XyCAM RI, XyCAM RI System, XyCAM Retinal Imaging System
K193319 · Vasoptic Medical, Inc. · HKI · Jun 1, 2020 · Ophthalmic
Device Facts
Record ID
K193319
Device Name
XyCAM RI, XyCAM RI System, XyCAM Retinal Imaging System
Applicant
Vasoptic Medical, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Jun 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The XyCAM RI System is intended to capture, store, and display images of dynamic blood flow changes in the human retina, and to provide quantitative outputs of relative blood flow information.
Device Story
XyCAM RI is a non-invasive retinal blood flow imaging system. It uses Laser Speckle Contrast Imaging (LSCI) with a 785 nm near-infrared laser diode to illuminate the retina and a high-speed camera to capture image stacks. The system PC processes these images by analyzing spatio-temporal pixel intensity variations to calculate a Blood Flow Velocity index (BFVi). The device is used in a clinical setting by an operator to generate maps of retinal blood flow velocity. It includes gaze fixation targets for the patient. The output is displayed on a monitor for the clinician to assess spatial and temporal trends in blood flow. It does not provide automated diagnostics or clinical decision support; it serves as an imaging tool to assist clinicians in monitoring retinal blood flow. Optional accessories include a pulse oximeter.
Clinical Evidence
Clinical validation involved 20 healthy subjects (9 male, 11 female; mean age 46.9 ± 16.6 years). Study assessed repeatability and reproducibility of blood flow measurements across different operators and imaging sessions (Optic Nerve Head and macula). Results showed mean intra-session coefficient of variation (CV) of 2.883% (ONH) and 2.782% (macula), and mean inter-session CV of 7.069% (ONH) and 6.447% (macula). Bench testing using a calibrated flow pump confirmed reliability with intra-session CV < 10%.
Technological Characteristics
Non-invasive ophthalmic camera using Laser Speckle Contrast Imaging (LSCI). Illumination: 785 ± 10 nm near-infrared laser diode. Field angle: 20-25 degrees. Working distance: 21 mm. Connectivity: System PC with keyboard/mouse/monitor. Standards: ANSI/AAMI ES60601-1 (safety), IEC 60601-1-2 (EMC), ANSI Z80.36-2016 (ophthalmic), IEC 60825-1 (laser safety). Software: Moderate level of concern.
Indications for Use
Indicated for capturing, storing, and displaying images of dynamic blood flow changes in the human retina and providing quantitative relative blood flow information in patients.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 1, 2020
Vasoptic Medical Inc. Abhishek Rege President 1215 East Fort Avenue, Suite 304 Baltimore, Maryland 21230
Re: K193319
Trade/Device Name: XyCAM RI, XyCAM RI System, XyCAM Retinal Imaging System Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: November 27, 2019 Received: December 2, 2019
Dear Abhishek Rege:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193319
Device Name XyCAM™ Retinal Imager - XyCAM RI
Indications for Use (Describe)
The XyCAM RI System is intended to capture, store, and display images of dynamic blood flow changes in the human retina, and to provide quantitative outputs of relative blood flow information.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "K193319" on the left side of the image. To the right of the text is a blue logo that looks like a stylized "W". Below the logo is the text "VASOPTIC MEDICAL INC.". The logo and text are likely the logo for the company Vasoptic Medical Inc.
# 510(k) Summary
#### I. Submitter Details
| Submitter Name | Vasoptic Medical Inc. |
|------------------------|-------------------------------------------------------|
| Submitter Address | 1215 East Fort Avenue, Suite 304, Baltimore, MD 21230 |
| Submitter Phone Number | 443-961-3591 |
| Primary Contact Person | Abhishek Rege |
| Primary Contact Email | Abhishek.Rege@vasoptic.com |
| Date Prepared | November 27, 2019 |
#### II. Device Information
| Device Name | XyCAM Retinal Imager (XyCAM RI) |
|---------------------------|---------------------------------|
| Device Identification | Ophthalmic Camera Device |
| Product Code | HKI |
| Regulation Number | 21 CFR 886.1120 |
| Regulatory Classification | II |
#### III. Predicate Device
| Predicate Device | Retinal Functional Imager 3000 (RFI 3000), K080180 |
|------------------|----------------------------------------------------|
| Reference Device | Laser Speckle Flowgraphy LSFG-NAVI, K153239 |
#### IV. Device Description
The XyCAM™ RI System (or XyCAM RI) is a non-invasive retinal blood flow imaging device. It generates images of one or more regions of the retina by using an imaging technique called Laser Speckle Contrast Imaging (LSCI). This technique utilizes a lowintensity laser beam to illuminate the back of the eye and a high-speed, high-sensitivity camera to acquire a stack of images. The captured image stack is then processed such that the intensity at every image pixel is processed in conjunction with the intensities of other spatio-temporally adjacent pixels to obtain a speckle contrast value at the said pixel. which is indicative of blood flow velocity at the said pixel. Accordingly, the XyCAM RI is able to generate and display a map of Blood Flow Velocity indices within the imaged field of view.
The XyCAM RI System consists of following modules:
- The XyCAM RI contains a module to provide illumination to the back of the eye, ● various optical components to manipulate light along the illumination and imaging path, a high-speed camera to capture images, visual targets for the imaged subject to fixate his/her gaze on during imaging, and electronic and mechanical components for control and operation.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Vasoptic Medical Inc. The logo features a stylized letter "W" in blue gradient, with a small gray circle to the right of the "W". Below the "W", the text "VASOPTIC MEDICAL INC." is written in a simple font.
- System PC contains the XyCAM RI Software. The Operator uses the XyCAM RI ● System PC keyboard, mouse, and monitor to interact with the XyCAM RI Software, which controls the XyCAM RI operation including management of user and patient information, control of image acquisition, facilitation of image analysis, and generation of imaging results.
- Accessories include an imager station and an external hard drive.
- A pulse oximeter is included as an optional third-party device.
The XyCAM RI Software outputs Blood Flow Velocity indices (arbitrary units) that can be assessed for spatial and temporal trends in user defined regions.
- V. Intended Use
The XyCAM RI System is intended to capture, store, and display images of dynamic blood flow changes in the human retina, and to provide quantitative outputs of relative blood flow information.
| Feature | XyCAM RI System | RFI 3000<br>(Predicate Device) | LSFG-NAVI<br>(Reference Predicate) |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K)<br>Number | K19xxxx | K080180 | K153239 |
| Product Code | HKI (Primary) | HKI (Primary),<br>HLI (Secondary) | HKI (Primary),<br>HLI (Secondary) |
| Classification<br>Name | Ophthalmic Camera<br>Device; AC Powered<br>(21 CFR 886.1120) | Ophthalmic Camera Device; AC Powered (21 CFR 886.1120) Ophthalmoscope (21 CFR 886.1570) | Ophthalmic Camera Device; AC Powered (21 CFR 886.1120) Ophthalmoscope (21 CFR 886.1570) |
| Class | II | II | II |
| Method | Imaging is performed<br>using coherent<br>illumination (laser<br>diode). | Imaging is performed<br>using illumination<br>similar to<br>conventional fundus<br>camera. | Image using Diode<br>Laser.<br>A mydriatic agent is<br>not required in a<br>darkroom (displayed<br>on PC monitor). |
| Feature | XyCAM RI System | RFI 3000<br>(Predicate Device) | LSFG-NAVI<br>(Reference Predicate) |
| | A mydriatic agent is not<br>necessary, but may be<br>used if needed. | A mydriatic agent is<br>required. | |
| Indication for Use | The XyCAM RI system<br>is intended to capture,<br>store and display images<br>of real time dynamic<br>blood flow changes in<br>the human retina, and to<br>provide quantitative<br>outputs of relative blood<br>flow information. | The RFI 3000 is<br>intended to observe,<br>capture, display, and<br>store images of<br>patients' fundus<br>(retina) under<br>mydriatic conditions<br>to aid in diagnosing<br>or monitoring<br>diseases of the eye<br>that may be observed<br>and photographed. | The LSFG-NAVI<br>system is intended to<br>capture and display the<br>blood flow distribution<br>in the human retina in<br>real time, and to<br>monitor the blood flow<br>in retinal vessels for<br>their quantitative<br>evaluation. |
| Observation<br>System | Software live view | Through eyepiece<br>with reticle or<br>through software live<br>view | Software live view |
| Imaging Light<br>Source | Near-Infrared (785 ± 10<br>nm) Laser Diode | Xenon flash lamp<br>420WS (multi-flash) | Laser Diode (830 nm) |
| Flash<br>Exposure | Continuous exposure | Automatic, 1 ms per<br>pulse, train of up to 8<br>pulses, repetition rate<br>17.5ms. | Continuous exposure |
| Focusing<br>Light Source | Focusing illumination is<br>the same as imaging<br>illumination | 12 V Halogen lamp | LED (940 nm) |
| Focusing | On-system gaze fixation<br>targets and manual<br>focusing by the operator | Matching two<br>oscillating points | Matching two laser<br>spots |
| Field Angle | 20 to 25 degrees | 50, 35, 20 degrees | 21 degrees<br>(Horizontal: 21<br>degrees; Vertical: 12<br>degrees) |
| Feature | XyCAM RI System | RFI 3000<br>(Predicate Device) | LSFG-NAVI<br>(Reference Predicate) |
| Working<br>Distance | 21 mm | 40 mm | 44 mm |
| Principle of<br>Blood Flow<br>Measurement | Laser Speckle Contrast<br>Imaging | Frame mapping of<br>xenon flash lamp<br>illumination | Laser Speckle<br>Flowgraphy |
| Measurement<br>Duration | 3 to 6 seconds at 82 Hz | 125 ms for data<br>acquisition | 30 Hz (typically 4 to 6<br>seconds) |
| Measurement<br>Results | Relative value | Given in mm/sec | Relative value |
| Numerical<br>Data | Blood flow velocity<br>index (BFVi) | Velocity flow rate | Blood flow Index |
| Graphical<br>Data | Blood Flow Velocity<br>map of retina<br>Blood Flow Velocity<br>map of select<br>regions/vessels<br>Spatial trend of BFVi<br>Temporal trend of BFVi | Path of flow | Perfusion map of<br>retina<br>Time variation of<br>blood flow index |
| Essential<br>Performance<br>and Electrical<br>Safety<br>Standard | • ANSI/AAMI ES60601-1:<br>2005 (3rd Edition)<br>• IEC 60601-1-2: 2014<br>(4th Edition) | • IEC 60601-1: 1st<br>Edition<br>• IEC 60601-1-2: 1st<br>Edition | • IEC 60601-1:2012<br>• IEC 60601-1-2: 2007 |
| Other<br>Performance<br>Standard | • ANSI Z80.36-2016<br>• IEC 60825-1: 2007<br>• ISO 14971: 2007 | • ISO 15004-2: 2007<br>• ISO 10940: 1998<br>• ANSI RP-27-1-96:<br>1st Edition<br>• ISO 14971: 2000 | • ISO 15004-2:2007<br>• ISO 10940: 1998<br>• ISO 60825-1: 2007<br>• ISO 14971: 2007 |
- VI. Comparison of Technological Characteristics with Predicate Device and Reference Predicate
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "K193319" on the left side of the image. To the right of the text is a blue logo that looks like a stylized letter "W". Below the logo is the text "VASOPTIC MEDICAL INC."
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the text "K193319" on the left side of the image. To the right of the text is a blue logo that looks like a stylized "W". Below the logo is the text "VASOPTIC MEDICAL INC.". The logo and text are likely associated with the company Vasoptic Medical Inc.
While the underlying technologies of XyCAM RI (assessment of blurring in the speckle pattern) and of the predicate (temporal changes in reflectance photographic frames) are different in some aspects, they are substantially equivalent in their approach, the risk profile, and efficacy of producing blood flow information. The measurement output of
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "K193319" on the left side of the image. To the right of the text is a blue logo that looks like a stylized letter "W". Below the logo is the text "VASOPTIC MEDICAL INC."
the XyCAM RI is a blood flow velocity index (BFVi), a quantitative measure of relative blood flow information, that has been shown to possess repeatability and reproducibility characteristics that are on par or better than the predicate, and conveys substantially equivalent information pertaining to the flow of blood in the human retina. The technological differences do not raise new concerns of safety or effectiveness in the intended use of the XyCAM RI to provide quantitative outputs of relative blood flow information in the human retina. The XyCAM RI has established safety and effectiveness through design verification of the system including software, electromagnetic compatibility, medical product safety, substantial equivalence in output performance, and system validation via clinical testing.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Illumination Classification and Safety
The XyCAM RI is classified as a Class 1 Laser Product and Group 1 Ophthalmic Instrument as defined by IEC 60825:2007 and ANSI Z80.36-2016. respectively. Refer to Section 18.1 for a complete illumination classification and safety analysis.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the XyCAM RI System. The system complies with the ANSI/AAMI ES60601-1 standard for product safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered to be a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator. The XyCAM RI System is designed to provide quantitative outputs of relative blood flow velocity information to clinicians and is not intended to provide any diagnostics or clinical decision support. However, the software is considered to possess a moderate level of concern because a failure or latent flaw could indirectly result in minor injury through the action of a care provider.
### Bench Testing
XyCAM RI bench testing involved the use of a calibrated flow pump which provided baseline velocity measurements to analyze against. The results of bench testing demonstrated the XyCAM RI is able to reliably capture retinal blood flow with an intrasession coefficient of variation significantly less than 10%.
Comparative assessment was conducted between the XyCAM RI output of blood flow velocity index (BFVi, in arbitrary units) and RFI 3000 output of blood velocities (in mm/s) when imaging the same tubular blood flow. The results demonstrated substantial
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "K193319" on the left side of the image. To the right of the text is a blue logo that looks like a stylized letter "W". Below the logo is the text "VASOPTIC MEDICAL INC."
equivalence between the two outputs and consistent repeatable measurements of physiologically relevant blood flow velocities obtained using the XyCAM RI.
## Clinical Studies
Clinical testing was conducted to validate that the XyCAM RI is consistently able to generate images of the back of the eye in a diverse range of subjects, and to characterize the repeatability and reproducibility of blood flow measurements when used in the clinic as intended.
Twenty (20) healthy subjects were recruited and imaged, with strict adherence to inclusion and exclusion criteria which were developed to obtain data from a cohort that is representative of the general population. There were 9 male subjects and 11 female subjects imaged, with a combined average age of 46.9 ± 16.6 years. The demographic distribution of subjects is displayed in the table below.
| Number of subjects | Age Range | | | |
|--------------------|-----------|----------|------|-------|
| Race | 21 to 40 | 40 to 60 | > 60 | Total |
| African American | 2 | 4 | 1 | 7 |
| Caucasian | 4 | 0 | 5 | 9 |
| Asian | 2 | 2 | 0 | 4 |
| Total | 8 | 6 | 6 | 20 |
To assess measurement errors across imaging sessions and operators, all subjects were divided into three groups: A1 (n=10), A2 (n=5), and A3 (n=5). All subjects in Group A1 was imaged with the field of view centered on the Optic Nerve Head (ONH) and on the macula by one operator. All subjects in Group A2 and A3 were imaged by three operators with the field of view centered on the ONH in Group A2 and centered on the macula in Group A3. To assess repeatability, multiple imaging sessions were conducted by each operator. For each imaging session, five regions of interest (ROIs) were assessed using the XyCAM RI Software, and analyzed across multiple cardiac cycles within an imaging session for intra-session repeatability and across multiple imaging sessions for intersession repeatability.
Over all regions analyzed, the XyCAM RI's mean intra-session CV was 2.883 ± 1.049% and the mean inter-session CV was 7.069 ± 2.018% in imaging centered around the ONH region and the mean intra-session CV was 2.782 ± 2.044% and the mean inter-session CV was 6.447 ± 3.778% in imaging centered around the macular region.
#### VIII. Conclusion
The XyCAM RI is substantially equivalent to the predicate device's intended use, functionality and performance. The XyCAM RI successfully passed required verification, validation and product safety testing to demonstrate overall device safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.