The VISX WaveScan™ Wavefront System is a diagnostic instrument indicated for the automated measurement and analysis of refractive errors of the eye including hyperopia and myopia from +6.00 to -8.00 diopters spherical, and astigmatism from 0.00 to -6.00 diopters.
Device Story
WaveScan™ Wavefront System Model HS 1 is a diagnostic autorefractor; measures refractive error via wavefront technology. Device directs light into eye; sensors detect gradient/slope of wavefronts exiting eye after reflection off retina. Internal sensors, lenses, and apertures capture irregularities caused by cornea and lens. Computer processes data via mathematical algorithms to calculate refractive error. Used in clinical settings by professionals or ancillary personnel to obtain objective measurements of spherical, cylindrical, and axis refractive error. Output provided as printed or electronic read-out. Assists clinicians in preliminary diagnostic refractive error determination. Benefits patient by providing precise, automated refractive measurements.
Clinical Evidence
Bench testing only. Comparison conducted against Canon R-50m using a Gullstrand Standard Test Eye Model. Tested across various refractive error ranges (myopia, hyperopia, astigmatism). Each condition repeated five times. Results showed performance within 95% confidence level; WaveScan demonstrated equivalent or superior accuracy and repeatability with lower variability than the control device.
Technological Characteristics
Diagnostic autorefractor; uses near-infrared light source. Measures wavefront gradients via sensors, lenses, and apertures. Range: Sphere -8 to +6D, Cylinder 0 to -6D, Axis 0-180 degrees. Up to 100 measurement points. Power: 110/120V, 50/60Hz. Data output: Print or electronic. Software-based mathematical analysis of wavefront distortions.
Indications for Use
Indicated for automated measurement and analysis of refractive errors (hyperopia, myopia, astigmatism) in patients requiring refractive assessment. Range: +6.00 to -8.00 diopters spherical; 0.00 to -6.00 diopters astigmatism.
Regulatory Classification
Identification
An ophthalmic refractometer is an automatic AC-powered device that consists of a fixation system, a measurement and recording system, and an alignment system intended to measure the refractive power of the eye by measuring light reflexes from the retina.
Predicate Devices
Canon R-50m
Submission Summary (Full Text)
{0}------------------------------------------------
MAY - 2 2000
Image /page/0/Picture/1 description: The image shows the logo for Visx. Above the name is a curved line with a circle in the middle. Below the name is the tagline "WE MAKE THINGS CLEAR."
K000327
## Summary Statement
SUBMITTER:
Submitted by:
Company Name: VISX, Inc. 3400 Central Expressway Santa Clara, California 95051 (408) 773-7327 (408) 773-7324
CONTACT PERSON:
David M. Patino Vice President, Regulatory and Clinical Affairs
DATE SUMMARY PREPARED:
Address:
Phone:
Fax:
WaveScan™ Wavefront System Model HS 1
TRADE NAME:
COMMON NAME:
Autorefractor
January 31, 2000
# SUBSTANTIALLY EQUIVALENT TO:
The WaveScan™ Wavefront System Model HS 1 is substantially equivalent to the predicate device, the Canon R-50m. Both devices have the same indication for use, in that they are devices which are diagnostic tools for obtaining objective output measurements of refractive error of the eye. These devices obtain this information by means of light ray deflection principles to measure the refractive elements of the eye and provide automated read-outs of refractive error. The principle of light waves incorporates the known elements of light energy moving in successive flat sheets, or wavefronts. The measurement principles of both instruments utilize the element of wavefront measurements from the refractive elements of the eye, the lens and cornea as light interference changes the wavefront as affected by these elements. Each instrument utilize condensing lenses within the body of the instrument, and a fogging technique which blurs a fixation target for the patient who is being measured. The output measures for each instrument consist of the spherical, cylindrical and axis of the subject's refractive error which are provided in a printed read-out format.
{1}------------------------------------------------
## DESCRIPTION of the DEVICE:
The WaveScan™ Wavefront System Model HS 1 autorefractor device is a diagnostic instrument designed to measure refractive error of the eye automatically by use of wavefront technology. Light travels in a procession of flat sheets known as wavefronts. As these wavefronts pass through an imperfect refractive medium including the cornea and the lens, the aberrations which are created by the irregular surfaces "wrinkle" the light rays and create wavefront errors or distortions. The instrument contains tiny sensors which measure the gradient, or slope, of the wavefront which emanate from the eye. After light travels through the eye's optical system and out again, the sensors accurately detect slight variations of wavefront irregularities as they exit the eye. The sensors then provide additional information within the confines of the instrument through a series of lenses and apertures which are subject to mathematical algorithms and software. Once analyzed by the computer, a refractive error read-out is provided to the user. This analysis is made from multiple points of light which precisely pinpoint variations in refractive status across the entrance pupil of the eye. This allows for the high level of accuracy of the instrument thus providing the user with very precise readings of refractive error.
## INDICATIONS FOR USE:
The VISX WaveScanted System Model HS 1 is a diagnostic instrument indicated for the automated measurement and analysis of refractive errors of the eye including hyperopia and myopia from +6.00 to -8.00 diopters spherical, and astigmatism from 0.00 to -6.00 diopters. I
| | VISX WaveScan |
|-----------------------|-------------------------------------------------------------------------------------------------|
| Accuracy | No Specification |
| Reproducibility | No Specification |
| Power<br>Requirements | 110/120V, 50/60Hz |
| Measurable<br>Range | Sphere -8 to +6 in 0.1D increments<br>Cylinder 0 to -6 in 0.1D increments<br>Axis 0-180 degrees |
| Data Output | Print or Electronic |
| Approvals | ETL |
| Measurement<br>Points | Up to 100 points |
#### SPECIFICATIONS OF THE DEVICE:
{2}------------------------------------------------
## TESTING OF THE DEVICE:
Comparison testing was conducted under a protocol designed to demonstrate the equivalence of the WaveScan™ Wavefront Model HS 1 to the Canon Model R-50m autorefractor. These devices which are basically equivalent, measure refractive error of the eye automatically by use of scanning light deflection through the optics of the eye. The near infrared light acts as a secondary light source reflected off the back of the eye at the retina. The emerging light passes through the optics of the lens and cornea and is affected by a series of apertures and lenses which are then systematically analyzed through mathematical models and reveal the measurement of refractive error. The tests were conducted using a model test eve developed by VISX. Inc. and modeled after the Gullstrand Standard Test Eye Model established for bench testing in the ophthalmic community. Tests were conducted across a range of refractive errors including myopia, hyperopia, and astigmatic errors of refraction at various and orientations. Combinations of myopic and hyperopic astigmatism were employed. Each of these test conditions was repeated five times in each instrument, and subject to statistical analysis. The results indicated that the WaveScan™, Wavefront System performed within statistical 95% level of confidence in all parameters measured and was equivalent or superior to the control instrument in accuracy and repeatability. Overall, the WaveScan device provided estimates of refractive error with less variability than the control device as seen in the data, from standard deviation estimates.
## LABELING
The WaveScan™ Wavefront System Model HS 1 is provided to the user with a User Manual of Operator's Instructions. The device is a diagnostic tool used by professionals and ancillary personnel in the determination of preliminary diagnostic refractive error information.
> VISX. Inc. 3400 Central Expressway Santa Clara, CA 9505 1 (408) 733-2020
> > GOODS B
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a bird-like figure, possibly representing an eagle or other national symbol. The image is in black and white.
MAY - 2 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David M. Patino Vice President, Regulatory and Clinical Affairs VISX, Inc. 3400 Central Expressway Santa Clara, CA 95051-0703
Re: K000327
Trade Name: WaveScan™ Wavefront System Regulatory Class: I Product Code: HKO Dated: January 31, 2000 Received: February 2, 2000
Dear Mr. Patino:
We have reviewed your Section 510(k) notification of intent to market the device we have leviewed your bowe determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device in interstace commerce prior to thay been reclassified in accordance with the provisions of Amendinents, of to devices that have ester (Act). You may, therefore, market the device, the rederal 1 000, Drug, and Cosmisions of the Act and the limitations described below. subject to the general controls pool the Act include requirements for annual registration, I no golf devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that This device will be used for an intended use not identified in the proposed labeling and that this device will be ased for an inteliare, in accordance with Section 513(i)(1)(E) of the Act, such use coura cause na must appear in the Warnings section of the device's labeling:
The safety and effectiveness of the WaveScan™ Wavefront System have not been I the Jacesy and encess release as an accessory interfaced to a refractive laser for the established for ase of the aberrations of the eye by photorefractive keratectomy (PRK), ercatinent of might or orectory (PTK) or laser-assisted in situ keratomileusis (LASIK).
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major ( remaines affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the
{4}------------------------------------------------
## Page 2 - David M. Patino
Current Good Manufacturing Practice requirements, as set forth in the Quality System Current Good Manufacturing Fress: General regulation (21 CFR Part 820) and that, Regulation (Q) for Mechear Devices. Sood and Drug Administration (FDA) will verify through periodic Q5 inspections, the GMP regulation may result in regulation may result in regulatory such assumptions. Tandre to comply while announcements concerning your device in action. In addition, I DA may publish response to your premarket notification submission in the Federal Register. Thease noter thight have under sections 531 through 542 of the Act does not allect any obligation you might nation Control provisions, or other Federal laws or regulations.
The FDA finding of substantial equivalence of your device to a legally marketed predicate I he FDA inding of substantial equit device and permits your device to proceed to the device results in a Classification for you to begin marketing your device as described in your market. This letter will anow you to bight marketing ,
510(k) premarket notification if the limitation statement above is added to your labeling, as described.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Prease note that the above labeling inned before these limitations are modified in any way or removed from the device's labeling.
If you desire specific information about the application of other labeling requirements to If your desire specific information abditionally 809.10 for in vitro diagnostic devices).
your device (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices your device (21 CFR Part of Compliance at (301) 594-4613. Additionally, for questions on please contact the Office of Compliance av ise, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to (201) 594-607. Thise, picase note 97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers responsibilities under the Net may De 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Robert R. Satting/
రా
David W. Feigal, Jr., M.D., M.P.H. Acting Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page / of
510(k) Number (if known): K000327
Device Name: _ WaveScan™ Wavefront System
Indications For Use:
The VISX WaveScan™ Wavefront System is a diagnostic instrument indicated for the automated measurement and analysis of refractive errors of the eye including hyperopia and myopia from +6.00 to -8.00 diopters spherical, and astigmatism from 0.00 to -6.00 diopters.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eurenett Bellum
510(k) Number
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.