile only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Gulden Ophthalmics Eye Sphere and Conformer (K972661)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Product and Educational Services LLC. The logo consists of the letters "PES" in a stylized font, with the words "Product and Educational Services LLC" written to the right of the letters. The text is in black and the background is white.
3 2 2009
## 510(k) Summary
Submitter's Name
Address
Product and Educational Services LLC
Bucks County Biotechnology Center 3805 Old Easton Road Doylestown, PA 18902 Phone: (215) 230 7515 Facsimile: (215) 230 7517
Edward T. Griffith President
## Date of Summary:
Contact Person
26 September 2008
Proprietary Name of Device:
None
Common/Usual Name:
Classification Name:
Eye Sphere
Conformer
Implant, Eye Sphere
Conformer, Ophthalmic
Legally Marketed Equivalent Devices:
Gulden Ophthalmics Eye Sphere and Conformer (K972661)
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Product and Educational Services LLC. The logo consists of the letters "PES" in a stylized font, with the words "Product and Educational Services LLC" to the right of the letters. The text is in black and the background is white.
## Summary of Device:
When an eye is severely damaged by trauma or disease, it may be necessary to remove it surgically. The most common procedures are evisceration (removal of the contents of the eye) or enucleation (removal of the eye itself). Eye spheres are permanent implants to occupy the cavity that results from surgery. Conformers are temporary devices, which maintain the shape of the eye and prevent closure or adhesion during the healing process. When healing is complete, conformers are replaced with prosthesis. Eye Spheres and Conformers are made of polymethylmethacrylate (PMMA), a hard, clear plastic that is lightweight, inert and virtually unbreakable. Eye spheres are available in even diameters from 10mm through 22mm. Conformers are available in three sizes: small, medium, large. Conformers may be ordered with or without holes. Eye spheres and Conformers are FDA Class II products.
#### Intended Use:
Eye Spheres are permanent implants that occupy the eye cavity when it has become necessary to surgically remove the eye (enucleation) or the contents of the eye sac (evisceration)
Conformers are temporary devices which maintain the shape of the eye and prevent closure or adhesion during the healing process
{2}------------------------------------------------
Product and
Educational Services LLC
ருப்
.
.
echnological Characteristics of the Device Compared to the Predicate Devices
| Manufacturer | Gulden Ophthalmics<br>Eye Spheres - Non Sterile<br>Conformers - Non Sterile<br>Gulden Ophthalmics | Gulden Ophthalmics<br>Eye Spheres – Sterile<br>Conformers – Sterile<br>Gulden Ophthalmics | Product and Educational<br>Services LLC<br>Eye Spheres - Sterile<br>Conformers - Non Sterile<br>Product and Educational<br>Services LLC |
|---------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| 510K Number | Pre-amendment | K972661 | New |
| FDA Product Class | Class II | Class II | Class II |
| Intended Population | Pediatric and Adult | Pediatric and Adult | Pediatric and Adult |
| Material | Polymethylmethacrylate<br>(PMMA) | Polymethylmethacrylate<br>(PMMA) | Polymethylmethacrylate<br>(PMMA) |
| Size Range (mm) –<br>Eye Sphere | 10, 12, 14, 16, 18, 20, and<br>22 mm | 10, 12, 14, 16, 18, 20, and<br>22 mm | 10, 12, 14, 16, 18, 20, and<br>22 mm |
| Size Range –<br>Conformers | Small, Medium, Large<br>(with and without holes) | Small, Medium, Large<br>(with and without holes) | Small, Medium, Large<br>(with and without holes) |
| Method of<br>sterilization | None | EtO (100%) and Gamma<br>Sterilization | Gamma Sterilization only |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Product and Educational Services LLC. The logo consists of the letters 'FES' in a stylized font, with the words 'Product and Educational Services LLC' to the right of the letters. The text is in a simple, sans-serif font and is black in color. The logo is simple and professional in appearance.
## Non-Clinical Tests:
Product and Educational Services LLC (PES) Eye Spheres and Conformers are the same Eye Spheres and Conformers manufactured and marketed by Gulden Ophthalmics as "non sterile". The only difference is that PES will be providing the same Eye Spheres and Conformers as sterile, single use devices. Gulden Ophthalmics Eye Spheres and Conformers are "preamendment" devices.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Product and Educational Services, LLC PA Biotechnology Center of Bucks County Mr. Edward T. Griffith President 3805 Old Easton Road Doylestown, PA 18902
FEB 2 7 2009
Re: K082850
Trade Name: Eye Spheres and Conformers Regulation Number: 21 CFR 886.3320 Regulatory Class: II Product Code: HPZ, HQN Dated: February 12, 2009 Received: February 17, 2009
Dear Mr. Griffith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 -- Mr. Edward T. Griffith
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from 1 the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
M.B. Egleston, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K082850
Device Name: Eye Spheres and Conformers
Indications for Use:
Eye Spheres are permanent implants that occupy the eye cavity when it has become necessary to surgically remove the eye (enucleation) or the contents of the eye sac (evisceration)
Conformers are temporary devices which maintain the shape of the eye and prevent closure or adhesion during the healing process
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Qoncurrence of CDRH, Office of Device Evaluation (ODE)
Z
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
KO 82850 510(k) Number
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.