K180820 · Crystalvue Medical Corporation · HKX · Dec 14, 2018 · Ophthalmic
Device Facts
Record ID
K180820
Device Name
Tono Vue Non-Contact Tonometer
Applicant
Crystalvue Medical Corporation
Product Code
HKX · Ophthalmic
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The TonoVue is a non-contact tonometer that is intended to measure the intraocular pressure of the human eye in vivo.
Device Story
TonoVue is a desktop non-contact tonometer (NCT) for measuring intraocular pressure (IOP). It operates by delivering a soft air puff to the eye; an electro-optical system monitors corneal deformation. Based on the Imbert-Fick principle, the device calculates IOP by dividing the air pressure by the area of the applanated corneal surface. The system features full auto-alignment and includes a built-in thermal printer. It is intended for use by clinicians in a clinical setting. The device provides automated IOP measurements, which assist healthcare providers in assessing ocular health and managing conditions related to intraocular pressure. The non-contact nature of the measurement provides a patient benefit by eliminating the need for physical contact with the eye during the procedure.
Clinical Evidence
Clinical study conducted per ISO 8612:2009 with 120 patients (corneal astigmatism < 3 diopters). Primary endpoint: validity of IOP measurements compared to a reference tonometer. Results showed no more than 5% of paired differences between the reference and test tonometer exceeded the ±5mmHg tolerance, meeting ISO 8612:2009 requirements.
Technological Characteristics
Desktop non-contact tonometer; dimensions 500mm x 260mm x 500mm. Sensing: electro-optical system for corneal deformation monitoring and pressure sensor. Energy: AC-powered. Biocompatibility: ISO 10993-1 compliant. Safety/EMC: IEC 60601-1 and IEC 60601-1-2 compliant. Software: automated alignment and calculation algorithm.
Indications for Use
Indicated for measuring intra-ocular pressure in patients with less than 3 diopters of corneal astigmatism.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
July 22, 2021
Crystalvue Medical Corporation Oliver Lin Director Of Quality Assurance No. 116, Ln.956, Zhongshan Rd., Taoyuan Dist., Taovuan, 33072 Taiwan
Re: K180820
Trade/Device Name: Tono Vue Non-Contact Tonometer Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and accessories Regulatory Class: Class II Product Code: HKX
Dear Oliver Lin:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 14, 2018. Specifically, FDA is updating this SE Letter as an administrative correction because we identified the typo in the tradename of your device referenced in your 510(k) Summary.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Elvin Ng, OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, (240) 402-4662, elvin.ng(@fda.hhs.gov.
Sincerelv.
# Charles Chiang -S
for Elvin Ng
Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue letters.
December 14, 2018
Crystalvue Medical Corporation Oliver Lin Director of Quality Assurance No. 116, Ln.956, Zhongshan Rd., Taoyuan Dist., Taoyuan City 33072 Taiwan
Re: K180820
Trade/Device Name: Tono Vue Non-Contact Tonometer Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and accessories Regulatory Class: Class II Product Code: HKX Dated: November 2, 2018 Received: November 7, 2018
Dear Oliver Lin:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
## Bradley S. Cunningham -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K180820
Device Name TonoVue Non-Contact Tonometer
Indications for Use (Describe)
The TonoVue is a non-contact tonometer that intra-ocular pressure of the eye in patients with less than 3 diopters of corneal astigmatism.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
|× Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Crystalvue" in a stylized font. The "C" in "Crystal" is replaced with a circular graphic that resembles a crystal or a lens. The word "Crystal" is written in black, while the "vue" portion of the word is written in a blue gradient.
TonoVue 510(K) Premarket Notification
### 510(k) Summary
Crystalvue Medical Corporation TonoVue Non-Contact Tonometer
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
#### Submitter
| Company name: | Crystalvue Medical Corporation |
|------------------|-----------------------------------------------------------------------------|
| Company Address: | No.116, Ln.956, Zhongshan Rd.,<br>Taoyuan Dist., Taoyuan City 33072, Taiwan |
| Contact Person: | Oliver Lin<br>Director of Quality Assurance |
| Phone: | +886 3 360 7711 Ext.2051 |
| Fax: | +886 3 360 7722 |
| E-mail: | Oliver.Lin@crystalvue.com.tw |
#### Device Information
| Classification: | Class II |
|----------------------|-------------------------------|
| Trade Name: | TonoVue Non-Contact Tonometer |
| Common Name: | Non-Contact Tonometer |
| Classification Name: | Tonometer, AC-Powered |
| Product Code: | 21 CFR § 886.1930 |
#### Predicate Devices
Trade Name: Canon Full Auto Tonometer TX-20. Classification Name: HKX, Tonometer and accessories 510(k) Number: K111710
#### Intended Use
The TonoVue is a non-contact tonometer that is intended to measure the intraocular pressure of the human eye in vivo.
#### Indication for Use
The TonoVue is a non-contact tonometer that intended to measure the intra-ocular pressure of the eye in patients with less than 3 diopters of corneal astigmatism.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Crystalvue. The logo features a stylized snowflake to the left of the word "Crystalvue". The word "Crystal" is in black, and the word "vue" is in blue.
#### Device Description
TonoVue Tonometer is designed to non-contact tonometer (NCT) that measures the intraocular pressure (IOP) by delivering a soft air puff without contacting eyes directly. It's designed as a full auto-alignment and All-In-One desktop type medical device with built-in thermal line printer. The dimensions of whole device are about 500mm (H) x 260mm (W) x 500mm (L). The AC power input port and a USB port are set at the bottom side of the device, but the USB port only for engineering use. Based on the Imbert-Fick principle, the IOP is calculated by dividing the amount of air pressure into the area of applanated surface. TonoVue utilizes a rapid air puff to apply force for flattening the cornea of human eye, and an advanced electro-optical system to monitor its deformation. The puff force increases until the cornea is applanated over the predetermined area and detected by the pressure sensor inside TonoVue, and the IOP can be calculated.
#### Safety
Electrical safety and EMC testing were conducted on the TonoVue device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
#### Effectiveness
TonoVue tonometer is in general comparable to the predicate devices, regarding the use, design and technical characteristics.
The result of clinical testing is also proved the effectiveness of TonoVue tonometer in clinical use is same as other products available on the market.
#### Substantial Equivalence
The TonoVue is substantially equivalent to the predicate devices: CANON Full Auto Tonometer TX-20 (K111710). It has the same intended use, technological characteristics, and principles of operation as its predicate devices. Bench and clinical testing demonstrate substantially equivalent performance to the predicates. Performance data demonstrate that the TonoVue is as safe and effective as the predicate devices. Thus, the TonoVue is substantially equivalent with Canon TX-20.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Crystalvue" in a stylized font. The "C" in "Crystal" is replaced with a circular, crystalline design, while the rest of the word "Crystal" is in black. The word "vue" is in a blue color, contrasting with the black of "Crystal". The overall design is clean and modern.
#### Performance Data
#### (a) Software Validation
The Software verification and validation testing were conducted and documentation was provided as recommended by FDA Guidance- "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Application of risk management to medical devices to show the software used in the TonoVue Tonometer is conform the safety principles.
#### (b) Biocompatibility Testing
Biocompatibility test evaluation for TonoVue non-contact The tonometer was conducted in accordance with the ISO10993-1. The test result of biocompatibility test is complies the ISO 10993-1 standard.
#### (c) Bench Testing
According to the FDA guidance of Tonometers - Premarket Notification Submissions requirement, the bench testing is performed to evaluate the accuracy and reproducibility of the TonoVue.
According to the testing result of accuracy and reproducibility, its prove the TonoVue non-contact tonometer's IOP measurement result can meet our product's specification.
#### (d) Clinical Testing
This study is followed the requirement of ISO 8612:2009 Ophthalmic instruments-tonometers, to test the validity of NCT Tonometer. This clinical testing of TonoVue non-contact tonometer was study the 120 patients with less than 3 diopters of corneal astigmatism.
The test result of the TonoVue non-contact tonometer is met the requirement of ISO 8612:2009:
No more than 5% of the paired differences between the reference tonometer reading and the test tonometer reading for each pressure range are greater than the tolerance ±5mmHg.
Thus, the test instrument should be feasible when use in clinical.
#### Conclusion
As described in this 510(k) Summary, comprehensive testing and
{7}------------------------------------------------
#### TonoVue 510(K) Premarket Notification
### Orystalvue
analysis was conducted on the TonoVue to ensure that the device is safe and effective for its intended use when used in accordance with its instructions for use.
The Performance Data demonstrate that TonoVue is as safe and effective as predicate device, Canon Full Auto Tonometer TX-20. Based on the information in this submission, the TonoVue has the same intended use, technological characteristics, and principles of operation as its predicate devices. Therefore, the TonoVue is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.