SYNTEC, INC. DISPOSABLE BARE END FIBER

K971338 · Syntec, Inc. · MPA · May 6, 1997 · Gastroenterology, Urology

Device Facts

Record IDK971338
Device NameSYNTEC, INC. DISPOSABLE BARE END FIBER
ApplicantSyntec, Inc.
Product CodeMPA · Gastroenterology, Urology
Decision DateMay 6, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Syntec, Inc. Disposable Bare End Fiber is used to illuminate with visible spectrum light the intraocular portion of the eye for improved visualization during vitreo-retinal surgery.

Device Story

Disposable ophthalmic endoilluminator; used during vitreo-retinal surgery to provide visible light for intraocular visualization. Device consists of handpiece tube, fiberoptic cable, cable sheath, and connector. Operates by transmitting light from an external source through the fiberoptic cable to the surgical site. Used by ophthalmologists in an operating room setting. Provides improved visualization of ocular structures, facilitating surgical procedures. Biocompatible materials ensure safety during contact with ocular tissues and fluids.

Clinical Evidence

Bench testing only. Biocompatibility confirmed via USP Class VI criteria. Sterilization validated using ethylene oxide gas overkill method. Performance equivalence (light output intensity and spot size) demonstrated against predicate device.

Technological Characteristics

Materials: surgical grade stainless steel (handpiece), polystyrene core/polymethylmethacrylate cladding (fiberoptic cable), PVC (sheath), 6063 aluminum (connector). Sterilization: ethylene oxide gas (overkill method). Biocompatibility: USP Class VI compliant.

Indications for Use

Indicated for patients undergoing vitreo-retinal surgery requiring intraocular illumination.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K971338 , . • **MAY - 6 1997** র মাধ্যম ## Section 9 - Summary of Safety and Effectiveness | Date of Preparation: | March 15, 1997 | |------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Syntec, Inc. Disposable Bare End Fiber | | Classification Name: | Ophthalmic Endoilluminator, 86MPA | | Manufacturer: | Syntec, Inc. is located at 812 Truman Blvd., Crystal<br>City, MO 63019. The phone number is (314) 931-<br>2204 and the fax number is (314) 931-6029. | | 510(k) Submitter: | Syntec, Inc. is located at 812 Truman Blvd., Crystal<br>City, MO 63019. The phone number is (314) 931-<br>2204 and the fax number is (314) 931-6029. Contact<br>person: Nathan H. Lewis. | | Predicate Device: | Grieshaber Disposable Monofilament catalog<br>number 630.78 manufactured by Grieshaber & Co.,<br>Inc. located at 1945 Vaughn Road, Kennesaw, GA<br>30144. | | Device Description: | The Bare End Fiber is comprised of five basic<br>components: the handpiece tube, the fiberoptic<br>cable, the fiber optic cable sheath and the connector. | | Intended Use: | The Syntec, Inc. Disposable Bare End Fiber is used<br>to illuminate with visible spectrum light the<br>intraocular portion of the eye for improved<br>visualization during vitreo-retinal surgery. | | Clinical and Non-Clinical<br>Similarities and Differences: | The Syntec, Inc. Disposable Bare End Fiber and the<br>Grieshaber & Co. Disposable Monofilament are<br>substantially equivalent since they both are used for<br>the same clinical purpose, ie: to illuminate with<br>visible spectrum light the intraocular portion of the<br>eye for improved visualization during vitreo-retinal<br>surgery.<br><br>Both devices are of a similar design and are made<br>using the exact same materials. The handpiece tube<br>is made of surgical grade stainless steel. The<br>fiberoptic cable is made with a polystyrene core and<br>a polymethylmethacrylate cladding. The fiber optic<br>cable sheath is made of PVC tubing and the<br>connector is made of 6063 aluminum. | · : {1}------------------------------------------------ K971338 The device is biocompatable with the body tissue and fluids that it comes in contact with as it is made of the same materials as the predicate device. These materials meet US Pharmacopoeia Class VI criteria and are widely used in many other medical products. The device is sterilized using ethylene oxide gas which is then validated by the overkill method. The light output intensity and spot size is the same as the predicate device. The only device differences are cosmetic.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...