The normative database provides a statistical reference of retinal and optic disc measurements from known normal subjects to aid in the diagnosis and management of ocular diseases.
Normative Database; Clinical Reference; Diagnostic Aid
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Normative Database Collection; Multi-site clinical data collection
480 healthy individuals (640+ eyes) aged 18-84, diverse ethnic backgrounds; Sample Size: 480 individuals; Number of Sites: 11
Not applicable for this study
Retinal nerve fiber layer, optic disc, and retinal thickness measurements
Indications for Use
The RTVue with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, and optic disk as an aid in the diagnosis and management of retinal disease. The RTVue with Normative Database is also a quantitative tool for the comparison of retina, retinal nerve fiber layer, and optic disk measurements in the human eye to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.
Device Story
RTVue is a computer-controlled optical coherence tomography (OCT) system for posterior segment eye imaging. It captures cross-sectional images of retinal layers (ILM, RNFL, GCC, RPE, outer/total retinal thickness) and optic disc structures. The system uses optical coherence tomography to generate measurements; these are compared against an integrated normative database (NDB) of 480 healthy individuals (ages 18-84). Used in clinical settings by eye care professionals to assess retinal pathologies (e.g., AMD, DME, ERM) and glaucoma. Output includes statistical comparisons of GCC, RNFL, full retinal thickness, and optic disc parameters (cup/rim) against normal ranges. This aids clinicians in detecting and managing ocular diseases by providing quantitative reference data. The device utilizes blood vessel matching for image registration to ensure scan consistency.
Clinical Evidence
Clinical study evaluated precision and reproducibility of RTVue with Normative Database using 36 subjects (12 normal, 12 glaucoma, 12 retina pathology) across 2 sites. Scans included EMM5 (retina), ONH (optic disc/RNFL), and GCC. Primary endpoints were repeatability and reproducibility standard deviations for retinal thickness, cup/rim area, and GCC parameters. Results showed good precision across all groups. Image registration efficacy was validated via accuracy and precision tests. Literature references (Tan et al. 2009, Sehi et al. 2009, etc.) support the device's diagnostic ability and repeatability compared to other OCT systems.
Technological Characteristics
Computer-controlled OCT imaging system; posterior segment measurement. Features include EMM5, ONH, and GCC scan patterns. Software-based normative database (NDB) provides statistical reference for retinal/optic disc measurements. Connectivity: Standalone system. Energy source: Optical (OCT). Image registration: Blood vessel matching algorithm. Materials: Standard ophthalmic device components.
Indications for Use
Indicated for in vivo imaging and measurement of retina, retinal nerve fiber layer (RNFL), and optic disc in human eyes to aid diagnosis and management of retinal and ocular diseases. Used as a quantitative tool for comparing patient measurements against a normative database of healthy subjects.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
StratusOCT with RNFL & Macula Normative Database (K033123)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "optovue" in a stylized font. The "o" in "optovue" is replaced with an image of an eye. There is a curved line above the word "optovue".
# 1101505
## 510(k) Summary
#### Optovue, Inc. RTVue with Normative Database
SEP 1 5 2010
This 510(k) summary for the RTVue with Normative Database is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Manufacturer: | Optovue, Inc.<br>45531 Northport Loop West,<br>Fremont, CA 94538<br>Phone: (510) 623-8868<br>Fax: (510) 623-8668<br>Registration No.: 3005950902 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John Talarico<br>VP of Regulatory and Clinical Affairs<br>Optovue, Inc.<br>Phone: (510) 623-8868 x209<br>e-mail: john_talarico@optovue.com |
| Device Information | |
| Classification: | Class II |
| Trade Name: | RTVue with Normative Database |
| Common Name: | Optical Coherence Tomography (OCT) |
| Classification Name: | Ophthalmoscope, a-c powered (21 CFR§ 886.1570) |
#### Predicate Device
- (1) RTVue with CA (K071250) -- Manufactured by Optovue, Inc.
- (2) StratusOCT with RNFL & Macula Normative Database (K033123) manufactured by Carl Zeiss Meditec, Inc.
#### Intended Use
The RTVue with Normative Database is an optical coherence tomography system indicated for the in-The NY Tue With Norman of the retina, retinal nerve fiber layer, and optic disc as an aid in the diagnosis and management of retinal disease. The RTVue with Normative Database is also a
{1}------------------------------------------------
quantitative tool for the comparison of retinal nerve fiber layer, and optic disc measurements in the human eye to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.
#### Device Description
The RTVue is a computer controlled ophthalmic imaging and measurement system that employs optical coherence tomography to image and measure the posterior segment of the eye. The device is currently cleared for in vivo imaging and measurement of the various retinal layers (K062552). Imaging and measurements include but are not limited to the internal limiting membrane (ILM), the retinal nerve fiber layer (RNFL), the ganglion cell complex (GCC), the retinal pigment epithelium (RPE), the outer retinal thickness, the total retinal thickness and optic disk structures including the cup and neuroretinal rim as an aid in the diagnosis and management of retinal disease. The measurements for the ILM and RPE are height measurements relative to the RPE reference plane. The RNFL. GCC. the outer retinal thickness and total retinal thickness measurements where RNFL is the thickness of the RNFL layer, the GCC is the thickness from the ILM to the inner plexiform layer (IPL), the outer retinal thickness is the thickness from the IPL to the RPE, and total retinal thickness is the thickness from the ILM to the RPE. The current submission is for a software modification to add a normative database. With the additional normative database (NDB), the RTVue can compare the measured data from the GCC, the RNFL, the full retinal thickness, optic disc cup and optic disc rim measurements to the normative database. The RTVue with Normative Database provides a statistical reference of the GCC, the RNFL, the full retinal thickness, optic disc cup and optic disc rim measurements to a database of known normal subjects. The RTVue with Normative Database will provide the analysis information to be used as a clinical reference to aid in the diagnosis and management of ocular diseases.
#### Substantial Equivalence
The RTVue with Normative Database is substantially equivalent to the predicate device identified previously. The RTVue is similar to the previously cleared RTVue device with regard to intended use, operating principle, function, material, and energy source. Technological comparisons and clinical testing demonstrate that the RTVue with Normative Database is functionally equivalent to the predicate devices.
#### Performance Data
Clinical data was collected and evaluated to support the intended use for the RTVue with Normative Database and to demonstrate substantial equivalence to the predicate devices.
#### Description of Precision Study, Subject Selection Criteria and Pathologies
#### Instrument:
Three RTVue (K062552) with three different operators were used in this study. Measurements results of various structures within the eye including retinal thickness, optic disc measurements, and Ganglion Cell Complex (GCC) measurements were compared in normal eyes, and eyes with retina pathology and glaucoma.
#### Scans:
The RTVue device has three scan patterns where measurements are compared to a normative database. These scans include the 1) EMM5 scan, 2) the ONH scan, and 3) the GCC scan. The
{2}------------------------------------------------
EMM5 scan is centered on the fovea and provides measurements of the full retina thickness and the inner retina thickness. Only the full retina thickness measurements are compared to the normative database. This scan can be used to assess possible retina pathologies. The ONH scan is centered on the optic disc and provides measurements of the RNFL and optic disc. This scan can be used to assess possible glaucoma. The GCC scan is centered at 1 mm off from the fovea in order to cover more temporal retina area. It provides measurements of the inner retina layers that contain the ganglion cell axon (RNFL), cell body, and dendrite (inner plexiform layer). This scan can be used to assess glaucoma. Both eyes were scanned in each subject.
#### Subjects:
1) Normal healthy eyes with no ocular pathology, 2) glaucoma patients and, 3) retina patients were enrolled in the study at two clinical sites. Normal subjects were from ocular pathology as determined by the Principal Investigator (PI) at each clinical site. Glaucoma patients were diagnosed as having glaucoma by the PI at each site. Retina patients included subjects with any type of retina pathology diagnosed by the PI. Retina pathologies included but were not limited to, AMD, DME, and ERM. There were 4 normal subjects, 4 retina patients, and 4 glaucoma patients enrolled for each RTVue devices. There were 3 RTVue devices at 2 clinical sites. At one site, there were 2 RTVues systems and 2 operators at different times. In total, there were 12 subjects enrolled for each of the 3 RTVue devices; a total of 36 subjects were used from the study.
#### Selection Criteria:
All data was carefully reviewed for completeness and quality in two levels, namely, the subject level and the individual scan level. At the subject level, the CRF was carefully reviewed to qualify each subject against all inclusion and exclusion criteria by comparing the study protocol with the CRF. At the individual scan level, the data was reviewed for quality to ensure data meeting inclusion criteria was accepted.
In a clinical environment, only scans with acceptable quality should be used. In order to match our analysis with acceptable clinical results, we reviewed and excluded all scans in the study with poor image quality. Image quality is based on a number of factors, including overall signal strength, localized weak signals, eye blink, data out of boundary, and data off-center. Due to the selection criteria, the sample size for each scan type varies.
#### Precision Results
The following tables provide the precision results for the scans with RTVue with Normative Database with both eyes combined. The data in the tables include the number of scans per subject group, the overall mean, the standard deviation, the repeatability standard deviation (median value of subfields with the minimum subfield value and the maximum subfield value), and reproducibility standard deviation (median value of subfields with the minimum subfield value and the maximum subfield value).
- EMM5 Scan Results (Normal and Retina Patients) i)
The following tables show the overall precision results with the EMM5 scan for full retinal thickness in the fovea for normals and retina patients.
| EMM5 | | | |
|-------------------|-----------------|-----------------|-------------------|
| Full Retina Fovea | | | |
| | Normal Patients | Retina Patients | Glaucoma Patients |
{3}------------------------------------------------
| # of Scans | 71 | 71 | 71 |
|---------------------------------|--------------------|--------------------|------------------|
| Overall Mean (Overall SD) | 243.61 (26.00) | 267.15 (73.32) | 246.22 (22.14) |
| Repeatability SD* (Min, Max) | 2.57(1.66, 2.57) | 2.96(1.83, 3.15) | 3.01(1.95, 3.01) |
| Reproducibility SD** (Min. Max) | 17.72(1.66, 17.72) | 13.78(1.83, 13.78) | 3.01(2.05, 5.80) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the EMM5 scan for full retinal thickness in the periphery (average over 8 sectors outside the fovea) for normals and retina patients.
| EMM5<br>Full Retina Thickness Peripheral Results | | | |
|--------------------------------------------------|-------------------|-------------------|-------------------|
| | Normal Patients | Retina Patients | Glaucoma Patients |
| # of Scans | 71 | 71 | 71 |
| Overall Mean (Overall SD) | 295.05 (16.21) | 299.03 (30.70) | 282.30 (25.26) |
| Repeatability SD* (Min, Max) | 1.96(1.66, 2.57) | 2.69(1.83, 3.15) | 2.35(1.95, 3.01) |
| Reproducibility SD ** (Min, Max) | 2.01(1.66, 17.72) | 2.69(1.83, 13.78) | 2.66(2.05, 5.80) |
*estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
ii) ONH Scan Results (Normals and Glaucoma Patients)
The following tables show the overall precision results with the ONH scan for average RNFL thickness for normals and glaucoma patients.
| | ONH<br>Average RNFL | |
|---------------------------------|---------------------|-------------------|
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 72 |
| Overall Mean (Overall SD) | 100.82 (8.11) | 86.40 (14.99) |
| Repeatability SD* (Min, Max) | 1.76(1.76, 6.18) | 1.56(1.56, 5.86) |
| Reproducibility SD** (Min, Max) | 2.11(2.11, 9.78) | 7.87(3.92, 16.33) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the ONH scan for the average over 2 hemi-spheres, 4 quadrants, and 8 sectors for normals and glaucoma patients.
| | ONH | |
|---------------------------|----------------------------------------------------------|-------------------|
| | RNFL averaged over 2 Hemispheres, 4 Quadrants, 8 Sectors | |
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 72 |
| Overall Mean (Overall SD) | 100.82 (14.80) | 86.41 (18.55) |
{4}------------------------------------------------
| Optovue RTVue with NDB 510(K) Premarket Notification | | |
|------------------------------------------------------|--|--|
|------------------------------------------------------|--|--|
| Repeatability SD* (Min, Max) | 3.92(1.76, 6.18) | 4.48(1.56, 5.86) |
|---------------------------------|------------------|-------------------|
| Reproducibility SD** (Min, Max) | 5.40(2.11, 9.78) | 7.84(3.92, 16.33) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the ONH scan for the cup area measurement for normals and glaucoma patients.
| ONH<br>Cup Area | | |
|---------------------------------|-------------------|-------------------|
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 72 |
| Overall Mean (Overall SD) | 0.39 (0.18) | 1.04 (0.53) |
| Repeatability SD* (Min, Max) | 0.02 (0.02, 0.03) | 0.07(0.07, 0.21) |
| Reproducibility SD** (Min, Max) | 0.02 (0.02, 0.03) | 0.07(0.07, 0.21) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
**estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the ONH scan for the rim area measurement for normals and glaucoma patients.
| | ONH | |
|---------------------------------|-------------------|-------------------|
| Rim Area | | |
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 72 |
| Overall Mean (Overall SD) | 0.69 (0.17) | 0.89 (0.66) |
| Repeatability SD* (Min, Max) | 0.03 (0.02, 0.03) | 0.18(0.07, 0.21) |
| Reproducibility SD** (Min, Max) | 0.03 (0.02, 0.03) | 0.18(0.07, 0.21) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the ONH scan for the vertical CD ratio measurement for normals and glaucoma patients.
| ONH<br>Vertical CD Ratio | | |
|---------------------------------|-------------------|-------------------|
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 72 |
| Overall Mean (Overall SD) | 2.15 (0.65) | 0.76 (0.23) |
| Repeatability SD* (Min, Max) | 0.01 (0.01, 0.04) | 0.05(0.05, 0.08) |
| Reproducibility SD** (Min, Max) | 0.01 (0.01, 0.04) | 0.05(0.05, 0.08) |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
{5}------------------------------------------------
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
#### iii) GCC Scan Results (Normals and Glaucoma Patients)
The following tables show the overall precision results with the GCC scan for the inner retina average for normals and glaucoma patients.
| | GCC<br>Inner Retina Average | |
|----------------------------------|-----------------------------|-------------------|
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 71 |
| Overall Mean (Overall SD) | 98.28 (9.31) | 81.76 (8.90) |
| Repeatability SD* (Min, Max) | 1.86 (1.86, 2.07) | 1.18 (1.18, 1.59) |
| Reproducibility SD ** (Min, Max) | 7.56 (7.32, 7.92) | 2.55 (1.59, 3.36) |
*estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning.
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the GCC scan for the FLV thickness measurement for normals and glaucoma patients.
| GCC<br>GCC FLV | | |
|---------------------------------|-------------------|-------------------|
| | Normal Patients | Glaucoma Patients |
| # of Scans | 72 | 71 |
| Overall Mean (Overall SD) | 0.60 (0.67) | 4.30 (3.47) |
| Repeatability SD* (Min, Max) | 0.37 (0.37, 0.37) | 0.88 (0.88, 0.88) |
| Reproducibility SD** (Min, Max) | 0.46 (0.46, 0.46) | 1.04 (1.04, 1.04) |
*estimate of the standard deviation among measurements taken on the same operator and device in the same testing session with repositioning.
**estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
The following tables show the overall precision results with the GCC scan for the GLV thickness measurement for normals and glaucoma patients.
| GCC<br>GCC GLV | | | | | |
|----------------------------------|-------------------|-------------------|--|--|--|
| | Normal Patients | Glaucoma Patients | | | |
| # of Scans | 72 | 71 | | | |
| Overall Mean (Overall SD) | 3.99 (3.43) | 15.51 (7.81) | | | |
| Repeatability SD* (Min, Max) | 0.83 (0.83, 0.83) | 1.03 (1.03, 1.03) | | | |
| Reproducibility SD ** (Min, Max) | 2.44 (2.44, 2.44) | 1.71 (1.71, 1.71) | | | |
*estimate of the standard deviation among measurements taken on the same subject using the same operator and device in the same testing session with repositioning.
{6}------------------------------------------------
** estimate of the standard deviation among measurements taken on different subjects using different operators and devices, including repeatability; in some cases the maximum reproducibility estimates were impacted by mean thickness difference between subjects in different sites.
#### Registration Results
Image registration using blood vessel matching is used for both the EMM5 scan and the ONH scan. The image registration algorithm uses a reference image to register all scans based on matching detected blood vessel patterns. For this internal validation study, we performed both an accuracy test as well as a precision test in order to evaluate the efficacy of the blood vessel registration method.
The following tables show the values at 2 times the standard deviation values for each parameters for these two scan types.
| EMM5 | FRT Fovea | FRT Para Tempo | FRT Para Superior | FRT Para Nasal | FRT Para Inferior | FRT Peri Tempo | FRT Peri Superior | FRT Peri Nasal | FRT Peri Inferior |
|------|-----------|----------------|-------------------|----------------|-------------------|----------------|-------------------|----------------|-------------------|
| | 6.345 | 3.9121 | 4.6186 | 3.6347 | 2.795 | 5.568 | 5.669 | 4.772 | 3.63 |
| റപ്പ് | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>י זיח | am<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | CL T | No. one possible from the may<br>S. YO Y | | די | 6<br>1 | The first of the county of<br>1 |
|-------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|------------------------------------------|---|----|----------|---------------------------------|
| 0<br> | e | € | ---------<br>1<br>----------------- | | C | | S<br>0.0 | |
#### Normative Databases
The normative data was collected at 11 clinical sites following an IRB approved protocol and enrolled known healthy eyes from 480 individuals (over 640 eyes) with a wide age range (18-84) from various ethnic backgrounds including 33% Caucasians, 22% Asians, 20% African Americans, 12% Hispanics, 12% Indian, and 1% other, approximately. Normative databases are adjusted by age, signal strength and disc area (where applicable), not by axial length, refraction or any other parameters. The normative limits do not take into account differences that may be present due to ethnicity.
In our clinical studies, the repeatability and reproducibility were found to be very good and were found to be similar for normal groups and pathology groups. We believe the Normative Database is a good representation of the patient population because it was created from a large population of patients from 11 clinical sites worldwide. Ocular measurements taken by the RTVue have been compared to other OCT devices by renowned leaders in the field of Ophthalmology and found to have good repeatability and reproducibility 1-5.
#### Conclusion
As described in this 510(k) Summary, all testing and analysis were completed on the RTVue with Normative Database to ensure that the device is safe and effective for its intended.
Tan O, Chopra V, Lu A, Schunan J, Ishkawa H, Wollstein G, Varma R, Huang D. Detection of macular ganglion cell loss in glaucoma by Fourier l. Domain Optical Coherence Tomography. Ophthalmol. 2009; 116:2305-2314.
Donall Optical Collection Touristic 2007, 10:230, 10.230, 10.230, 10.12 Reproducibility of RTVer Layer Thickness 2. Optic Disc Measurements and Agreement with Straus Optical Coherence Tomography Measurements. Am J Ophthalment (1909 Mar 5
Opic Disc Measurements and Agreement Will Suatts Optical Contractions Frin Opiliations Print of Partin Province Province Printer Price Printer Print Province Price Portugio 3. tomography. Ophthalmol. 2010; 117: 738-746.
toniolality. Ophilimino. 2000, 11:17 Comparison of repeatability of retinal nerve fiber thickness measurement made with the RT 4. domain Optocal Collerence Tomograph and the GDx Scanning Laser Polarimeter with Variable or Enhanced Comea Compensation. J Glaucoma. Oct
Sehi M, Grewal DS, Sheets CW, Greenfield DS. Diagnostic ability of Fourier-domain optical coherence tomography for glaus 5. detection. Am J Ophthalmol. 2009; 148: 597-605.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Optovue, Inc c/o Mr. John Talarico VP Regulatory and Clinical Affairs 45531 Northport Loop W. Fremont, CA 94538
SEP 1 5 2010
Re: K101505
Trade/Device Name: RTVue with Normative Database Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLI Dated: August 12, 2010 Received: August 19, 2010
Dear Mr. Talarico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Mr. John Talarico
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kevia Alexander
Malvina B. Eydelman, M.D. Director
Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use
KI0ISOS
510(k) Number (if known):
SEP 1 5 2010
Device Name: RTVue with Normative Database
Indications For Use:
The RTVue with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, and optic disk as an aid in the diagnosis and management of retinal disease. The RTVue with Normative Database is also a quantitative tool for the comparison of retina, retinal nerve fiber layer, and optic disk measurements in the human eye to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sesn.
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K101505 Page 18 of 922
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.