K201953 · Sight Sciences, Inc. · MRH · Aug 11, 2020 · General Hospital
Device Facts
Record ID
K201953
Device Name
OMNI PLUS Surgical System
Applicant
Sight Sciences, Inc.
Product Code
MRH · General Hospital
Decision Date
Aug 11, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OMNI® PLUS Surgical System is an ophthalmic surgical tool for the delivery of controlled amounts of viscoelastic fluid into the anterior segment and the cutting of trabecular meshwork when a trabeculotomy is indicated.
Device Story
Handheld, manually operated ophthalmic surgical instrument; used by ophthalmologists in clinical settings. Device features stainless-steel cannula, polymeric microcatheter, and internal reservoir. Input: cohesive viscoelastic fluid (OVD) loaded via Luer fitting. Operation: surgeon rotates advancement wheels to advance/retract microcatheter; rack-and-pinion mechanism dispenses 21 µL of viscoelastic fluid during retraction. Trabeculotomy function: cannula tip enters Schlemm's canal; microcatheter advanced up to 180 degrees; cannula removal cuts trabecular meshwork. Ambidextrous design allows use in either eye/hand. Output: controlled viscoelastic delivery and mechanical tissue cutting. Benefits: facilitates trabeculotomy procedures by combining fluid delivery and tissue cutting in one device.
Clinical Evidence
Bench testing only. No clinical data. Testing included design verification, functional performance, sterilization validation (ANSI/AAMI/ISO 11137-2:2013), packaging/shelf-life (13 months), biocompatibility (ISO 10993-1), and bacterial endotoxin testing (USP <85>, ANSI/AAMI ST72). Results confirmed device meets specifications and functions as intended.
Indicated for ophthalmic surgery to deliver small amounts of viscoelastic fluid (e.g., Healon® PRO, Healon GV® PRO, Amvisc®, PROVISC®) and to cut trabecular meshwork tissue during trabeculotomy procedures.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services seal. To the right of the symbol is a blue square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Sight Sciences, Inc. Mr. Edward Sinclair Vice President, Regulatory Affairs 4040 Campbell Ave., Suite 100 Menlo Park, CA 94025
Re: K201953
Trade/Device Name: OMNI PLUS Surgical System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: MRH, HMZ Dated: July 13, 2020 Received: July 14, 2020
Dear Mr. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
## 510(k) SUMMARY
### Submitter Information
| 510(k) Owner: | Sight Sciences, Inc.<br>4040 Campbell Ave., Suite 100<br>Menlo Park, CA 94025<br>Tel: (877) 266-1144 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Edward J. Sinclair<br>Vice President, Regulatory Affairs<br>4040 Campbell Ave., Suite 100<br>Menlo Park, CA 94025<br>Tel: 650-218-9149 |
| Date Prepared: | June 23, 2020 |
### Device Name and Classification
| Trade Name: | OMNI® PLUS Surgical System |
|-------------------------|----------------------------|
| Common Name: | Ophthalmic Infusion Pump |
| Classification Name: | Infusion Pump |
| Regulation Number: | 21 CFR 880.5725 |
| Device Classification: | Class II |
| Primary Product Code: | MRH |
| Secondary Product Code: | HMZ |
## Predicate Device
| Device Name: | OMNI® Surgical System |
|-----------------|-----------------------|
| 510(k) Holder: | Sight Sciences, Inc. |
| 510(k) Number: | K173332 |
| Clearance Date: | December 21, 2017 |
#### Intended Use
The OMNI® PLUS Surgical System is an ophthalmic surgical tool for the delivery of controlled amounts of viscoelastic fluid into the anterior segment and the cutting of trabecular meshwork when a trabeculotomy is indicated.
## Indications for Use
The OMNI® PLUS Surgical System is a manually operated device for delivery of small amounts of viscoelastic fluid, for example Healon® PRO or Healon GV® PRO from Johnson & Johnson Vision,
{4}------------------------------------------------
Amvisc® from Bausch & Lomb, or PROVISC® from Alcon, during ophthalmic surgery. It is also indicated to cut trabecular meshwork tissue during trabeculotomy procedures.
# Device Description
The OMNI PLUS Surgical System is a sterile, single use, manually operated instrument used by ophthalmologists to deliver small, controlled amounts of viscoelastic fluid into the anterior segment of the eye during ophthalmic surgery and cut trabecular meshwork tissue during trabeculotomy procedures. The OMNI PLUS device is provided sterile and disposed after singlepatient use. The device is fabricated from biocompatible materials standard to the medical device industry. The OMNI PLUS Surgical System dispenses fluid on the principle of exchanging volumes much like a syringe and are designed to function with commercially available cohesive viscoelastic fluids (also known as an ophthalmic viscosurgical device, or "OVD") that are commercially available from companies such as Johnson & Johnson, Bausch & Lomb, and Alcon.
The OMNI PLUS device includes a stainless-steel cannula, polymeric microcatheter, removable priming lock, internal reservoir and plunger tube, a Luer fitting for direct connection with an OVD cartridge to prime the internal reservoir, and two advancement wheels. A single advancement wheel is located on each side of the handle. This allows the OMNI PLUS device to be used in either eye (OD or OS) and in either hand of the surgeon (left or right), by turning the device 180 degrees along its vertical axis. These wheels are used to advance and retract the microcatheter.
The microcatheter can be advanced/retracted up to 20 mm per cycle by manually rotating either advancement wheel. The microcatheter can be fully advanced/retracted multiple times, however, viscoelastic fluid can only be dispensed during the first two advancement/retraction cycles. When the OMNI PLUS device is being used to deliver viscoelastic fluid, retraction of the microcatheter causes a plunger tube to advance into the viscoelastic fluid reservoir thereby automatically dispensing viscoelastic fluid as the microcatheter is being retracted back into the stainless-steel cannula.
When the OMNI PLUS Surgical System is used to perform a trabeculotomy procedure, the beveled tip of the curved stainless-steel cannula is used to enter Schlemm's canal through the trabecular meshwork. The polymeric microcatheter is advanced into Schlemm's canal up to 180 degrees (one hemisphere) by rotating the advancement wheel forward until the wheel stops (about 20 mm). With the microcatheter resting in the canal, the cannula is removed from the corneal incision and out of the eye causing the microcatheter to cut through the trabecular meshwork. This process can be repeated in the second Schlemm's hemisphere.
The OMNI PLUS Surgical System is an additional model based on the current OMNI Surgical System design; OMNI PLUS is designed to dispense a nominal volume of 21 microliters of viscoelastic fluid.
# Comparison of Technological Characteristics with the Predicate Device
The technical features of the OMNI PLUS Surgical System are substantially equivalent to the OMNI Surgical System predicate device (cleared by FDA in K173332 on December 21, 2017). The OMNI
{5}------------------------------------------------
PLUS Surgical System is an additional model based on the OMNI Surgical System design. The primary functional difference between the predicate device and subject device is the volume of viscoelastic fluid that can be delivered during use. The predicate OMNI Surgical System delivers a nominal volume of approximately 9 microliters of viscoelastic fluid, whereas the subject OMNI PLUS design can deliver a nominal volume of approximately 21 microliters of viscoelastic fluid. In order to deliver the additional viscoelastic fluid, three internal components, consisting of the reservoir, plunger tube and distal O-ring, have modified dimensions to contain and dispense additional viscoelastic fluid in the OMNI PLUS device. Other changes were made to improve the aesthetics, ergonomics and manufacturability of the device. These include the handle shape and materials, reduced number of advancement wheels, addition of a proximal Luer fitting to aid priming and a packaging tray with a heat-sealed Tyvek lid. Other minor changes are included in the technological attributes comparison between the OMNI PLUS Surgical System (subject device) with the OMNI Surgical System (predicate device) in Table 1 below.
| Characteristic | OMNI PLUS Surgical System | OMNI Surgical System (K173332) |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SUBJECT DEVICE | PREDICATE DEVICE |
| Intended Use | Ophthalmic surgical tool for delivery of<br>controlled amounts of viscoelastic fluid into the<br>anterior segment and used to cut trabecular<br>meshwork when a trabeculotomy is indicated | Ophthalmic surgical tool for delivery of<br>controlled amounts of viscoelastic fluid into the<br>anterior segment and used to cut trabecular<br>meshwork when a trabeculotomy is indicated |
| Indications for<br>Use | The indications were updated to reflect a<br>change in manufacturer of the Healon<br>viscoelastic fluids from Abbott Medical Optics<br>to Johnson & Johnson, as well as to add the<br>OMNI PLUS Surgical System as follows:<br><br>The OMNI PLUS Surgical System is a manually<br>operated device for delivery of small amounts<br>of viscoelastic fluid, for example Healon® PRO<br>or Healon GV® PRO from Johnson & Johnson<br>Vision, Amvisc® from Bausch & Lomb, or<br>PROVISC® from Alcon, during ophthalmic<br>surgery. It is also indicated to cut trabecular<br>meshwork tissue during trabeculotomy<br>procedures | The OMNI Surgical System is a manually<br>operated device for delivery of small amounts<br>of viscoelastic fluid, for example Healon® or<br>Healon GV® from Abbott Medical Optics<br>(AMO), Amvisc® from Bausch & Lomb, or<br>PROVISC® from Alcon, during ophthalmic<br>surgery. It is also indicated to cut trabecular<br>meshwork tissue during trabeculotomy<br>procedures |
| Regulation | 880.5725 (Infusion Pump) | 880.5725 (Infusion Pump) |
| Device Class | Class II | Class II |
| Product Code | Primary: MRH<br>Secondary: HMZ | Primary: MRH<br>Secondary: HMZ |
| Prescription<br>Status | Prescription use only | Prescription use only |
| Target<br>Anatomy | Anterior Segment including Schlemm's<br>Canal/Trabecular Meshwork | Anterior Segment including Schlemm's<br>Canal/Trabecular Meshwork |
| Operating<br>Principle | Manual | Manual |
| | OMNI PLUS Surgical System | OMNI Surgical System (K173332) |
| Characteristic | SUBJECT DEVICE | PREDICATE DEVICE |
| Design/<br>Mechanism of<br>Action | • Flexible microcatheter with rounded,<br>atraumatic tip for dispensing of viscoelastic<br>• Proximal handle – changed to ovoid shape<br>• Handle has internal viscoelastic reservoir and<br>plunger tube with dimensional changes to<br>allow dispensing of additional viscoelastic<br>• Two advancement wheels (finger wheels) for<br>advancing and retracting microcatheter up to<br>20mm using a rack and pinion mechanism<br>• Tactile and audible clicks indicate precise<br>advancement<br>• Viscoelastic dispensed during retraction of first<br>two cycles after priming with viscoelastic fluid<br>• Flexible microcatheter introduced into<br>Schlemm's canal and pulled through to cut<br>trabecular meshwork | • Flexible microcatheter with rounded,<br>atraumatic tip for dispensing of viscoelastic<br>• Proximal handle – rectangular shape<br>• Handle has internal viscoelastic reservoir and<br>plunger tube<br>• Four advancement wheels (finger wheels) for<br>advancing and retracting microcatheter up to<br>20mm using a rack and pinion mechanism<br>• Tactile and audible clicks indicate precise<br>advancement<br>• Viscoelastic dispensed during retraction of first<br>two cycles after priming with viscoelastic fluid<br>• Flexible microcatheter introduced into<br>Schlemm's canal and pulled through to cut<br>trabecular meshwork |
| Dispensing<br>Control | After priming, viscoelastic fluid dispensing<br>control occurs through manual rotation of the<br>advancement wheels at the distal end of the<br>device. The ovoid shape of the handle allows a<br>single advancement wheel on each side of the<br>handle while maintaining the ability for<br>ambidextrous use in either patient eye.<br>Synchronization of the advancement wheels<br>and microcatheter movement was achieved for<br>ease of use by adding a gear in the rack and<br>pinion mechanism. | After priming, viscoelastic fluid dispensing<br>control occurs through manual rotation of the<br>advancement wheels at the distal end of the<br>device. The rectangular shape of the handle<br>required two advancement wheels on each side<br>of the handle for ambidextrous use in either<br>patient eye |
| Dispensing<br>Mechanism | Internal reservoir with plunger tube (syringe-<br>like volume exchange). Three components<br>consisting of the Reservoir, Plunger Tube and<br>Distal O-Ring have modified dimensions to<br>contain and dispense larger volume of<br>viscoelastic fluid (OVD) | Internal reservoir with plunger tube (syringe-<br>like volume exchange) |
| Viscoelastic<br>Fluid (OVD)<br>and Priming<br>Method | Cohesive viscoelastic fluid (OVD or ophthalmic<br>viscosurgical device) is supplied separately.<br>Viscoelastic loaded into device (primed) prior<br>to use by attaching OVD cartridge directly to<br>Luer fitting on proximal end of OMNI device<br>handle | Cohesive viscoelastic fluid (OVD or ophthalmic<br>viscosurgical device) is supplied separately.<br>Viscoelastic loaded into device (primed) prior<br>to use by attaching OVD cartridge to a supplied<br>nozzle and advancing into proximal end of<br>OMNI device handle |
| OVD Volume<br>Dispensed | $21 \pm 3$ µL (10.5 µL on first microcatheter<br>retraction cycle and 10.5 µL on the second<br>cycle) | $9 \pm 3$ µL (4.5 µL on first microcatheter<br>retraction cycle and 4.5 µL on the second cycle) |
| | OMNI PLUS Surgical System | OMNI Surgical System (K173332) |
| Characteristic | SUBJECT DEVICE | PREDICATE DEVICE |
| Materials | Medical grade materials, including ABS,<br>polycarbonate, stainless steel, silicone,<br>parylene coating, cyanoacrylate, acrylated<br>urethane, polyimide. Minor changes were<br>made to the materials in the handle, reservoir,<br>Luer fitting and a new bonding adhesive was<br>used | Medical grade materials, including ABS,<br>polycarbonate, stainless steel, silicone,<br>parylene coating, cyanoacrylate, acrylated<br>urethane, polyimide |
| User Interface | Handheld | Handheld |
| Microcatheter<br>Shaft Outer<br>Diameter | 200 microns | 200 microns |
| Microcatheter<br>Tip Outer<br>Diameter Range | 0.0090 to 0.0110 inches. The lower<br>specification limit was reduced by 0.0005<br>inches | 0.0095 to 0.0110 inches |
| Sterile and<br>Single Use | Provided sterile. Single patient use | Provided sterile. Single patient use |
| Sterilization<br>Method | Gamma radiation | Gamma radiation |
| Sterility<br>Assurance Level | 10-6 | 10-6 |
| Packaging | Thermoformed plastic tray with heat-sealed<br>Tyvek lid | Tyvek pouch with a polymer tray card |
| Shelf Life | 13 months | 6 months |
# Table 1: Technological Characteristics Comparison OMNI PLUS Surgical System and Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
# Risk Management
The risk management process at Sight Sciences complies with ISO 14971:2012 "Medical devices --Application of risk management to medical devices." As required by this standard, risk analyses are conducted according to defined procedures, using experienced, qualified personnel from multiple functions throughout the organization with prior experience in risk assessment.
Risk management activities were first documented in a Risk Management Plan. This plan specified the risk management activities to be conducted based on the cleared OMNI Surgical System that resulted in the OMNI PLUS Surgical System design. The following activities were completed:
- . The Clinical Hazard Analysis for the OMNI Surgical System changes and the OMNI PLUS Surgical System was re-assessed by qualified personnel with respect to all changes.
- . All identified hazards associated with the OMNI devices were evaluated and recorded along with potential harm and effects resulting from the hazard and the severity ranking of each hazard in Failure Modes Effect Analyses (FMEAs).
- . Potential hazards were mitigated through the device design and manufacturing processes and any mitigations were subsequently verified.
{8}------------------------------------------------
All the identified hazards were mitigated to an acceptable level of risk. The potential benefits to patients outweigh the low residual risk of the design changes and taking into consideration the indications for use of the OMNI devices.
## Performance Data – Bench Testing
Non-clinical bench testing included design verification and functional product testing, sterilization validation, packaging and shelf life testing, biocompatibility testing, bacterial endotoxin testing and simulated use (usability) testing in a bench model. Results of the nonclinical testing demonstrate that the OMNI PLUS Surgical System meets the defined specifications and functioned as intended.
## Design Verification and Functional Product Testing
Design verification testing was performed to demonstrate the OMNI PLUS Surgical System device meets the performance requirements described in the Product Specification after the following conditioning: worst-case dosage sterilization cycle, extreme conditions environmental exposure, simulated transit and accelerated and real-time aging. After conditioning and accelerated aging, test samples were subjected to a variety of tests including mechanical integrity assessment, sterile barrier packaging testing, physical requirements, product inspections and specification verifications. The results establish that the OMNI PLUS Surgical System functioned as intended and complies with the applicable requirements.
## Sterilization Validation
The OMNI PLUS Surgical System was adopted into the sterilization process that was validated for the predicate OMNI device in K173332. Sterilization is performed by the same approved contractor. The rationale for adoption was based on similarities in mass and product-related variables that affect bioburden. In addition to a written justification for adoption of the OMNI PLUS Surgical System, representative device samples were tested for bioburden and subjected to a verification dose experiment as a demonstration of process effectiveness. The tested units were found to be sterile after processing with the sub-lethal dose, confirming that the product may be labeled "Sterile" per ANSI/AAMI/ISO 11137-2:2013.
# Packaging and Shelf Life Testing
The OMNI PLUS Surgical System is currently validated for total shelf life of 13 months. The device packaging is labeled with an expiration date of 12 months, providing a one-month safety factor. The shelf life study evaluated the functional performance of the OMNI PLUS Surgical System, as well as the packaging integrity of the tray sealed with the Tyvek lid after test samples were subjected to worst-case dosage sterilization cycle, extreme conditions environmental exposure, simulated transit conditioning, and real-time aging. After conditioning and real-time aging for a minimum of 13 months, OMNI samples were subjected to a variety of tests related to shelf life including visual verification of packaging, labeling and device integrity, sterile barrier packaging testing, and functional performance testing.
# Biocompatibility and Bacterial Endotoxin Testing
Biocompatibility of the OMNI PLUS Surgical System was demonstrated through testing in accordance with ISO 10993- 1:2009/(R)2013 "Biological Evaluation of Medical Devices – Part 1:
{9}------------------------------------------------
Evaluation and testing within a risk management process." Testing was performed by an approved contract laboratory on all components that have direct or indirect contact with the patient. Test results demonstrated that the device materials have an acceptable biocompatibility profile and met the requirements of the ISO standard. Additionally, OMNI PLUS devices were tested for bacterial endotoxin according to USP <85>:2011 "Bacterial Endotoxin Test" and ANSI/AAMI ST72:2011/(R)2016 "Bacterial endotoxins – Test methodologies, routine monitoring, and alternatives to batch testing." Test results demonstrated that the samples do not have an unacceptable level of bacterial endotoxin and met the endotoxin limit requirement of ≤0.2 EU/device.
## Conclusions
The Sight Sciences OMNI PLUS Surgical System meets all product design requirements and applicable standards as the predicate OMNI Surgical System. The subject device shares the same principle of operation, intended use/indications and many of same key technological characteristics as the predicate device. Differences in technological characteristics were evaluated using the same recognized consensus standards and scientific methods used to clear the predicate device in K173332. The conclusions drawn from the non-clinical bench performance tests demonstrate that the OMNI PLUS Surgical System meets the defined specifications and performs as intended.
The OMNI PLUS Surgical System is substantially equivalent to the legally-marketed OMNI Surgical System predicate device cleared under 510(k) K173332.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.