The data was used to develop a reference database for the MAIA device, allowing for the quantitative comparison of patient macular sensitivity to a normative population.
Reference database; Normative data; Clinical site recruitment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Reference Database Development; Retrospective analysis of normative clinical data
494 eyes of 270 normal subjects (age 21-86); Sample Size: 494 eyes; Number of Sites: 4
Not applicable for this study
Threshold sensitivity
Indications for Use
The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference database that is a quantitative tool for the comparison of macular sensitivity to a database of known normal subjects.
Device Story
MAIA™ is a non-mydriatic, confocal, line-scanning, infrared ophthalmoscope combined with an automated perimeter for microperimetry. It captures 36° x 36° infrared retinal images; tracks eye movements to quantify fixation characteristics; and measures differential light sensitivity (threshold sensitivity) at macular locations via patient response to projected light stimuli. The device integrates a computer, touch-screen, and custom Linux OS. Operated by clinicians in a clinical setting, it compares patient threshold data against a normative reference database. Output includes retinal images, fixation analysis, and sensitivity maps, aiding clinicians in diagnosing and monitoring retinal pathologies like age-related macular degeneration and diabetic retinopathy. Benefits include objective, quantitative assessment of macular function and fixation stability.
Clinical Evidence
Clinical evaluation included a normative database study (n=494 eyes, 270 subjects, age 21-86) and a precision study. Precision study involved 24 subjects (12 normal, 12 with retinal pathology) tested across 3 devices and 3 operators. Results for normal eyes: mean 29.7 dB, repeatability SD 0.42 dB, reproducibility SD 0.96 dB. Results for pathology eyes: mean 23.5 dB, repeatability SD 0.75 dB, reproducibility SD 0.75 dB. Data supports the device's ability to measure macular sensitivity and fixation stability.
Technological Characteristics
Confocal, line-scanning, infrared ophthalmoscope; automated perimeter; non-mydriatic. Features include infrared illumination, visible light projection for perimetry, touch-screen interface, and integrated computer. Software runs on custom Linux OS. Connectivity includes power cord and push-button input. Dimensions/materials not specified. Class II device.
Indications for Use
Indicated for patients requiring assessment of macular sensitivity, fixation stability, and locus of fixation, and infrared retinal imaging. Applicable to patients with retinal pathologies, including age-related macular degeneration and diabetic retinopathy, as well as normal subjects.
Regulatory Classification
Identification
A perimeter is an AC-powered or manual device intended to determine the extent of the peripheral visual field of a patient. The device projects light on various points of a curved surface, and the patient indicates whether he or she sees the light.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white graphic that appears to be an abstract representation of a circular or spiral structure. The central part of the image features a solid black circle, surrounded by a white ring. The outer area of the image is filled with a pattern of irregular black and white shapes, which could be interpreted as buildings or other architectural elements. The overall impression is that of a stylized map or diagram.
mmary of safety and effectiveness
Image /page/0/Picture/3 description: The image contains the word "centervue" in a bold, sans-serif font. The letters are closely spaced together, creating a compact and unified appearance. The word is written in lowercase.
**SUBMITTER INFORMATION**
| A. Company name: | CenterVue SpA |
|---------------------------|---------------------------------------|
| B. Company address: | via Tommaseo 77, 35131, Padova, ITALY |
| C. Company phone: | +39 049 781 1828 |
| Company fax: | +39 049 781 1899 |
| D. Contact person: | Giuliano Barbaro<br>Managing Director |
| E. Date summary prepared: | 4 April 2010 |
#### DEVICE IDENTIFICATION
- Ophthalmoscope, AC-powered; Automated Perimeter, AC-powered Generic Device Name: A.
- MAIA™, Macular Integrity Assessment Trade/proprietary name: в.
- C. Classification: Class II
- HLI, HPT Product Code: D.
### DEVICE DESCRIPTION
The MAIA™ is a confocal, line scanning, infrared, ophthalmoscope, combined with a system for visible light projection to obtain perimetric measurements, using "fundus perimetry" (also "microperimetry").
MAIA™ integrates in one device an automated perimeter and an ophthalmoscope, providing:
- images of the central retina over a field of view of 36° x 36°, acquired under infrared . illumination;
- recordings of eye movements obtained by "tracking" retinal details in the live retinal images and . providing a quantitative analysis of fixation characteristics;
- measurements of differential light sensitivity (or threshold sensitivity) at multiple locations in . the macula, obtained by recording a patient's subjective response (see / don't see) to a light stimulus projected at a certain location over the retina;
- comparison of measured threshold sensitivity with a reference database obtained from normal . subjects, indicating whether measured thresholds are above or below certain percentiles.
MAIA 100 works with no pupil dilation (non-mydriatic).
MAIA™ integrates a computer for control and data processing and a touch-screen display and it is provided with a power cord and a push-button. MAIA™ works with a dedicated software application running on a custom Linux O.S.
#### INDICATIONS FOR USE
The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference
CenterVue SpA
ia Tommaseo. 7 35131 Padova, Italy www.centervue.com
0 +30 Mg 781 1829 fax +39 049 781 1899 e-mail: info@centervue.com 378243 CCIAA PD oc .: € 480.000,00 i.v.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a black and white graphic design. It features a circular shape in the center, surrounded by a more complex, abstract pattern. The pattern consists of various geometric shapes and lines, creating a visually intricate design around the central circle.
centervue
t the center of vision
that is a quantitative tool for the comparison of macular sensitivity to a database of known
JIVALENCE
device is substantially equivalent to the combination of the following predicate devices with regard to intended use, operating principle, function:
| Predicate Device | 510(k) Holder | 510(k) No. | Date cleared |
|---------------------|-------------------------|------------|---------------------|
| MP-1 MICROPERIMETER | Nidek Technologies Srl | K023719 | December 23rd 2002 |
| MP-1 MICROPERIMETER | Nidek Technologies Srl | K061768 | September 28th 2006 |
| CIRRUS HD-OCT(*) | Carl Zeiss Meditec Inc. | K063378 | January 26th 2007 |
(*) only with regard to its retinal imaging sub-system
#### CLINICAL EVALUATION
Clinical data was collected and evaluated to support the indications for use statement for MAIA™ and to demonstrate substantial equivalence to the above predicate devices.
In particular, the reference database was developed by obtaining threshold sensitivity data from 494 eyes of 270 normal subjects, enrolled at 4 different clinical sites. The age range of the measured population was 21-86 (mean 43, std. dev. 15). Subjects were recruited among the clinics' personnel and among the relatives of the clinics' regular patients. The following MAIA perimetric grid and other device settings were used for the measurements:
- Number of stimuli: 61; .
- Macular coverage: 10°; .
- Size of stimulus: Goldmann III; .
- Duration of stimulus: 200 ms; ●
- . Background luminance: 4 asb;
- Maximum luminance: 1000 asb; .
- Threshold algorithm: standard 4-2 staircase; ●
- . Measurement scale: 0 to 36 dB;
- . Fixation target: central circle of 1° diameter.
#### PRECISION
A precision study was conducted which measured both individual location thresholds (n = 37 per eye) and average threshold over all locations. Three devices were used, each operated by a different operator. The study included 24 subjects, each tested on one eye only: 12 without pathology and 12 with bathology. Each subject/eye was tested 3 times within a session (3 repeated measures for each subject). Subjects were nested within each configuration, that is, each subject appears in one, unique device-operator configuration only. The design was repeated twice: once for healthy eyes and once for eyes with pathologies.
CenterVue SpA
ia Tommasoo, 77 35131 Padova, Italy www.centervue.com
tel +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com
A 478243 COLA PO CF /P VA 04296580287 Cap. soc .: € 480.000.00 i.v.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular design with a black circle in the center. Surrounding the central circle is a pattern of black and white shapes that resemble buildings or structures. The overall design has a somewhat abstract and urban feel, with the contrasting colors creating a visually interesting composition.
t the center of vision
#### ubiect selection criteria
Twelve subjects with normal eyes and 12 subjects with retinal pathologies were enrolled at two different clinical sites based on the following criteria:
- 4 normal subjects and 4 subjects with retinal pathologies enrolled at site 1 and measured with device 1 by operator 1;
- 4 normal subjects (different from above) and 4 subjects with retinal pathologies (different from above)enrolled at site 1 and measured with device 2 by operator 2;
- 4 normal subjects and 4 subjects with retinal pathologies enrolled at site 2 and measured with device 3 by operator 3;
Within each pathology category, subjects with varying parameters including age and visual acuity were enrolled. For the pathology group, subjects with early and intermediate age-related macular degeneration as well as with mild, moderate, severe and proliferative diabetic retinopathy were enrolled. Diagnosis of retinal pathology was made by a complete eye examination by an ophthalmologist, including dilated funduscopic examination and pertinent history. Subjects met defined inclusion criteria, but none of the exclusion criteria.
#### Precision results
| | Normal | Pathology |
|----------------------------|---------|-----------|
| # subjects | 12 | 12 |
| Overall Mean | 29.7 dB | 23.5 dB |
| Overall Standard deviation | 1.14 dB | 4.23 dB |
| Repeatability SD* | 0.42 dB | 0.75 dB |
| Reproducibility SD** | 0.96 dB | 0.75 dB |
* estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on the same subject using different operators and devices, including repeatability.
#### Individual Grid Point Results
| Group/Parameter | Repeatability SD | Reproducibility SD | |
|-----------------|------------------|--------------------|------|
| Normal | Minimum | 0.94 | 1.06 |
| | Median | 1.40 | 1.80 |
| | Maximum | 2.43 | 2.70 |
| Pathology | Minimum | 1.33 | 1.33 |
| | Median | 2.36 | 2.43 |
| | Maximum | 3.16 | 3.24 |
#### CONCLUSION
As described in this 510(k) Summary, all testing deemed necessary was conducted on the MAIA™ to ensure that the device is safe and effective for its intended use when used in accordance with its Instructions for Use.
CenterVue SpA
via Tommaseo. 77 35131 Padova, Italy www.centervue.com
tel. +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com REA: 378243 CCIAA PD CFJP, NA: 04296580287 Cap. soc.: € 480.000,00 i.v.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
CenterVue SpA c/o Mr. Giuliano Barbaro Managing Director Via Tommaseo 77 Padova, Italy 35131
MAY -2 7 2010
Re: K092187
Trade/Device Name: Macular Integrity Assessment (MAIA ") Device Regulation Number: 21 CFR 886.1605 Regulation Name: Perimeter, Automatic, AC-powered Regulatory Class: Class II Product Code: HPT, HLI Dated: April 22, 2010 Received: April 26, 2010
Dear Mr. Barbaro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for
Ma
Dir
Div
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a black and white graphic that resembles an eye. The pupil is a solid black circle in the center. The iris is a white ring around the pupil, and the rest of the eye is made up of a pattern of black and white shapes that create a spiral effect.
centervue at the center of vision
# Indications for use statement
510(k) Number: K092187
Device Name: Macular Integrity Assessment (MAIATM)
## Indications for use:
The Macular Integrity Assessment (MAIA™) is indicated for measuring macular sensitivity, fixation stability and the locus of fixation, as well as providing infrared retinal imaging. It contains a reference database that is a quantitative tool for the comparison of macular sensitivity to a database of known normal subjects.
Prescription Use M
AND/OR Over-The-Counter Use
Callaway
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092187
via Tommaseo, 77 35131 Padova, Italy www.centervue.com tel. +39 049 781 1828 fax +39 049 781 1899 e-mail: info@centervue.com
REA: 378243 CCIAA PD C.F./P. IVA: 04296580287 Cap. soc.; € 480,000,00 i.v.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.